TR, TE and TI: Contrast Weighting, Timing Limits and Slice Capacity
MRIninja Knowledge Base | MRI Parameter Deep Dive Parent page: MRI Parameters — Overview and Classification (9501), Sections 4.1–4.4 Builds on: Echo Spacing, Turbo Factor and Echo Trains in TSE (9561) Version 1.1 — November 2026
MRI Parameter Deep Dive
TR, TE and TI: Contrast Weighting, Timing Limits and Slice Capacity
What the three timing parameters set, in the signal equations and on the console: the weighting each pair of TR and TE produces, how many slices fit in one TR, what interleaving and concatenations do, why the null inversion time depends on TR and not only on T1, which limits stop a value from being chosen, and which labels a Siemens console shows for them.
1. Scope, Boundary and Information Status
The overview page 9501 (Sections 4.1–4.4) states in a few paragraphs what TR, TE and TI are and gives a table of typical values. This page does not repeat that table. It adds what 9501 does not contain: the signal equations that make the three times matter, the contrast calculation with tissue values, the capacity of one TR (how many slices fit, how slices are ordered, what concatenations are), the dependence of the null inversion time on TR, the limits that fix a minimum TE or TR, and the console labels. Elsewhere: echo spacing, turbo factor and the effective TE of a turbo spin echo (TSE) are on 9561; partial echo and partial Fourier on 9563; the acceleration of slice acquisition by simultaneous multi-slice on 9564; the choice of flip angle, presaturation and black-blood preparation for fluid signal on 9598; the refocusing flip angle of a TSE train, its effect on effective TE and the flip-back pulse on 9566; the inversion-recovery sequence itself on 9020; the physics of the RF pulse and of the Ernst angle on 9204; the number of slices, the slice gap and the scan order on 9590, 9591 and 9592; SAR on 9210.
This page is the home of the timing parameters as controls: what they set, how they limit one another, and how slice capacity is organised. Typical protocol values for a given examination belong to the protocol page of the region.
Information status. No guideline, consensus statement or systematic review defines these parameters; they are textbook physics. The sources are of four kinds: (1) one experimental study on the null inversion time (phantoms and volunteers); (2) educational texts, lecture notes and one book-style web resource, each read through a summarising tool; (3) two manufacturer texts (a glossary and a job aid) and protocol listings seen only as search extracts; (4) one audit of a hospital department. Each reference states its reading level. No textbook, no IEC standard and no operator manual was read (Section 15). Statements marked (Reasoning) are the editor's inference or calculation from the sources, not statements of a source.
2. The Three Times: Definitions as the Sources State Them
The sources agree on the meaning of the three times, with a small difference about where TE starts.
| Parameter | Definition in the sources | Source |
|---|---|---|
| TR (repetition time) | The interval between successive excitations [6]. A manufacturer glossary adds that it is generally the time between two excitation pulses, that it is one of the parameters that determines contrast, that the acquisition time is directly proportional to it, and that within one TR the signals may come from one or several echo times or phase encodings, depending on the technique [11]. | [6], [11] |
| TE (echo time) | The time between the excitation pulse and the echo used as the MR signal, which determines image contrast [11]; the interval between excitation and measurement of the induced current [6]. For a spin echo the 180° refocusing pulse is placed at half the TE: in the example of one article, an 80 ms echo uses a refocusing pulse 40 ms after excitation [6]. For a gradient echo one online book defines TE as the time between the start of the RF pulse and the maximum of the signal [5]. | [5], [6], [11] |
| TI (inversion time) | The interval between a 180° inversion pulse and the 90° excitation pulse in an inversion-recovery sequence [11]. | [11] |
| Effective TE (TEeff) | In a train of echoes, the time between the excitation pulse and the echo at which the phase-encoding gradient has its smallest amplitude [11]. The consequences for a TSE are on 9561. | [11] |
The reference point of TE is not the same in all the texts: the book chapter measures it from the start of the RF pulse for a gradient echo [5], the glossary from the excitation pulse without saying which point of it [11]. The difference is of the order of half the pulse duration and does not change any statement on this page (see Section 13).
3. Signal Equations and What "Weighted" Means
3.1 The four equations
One online book gives the following signal equations, with S the signal, k a constant of the detector, ρ the spin density, and θ the rotation angle [4]:
| Sequence | Signal equation as printed [4] |
|---|---|
| Spin echo | S = kρ (1 − e^(−TR/T1)) e^(−TE/T2) |
| Inversion recovery (180–90) | S = kρ (1 − 2e^(−TI/T1) + e^(−TR/T1)) |
| Inversion recovery with spin-echo readout (180–90–180) | S = kρ (1 − 2e^(−TI/T1) + e^(−TR/T1)) e^(−TE/T2) |
| Gradient echo | S = kρ (1 − e^(−TR/T1)) sin θ e^(−TE/T2*) / (1 − cos θ e^(−TR/T1)) |
The same chapter states that the values of T1, T2 and ρ are specific to a tissue or a pathology [4]. Three readings follow directly from the spin-echo equation and are the content of every statement about "weighting" (Reasoning, from [4]): the term (1 − e^(−TR/T1)) is the fraction of the longitudinal magnetisation that has recovered by the next excitation, so TR controls how much T1 differences show; the term e^(−TE/T2) is the fraction of transverse magnetisation left at the echo, so TE controls how much T2 differences show; and ρ multiplies everything, so it is always present.
3.2 Conditions for the three weightings
The same chapter gives a table of conditions in which the symbols were garbled in the page as read; as read, T1 weighting needs a TR of the order of T1 or shorter and a TE much shorter than T2; T2 weighting needs a TR much longer than T1 and a TE of the order of T2 or longer; spin-density weighting needs a TR much longer than T1 and a TE much shorter than T2 [4]. Other sources state the same rules in words and give numbers:
| Source | T1-weighted | Proton-density-weighted | T2-weighted | Remark |
|---|---|---|---|---|
| Pipe, Barrow Quarterly 2000 [6] | TR 500 ms, TE 20 ms | TR 4 s, TE 20 ms | TR 4 s, TE 80 ms | Figure 5 of the article; field strength not stated in the text read |
| Clarke, AAPM 2004 slides [9] | Short TR, short TE (no numbers) | TR 2350 ms, TE 30 ms (multiple sclerosis example) | TR 2350 ms, TE 80 ms (same example) | Field strength of the example not stated; the slides define PD as long TR with short TE and T2 as long TR with long TE |
| Sung, UCLA M219 lecture 2022 [8] | Short TE, intermediate TR | Short TE, long TR | Intermediate TE, long TR | Qualitative table; a TSE example with TR 6500 ms and an echo train of 12 is shown at five TEs between 12 and 235 ms |
| Radiographics tutorial 1994 [2] | — | At a TR of about 10 times the tissue T1 and a minimum TE, differences from relaxation vanish and contrast arises from proton density only | — | Abstract only |
The article of Pipe adds two statements useful to a user of the console: setting TR near the average T1 of the tissues of interest adds T1 contrast, and setting TE near their average T2 adds T2 contrast; and T1 and T2 weighting are rarely combined, because tissues bright on a T1-weighted image tend to be dark on a T2-weighted one [6]. The lecture of Sung states that a longer TR minimises T1 contrast and a short TE minimises T2 contrast, while an intermediate TR maximises T1 contrast and an intermediate TE maximises T2 contrast [8]. For the specific numbers of a protocol see the protocol pages; the table of typical values of 9501 §4.1 has no source read for this page.
4. Tissue Relaxation Values Behind the Contrast
A calculation of contrast needs T1, T2 and ρ. The published values differ, and the differences are as large as the contrast being calculated.
| Tissue | Hornak, 1.5 T ranges [4]: T1 (s) / T2 (ms) | Sung, 1.5 T [8]: T1 / T2 (ms) | Table of a 2014 article [17]: 1.5 T T1 / T2; 3 T T1 / T2 (ms) |
|---|---|---|---|
| White matter | 0.76–1.08 / 61–100 | 790 / 92 | 884 / 72; 1,084 / 69 |
| Grey matter | 1.09–2.15 / 61–109 | 925 / 100 | 1,124 / 95; 1,820 / 99 |
| CSF | T1 printed as 0.8–20 s; T2 110–2,000 ms | 2,400 / 180 | — |
| Muscle | 0.95–1.82 / 20–67 | 875 / 47 | 1,008 / 44; 1,412 / 50 |
| Fat (adipose) | 0.2–0.75 / 53–94 | 260 / 85 | — |
| Liver | — | 500 / 43 | 576 / 46; 818 / 42 |
Three points follow from the table itself. (1) The three sources do not agree for the same tissue and field strength: the grey-matter T1 at 1.5 T is 925 ms in one, 1,124 ms in another and a range of 1,090 to 2,150 ms in the third. (2) T1 is longer at 3 T than at 1.5 T in the 2014 table for every tissue listed (for example 1,124 to 1,820 ms in grey matter), whereas T2 changes little [17]. (3) A review of in-vivo values at 3 T states that wide ranges have been reported for similar tissues, for example from 699 to 1,735 ms for the T1 of white matter and from 41 to 371 ms for the T2 of fat, and that the true values are still unclear [3]. The review explains the spread by the methods used to measure the times [3]; only its abstract was read. The consequence for the console (Reasoning): a TR or TI taken from a table is a starting point for a field strength and a sequence, not a physical constant; the protocol must be checked on an image.
4.1 A worked calculation of spin-echo contrast (editor's calculation)
The following values were calculated by the editor from the spin-echo equation [4] with the 1.5 T relaxation values of Sung [8], assuming equal spin density in every tissue, which is not true (Hornak lists different ρ ranges [4]), so the numbers show the effect of relaxation only. Signal is relative to a fully relaxed tissue with no T2 decay (Reasoning, calculation).
| Tissue | "T1-weighted": TR 500 ms, TE 15 ms | "PD-weighted": TR 2,500 ms, TE 15 ms | "T2-weighted": TR 4,000 ms, TE 100 ms |
|---|---|---|---|
| Grey matter | 0.36 | 0.80 | 0.36 |
| White matter | 0.40 | 0.81 | 0.34 |
| CSF | 0.17 | 0.60 | 0.47 |
| Fat | 0.72 | 0.84 | 0.31 |
| Muscle | 0.32 | 0.69 | 0.12 |
| Liver | 0.45 | 0.70 | 0.10 |
Reading the table: at TR 500 ms CSF is the darkest and fat the brightest tissue, which is the T1-weighted pattern described in the sources [6][8]; at TR 2,500 ms and TE 15 ms the tissues lie between 0.60 and 0.84, so the relaxation contrast is small and proton density decides, as the tutorial abstract states [2]; at TE 100 ms CSF is the brightest and muscle and liver the darkest. The grey-white difference at the T1-weighted setting is 0.04 on a scale of 0.4, about 10 %, and it reverses sign at the T2-weighted setting. With the 1.5 T grey- and white-matter values of the 2014 table [17] instead of Sung's, the same calculation gives 0.31 and 0.35 at the T1-weighted setting (white matter still brighter), 0.76 and 0.76 at the PD setting, and 0.34 and 0.25 at the T2-weighted setting: the ordering is the same, but the T2-weighted grey-white difference changes from 0.03 to 0.09, which is the point of Section 4.
5. What TR Sets: Contrast, Time, Slice Capacity and SAR
5.1 Scan time
The acquisition time of a conventional (one echo per TR) two-dimensional sequence is TR × number of phase-encoding steps × number of averages [8]; the lecture of Sung writes it as T_scan = TR · PE · N_avg and gives TR 1,000 ms, 256 steps and one average as 4 min 16 s [8]. The book chapter states that the time to produce an image is the product of TR and the number of phase-encoding steps [5]. The manufacturer glossary gives the same product for a 2D measurement (number of scans × TR × number of acquisitions) and states that the acquisition time is directly proportional to TR [11]. For a sequence that measures one slice at a time, the lecture adds a factor of the number of slices, and divides by the number of interleaved slices when slices are interleaved [8]. In a turbo spin echo the lines measured per TR equal the number of echoes in the train: the glossary states that within one TR raw-data rows equal to the number of echoes of the train are acquired, and that at a turbo factor of 7 the sequence measures 7 times faster than a spin echo with comparable parameters [11]; in the lecture an example with an echo train of 16 and TR 2,500 ms takes 2 min 51 s against 22 min 21 s for the conventional spin echo with the same TR [8]. The echo-train arithmetic (trains per slice, ceiling, effective TE) is on 9561.
5.2 TR and SAR
A manufacturer job aid on SAR states that a longer TR spreads the RF pulses over a longer time and so lowers the average power absorbed; that SAR scales with the second power of the flip angle; that reducing the number of slices at constant TR reduces SAR because the number of slices is directly tied to the number of RF pulses; and that reducing the turbo factor at constant TR reduces SAR and increases scan time [12]. It also states that the system gives a SAR prediction before each scan, that the prediction can differ from the measured value because it rests on assumptions and because an accurate prediction needs the adjustment of the transmit chain (done just before the measurement), and that the recommended workflow is to check the prediction after the adjustment scans [12]. The Clarke slides state that SAR limits force a smaller number of slices per study and a longer TR [9]. The mechanism and the limits are on 9210 and the transmit calibration on 9708. A discrepancy within the Siemens material is recorded in Section 13.
5.3 What limits a short TR
Sources read state three limits (the first and third are not quantified in them):
- The duration of everything that must be played inside one repetition: RF pulses, gradients, readout. An abstract on oblique imaging states that slew rates and maximum currents of the gradient amplifiers limit the minimum TR and TE by adding dead periods (intervals in which neither RF is applied nor data acquired) to the sequence [19]. The abstract was read as an extract only, without its authors.
- The number of slices to be fitted in the TR (Section 6.2).
- SAR (Section 5.2). A glossary entry for TrueFISP states that TR should be as short as possible for that sequence [11]; the other sources do not give a numerical lower limit.
6. Slice Capacity: Interleaving, Slice Order and Concatenations
6.1 How multi-slice acquisition uses the TR
The manufacturer glossary describes multi-slice imaging as a variant of sequential imaging that uses the recovery period of the first excited slice to measure further slices, so that the slices are interleaved and time is saved [11]. The Rad229 lecture states that interleaving is time-efficient when the sequence has dead time and that it improves acquisition efficiency [7]. The Clarke slides state that multi-slice imaging improves efficiency by scanning several planes during the TR interval [9].
6.2 The capacity of one TR
The Rad229 lecture gives the maximum number of interleaved slices as N_max = TR / T_seq or TR / T_min, where T_seq is built from the pulse durations and T_min from the RF power limit; if the number of slices exceeds N_max, the scan takes two, three or more times as long as needed [7]. A lecture slide of Sung states that the speed-up of interleaving is limited by N_interleaves ≈ TR/TE and by SAR [8]; the two statements are consistent only when the time one slice needs is about its TE (Section 13). For FLAIR and STIR the Rad229 lecture notes that the TI interval gives extra dead time in which other slices can be acquired, which adds constraints on TR, TI and T_seq without giving formulas [7].
In words (Reasoning, from [7]): the number of slices per TR is the TR divided by the time one slice occupies the sequence; shortening TR for T1 contrast therefore removes slices from each pass, and the remaining slices need another pass.
6.3 Slice order and cross-talk
The Rad229 lecture lists three orders for eight slices: sequential (0, 1, 2, 3, 4, 5, 6, 7), odd/even (0, 2, 4, 6, 1, 3, 5, 7) and bit-reversed (0, 4, 2, 6, 1, 5, 3, 7), and says that different orders reduce the saturation of adjacent slices [7]. The glossary lists the options of the slice sequence as ascending, descending, interleaved (1, 3, 5 … then 2, 4, 6 …) and freely defined [11]. It defines cross-talk as the influence of adjacent slices on one another when slices are too close, especially at zero distance, and states that it is remedied primarily by an interleaved slice sequence [11]. The Sung lecture states that interleaving is difficult for adjacent slices because a good 180° slice profile is hard to match to the 90° profile [8]. The effect of the gap and the choice of order are the subject of 9591 and 9592; this page records only that the slice order is one of the settings that interacts with TR.
6.4 Concatenations
The manufacturer glossary defines concatenation as distributing the slices to be measured over several measurements and gives two applications: at a short TR, increase the number of concatenations to be able to measure more slices; and, to prevent cross-talk when the slice distance is short, set two concatenations and use an interleaved slice sequence [11]. A web resource states the same two uses, adds that scan time rises with the concatenation level ("doubling the concatenation factor" doubles scan time, in its words) and that for covering a larger volume raising TR is often preferred, and that shorter breath-hold scans can result from a higher concatenation [14]. A patent text states that the echo trains of two adjacent slices in TSE imaging are often produced in two concatenations to prevent slice cross-talk [18].
The job aid states that two concatenations will reduce the TR by a factor of two at the same scan time, and that the interplay of TR and concatenations defines the number of pulses per unit time and so the SAR [12]. The two statements on scan time are consistent: they hold a different quantity constant (Reasoning). With TA ≈ n_concatenations × (number of TR repetitions per concatenation) × TR × averages (from [11] and [14]), doubling the concatenations at the same TR doubles TA, while doubling them and halving TR leaves TA unchanged.
| If the slices do not fit in one TR | What it does to contrast | What it does to scan time and SAR | Source |
|---|---|---|---|
| Raise TR until all slices fit | T1 weighting decreases | Scan time rises in proportion to TR; average power falls | [5], [11], [12] |
| Add concatenations at the same TR | Weighting kept | Scan time rises in proportion to the number of concatenations; cross-talk reduced when the slice distance is small | [11], [14], [18] |
| Reduce the number of slices at the same TR | Weighting kept; coverage reduced | SAR reduced; scan time unchanged | [12] |
| Accelerate slice acquisition by simultaneous multi-slice (SMS) | Depends on the new TR | An audit reports shorter scan times with fewer concatenations or shorter TR; SAR limits must be watched | [13] |
6.5 What one audit reports on concatenations and TR
An audit of a hospital department (Leeds, 1.5 T Siemens system, SMS factor 2, work done in February and March 2022) lists the changes made to musculoskeletal TSE protocols and the resulting scan times [13]. Rows in which concatenations were changed are given here; all rows combined SMS with the change, so the saving is not due to the concatenations alone.
| Sequence | Scan time before | Scan time after | Change as listed [13] |
|---|---|---|---|
| Ankle T1 TSE sagittal | 2:32 | 1:10 | SMS 2; concatenations 2 → 1 |
| Hip T1 TSE coronal | 4:38 | 2:15 | SMS 2; concatenations 2 → 1 |
| Scaphoid T1 TSE coronal | 3:42 | 1:41 | SMS 2; concatenations 2 → 1 |
| Hand T1 TSE coronal | 3:02 | 1:48 | SMS 2; concatenations 2 → 1 |
| Hand T1 TSE fat-suppressed transverse | 5:19 | 2:59 | SMS 2; concatenations 5 → 3; TR 983 → 860 ms |
| Knee PD TSE fat-suppressed transverse | 4:47 | 2:27 | SMS 2; TR 4,160 → 3,000 ms |
The audit states that the ankle T1 sequence showed no noticeable change in image quality while saving over 50 %, that one knee sequence was rejected by the radiologists for image quality, and that care must be taken with SAR limits [13]. It is one department, one system and one set of reviewers, and the audit does not report an objective image-quality measure in the text read; it is useful as an example of how TR and concatenations are traded, not as evidence that a given change is safe on another system.
7. What TE Sets and What Limits It
7.1 Effects of TE
TE sets the T2 (or T2*) weighting through the factor e^(−TE/T2) (Section 3). For a spin echo the refocusing pulse reverses the dephasing caused by field inhomogeneity but not the T2 decay, which arises from molecular-scale field changes; a gradient echo without a refocusing pulse is sensitive to field homogeneity at long TE, which is the basis of functional imaging [6]. The longer the TE, the more signal is lost: the strongest signal comes from a long TR and a short TE [6]. The choice of TE for susceptibility contrast and for fluid signal is on 9501 §4.2 and 9598.
7.2 What fixes the minimum TE
- Everything that has to be played between the excitation and the echo: the end of the RF pulse, the slice-rewinding and phase-encoding gradients, the dephasing lobe and the first part of the readout. The abstract on oblique imaging attributes the minimum TE and TR to gradient amplifier limits and dead periods [19].
- For a spin echo, the 180° pulse sits at TE/2, so TE cannot be shorter than twice the time needed between the excitation and the refocusing pulse [6] (Reasoning, from the placement at TE/2).
- Partial (fractional) echo reduces the minimum TE: if the entire echo does not need to be recorded, the minimum TE is decreased and more signal may be achieved [5]. The details, the limit and the Siemens label are on 9563.
- In a turbo spin echo, TEeff is one of the echoes of the train and is limited below by the echo spacing (9561).
- On the console the minimum is displayed: in the owner's Siemens screenshots (3 T) the TE of an echo-planar diffusion series is shown greyed at its minimum of 62 ms [20]. The rule that fixes this minimum was not read.
8. TI: The Null Point, Finite TR and Fast Readout
8.1 Long TR: TI = T1 · ln 2
The educational site of Elster states that when TR is much longer than T1 the null inversion time is T1 · ln 2, about 0.69 × T1, and gives approximate null values at 1.5 T for a TR of 4,000 ms or more: fat about 180 ms, white matter about 400 ms and grey matter about 650 ms; because T1 rises with field strength these values must be adjusted upward at 3 T or higher [10]. The result follows from the inversion-recovery equation of Section 3: with TR much longer than T1 the term e^(−TR/T1) is negligible, and S = 0 gives 1 − 2e^(−TI/T1) = 0, that is TI = T1 ln 2 (Reasoning, derived from [4]). With the fat T1 of 260 ms of the Sung table [8] the relation gives 180 ms, in agreement with the value quoted by Elster (Reasoning, calculation).
8.2 Finite TR: the null TI is shorter
Setting S = 0 in the 180–90 equation of Section 3 without dropping e^(−TR/T1) gives TI_null = T1 · [ln 2 − ln(1 + e^(−TR/T1))] (Reasoning, derived from [4]). The same relation is printed in an experimental study of fast inversion recovery, which states that for a conventional readout the null TI of a tissue depends only on its T1 and on the selected TR [1]. Because ln(1 + e^(−TR/T1)) is positive, the null TI at a finite TR is shorter than 0.69 × T1. The following values are calculated by the editor (Reasoning, calculation) with the T1 of the two phantoms of that study (corn oil 218 ± 5 ms; distilled water 3,175 ± 70 ms [1]); the measured values are those the study reports with a hybrid-RARE (fast spin echo) readout and an echo train of 4 at 1.5 T [1].
| Phantom (T1) | TR (ms) | Null TI from the relation, conventional readout (ms) | Null TI measured with the fast readout, echo train 4 [1] (ms) |
|---|---|---|---|
| Corn oil (218 ms) | 700 | 143 | 127 |
| Corn oil (218 ms) | 4,000 or more | 151 (= 0.69 × T1) | no longer differs from the conventional value (the study states the effect on fat disappears at a TR of about 4,000) |
| Water (3,175 ms) | 2,000 | 845 | 800 |
| Water (3,175 ms) | 6,000 | 1,754 | 1,615 |
| Water (3,175 ms) | 20,000 | 2,195 (limit 2,201 = 0.69 × T1) | no effect reported at 20,000 (the effect on water is stated to disappear at a TR of about 20,000) |
The study states that with a fast readout the time available for signal recovery is approximately TR − TI − (echo train length × echo spacing); that the effect of the readout on the null TI is negligible when this quantity exceeds four to five times T1 and is amplified when it falls below that; and that users of fast STIR and fast FLAIR should account for it [1]. For CSF the Elster page states that its T1, of the order of 3,000–4,000 ms, makes the long-TR simplification fail, that modern inversion recovery often uses a fast spin echo with 16–32 echoes so that the time of the last echo cannot be ignored, and that the CSF null point is therefore not one number but depends on the TR and on the time of the last echo [10]. In a human-brain figure caption of the study, CSF was nulled with an echo train of 16 at an echo spacing of 17 ms, TR 6,000 ms and TI 2,000 ms [1].
8.3 What this means at the console
- A TI is a value for a tissue, a field strength and a TR (and, for a fast readout, an echo-train duration). Changing the TR of an inversion-recovery protocol moves the null point (Reasoning, from [1] and [10]).
- The glossary states that STIR is an inversion recovery with a short TI used for fat suppression and that the TI selection depends on field strength; it describes the turbo inversion recovery as a turbo spin echo with a long effective TE used to suppress fluids, with the magnitude variant (TIRM) showing only the magnitude [11].
- In the owner's Siemens screenshots (3 T) the diffusion series uses a slice-selective inversion with TI 210 ms as its "magnetization preparation", with TR 19,900 ms [20]. Why that TR is so long was not found in a source; a likely contributor is the number of slices that each need their own inversion and recovery (Reasoning, from the capacity argument of Section 6.2). The preparation is a diffusion page matter (9012).
9. Siemens Console Labels
The labels below are those seen in Siemens documents and in the owner's screenshots of one 3 T examination (inventory of 2026-11-23, taken from the project record, not re-opened for this page). They are given as a mapping; other makers use other words, and the meaning of a label is stated only where a source defines it.
| Console label | Where seen | What the sources say | Source |
|---|---|---|---|
| TR, TE, TI | Routine and Contrast cards; the protocol listings show TR and TE on the Routine card | Definitions in Section 2; TE of the diffusion series greyed at its minimum (62 ms) | [11], [15], [20] |
| Concatenations | Routine card (for example "Concatenations 2" in a T1 TSE shoulder listing; 6 in the owner's T1 TSE transverse series) | Number of measurements over which the slices are distributed (Section 6.4). The same parameter is called "acquisitions" by one other maker and "packages" by another, according to a trial protocol | [11], [15], [16], [20] |
| Multi-slice mode | Geometry card | Values seen: Sequential, Interleaved, Single shot (listings of a research site and a journal supplement; a T1 TSE listing shows Interleaved). The glossary describes sequential and interleaved multi-slice imaging; "single shot" in the glossary refers to the acquisition of all image information in one pulse train | [11], [15] |
| Series | Geometry card | Values seen: Ascending, Interleaved. The glossary lists ascending, descending, interleaved and freely defined as slice sequences. A web page also states that the Siemens "distance factor" is a percentage of the slice thickness while other makers use a gap in millimetres (see 9591) | [11], [15], [21] |
| Magn. preparation (Magnetization preparation) | Contrast card | "None" in a T1 TSE listing; "slice-selective IR" with TI 210 ms in the owner's diffusion series | [15], [20] |
| TD | Contrast card ("TD 0.0 ms" in a T1 TSE listing; "TD 0" in the owner's T1 TSE series) | Meaning in TSE not found. The glossary defines the trigger delay time (TD) as the interval between the trigger and release of the measurement, and mentions a delay time in a 3D inversion-prepared sequence that is "necessary to prevent saturation effects"; neither is shown to be the TD of the Contrast card of a TSE | [11], [15], [20] |
10. What Changing One Timing Parameter Does
The table gathers the directions of change that the sources state or that follow from the equations of Section 3; cells that rest on the editor's reading are marked.
| Change | Contrast | Signal and noise | Scan time | Slice capacity and SAR |
|---|---|---|---|---|
| TR longer | Less T1 weighting [8] | More recovered magnetisation, more signal (Reasoning, from [4]) | Longer, in proportion [11] | More slices per TR [7]; average power lower [12] |
| TR shorter | More T1 weighting [6] | Less signal (Reasoning, from [4]) | Shorter, in proportion [11] | Fewer slices per TR [7]; average power higher [12]; limited below by gradient and RF timing [19] |
| TE longer | More T2 (or T2*) weighting [6] | Less signal: the factor e^(−TE/T2) falls [4] | Not changed in a conventional sequence | Fewer slices when the time per slice grows (Reasoning, from [7]) |
| TE shorter | Less T2 weighting [8] | More signal [4] | Not changed | Limited below by the sequence timing [19]; partial echo lowers the minimum [5] |
| TI changed (inversion recovery) | Moves the null point [10] | Signal of the tissue near the null point goes to zero [4] | Not changed by TI itself; TR must stay long enough for recovery [1] | TI adds dead time in which other slices can be measured [7] |
| Concatenations increased | Weighting kept at the same TR | No change stated | Longer, in proportion at the same TR [14] | More slices measured at a short TR [11]; cross-talk reduced [11] |
11. Two Worked Examples from the Owner's Screenshots
The owner's screenshot inventory lists a T1 TSE transverse series (TR 400 ms, TE 8.70 ms, 50 slices of 3.5 mm, concatenations 6, averages 1, TA 2:45) and a T2 TSE coronal series (TR 8,760 ms, TE 126 ms, 40 slices, averages 2, concatenations 1, TA 3:05), both from one 3 T head-and-neck examination [20]. The turbo factor and the number of phase-encoding lines of these series were not captured, so the calculations below are partial and cannot be reconciled further (as already noted for 9561).
| Series | Arithmetic (Reasoning, calculation) | Result | What it shows |
|---|---|---|---|
| T1 TSE transverse | TA 165 s ÷ (6 concatenations × TR 0.4 s) | about 69 TR repetitions per concatenation | Scan time follows concatenations × repetitions × TR (Section 6.4); includes any fixed overhead, so only an upper bound of the repetitions |
| T1 TSE transverse | 50 slices ÷ 6 concatenations | about 8 to 9 slices per pass; with TR 400 ms the sequence would occupy at most about 44 ms per slice (400 ÷ 9) | Capacity argument of Section 6.2; the actual time per slice was not read |
| T2 TSE coronal | TA 185 s ÷ (TR 8.76 s × 2 averages) | about 10.6 echo trains per average | With one concatenation and a long TR the repetitions are few; a turbo factor would then give the number of lines (not read) |
The first row illustrates the point of Section 6.4: the T1 series keeps a TR of 400 ms (T1 weighting) and pays for coverage with six passes, so its scan time is six times what one pass would take at the same TR, whereas the T2 series needs one pass because its long TR leaves room for all slices (Reasoning, from [7], [11], [14]).
12. The Technologist's View (Reasoning)
- Before lengthening TR to fit slices, check what weighting the protocol is meant to have: in a T1 series a longer TR reduces T1 contrast, so more concatenations (with more scan time) or fewer slices are the choices that keep the weighting.
- When the scan time suddenly doubles after a change of coverage or TR, look at the concatenations first.
- When copying an inversion-recovery protocol between field strengths, or after changing TR or the echo train, do not copy the TI: it was tuned for a tissue T1, a TR and a readout (Section 8).
- A minimum TE shown greyed is a result of the other parameters (bandwidth, resolution, pulse type); to shorten it, the other parameters must change (the rule was not read for this page).
- Names differ between makers: concatenations, acquisitions and packages are the same idea [16]; distance factor and gap are not the same unit [21].
- The SAR prediction is checked after the adjustment scans, as the manufacturer advises [12].
13. Discrepancies Between Sources
| Topic | Statement A | Statement B | Status |
|---|---|---|---|
| SAR remedies | Job aid (2021): a longer TR lowers the average power; fewer slices at constant TR lowers SAR [12] | Glossary (undated), as read through the tool: remedies are other RF pulses, smaller flip angles, lower TR and fewer slices [11] | Not resolved. Physics (power averaged over TR) supports the job aid, and 9501 §4.1 states that SAR rises as TR falls; the glossary text may have been summarised wrongly. Check the original glossary entry. |
| Concatenations and scan time | Job aid: two concatenations reduce TR by a factor of two at the same scan time [12] | Web resource: doubling the concatenation factor doubles scan time [14] | Consistent when read as different quantities held constant (Section 6.4, Reasoning) |
| Start of TE | Book chapter: for a gradient echo from the start of the RF pulse [5] | Glossary: from the excitation pulse, point not stated [11] | Difference of the order of half a pulse duration; not resolved |
| Limit on interleaved slices | Rad229: N_max = TR / T_seq [7] | Sung slide: N_interleaves ≈ TR/TE [8] | Consistent only if the time per slice is about the TE (Reasoning); the first is the general form |
| Tissue relaxation values | Grey-matter T1 at 1.5 T: 925 ms [8]; 1,124 ms [17]; 1,090–2,150 ms [4] | White-matter T1 at 3 T: 699–1,735 ms across the literature [3] | Not reconcilable; methods differ [3]; values are starting points (Section 4) |
| FLAIR TI at 3 T | 9501 §4.3 table: FLAIR TI 2,200–2,400 ms at 1.5 T and 1,700–1,900 ms at 3 T (no source read for this page) | Elster: TI values must be adjusted upward at 3 T because T1 is longer [10]; the null relation of Section 8.2 gives a longer TI for a longer T1 at the same TR | Not resolved. A shorter TI at 3 T could arise only if TR or the readout differ (Reasoning). The 9501 values need a source and a stated TR; flagged for the next 9501 revision. |
| Null TI of fat at 1.5 T | 9501 §4.3: 150–175 ms [no source read here] | Elster: about 180 ms at TR of 4,000 ms or more [10]; the editor's calculation gives 175 ms at TR 1,000 ms and 180 ms at 2,000 ms or more (T1 260 ms [8]) | The ranges are compatible if the 9501 values assume a shorter TR; to be confirmed with a source |
| Meaning of TD | Trigger delay time [11] | Delay time in a 3D inversion-prepared sequence [11]; TD shown on the Contrast card of a TSE [15], [20] | Not found for TSE (Section 15) |
14. Open Questions
- What rule fixes the minimum TE and the minimum TR on a given console, and how does it change with bandwidth, matrix and RF pulse type? (Operator documentation not read.)
- What does the TD of the Contrast card do in a TSE or a diffusion series?
- For which TR and readout were the FLAIR and STIR values of 9501 §4.3 chosen at 1.5 T and 3 T?
- How large is the difference between the scan-time saving expected from a change of concatenations and the saving measured on a given scanner (the audit gives only combined changes)?
- Is the TSE effective TE reported identically by all makers (the glossary defines it by the smallest phase-encoding step)?
15. Evidence Gaps and Attempts Made
- No guideline, consensus or systematic review exists on these parameters in what was found; the page rests on educational texts and one experimental study.
- No textbook read. A teaching site cites a handbook of pulse sequences for the null-TI relations; the handbook was not available. A RadioGraphics physics tutorial of 2005 was tried at its publisher address, which returned a client error; only its figure legends were seen in search extracts. The 1994 RadioGraphics tutorial was read as an abstract.
- Relaxation values: only the abstract of a review of 3 T values [3] was read; the full text and the 2005 in-vitro study of Stanisz and colleagues (search record and abstract only) were not read. The 2014 table [17] was seen only as a search extract of a table, without authors, and its page was refused (not on the session allowlist).
- Manufacturer documents: the glossary and the job aid were read in full through a summarising tool; protocol listings only as search extracts; no operator manual. The glossary has no date. The labels TD and the options of "Multi-slice mode" are therefore only partly documented.
- Minimum TE and TR: only an extract of an abstract on oblique imaging; no source gives the rule or numbers for a clinical system.
- Slice capacity formulas are from lecture slides (Rad229, year not shown in the text read; UCLA 2022); no peer-reviewed derivation was read.
- Concatenations: the definitions are manufacturer and web-resource statements; no study isolates the effect of concatenations on image quality or cross-talk; the audit [13] combines changes.
- Not covered on purpose: flip angle and Ernst angle (9204, 9598), echo spacing and turbo factor (9561), and protocol values by region.
16. Where to Go Next
For the echo train, effective TE and scan time of turbo spin echo see 9561; for partial echo and the minimum TE see 9563; for SMS and its effect on TR and slices see 9564; for the inversion-recovery sequence, FLAIR and STIR see 9020; for fluid signal and the flip angle see 9598 and 9204.