Echo Spacing, Turbo Factor and Echo Trains in Turbo Spin Echo
MRIninja Knowledge Base | MRI Parameter Deep Dive Parent pages: MRI Parameters — Overview and Classification (9501); Turbo Spin Echo (TSE) Sequence (9005) Builds on: Acquisition Bandwidth (9597); Acquisition Matrix (9582) Version 1.0 — November 2026
MRI Parameter Deep Dive
Echo Spacing, Turbo Factor and Echo Trains in Turbo Spin Echo
How the three numbers on a TSE console — echo spacing, turbo factor and echo trains per slice — are linked to each other and to bandwidth, scan time, T2 blurring and the effective TE, with the relations written out, a published quantitative model applied to console values, and the places where the console numbers could not be reconciled.
1. Scope, Boundary and Information Status
The choice of echo-train length for each type of weighting (short for T1, 8–24 for T2, the SAR and J-coupling consequences) is on Turbo Spin Echo (TSE) Sequence (Section 8.1) and in the overview MRI Parameters (Sections 4.4 and 6.1); the receiver-bandwidth conventions, the SNR cost and the chemical-shift effect are on Acquisition Bandwidth; variable refocusing flip angles and 3D TSE are on 9005 (Section 8.4). This page does not repeat them. It is the home of the arithmetic that connects the three console numbers: what echo spacing is made of, how the number of echo trains follows from the number of phase-encoding lines and the turbo factor, how the duration of the echo train enters a published point-spread model of T2 blurring, how the effective TE follows from echo spacing, and how far these relations hold on real console values.
Information status. Six publications and one set of console screenshots were used. Only one clinical study with patients was found that varies echo-train length and echo spacing separately [1]. The two most quantitative sources [2, 3] were read through a tool that returns a model-written answer about the page, not the raw text. Three landmark or technical papers [4, 5, 6] were available as abstract or first-page excerpt only. No guideline, consensus statement or systematic review addresses these parameters. Statements marked (Reasoning) are derived by the editor; gaps are marked (to be sourced). Console values come from screenshots of one 3 T head-and-neck examination supplied by the site owner; the series to which several captures belong is inferred, not stated, and this is said wherever it matters.
| Source | Type | Reading level and weight |
|---|---|---|
| [1] Sze et al., AJNR 1993 | Prospective clinical study with flow phantom, 1.5 T, cervical spine | Full text through a summarising tool; small subgroups; contrast and artefact outcomes, not diagnostic accuracy; patient count inconsistent in the paper (43 stated, 41 in the results) |
| [2] Mugler, JMRI 2014 | Peer-reviewed technical review (3D TSE) | First 100,000 of 122,921 characters through a summarising tool; reference list cut off; the examples are 3D, the definitions general |
| [3] Qin, 2012 | Peer-reviewed technical derivation | First 100,000 of 156,781 characters through a summarising tool; analytical and simulated, no phantom or in-vivo validation seen; spiral part not read |
| [4] Busse et al., MRM 2008 | Technical article (3D TSE, flip-angle modulation) | Abstract only (full text blocked); not used for numbers |
| [5] Constable and Gore, MRM 1992 | Technical article (point-spread function) | Abstract only; used for one qualitative statement |
| [6] Hennig et al., MRM 1986 | Landmark (original RARE) | Abstract and first page excerpt only; used for the historical definition |
| [7] Console screenshots | Unpublished observations, Siemens 3 T (and one Philips 3 T capture set) | Read visually; not vendor documentation; series assignment inferred where stated |
2. Definitions and Console Labels
| Quantity | Definition | Console label seen |
|---|---|---|
| Echo-train length (ETL) / turbo factor | Number of 180° (or lower-angle) refocusing pulses, and therefore of echoes, played after one excitation; each echo is phase-encoded differently [1, 6] | Siemens: Turbo Factor. Philips: TSE factor (with separate startup echoes). Other vendors (to be sourced) |
| Echo spacing (ES) | Interval between successive echoes [1] | Siemens: Echo Spacing (ms, read-only). Philips: shown on the Info panel (TSE echo spacing and shot duration) |
| Echo trains per slice | Number of excitations (shots) needed to fill the k-space of one slice for one average (Reasoning: standard definition, not found in a source read) | Siemens: Echo Trains per Slice (read-only). Philips: number of shots (to be sourced) |
| Echo-train duration Tk | Time from the first to the last echo of one train; used in the blurring model [3]. Approximately turbo factor × ES when no echoes are discarded (Reasoning) | Not shown directly |
| Effective TE (TEeff) | Echo time of the echo that fills the centre of k-space [2] | Siemens: TE |
Console labels were read from the owner's screenshots, not from vendor documentation.
3. What Echo Spacing Is Made Of
3.1 Readout, RF and gradient components
Between two echoes the sequence must play, in order: half of one refocusing pulse, the phase-encoding gradient, the readout gradient with its sampling window, the rewinding phase-encoding gradient, spoiler (crusher) gradients, and half of the next refocusing pulse. The echo spacing is the sum of their durations (Reasoning: written from the sequence diagram; no closed-form equation is given in the sources read, which state the relation qualitatively [2]):
ES ≈ tread + tRF + tgrad, with tread = Nread × dwell time = 1 / BWpx (see 9597, Section 2.1).
Two worked examples from a review of 3D TSE with a receiver bandwidth of 500 Hz/pixel (sampling time 1/500 = 2 ms) show how small the readout can be in the total [2]: with slice-selective refocusing pulses of 3.84 ms the ES was 8.5 ms, the RF pulse occupying about 45 % of it; with non-selective pulses of 0.6 ms the ES was 3.9 ms, the RF pulse about 15 %. In the multi-slab case the minimum ES was set by the time between excitation and the first refocusing pulse, which is half the ES, and phase-encoding and spoiler gradients also have to fit in the interval; shorter, stronger crushers were used when no slice selection was needed [2]. These are 3D examples; the values for clinical 2D TSE are (to be sourced).
3.2 What the console shows
Five Sequence-tab captures of the Siemens head-and-neck examination give a bandwidth and an echo spacing. Subtracting the readout time (1000 / BW in ms) leaves the time taken by everything else (Reasoning, arithmetic only):
| Capture (sequence name) | BW (Hz/Px) | 1000 / BW (ms) | ES shown (ms) | ES minus readout (ms) | Flow compensation |
|---|---|---|---|---|---|
| tse | 255 | 3.92 | 8.52 | 4.60 | None |
| tseR | 207 | 4.83 | 9.66 | 4.83 | None |
| qtseR_rs | 255 | 3.92 | 9.55 | 5.63 | Slice (greyed) |
| tseR_rr (a) | 252 | 3.97 | 11.0 | 7.03 | Read |
| tseR_rr (b) | 525 | 1.90 | 9.24 | 7.34 | Read |
The non-readout part is 4.6–7.3 ms, far from constant. The two captures with flow compensation set to Read have the largest remainders (7.0 and 7.3 ms), which is what extra gradient lobes would produce (Reasoning; five captures from unidentified series with different RF-pulse and gradient-mode settings: a suggestion, not a test). The effect of the RF-pulse type and of the gradient mode on ES could not be separated with these data (to be sourced).
The practical consequence is that at a fixed bandwidth ES cannot be reduced below the sum of the non-readout components, and that raising the bandwidth shortens only the readout term: in the last capture, a bandwidth of 525 Hz/Px leaves 7.3 ms of non-readout time in a 9.24 ms echo spacing.
4. Echo Trains per Slice and Scan Time
4.1 From phase-encoding lines to echo trains
Each echo fills one k-space line, so the number of trains is the number of acquired phase-encoding lines divided by the turbo factor, rounded up (Reasoning):
Ntrains = ceil( Nlines / TF )
Illustration with values typical of the captured series (base resolution 272, phase resolution 70 %): Nlines = 272 × 0.70 = 190. A turbo factor of 21 gives ceil(190/21) = 10 trains, in which 210 echo slots are available for 190 lines (20 slots not needed). A turbo factor of 15 would give ceil(190/15) = 13 trains: the train is 29 % shorter, and the acquisition 30 % longer (13 against 10 excitations). Partial Fourier and parallel imaging reduce Nlines; their rules are on 9588 and in 9501 (Section 6.2). In multi-slab or single-slab 3D TSE the mapping of trains to k-space planes is a design choice, and the number of trains is not given by a formula in the source read [2].
4.2 Scan time
For a multi-slice 2D TSE the time is the repetition time multiplied by the number of excitations (Reasoning):
TA ≈ TR × Ntrains × averages × concatenations, plus a preparation overhead of the order of one TR on the console used (below).
The reduction relative to conventional spin echo is by the factor ETL: for 3D acquisitions the time falls in proportion to ETL [2]. In the original description of the method a 256 × 256 image was acquired in 2 to 40 s [6].
4.3 Console check
Two series of the Siemens examination could be paired with a Sequence-tab capture (the pairing rests on the TE/ES ratio in Section 6, so it is inferred). The overhead of one TR fits both within about a second:
| Series | TR (s) | Averages | Trains (paired capture) | TR × trains × averages (s) | Plus one TR (s) | Acquisition time shown |
|---|---|---|---|---|---|---|
| t2_tse_cor | 8.76 | 2 | 10 | 175.2 | 184.0 | 3:05 (185 s) |
| t2_tse_sag | 5.68 | 2 | 11 | 125.0 | 130.6 | 2:12 (132 s) |
Two agreements do not establish the one-TR overhead as a rule. The other series (axial T2, T2 Dixon, T1) could not be reconciled with any capture: for the axial T2, for example, the two candidate captures give 207 s or 232 s with the same one-TR overhead, against 230 s shown. They are not used here, and they remain an open calibration item (to be sourced: a set of Routine, Sequence and Resolution captures of the same series taken together).
5. T2 Decay Along the Echo Train: Blurring and Signal Loss
5.1 Mechanism
The echoes at the end of the train have decayed by T2 for longer. If they are mapped to the outer, high-spatial-frequency k-space lines, the edges are attenuated relative to the centre, and the point-spread function broadens [2]. For the original FSE, RARE and EPI the analysis of the point-spread function showed that this blurring can reduce the contrast of small objects and that the signal from small lesions with short T2 can be lost; it may not be obvious in the image because the noise is unaffected by the point-spread function [5]. Guidance from a 3D TSE review (180° refocusing): for a short effective TE the echo train should be shorter than the T2 of interest; for a long effective TE, below about two to three times T2; for routine high-resolution 2D imaging ETL is about 30 or fewer; and T2-weighted brain trains are typically under 300 ms, since white- and grey-matter T2 is roughly 100 ms at 1.5 T and at 3 T [2].
5.2 A quantitative model
A derivation for Cartesian acquisition with constant echo spacing, a single mono-exponential T2 and one dimension (T2* and off-resonance ignored) expresses the loss in terms of the ratio x = Tk/T2 [3]. The amplitude loss at the centre of the point-spread function, for either ordering, is (1 − exp(−x)) / x. The blurring (full width at half maximum, in pixels) was obtained by fitting second-order polynomials to plotted point-spread functions, and is therefore empirical [3]:
- linear (sequential) ordering: FWHM = 1.23 − 0.05 x + 0.08 x²;
- low–high (centric) ordering: FWHM = 1.20 + 0.17 x + 0.04 x²;
- truncation alone (Tk much shorter than T2): FWHM = 1.21 pixels.
The SNR per unit time (loss × √Tk) is maximal at x ≈ 1.2 for the Cartesian case; at x = 4 it is 77 % of that maximum [3]. Evaluating the published expressions (Reasoning: arithmetic):
| x = Tk/T2 | Centre amplitude | FWHM, linear (pixels) | FWHM, low–high (pixels) |
|---|---|---|---|
| 0.25 | 0.88 | 1.22 | 1.24 |
| 0.5 | 0.79 | 1.23 | 1.29 |
| 1 | 0.63 | 1.26 | 1.41 |
| 2 | 0.43 | 1.45 | 1.70 |
| 4 | 0.25 | 2.31 | 2.52 |
The amplitude values (0.88, 0.79, 0.63, 0.43, 0.25) are those stated in the article [3].
5.3 Worked example with console numbers
Take the capture with turbo factor 21 and ES 9.66 ms: Tk ≈ 21 × 9.66 = 203 ms (Reasoning, no startup echoes assumed). For tissue with T2 = 100 ms, x = 2.0: centre amplitude 0.43, FWHM 1.46 pixels (linear) or 1.71 (low–high). For tissue with T2 = 50 ms, x = 4.1: amplitude 0.24, FWHM 2.3–2.6 pixels. The same train is therefore harmless for long-T2 fluid and a wide point-spread function for short-T2 structures, which is the reason why the T2 of the target structure, not the protocol default, should set the train length (Section 8). The model assumes a single T2 and ideal 180° refocusing; with reduced refocusing angles the signal evolution differs [2] and the model does not apply (Reasoning).
5.4 Clinical evidence: cervical spine at 1.5 T
The only clinical study found that varied ETL and ES separately examined 43 patients with suspected cervical degenerative change at 1.5 T (23 sagittal, 18 axial in the results), 3 mm sections, and a flow phantom [1]. Converting the tested conditions to train duration (Reasoning: ETL × ES) and placing the authors' observations beside them:
| ETL × ES (ms) | Tk (ms) | Observed blurring (sagittal) [1] |
|---|---|---|
| 4 × 20 | 80 | None |
| 8 × 20 | 160 | Minimal |
| 12 × 20 | 240 | Significant on early echoes |
| 16 × 20 | 320 | Significant on early echoes (ETL series); severe in 8 of 11 patients at ES 20–22 ms in the ES series |
| 16 × 14 | 224 | Mild (short TR/TE) |
| 16 × 16 to 16 × 18 | 256–288 | Moderate (short TR/TE) |
Other findings [1]: a longer ETL raised the CSF-to-cord and CSF-to-disc contrast, mainly by relative increase of CSF signal, improved CSF homogeneity and shortened the acquisition; a longer ES slightly reduced contrast and SNR, reduced the myelographic effect and increased motion, edge-enhancement, truncation-type and susceptibility artefacts; the maximum usable TE is about ETL × ES; lesions were seen regardless of ETL. The authors recommended ETL 8 for the short-TE and 16 for the long-TE sagittal sequence, with ES 14 ms, and ETL 12 with ES 14 ms for the axial 3D sequence. The ordering of blurring by Tk is consistent across the rows, but the subgroups were different, small (8–12 patients), and the study measured contrast and artefact, not diagnostic accuracy; it was a single 1.5 T system with ES values (14–30 ms) much longer than the 8–11 ms of the current console captures. It gives no threshold that can be transferred to other systems.
6. Effective TE
The effective TE is the echo time of the echo that fills the centre of k-space: TEeff = (index of the central echo) × ES; for example the 5th echo with ES = 20 ms gives 100 ms [2]. In the clinical study the maximum usable TE was about ETL × ES [1]. With reduced refocusing angles the contrast can correspond to a shorter TE than TEeff (one example in a 3D TSE review: TEeff 585 ms with the contrast of about 140 ms), and the review notes that manufacturers use "effective TE" for either meaning [2]; the refocusing-angle effect is on 9005 (Section 8.4).
Consequence for the console (Reasoning): the TE that can be set is quantised in steps of ES, because the centre of k-space must coincide with one echo. Three captures support this when paired by ratio:
| Series | TE shown (ms) | ES of the paired capture (ms) | TE / ES |
|---|---|---|---|
| t2_tse_cor | 126 | 9.66 | 13.04 |
| t2_tse_sag | 124 | 9.55 | 12.98 |
| t1_tse_tra | 8.70 | 8.52 | 1.02 |
All three ratios are within about 2 % of an integer. The pairing was chosen with this ratio in mind, so the table shows consistency, not an independent confirmation. The residual differences (up to 0.5 % for the T2 series, 2 % for the T1) are not explained (to be sourced); the displayed ES is rounded to 0.01 ms. A ratio near 1 for the T1 series (turbo factor 5) implies that the centre of k-space is filled by the first echo.
7. Interactions and Trade-offs
| Change | ES | Tk | Trains and time | Other consequences |
|---|---|---|---|---|
| Raise bandwidth per pixel | Shorter, by the readout term only | Shorter | Unchanged | SNR falls with the square root of bandwidth and chemical shift decreases (9597); more gradient demand (Reasoning) |
| Raise turbo factor | Unchanged | Longer | Fewer trains, shorter scan; the rounding up of Nlines/TF limits the gain | More blurring for short-T2 tissue; more RF power per TR (SAR scales with the number of refocusing pulses and with the square of the flip angle [2]); T1 weighting diluted (9005) |
| Lower turbo factor | Unchanged | Shorter | More trains, longer scan (Section 4.1 example: +30 %) | Less blurring; lower SAR; allows a TR to contain fewer slices when TR is short (Reasoning) |
| Enable flow compensation | Longer (Reasoning, Section 3.2) | Longer | Unchanged at fixed TF | Reduces flow-related dephasing; increases minimum TE |
| Lower flip angle of the refocusing pulses | Not changed by the angle itself (Reasoning) | Usable train longer [2] | Fewer trains possible | SAR falls with the square of the angle; signal and contrast evolution change (9005 Section 8.4) |
| Raise phase matrix (more lines) | Unchanged | Unchanged | More trains, longer scan | Smaller pixels; the effect on blurring depends on the view ordering (to be sourced) |
The echo spacing and turbo factor are not independent: for a given echo-train duration, halving the ES allows twice the number of echoes, (Reasoning). Shorter RF pulses and non-selective refocusing pulses are among the means used to shorten the ES in single-slab 3D TSE [2].
8. Technologist Decision Guide (Reasoning)
These steps are derived from Sections 3–7; they are not a protocol.
- Compute Tk before judging a train. Tk ≈ turbo factor × ES. Compare it with the T2 of the structure that must stay sharp: the sources read give "shorter than T2" for short effective TE and "two to three times T2" for long effective TE [2], and an SNR-per-time optimum near Tk/T2 ≈ 1.2 [3]. These criteria optimise different things and do not agree on one value.
- Check the effective TE against the ES. If the TE the console proposes differs from the one wanted, the nearest allowed value is a multiple of ES.
- To cut blur without losing time, shorten the ES first (bandwidth, non-selective or shorter refocusing pulses where the platform allows) rather than lowering the turbo factor; the second costs trains.
- To cut time, check first whether the last train is mostly empty: Nlines/TF rounded up wastes slots when the remainder is small (190/21 gives 10 trains although 9.05 would suffice).
- Read the echo-train count back from the console after any change of matrix, phase resolution, partial Fourier or acceleration; a changed count is the earliest sign that the line count changed.
9. Discrepancies, Controversies and Limits of the Evidence
- Which echoes count in Tk. Some consoles add startup (dummy) echoes before the first acquired echo. On a Philips 3 T capture set of a 3D TSE, a TSE factor of 20 with three startup echoes and ES 7.1 ms gave a shot duration of 170 ms; (20 + 3) × 7.1 = 163 ms leaves about 7 ms of unexplained overhead (Reasoning, from console captures; not vendor documentation). Tk = TF × ES is therefore a lower bound.
- Different optimisation targets. The model optimum for SNR per time (x ≈ 1.2) [3], the "shorter than T2" rule [2] and the empirical contrast and artefact preferences of the clinical study [1] answer different questions; no source read states a consensus value.
- Model limits. The blurring fits are empirical, one-dimensional, for one T2 and constant ES; no validation in phantoms or volunteers was seen in the text read [3]. Real tissues have several T2 components, and tissue T2 also varies with field strength (to be sourced).
- Clinical evidence. One study of 43 patients (41 in its results) at 1.5 T with ES of 14–30 ms [1]; no study found compares ES and turbo-factor settings against diagnostic accuracy on current systems (to be sourced).
- Not covered by any source read: vendor formulas for minimum ES, the dependence of ES on RF-pulse type and gradient mode, the effect of ES on acoustic noise and nerve stimulation (the gradient demand rises as the ES shortens (Reasoning); see 9807), the J-coupling contribution to fat brightness in 2D TSE (cited on 9005, not read here), and the console check of the axial T2, Dixon and T1 series.
10. Evidence-Based References
No guideline, consensus statement or systematic review addresses echo spacing, turbo factor or echo trains. Sources are ordered by evidence level and then by relevance; the reading level of each is stated.
A. Guidelines / Consensus / Systematic Reviews
None found.
B. Prospective Clinical Study
C. Technical Papers
D. Landmark Historical Reference
E. Console Observations (Unpublished)
End of document — Echo Spacing, Turbo Factor and Echo Trains in Turbo Spin Echo — MRIninja v1.0 — November 2026