MRI RF Transmit System and Body Coil — Transmit Chain, Power Amplifier, Birdcage Body Coil, Quadrature and Two-Channel Drive, Transmit Calibration and Power Monitoring
MRIninja Knowledge Base | Foundations — Scanner Hardware Parent page: MRI Physics — Fundamentals and Principles (9201) Builds on: MRI Scanner Structure and Components — Overview (9703); Magnetic Field, Hardware and Homogeneity (9208) Version 1.1 — November 2026
Foundations — Scanner Hardware
MRI RF Transmit System and Body Coil
The hardware that turns a digital pulse into the B1 field: the transmit chain from waveform synthesis to the coil ports, the power amplifier and its reported values, the birdcage body coil, quadrature and two-channel drive, transmit calibration, and the hardware that monitors delivered power.
1. Scope, Boundary and Information Status
The physics of the RF pulse and of the flip angle is on RF Pulses and Pulse Sequence Physics. The physics of transmit and receive coils, the B1 non-uniformity at high field and the concept of multi-transmit and B1 shimming are on Magnetic Field, Hardware and Homogeneity (Sections 6.1 and 7); this page does not repeat them. The position of the RF system among the other blocks is on 9703 (Section 5). The physics of SAR and its limits is on SAR, Bioeffects and MRI Safety Physics, and RF heating and burns are on 9804. Receive coils as devices are on MRI Receive Coils and are not covered here, except where the transmit side forces a design (Section 5).
This page is the home of the hardware detail and of the numerical values of the transmit side: the stages of the chain, the amplifier, the birdcage construction, the drive modes of the body coil, the transmit calibration and the power monitoring.
Information status. Fourteen sources were read, all through a tool that returns a model-written answer about the page, not the raw text. Seven are peer-reviewed: three reviews [1, 2, 3], one measurement study on real scanners [4], and three studies based on simulation or on small volunteer groups [5, 6, 7]. Five are ISMRM educational items [8, 9, 10, 11, 12]. Two are manufacturer documents [13, 14]. No standard (including IEC 60601-2-33), textbook or platform documentation was read. No guideline, consensus statement or systematic review exists for this topic in the sources found. Several numbers come from simulations, from 3 T research systems or from scanner models of unstated age; they are labelled where they appear. Statements marked (Reasoning) are derived by the editor; gaps are marked (to be sourced).
| Source | Type | Reading level and weight |
|---|---|---|
| [1] Padormo et al., NMR Biomed 2016 | Peer-reviewed review (parallel transmission) | First 100,000 of 109,099 characters through a summarising tool; reference list partly missing; ultrahigh-field emphasis; used for pTx hardware and SAR supervision |
| [2] Gruber et al., JMRI 2018 | Peer-reviewed review (RF coils for non-physicists) | First 100,000 of 122,782 characters; birdcage construction, rung and end-ring design and balun details were not in the portion read |
| [3] Ahmad et al., Diagnostics 2020 | Peer-reviewed review (birdcage coils) | First 100,000 of 163,452 characters; tables and figures not captured; no body-coil design discussion in the text read; used for birdcage types and modes |
| [4] El-Sharkawy et al., Med Phys 2012 | Peer-reviewed measurement study | Full text through a summarising tool; figures not available; two 3 T scanner models, whole-body coils, 17 volunteers |
| [5] Wu et al., NMR Biomed 2015 | Peer-reviewed simulation study | First 100,000 of 104,739 characters; simulation only; funding included the manufacturer of one platform |
| [6] Childs et al., MAGMA 2013 | Peer-reviewed volunteer study (n = 7) | Full text through a summarising tool; some figure values garbled |
| [7] Brink et al., 2015 | Peer-reviewed simulation and volunteer study | Full text as supplied; journal not named in the text read; "data not shown" for several results |
| [8] Gudino, ISMRM 2022 | Educational lecture summary | Full text; the two chain diagrams were captions only |
| [9] Twieg, ISMRM 2017 | Educational abstract | Full text; no duty-cycle values despite the title |
| [10] Scott, ISMRM 2015 (entry 8213) | Educational tutorial, condensed web extract | Entire extract read; no figures, tables or equations; may not match the original tutorial in full |
| [11] Boskamp, ISMRM 2015 (entry 8337) | Educational syllabus | Full text; the author states that the lecture covers more than the text; no rung count, dimensions or capacitor values |
| [12] King, ISMRM 2016 | Educational synopsis | Full text; no figures, no numerical tolerances |
| [13] Blasche, Riffel, Lichy, MAGNETOM Flash 2012 | Manufacturer article (two of three authors employed by the manufacturer) | Full text; few volunteer examples; no statistics; product described as under development at the time |
| [14] Harvey, Hoogeveen, Philips white paper 2025 | Manufacturer white paper | Full text; qualitative claims; no numerical uniformity or SAR values |
2. The Transmit Chain from Waveform to Coil
The transmit chain has to synthesise a complex RF waveform, amplify it with high fidelity, monitor the power and deliver as much as possible to the coil [10]. The stages below follow the sources; the exact order and the position of the protective and monitoring elements differ between systems (Reasoning).
| Stage | Statement | Source |
|---|---|---|
| Waveform synthesis | All systems use direct digital synthesis; a mixer with a phase-locked-loop oscillator often performs the final conversion to the 1.5 T, 3 T or 7 T frequency; arrays commonly use ganged synthesis blocks. Resolution and update rate are not stated | [10] |
| RF power amplifier | See Section 3 | [8, 9, 10] |
| Isolator and circulator | Isolators and/or directional couplers protect the amplifier against varying load; circulators are often included to protect against severe load conditions; patient size varies widely, so amplifier protection depends on circulators | [9, 10] |
| Directional couplers | Placed between the amplifier and the coil to determine forward and reverse power flow; they flag extreme reverse power from coil failure or adverse patient loading. Accuracy is not stated | [10] |
| Cable and connection | The amplifier output reaches the coil through a long, low-loss, high-power 50 Ω coaxial cable; in one measured layout the cables pass through a filter box in the Faraday cage before the transmit switch | [8, 9]; [4] |
| Transmit/receive switch | Disconnects the transmit circuitry from the coil and connects the receive circuitry, to avoid noise injection and loading; almost always a PIN-diode circuit with an on-resistance under 0.5 Ω at a bias of 100 mA or more and a very high impedance when reverse biased | [10] |
| Quadrature hybrid | Acts as a two-way 90° splitter that produces the quadrature voltages for the two coil ports; a single source is split by the hybrid and the body coil is driven at two ports 90° apart | [8]; [14] |
| Location of the amplifier | In a remote equipment room, because parts of the amplifier may be ferromagnetic and because of the large water-cooled heat sinks; a manufacturer article places two independent low-level transmitters in a box at the magnet next to the receivers and does not mention the power amplifiers | [8, 9]; [13] |
3. The RF Power Amplifier and the Reported Power Values
3.1 Amplifier design
| Topic | Statement | Source |
|---|---|---|
| Class and efficiency | Commercial scanners mostly use voltage quasi-linear amplifiers (class AB) with an efficiency below 60 %; a second source gives class AB or class C. Linear classes trade linearity and bandwidth against efficiency and need large heat spreaders and exchangers | [8]; [9] |
| Pulsed operation, continuous DC | The amplifier works in pulsed mode but continuously draws DC current, hence large water-cooled heat sinks | [8] |
| Construction | Power-combined transistor arrays; many smaller amplifiers (described as under 2 kW each) are combined with splitters and combiners to reach tens of kW; VHF amplifiers mostly use LDMOS transistors | [10]; [9] |
| Switch-mode alternative | Theoretical efficiency 100 % and about 90 % in practice at VHF, but nonlinear, needing resonant filters and more complex power modulation | [9] |
| Bandwidth | MRI needs little amplifier bandwidth (under 100 kHz) | [9] |
| Nonlinearity | Memory effects are the least understood source of nonlinearity: with a complex envelope the power supply cannot track the demanded current, producing distortions that are hard to pre-calibrate. Droop and gain-compression figures are not stated | [10] |
| Duty cycle | The pulses occupy a small fraction of the repetition time; no percentage is given in either source that discusses it | [8, 9] |
3.2 Reported power, current, voltage and field values
| Quantity | Value as reported | Context stated by the source | Source |
|---|---|---|---|
| Peak electrical power to the coils | Up to 35 kW | At frequencies up to 300 MHz; depends on application, subject and field strength | [9] |
| Amplifier power | 15–35 kW | "Typically employed" in 1.5 T and 3 T scanners | [10] |
| Pulsed power | Tens of kW | Whole-body transmit efficiency stated as the order of tenths of µT/√W; power needs rise with field strength because of higher energy requirements and higher cable and circuit losses | [8] |
| Power for a 200 lb patient | About 30 kW | At 128 MHz; the field strength is not stated in the text | [11] |
| Peak RF current and voltage in the body coil | About 65 A; about 2 kV | Same passage | [11] |
| B1 for excitation | Order of 20 µT; pulse width limited by bandwidth to a few ms | Typical value, not tied to a sequence | [11] |
| B1+ for a short inversion pulse | Above 12 µT for a 180° pulse shorter than 1 ms | Statement in the lecture summary | [8] |
| Fraction of amplifier power at the coil ports | 56.5 ± 2.5 % (measured; about 59 % predicted from the losses) on one 3 T scanner model; 90 ± 2 % on another | Measured at the outputs of the quadrature hybrid relative to the amplifier output; the cable, filter box, switch and hybrid dissipate the rest | [4] |
| Coil loss | 8.8 W (short pulses) and 11.1 W (long pulses) | One 3 T scanner model, TR 50 ms, measured with a 1 L mineral-oil phantom | [4] |
Reading of the figures (Reasoning). The values do not describe one system. They come from different field strengths, patient sizes and measurement points, and the sources do not say whether "amplifier power" means installed capability or the power needed for a typical sequence. As an arithmetic check only: a B1 of 20 µT at 30 kW corresponds to about 0.12 µT/√W (20 µT divided by the square root of 30,000 W), which is of the same order as the "tenths of µT/√W" quoted in [8]; the two sources do not state that their conditions match, so this is a plausibility check and not a confirmation. The figure of 8–15 kW on the overview page and the values above are compared in Section 11.
Because the coil is loaded by the patient, the power needed for a given B1 depends on the patient. In coils in general the loaded quality factor is stated as 10–100 against 50–600 unloaded [2]. SAR may rise about fourfold when going from 1.5 T to 3 T [2]; the SAR physics is on 9210.
4. The Body Coil
4.1 Birdcage construction
| Topic | Statement | Source |
|---|---|---|
| Structure | Two conducting end rings joined by an even number of legs (rungs) running along the axis | [3] |
| Low-pass, high-pass, band-pass | Low-pass: identical capacitors in the legs, end rings short-circuited. High-pass: identical capacitors in the end-ring segments between legs. Band-pass (hybrid): capacitors in both legs and end rings, so it can be run in either mode | [3] |
| Resonant modes | An N-leg coil supports up to N/2 + 1 modes (Hayes) and up to N/2 + 2 when a co-rotating end-ring mode is counted (Leifer), as reported by the review | [3] |
| Useful mode | The dominant leg-current mode is responsible for the homogeneous B1 distribution in the transverse plane; higher-order modes have non-homogeneous fields and are avoided for imaging; end-ring modes have a field parallel to the axis and are unfavourable for imaging | [3] |
| Role in clinical systems | The review calls the birdcage the most successful RF coil for whole-volume MRI in clinical and research use; the body coil is built into the bore, is a volume coil used mostly for transmit, and can be connected for receive through a transmit/receive switch | [3]; [2] |
| Ports | Two feed ports 90° apart in the modelled body coil; a two-port high-pass birdcage in one 3 T system | [5]; [7] |
| Example geometry (simulation model) | 32 rungs, 500 mm long and 20 mm wide, on a cylinder of 704 mm diameter, inside a copper RF shield of 744 mm diameter and 1,200 mm length | [5] |
The geometry in the last row is the model used in a simulation study; it is not a product specification, and the rung count, diameters and capacitor values of commercial body coils were not found in the sources read (to be sourced).
4.2 Quadrature drive
In conventional single-source operation the two ports are fed with equal amplitude and a 90° phase shift, which generates a circularly polarised field in an empty coil [13, 14]. Circular polarisation needs half the transmit power of a linearly polarised coil for the same B1+, and gives a net SNR gain of about √2 on receive [2]; the review of birdcage coils cites the same √2 factor for quadrature detection [3]. The standard body coil can also be seen as two independent volume-coil modes located 90° apart [14]; the next section uses exactly this property.
4.3 RF shield, gradient interaction and cable currents
| Topic | Statement | Source |
|---|---|---|
| RF shield | An RF shield separates the RF space from the gradient space | [11] |
| Slits in the shield | The shield is slitted so that its impedance is low at the RF frequency (preserving the Q of the coil) and high in the 1–10 kHz gradient band (limiting eddy-current heating); slit geometry and number are not given | [11] |
| Electric field | The electric fields are highest between the end ring and the shield | [11] |
| Baluns | Narrow-band tank circuits connected to the cable shield at several points; they attenuate the currents that the electric field of the body coil induces on cables; such currents can disturb reception and can heat the patient if cables approach the patient | [11] |
The gradient side of this interface (eddy currents in the shield and the coil layers) is on MRI Gradient System Hardware.
4.4 Why the single body coil stops at 3 T
| Topic | Statement | Source |
|---|---|---|
| Frequency and wavelength | The ¹H frequency is 128 MHz at 3 T and about 300 MHz at 7 T; the wavelength in tissue is about 12 cm at 7 T, shorter than the body | [2] |
| Wavelength at 3 T | About 25 cm in tissue in one study and about 26 cm in the manufacturer article; at 1.5 T quadrature drive gives a sufficiently homogeneous B1+ field | [7]; [13] |
| Consequence at 3 T | Body coils are standard up to 3 T; at 3 T the abdomen can degrade, so independent control of the two ports was added in clinical systems; B1+ over the heart can vary by up to 50 % at 3 T (a figure the study cites from earlier work) | [8]; [7] |
| Birdcage limits at high frequency | Shorter wavelengths disturb the sinusoidal current distribution and form standing waves; stray capacitance can shift the resonance; longer legs radiate electric fields that raise SAR; remedies named are shorter legs, series capacitors and distributed capacitance | [3] |
| Ultrahigh field | Combined transmit/receive local coils are used because a body transmitter is lacking; multiple transmit channels are the most flexible approach | [2]; [1] |
5. Decoupling Between Transmit and Receive Coils
The receive elements sit inside the transmit field and must not absorb or distort it. The sources describe the following measures; the receive-coil structure itself is outside this page.
| Measure | Statement | Source |
|---|---|---|
| Active detuning | Receive elements are switched off during the transmit pulse by tank circuits and switches, usually PIN diodes; with a forward DC bias the diode places a parallel LC trap in series with the loop, blocking current at the Larmor frequency; one or more blocking networks are used per element depending on size and frequency | [11]; [2] |
| Passive second stage | An adjustable capacitor with crossed diodes, and an RF fuse, act as further safety stages for the receive loop and the patient; in the example described, three redundant safety features protect one loop | [2] |
| Preamplifier protection | The preamplifiers must be protected from transmit power; the transmit and receive coils must not couple | [2] |
| Cable currents | Baluns (Section 4.3) | [11] |
Consequence for practice (Reasoning). The protection of the receive loop and the cables depends on the correct connection and on the cable path; the rules on loops and contact with the patient are on 9804.
6. One Source, Two Ports, Several Channels
6.1 Configurations
| Configuration | Hardware as described | Source |
|---|---|---|
| Single source, quadrature | One amplifier, one hybrid, body coil driven at two ports 90° apart with fixed equal amplitude | [8, 14] |
| Two-channel (dual-source) body coil, static | Two dedicated transmit chains give independent control of the amplitude and phase driving the two ports of the same two-port birdcage; before each scan a calibration of under one minute maps the B1+ field of each channel and the shim is calculated. Described as installed as standard in the latest clinical 3 T systems of the major vendors at the time of the review | [7]; [1] |
| Two-channel, dynamic | Each of the two channels can carry its own arbitrary waveform, simultaneously with gradient shapes; the manufacturer article names this as the basis of zoomed excitation | [13]; [14] |
| Multi-element body arrays | An eight-element transmit body coil distributed azimuthally around the bore, with its own six-element receive array, on a 3 T system; elements need better than −15 dB isolation and a 50 Ω match; complexity grows with the channel count | [6]; [8] |
| Parallel transmission at ultrahigh field | Each channel needs its own amplifier and its own synthesiser output; example arrays: 8-channel dipole array at 7 T, 12-channel TEM head coil, 16-channel arrays; as the channel count rises the cables become thinner, which increases attenuation and needs more powerful amplifiers; amplifiers inside or next to the bore are proposed but need B0 compatibility and limited heating | [1]; [9] |
Static parallel transmission applies one waveform scaled by fixed channel weights and is called B1+ shimming; dynamic parallel transmission gives each channel its own waveform [1]. The weights or waveforms require some knowledge of the B1+ field of each channel, usually acquired in the subject; most mapping methods are slow and the channels are mapped sequentially [1].
6.2 How two manufacturer sources describe the two-channel shim
| Item | Manufacturer article A [13] | Manufacturer white paper B [14] |
|---|---|---|
| What is controlled | Amplitude and phase of the two channels of the body coil; the 90° phase of quadrature is abandoned | Two sources on the two body-coil modes; per source an independent waveform, phase offset, amplitude scaling and frequency offset |
| How the setting is chosen | Anatomy-specific settings, so no B1 mapping is needed; patient-specific mapping can take about 1 minute per body region | An in-vivo calibration determines the shim parameters; the paper states that the coefficients must be adjusted per patient; calibration duration not stated |
| Evidence given | A few volunteer examples; no statistics | Single examples, e.g. a lumbar spine T2-weighted TSE of 4:25 min with single transmit against 2:23 min with MultiTransmit; no numerical uniformity or SAR values |
| More than two channels | Mentioned as a possible extension | The benefit of more than two sources is stated as still to be defined and validated |
Both documents are written by the manufacturers, and neither is an independent evaluation. They describe different ways of choosing the shim (preset by anatomy against calibrated per patient); whether this reflects a real difference between platforms or a difference of emphasis cannot be decided from these texts (to be sourced).
6.3 Evidence on the effect of the extra channels
| Study | Design | Result as reported | Main limits |
|---|---|---|---|
| Childs 2013 [6] | 7 healthy volunteers; 8-channel transmit body coil on a 3 T system; 1, 2, 4 and 8 channels emulated from the 8-channel coil; pelvis and thigh | B1+ homogeneity improved significantly with each step from quadrature to 2, 4 and 8 channels (all p < 0.05); radiologist image-quality score (1–5) for thigh T1-weighted rose from 2.50 ± 0.55 in quadrature to 4.67 ± 0.52 with 8 channels, for pelvis T1-weighted from 4.00 ± 0.63 to 4.67 ± 0.52; T2-weighted images were less sensitive | Small sample; one radiologist; reduced-channel configurations emulated, not built; total power used as a proxy for global SAR; 3 T only |
| Wu 2015 [5] | Electromagnetic simulation at 3 T on one adult male body model; 32-rung birdcage in quadrature and in two-channel mode compared with stripline arrays | At matched homogeneity (T-spine, one-spoke design) the two-channel birdcage mode reduced local SAR to 28 % of quadrature; a two-ring array of 2 × 8 elements gave about an eightfold reduction of local and global SAR relative to quadrature with a two-spoke design; in the pelvis the two-channel birdcage gave lower SAR than the arrays | Simulation only; one body model; matching circuitry not modelled; mutual coupling not included in the pulse design |
| Brink 2015 [7] | Simulation on five body models and 33 volunteers (11 on a 60 cm bore, 22 on a 70 cm bore) with two-channel transmit and high-permittivity pads | Simulated B1+ inhomogeneity (coefficient of variation) in quadrature 10.7–15.0 % without pads against a minimum of 6.8–8.6 % after shimming; with pads the optimal shim lay near quadrature; for some non-optimal shim settings with pads the actual SAR was up to about 20 % higher than the scanner estimate | Not a patient study; five models; several results "data not shown"; the scanner SAR model does not account for the pads |
The common message of these studies is that additional transmit degrees of freedom improve B1+ homogeneity or reduce power, and that the gain depends on the anatomy. No guideline, consensus or systematic review quantifying the clinical effect was found. The safety issue is not the shimming itself but the SAR estimate under non-standard settings (Section 8).
7. Transmit Calibration
| Topic | Statement | Source |
|---|---|---|
| What is calibrated | Transmit power and transmit/receive frequency are adjusted for each patient exam. Before the scan the transmit power is set by measuring the flip angle at the isocentre | [12]; [8] |
| Why per patient | Patient size and composition within the sensitive volume of the transmit coil change the absorbed RF energy, so the power needed for the desired flip angle must be calibrated at the scanning location | [12] |
| Method | A calibration sequence finds the flip angle produced by a given transmit gain for a standard pulse; the gain for any other pulse follows by scaling; only a small part of the volume is mapped, usually within a slice at the isocentre; B1-mapping methods named as usable for this purpose are double-angle, actual flip-angle imaging, phase-sensitive imaging, DREAM and Bloch–Siegert | [12] |
| Repetition | The process need not be repeated unless the patient loading is substantially changed | [12] |
| Limit | B1 non-uniformity prevents the nominal flip angle from being reached everywhere; the optimum gain is hard to find when B1 varies strongly across the patient, for example at high field or with complex anatomy, and a trade-off with speed is needed | [12] |
| Frequency calibration | Must be set for each patient and sometimes several times per exam; errors cause slice-localisation errors, phase-encode shifts in EPI, banding in balanced SSFP, blurring in spiral imaging and poor fat saturation; it is hard in poor-B0 regions such as the neck | [12] |
| Tolerances and recalibration criteria | No numerical tolerances for gain or frequency are given in the source | [12] (to be sourced) |
Consequence for practice (Reasoning). Because the calibration is made in a slice at the scanning location and is tied to the loading, a large change of table position, of the patient or of the coil set-up is a reason for the system to repeat it. The source states only the loading criterion; how each platform decides when to recalibrate is (to be sourced). The effect of the flip-angle error on contrast is on 9204.
7.1 Console labels seen on one system (mapping view)
The following labels were read from screenshots of one 3 T head-and-neck examination on a Siemens console; the reading level is the screenshot, with no vendor document consulted. The meaning of each label is therefore a hypothesis (to be sourced).
| Console tab and label | Value seen | Possible meaning (Reasoning) |
|---|---|---|
| System / Tx-Rx: frequency 1H | 123.261512 MHz and 123.261494 MHz in two captures | Per-series centre frequency; the difference of 18 Hz is about 0.15 ppm of 123.26 MHz (arithmetic); consistent with a per-series frequency adjustment (Section 7) |
| System / Tx-Rx: reference amplitude 1H | 436.083 V and 442.632 V in two captures | Probably the result of the transmit calibration for the series; the difference is about 1.5 % (arithmetic); the definition of the quantity is not sourced |
| System / Adjustments: B1 shim | TrueForm | Name of the two-channel shim mode of that platform; see [13] for the manufacturer description |
| System / Adjustments: adjust with body coil | Off | Meaning not sourced |
8. Power Monitoring and SAR Supervision Hardware
| Topic | Statement | Source |
|---|---|---|
| What is measured | Forward and reverse power at directional couplers placed between the amplifier and the coil; in arrays, non-magnetic directional couplers near the coil terminals and magnetically coupled current sensors give the amplitude and phase of each channel | [10]; [8] |
| Global SAR | Estimated from forward and reflected power measurements, from the power monitors; reducing it requires lower duty cycle or power, at the price of longer scans and altered contrast | [1]; [2] |
| Local SAR | Cannot be measured directly; calculated from simulated electric fields on digital body models (grids of 1–5 mm³); evaluating every location involves roughly 10⁶–10⁸ matrices, reduced by virtual observation points (compression factors of the order of 5,000 for eight channels, with the overestimation of the maximum capped at 5 %) | [1]; [2] |
| Action on the sequence | Clinical scanners halt any sequence that exceeds the SAR limits; estimates are conservative, which can limit hardware and sequence performance | [8] |
| Standard | Limits come from IEC 60601-2-33; no numerical limits are given in the educational sources | [1, 8] (to be sourced) |
8.1 Independent measurement of delivered power
One study measured the power delivered by two 3 T scanner models independently, with a six-channel monitor built from three 50 dB directional couplers [4]. The scanners were a 127.8 MHz system with whole-body coil (11 volunteers: 9 men and 2 women, aged 22–65) and a 123 MHz system (6 men, aged 23–66), using field-echo sequences. Whole-body SAR was calculated as the power deposited in the subject divided by body weight.
| Scanner and source of the reported value | Maximum overestimate by the scanner |
|---|---|
| Model 1, short (1 ms) pulses | 78 % (1.8-fold) |
| Model 1, long (7 ms) pulses | 123 % (2.2-fold); the discussion of the paper gives "as much as 120 %" |
| Model 2, log file | 103 % (2-fold) |
| Model 2, console | 71 % (1.7-fold) |
The authors infer that one platform uses a worst-case estimate independent of the loading and the other seems to depend on subject weight; the vendor models are proprietary, so causes could not be established, and the two values shown by the second platform (log file and console) differ, with the manufacturer not stating which one is used [4]. Limits: two scanner models, 3 T, whole-body coils and a few sequences; only total and body-average SAR (local SAR requires electromagnetic modelling); the age and software version of the scanners are not stated, so the result cannot be transferred to current platforms (Reasoning). What the figures show is that the displayed SAR is a model output with a built-in margin, not a measurement of the energy actually absorbed.
9. Differences by Field Strength and System Type
| System | Transmit hardware as described | Source |
|---|---|---|
| 1.5 T | Single-source quadrature body coil gives a sufficiently homogeneous B1+ field | [7] |
| 3 T | Body coil still standard; B1+ non-uniformity in the trunk; two-channel drive installed in clinical systems; eight-channel body arrays and dielectric pads in research | [8]; [1]; [6]; [7] |
| 7 T and above | No body transmit coil; local transmit/receive or transmit arrays; parallel transmission with channel-specific amplifiers; local SAR supervision with virtual observation points | [2]; [1] |
| Low field | Excitation can be highly homogeneous; the main challenge is receive-side SNR (see the overview page); transmit hardware details of low-field and open systems | 9703 summary; (to be sourced) |
The field-strength dependence of the B1 problem is developed on High-Field Physics; the low-field magnets are on 9705.
10. Technologist View (Reasoning)
The statements below connect the sources to what is seen at the console. They are derived by the editor from the sources named and are not recommendations.
| What the technologist sees or does | Component behind it | Basis and limit |
|---|---|---|
| A prescan runs before the first series and sometimes again | Transmit gain and frequency calibration | [12]: per patient and location; recalibration criteria by platform (to be sourced) |
| A large patient reduces what the sequence can do (longer TR, fewer slices) | Higher power need and the SAR supervision | [11] gives about 30 kW for a 200 lb patient; [8] states that estimates are conservative and can limit performance; the link to a given protocol is not sourced |
| Image shading in the trunk at 3 T | B1+ non-uniformity of the body coil | [7, 13]; remedies: two-channel shim, pads [7] |
| High-permittivity pads placed on the patient | Passive B1 shimming | [7]: offsets under 5 cm made little difference in the simulations; the scanner SAR model does not account for the pads |
| Sequence stops or is refused for SAR | Power monitoring and SAR model | [8]; limits and modes on 9210 and 9800 |
| Coil cables placed away from the patient and without loops | Baluns and detuning of the receive loops | [11, 2]; rules on 9804 |
11. Discrepancies Between Sources
| Item | Values or statements | Treatment |
|---|---|---|
| Amplifier output power | 8–15 kW (one source on the overview page); tens of kW [8]; up to 35 kW [9]; 15–35 kW [10]; about 30 kW for a 200 lb patient at 128 MHz [11] | All stated; not reconciled; likely to depend on field strength, patient size, installed capability against sequence need and the point of measurement (Reasoning) |
| Power at the coil against amplifier output | 56.5 ± 2.5 % on one platform and 90 ± 2 % on another [4] | Measured on specific models; the amplifier rating alone does not give the power delivered |
| Wavelength in tissue at 3 T | About 25 cm [7]; about 26 cm [13] | Both stated; the difference is small and the basis of each value is not given |
| Location of the transmit electronics | Amplifiers in an equipment room [8, 9]; two independent transmitters in a box at the magnet [13] | Probably different stages (low-level transmitters against power amplifiers), since [13] does not mention the amplifiers (Reasoning) |
| Shim of the two-channel body coil | Preset by anatomy, no mapping needed [13]; calibrated per patient [14]; calibration under one minute [7] | Manufacturer descriptions; not independently compared |
| Global SAR limit | 4 W/kg quoted as the maximum for standard systems in most countries [2] | The source does not mention operating modes; the limits and modes are on 9210 and 9800 and were not verified here |
| Scanner-reported against measured SAR | Overestimates of 1.7- to 2.2-fold in one study [4]; underestimates up to about 20 % for some pad settings in a simulation [7] | Different designs, different platforms; both show that the displayed value is a model output |
12. Evidence Gaps
| Gap | Status |
|---|---|
| Requirements of IEC 60601-2-33 for RF power, B1+rms limits, monitoring and redundancy | (to be sourced): standard not read |
| Geometry of commercial body coils (rung count, diameters, wide-bore variants, shield design) | (to be sourced): only a simulation model was found |
| Amplifier ratings and duty-cycle limits of current clinical platforms | (to be sourced) |
| Interlocks, fail-safe behaviour and the action taken on reflected-power alarms | (to be sourced): only the flagging of extreme reverse power is stated [10] |
| Quality-assurance tests of the transmit system (gain stability, B1 uniformity, power calibration) | (to be sourced) |
| Clinical evidence for two-channel shimming outside cardiac, pelvic and thigh imaging, and independent comparison of platform implementations | (to be sourced) |
| Transmit hardware of low-field and open systems | (to be sourced) |
| Definition of the console labels "reference amplitude" and "adjust with body coil" and recalibration criteria by platform | (to be sourced): vendor documentation needed |
| Replication of the independent SAR measurements on current scanners | Not found; the 2012 study used two scanner models of unstated age [4] |
13. Evidence-Based References
No guideline or study with an EBM level applies to this descriptive engineering page. Sources are ordered by publication quality; the reading level of each is stated.