MRI Receive Coils: Elements, Circuits, Arrays and Coil Classes
MRIninja Knowledge Base | Foundations — Scanner Hardware Parent page: MRI Physics — Fundamentals and Principles (9201) Builds on: MRI Scanner Structure and Components — Overview (9703); MRI RF Transmit System and Body Coil (9708) Version 1.0 — November 2026
Foundations — Scanner Hardware
MRI Receive Coils: Elements, Circuits, Arrays and Coil Classes
How the radiofrequency coil that detects the signal is built and why it matters for the image: the single receive element (size, depth, noise, quality factor, quadrature), the circuitry around it (tuning, preamplifier, decoupling, detuning, cable traps), the phased array and what extra channels buy, the coil classes used for the head, spine, body, breast, prostate and knee as far as the sources describe them, flexible and special coils, and how coils are checked.
1. Scope, Boundary and Information Status
The transmit side of the radiofrequency (RF) system, including the body coil and the decoupling between transmit and receive coils, is on 9708. The physics of the RF pulse is on 9204, and a short overview of receive coils and of the phased array as the basis of parallel imaging is in Sections 6.2 and 6.3 of 9208; this page does not repeat them and goes into the structure of the receive coil itself. The reconstruction methods that use the coil sensitivities are on 9588 and 9564; the heating hazard of RF coils and cables is on 9804 and the physics of SAR on 9210.
This page is the home of the structure and the classes of receive coils and of how they are checked. It is the home neither of coil-specific positioning for any examination (that belongs to the protocol of the region) nor of the choice between two products of different makers.
Information status. No guideline or consensus statement describes how a receive coil is built. What exists is of three kinds: (1) society and accreditation documents that state requirements for the coil in one examination (breast, prostate) and how coils must be tested; (2) review articles and educational syllabi on coil design, read through a summarising tool; (3) experimental comparisons of coils, mostly on volunteers or small patient groups. Each reference states its reading level. No IEC standard, no textbook and no manufacturer coil specification was read (Section 15). Statements marked (Reasoning) are the editor's inference from the sources, not statements of a source.
2. What a Receive Coil Is and How Coils Are Classified
A receive coil is the antenna that detects the oscillating voltage induced by the precessing transverse magnetisation (physics on 9208, Section 6.2). The review written for non-physicists classifies RF coils by their mode of operation as transmit-only, receive-only or combined transmit/receive, and by geometry as volume coils or surface coils [10]. Arrays of surface elements are treated as a further class [10]. The table gives the definitions as that review states them.
| Term | What the source states | Source |
|---|---|---|
| Volume coil | Encloses the anatomy (examples named: body coil, birdcage, transverse electromagnetic (TEM) coil); mostly used for transmit but can transmit and receive; designed for a very homogeneous B1 field | [10] |
| Surface coil | Partial wire loop sized to the area of interest and tuned to the Larmor frequency; its inhomogeneous field restricts it "primarily to the Rx-only mode"; high SNR near the surface | [10] |
| Array | Several surface elements (or dipoles) combined for high SNR over a larger region; the elements must be decoupled to limit crosstalk | [10] |
| Receive-only vs transmit/receive | Separate transmit and receive coils allow each function to be optimised on its own; combined coils suit X-nuclei spectroscopy and ultrahigh field, where a body transmitter may be missing; in a combined coil a transmit/receive switch routes the RF line | [10] |
| Element, channel | An element is one coil loop or dipole; a channel is one receiver chain. The sources use "channels" for the number of independent signals sent to the receiver; the review for non-physicists says Roemer proposed arrays of small elements "grouped or fed to separate receive channels" and the sources read do not define the two terms further | [10] |
In current clinical practice most examinations use a receive array placed on or around the patient together with the body coil for transmit. A 2013 review states that "virtually every clinical coil sold" was an array and that arrays had grown from four or eight elements to 16–32 on most new scanners, with 64–128 channels under exploration [11]; a 2014 industry overview gives 32 channels as the commercial maximum for head and torso at that time, against research prototypes of 96 elements for the head and 128 for the torso [12]. These are counts of the dates of the sources and are not a statement about the coils sold today (Section 15).
3. The Single Receive Element: Size, Depth and Noise
3.1 Size, Depth and Field of View
A smaller loop close to the target detects a larger signal relative to the electronic noise, but its sensitivity falls with distance. The review for non-physicists defines penetration depth as the depth where sensitivity falls to 37 % of its value at the coil centre and gives the rule of thumb that for a circular, sample-noise-dominated loop the penetration depth is approximately equal to its diameter [10]. An industry overview states the same design rule in other words: the element size should ideally be about equal to the coil-to-target distance, because oversized elements add loss and undersized elements lose penetration depth [12]. A survey of array design adds that surface elements give high SNR whose sensitivity falls off with depth, so that designers balance peripheral SNR against uniformity [14]. The same survey states what the array adds: simultaneous detection from several closely positioned elements offers high SNR from the surface elements together with a large field of view from the data of the independent coils [14] (Section 5).
3.2 Where the Noise Comes From: Coil Loss, Sample Loss and the Quality Factor
The noise of a receive element has a part from the coil itself (resistance of conductors, capacitors and solder joints) and a part from the patient (currents induced in the conducting body), which behaves as a resistance coupled to the coil. The signal-to-noise ratio (SNR) is determined by the coil resistance and the patient-equivalent resistance; coil losses dominate at low field and patient losses at high field [12]. A survey of array design writes the noise as an equivalent resistance that combines coil, sample and radiation loss [14].
| Quantity | Definition or statement in the source | Source |
|---|---|---|
| Quality factor ratio | Q-ratio = Qunloaded / Qloaded = (Rcoil + Rsample) / Rcoil. A ratio below 2 means coil noise dominates, above 2 sample noise dominates. Typical loaded Q is 10–100 and unloaded Q 50–600 | [10] |
| Body-noise dominance | Rbody / Rcomp = QUL / QL − 1. At a ratio QUL/QL of 6, removing all component losses would improve SNR by less than 10 %; at 10 the factor is given as 1.005. For a loop near a conductive sample, the review states that induced eddy-current losses in the body typically dominate the component losses | [11] |
| Filling factor | Should be as close to 1 as possible, that is, the coil should fit the anatomy closely; the review states that its formula does not apply above 3 T for body imaging or at 4.7 T and higher for head imaging | [10] |
| Signal and noise share a mechanism | Body noise and the MR signal share the same detection mechanism, so increasing body noise also increases signal; keeping small elements body-noise dominated is named as the main barrier beyond 128 channels | [11] |
| SNR and scan time | A factor of √2 in SNR from the coil can be used to halve the scan time (through the averaging relation; see 9589) | [10] |
(Reasoning) The two Q-ratio statements agree: a ratio of 6 means the body resistance is five times the component resistance, so coil losses are a small part of the total, which is why careful engineering of a loop that already sits close to the patient gains little, whereas a loop that is too small or too far from the body is limited by its own losses.
3.3 Linear and Quadrature Reception
A single receive loop is linearly polarised and picks up one component of the rotating magnetisation. A pair of aligned coils whose outputs are summed through a quadrature hybrid with a 90° phase difference is circularly polarised and gains a factor √2 in SNR over a single coil [10]. In the transmit direction, circular polarisation needs half the power of a linearly polarised coil for the same B1+ [10]; this is the quadrature drive described for the body coil on 9708. An educational example in an ISMRM syllabus shows two coil configurations forming a quadrature pair that can feed one preamplifier through a quadrature combiner [13].
4. The Circuitry Around the Element
A bare loop is not a receive coil. The sources describe the following parts, in the order of the signal path from the element to the receiver.
| Part | What it does | Statements and figures in the sources | Source |
|---|---|---|---|
| Tuning capacitor | Sets the resonance of the loop at the Larmor frequency | Resonance approximately fr ≈ 1/(2π√(L·C)), with L the loop inductance and C the tuning capacitance | [10] |
| Matching network | Transforms the loop impedance to what the preamplifier needs | Conventional practice matches to 50 Ω; the preamplifier-decoupling method of Roemer instead matches the loop to the preamplifier's noise-optimal load and removes the need for a 50 Ω power match; a conventional power match halves the Q; matching losses must be minimised because they degrade the noise figure | [10]; [11]; [12] |
| Low-noise preamplifier | First amplification of the weak signal, close to the coil | Gain of about 27 dB for low-noise preamplifiers; gallium arsenide field-effect transistors (GaAs FET) give a high input impedance and a low noise figure; the static field strength and orientation affect noise figure and gain; for preamplifier decoupling the input impedance is about 1–2 Ω (figure caption in a review); the ratio of input impedance to optimal noise impedance must be very high or very low | [10]; [12]; [11] |
| Preamplifier decoupling | Reduces the current that flows in a loop because of its neighbours | The preamplifier together with the matching circuit presents a high impedance on the loop, which blocks coupled current; the review for non-physicists states that this "does not directly reduce the mutual inductance". For a loop of Q = 100 at 125 MHz a mistuning of 1.25 MHz costs 3 dB without decoupling, whereas with preamplifier decoupling a mistuning of a few MHz causes almost no SNR loss | [10]; [12]; [11] |
| Detuning trap | Makes the coil harmless while the body coil transmits | An active trap (PIN diode, inductor, capacitor) inserts a high impedance during transmit; one or more traps per element depending on size and frequency; a passive detuning circuit and an RF fuse add redundant protection ("three redundant safety features" in one figure). Serial detuning is broadband but needs a high bias voltage; parallel detuning is narrowband and uses a resonant circuit that becomes a high impedance when the diode is forward biased. Passive safeguards such as fuses protect against failed detuning components and are required for compliance. A test described in a review: isolation (S21) of more than 35 dB between the detuned and the active state | [10]; [12]; [11] |
| Cable trap and balun | Stops RF current flowing on the cable shield | Cable traps block common-mode current coming from the body coil; they are resonant circuits that create a high impedance and block standing waves; trap efficiency depends on Rp = Q·Z0 (a higher Z0 gives more bandwidth, a higher Q more attenuation). Isolating transmission lines from the coils is called "very important in practice"; baluns are named but their design is not given in the sources read | [12]; [13] |
| Digitisation | Converts the signal to digital form | The overview of 2014 describes a move from analogue receive chains (with the spectrometer in the technical or examination room) to fully digital coils with a converter at each element and optical-fibre transport, which would make the channel count scalable and separate coil design from cable management; the state of current commercial coils was not established from the sources read | [12] |
The decoupling between the receive elements and the transmit body coil is the detuning trap described above; the other half of the same problem, the effect of the receive elements on the transmit field, is on 9708, Section 5. The review for non-physicists states that the receive preamplifiers need protection from the transmit power and names the trap, the passive detuning and the fuse as the means [10].
5. The Phased Array
5.1 The Concept
The abstract of the 1990 paper by Roemer and colleagues (read only as an abstract on aggregator pages, Section 15) states that the method acquires and combines data from many closely positioned receiving coils and that the array gives the SNR and resolution of a small surface coil over the large field of view usual in body imaging, without a longer imaging time; the abstract names the analogy with phased-array radar and ultrasound, and states that the method applies to imaging and spectroscopy with any pulse sequence [23]. In the demonstration reported in the abstract, high-resolution spin-echo images of the thoracic and lumbar spine were acquired with four-element linear arrays made of 12 cm and 8 cm square coils; compared with a single 15 × 30 cm rectangular coil with identical parameters, the arrays gave roughly twice and three times the SNR at spinal depth (about 7 cm) [23]. The 2013 review of massively parallel arrays credits the same paper with showing that simultaneous detection with an array of surface coils can outperform a single larger surface coil covering the same area, with giving a method for optimal combination, and with introducing preamplifier decoupling [11]. The syllabus of 2013 adds that the distinct sensitivity patterns of the elements were later exploited for spatial encoding, which led to parallel imaging [13]; that use is on 9588.
5.2 Decoupling Between Elements
Neighbouring elements couple through their mutual inductance, which couples signal and noise between elements and lowers SNR and parallel-imaging capability [14]. The sources describe four ways to limit it.
| Method | Statement in the sources | Source |
|---|---|---|
| Overlap | The mutual inductance of adjacent coils ideally reaches zero at one overlap position that depends on loop shape and size; for circular loops the critical overlap is a centre-to-centre distance of 0.75 times the diameter, giving zero shared flux. With stronger coupling the resonance splits into two modes. | [10]; [11]; [13] |
| Shared capacitor or inductor | Used in a two-element example array where the elements share a common leg and capacitor; also named by the overview as capacitive or inductive decoupling "via passive components or overlap" for neighbouring elements | [13]; [12] |
| Preamplifier decoupling | Used in addition, mainly for elements that are not neighbours (Section 4) | [12]; [13] |
| Other circuits and structures | Inductive-coupling networks, inductive and "magnetic wall" decoupling and self-decoupled geometries are listed by title in a 2024 syllabus; the page names them without detail | [14] |
A bench study at 3 T that compared capacitive and overlap decoupling of adjacent elements as the sample load varied is cited here only as a pointer [27]: its abstract reports that the two methods gave similar relative SNR and similar parallel-imaging performance (search extract; Section 15). Industrial practice described in the overview is to design multi-element coils as single-element matches, each tuned to the preamplifier's optimal impedance, and notes that matching the preamplifier to the effective array impedance including coupling can improve SNR [12].
5.3 Combining the Element Signals
| Method | Statement in the sources | Source |
|---|---|---|
| Root-sum-of-squares (rSoS) | Needs no pre-scan measurement; coil sensitivity is approximated by the signal itself; "generally adequate" when the SNR of each element is high | [11]; [10] |
| Optimal combination | Weights w = Ψ−1C, which need the coil sensitivities C and the noise covariance Ψ; weights have amplitude and phase; the separate scan needed for high-SNR weights is mentioned | [11]; [10] |
| Covariance-weighted rSoS | Adds a noise-covariance measurement, which takes about one second; the resulting SNR equals the image intensity | [11] |
| Bias of magnitude rSoS | The SNR measured on magnitude rSoS images is biased upward, more with more channels and less with higher SNR; for a 32-channel image at SNR = 13 the correction lowers the measured SNR by a factor of 0.6. Noise-only regions in the background are unreliable for rSoS because the weights there are noise-weighted | [11] |
The practical consequence stated by AAPM Report 100 is that the single-image SNR equation applies only to magnitude images from quadrature or linear volume coils and that phased-array coils need a correction based on the element count [6]; SNR measurements on arrays are therefore comparable only when the same method is used (Section 11).
5.4 What an Array Does and Does Not Add
The review for non-physicists states that arrays give no SNR gain at the centre of the field of view over a size-matched volume coil, that the gains are greatest at the periphery, and that conformal helmet arrays do improve central SNR through a better filling factor [10]. The 2013 review gives modelling results (Wiesinger and colleagues) that at 1.5 T as few as 8–16 elements reach near-ultimate central SNR, whereas at 7.5 T 32 channels reach the ultimate SNR only in a small central region, with a factor of seven remaining to the theoretical limit at 80 % of the coil radius; a modelled 32-channel brain-like sphere reached 97 % of ultimate SNR at the centre and 44 % at the periphery [11]. Ideal averaging would give a central gain that exceeds a factor of 10 for 128 channels, which the review says is reduced in real arrays by noise correlation between neighbouring coils (the expression itself was not captured; √128 ≈ 11.3 is consistent with the statement) [11]. For the number of channels per region see Section 7.
6. What More Elements and Channels Buy: Comparisons in the Sources
The comparisons below are the experimental results read. They differ in field strength, in the former on which the array was built and in the outcome measured, and several compare coils that differ in more than the channel count; the table states what was compared.
| Comparison | Design | Result as reported | Source and level |
|---|---|---|---|
| 32-channel helmet array vs a larger commercial 8-channel head coil, 3 T | Receive-only head array with 32 elements in a "soccer-ball" layout built on a close-fitting helmet | SNR gain up to 3.5-fold in the cortex and 1.4-fold in the corpus callosum; the peak g-factor was 59 % and 26 % of the 8-channel values at four- and five-fold acceleration | [15]; abstract on an institutional page |
| 96-channel vs 32-channel on the same helmet former vs a commercial 12-channel head coil, 3 T | Phantom and human imaging; several element and matching designs tested | Substantial SNR gain in the distal cortex over both; at the centre the SNR matched the 32-channel coil with optimal combination and was about 20 % lower with rSoS combination; the maximum g-factor at acceleration 4 was 65 % of the 32-channel value | [16]; abstract via a summarising tool |
| 64- vs 32-channel brain arrays on the same former | Comparison described in the review of massively parallel arrays (figures not available to the reader) | 1.2-fold central SNR gain at an acceleration of 4 and no gain without acceleration; the advantage of the 64-channel coil grew with the acceleration factor (3 to 5 measured) | [11]; review text |
| 12-, 32- and 96-channel head arrays | Measured SNR maps in the same review; the 12-channel array used a larger former, which confounds the comparison | "Nearly a 10x SNR increase" in the distal cortex; small central gains | [11]; review text |
| 32-channel vs 12-channel head coil, 3 T | Five healthy volunteers, proton-density images, acceleration factors 2–6 | Up to threefold higher SNR with the 32-channel coil, largest near the brain surface and about 40 % at the centre; the authors judge the 32-channel coil superior with and without acceleration | [18]; abstract via a summarising tool |
| Paediatric 32-channel helmet array vs adult 32- and 12-channel coils, 3 T | Array sized to the head of a 4-year-old | SNR gains of 4 and 2 times in the cortex and 1.2 and 2 times in the centre of the head, in the order of the two comparison coils named in the abstract | [19]; search extract |
| 32-channel vs 12-channel head coil, 1.5 T, patients | 21 patients scanned with both coils in randomised order; T1, T2, FLAIR and diffusion-weighted images rated side by side by three neuroradiologists; 41 lesions | FLAIR and diffusion-weighted images showed significant improvement in nearly all parameters; T2 improved for overall quality and white-matter smoothness; T1 improved in only a few parameters with no preference for overall quality; no difference in artefacts; the 32-channel coil was preferred for lesion evaluation (mean 0.55 on a −2 to +2 scale), but only 4 of 41 lesions were rated as improved in diagnostic value; central signal loss on 7 T1 stacks and 1 FLAIR stack was attributed by the authors to imperfect intensity normalisation | [17]; full text via a summarising tool |
| Quadrature transmit/receive knee coil vs quadrature transmit with an 8-channel receive array, 3 T | 10 knees scanned twice per coil | The array roughly doubled cartilage SNR (medial tibia 2.8 to 5.6, lateral tibia 2.5 to 6.0); flip-angle uniformity did not change (<1.5 % variation); cartilage T2 values were significantly longer with the array (central medial femur 50.7 vs 45.9 ms; medial tibia 48.2 vs 41.6 ms), and the authors describe an effect of the order of a change in field strength | [20]; full text via a summarising tool |
Four points follow from the table without going beyond it. First, the gain from more elements is concentrated at the periphery of the object and at high acceleration; at the centre it is small [10], [11], [16], [18]. Second, an SNR gain is not the same as a diagnostic gain: in the only patient study read, most lesions were not rated as better visualised [17]. Third, a change of coil can change a quantitative measure, as the knee study shows for cartilage T2 [20]; a series measured with one coil is therefore not directly comparable with a series measured with another. Fourth, several comparisons are confounded by the size and shape of the former [11], [18], [19]. The edition and the model of the coils in these studies are those of 2006–2013 and not those on the market today.
7. Coil Classes by Region as the Sources Describe Them
The review for non-physicists tabulates the coil types and the channel counts for coils of each region described in the literature; the extract read does not separate research from commercial coils, and the ranges are therefore the span of published coils and not what a clinical scanner offers [10]. The table adds what the society and accreditation documents and the studies read state for the regions where they say anything. Regions for which no source was read are named at the end and not described.
| Region | Coil types and channels in the review's table [10] | What the other sources state | Source |
|---|---|---|---|
| Head, neck, brain | Birdcage, surface arrays, dipoles, local shim arrays; 4–128 channels | Examples of 12-, 32-, 64- and 96-channel head arrays (Section 6); a clinician may prefer a larger, looser helmet to the tightest fit, accepting an SNR loss, because of comfort and workflow | [10]; [11]; [15]–[18] |
| Spine | Surface arrays; 15–75 channels | No guideline or study on spine coils was read | [10] |
| Chest and abdomen (torso) | Surface arrays; 6–32 channels (one row of the review for breast, chest and abdomen) | For the torso, a research 128-channel body array and 32- and 128-channel cardiac and chest arrays are cited by title or in the text of the review of massively parallel arrays and in the overview of 2014. Breast: see the next row | [10]; [11]; [12] |
| Breast | Same row of the review as the chest and abdomen | ACR practice parameter (2023): examinations "should be performed with a dedicated bilateral breast MRI coil", the field of view must include the entire bilateral breasts from the axillae to the inframammary folds, skin folds should be minimised and the nipples positioned to point down if feasible. The ACR accreditation requirements state a dedicated bilateral breast coil and simultaneous bilateral imaging, without a minimum field strength. EUSOBI (2008): a dedicated bilateral breast coil is mandatory, the patient is prone with the breasts hanging in the coil loops, supported but not compressed, placed as deeply as possible in the coil, arms at the sides give better coverage; no channel count is given. UK technical guidelines for surveillance of high-risk women: dedicated bilateral breast coils with at least 8 elements (channels) are required to improve SNR and allow parallel imaging | [1]; [4]; [3]; [2] |
| Prostate, pelvis, rectum | For rectum, prostate and other inner surfaces: surface, meanderline, Litz, spiral, catheter and micro coils; 1–6 channels | The PI-QUAL v2 consensus text states that the use of endorectal coils is discouraged and phased-array surface coils are preferred for signal reception (wording of the PI-RADS technical recommendations as quoted); endorectal coils can increase SNR, mainly at 1.5 T, and may help patients with a large body mass index; a meta-analysis found no benefit for extraprostatic extension and marginal sensitivity gains for seminal vesicle invasion; discomfort, time and cost are drawbacks; surface coils can give poor SNR in patients with a high body mass index. In 33 men at 3 T, a rigid two-channel endorectal coil with an external array did not differ significantly in lesion-detection sensitivity from the external array alone (79.2 % vs 75 % and 79.1 % vs 66.7 % for the two observers) | [10]; [8]; [22] |
| Extremities, knee | Resonator, volume, flexible and adaptive arrays; 2–32 channels | Knee: an 8-channel receive array roughly doubled cartilage SNR over a quadrature transmit/receive coil (Section 6); a conference abstract describes a 15-channel receive knee coil with a local 12-rung birdcage transmitter at 3 T, built with three rows of elements along the long axis to allow parallel imaging in that direction, and states that a local transmit coil needs less power than the body coil | [10]; [20]; [21] |
| Whole body | Body coil, surface arrays, dipoles, flexible and adaptive arrays; 12–128 channels | The review states that scanner hardware could not accept more than 128 channels at the time of writing | [10] |
No source read describes dedicated coils for the shoulder, wrist, hand, foot and ankle, temporomandibular joint, orbit, cardiac imaging in a clinical protocol, or neonatal and paediatric imaging beyond the head coil in Section 6; these are not described here and are listed in Section 15. The clinical protocol pages state where a particular coil is used (for example the protocols for the cervical spine, prostate, knee and heart).
8. Flexible, Adaptive and Special Coils
| Coil | Statement in the sources | Source |
|---|---|---|
| Flexible and screen-printed coils | Described as low-cost and well suited to close fits, especially in paediatric patients | [10] |
| Stretchable and adaptive arrays | Raise SNR across a range of patient sizes; adaptive coils change their electromagnetic properties as they reshape and therefore need automatic tuning and matching | [10] |
| Conformal helmet arrays | Improve central SNR through a better filling factor | [10] |
| Wireless and digital coils | Wireless coils and a wireless digital capacitor module for tuning receive arrays appear in the reference list as conference abstracts; digital coils: Section 4 | [10]; [12] |
| Cryogenic and superconducting elements | Proposed in a 2024 syllabus as a way to reduce coil loss where sample loss does not dominate; no clinical use was established from the sources read | [14] |
For flexible coils the practical constraint is stated in the review: because the shape changes, the sensitivity pattern and the tuning change with it [10]. (Reasoning) This is why a flexible coil that is not applied as intended can give a different result from the same coil applied flat; the sources read do not quantify it.
9. Consequences for the Image
9.1 Intensity Shading and Its Correction
Because the sensitivity of an array falls with distance from its elements, the raw image is brighter near the coil than far from it. Modern scanners estimate the sensitivities from a pre-scan and correct the image. A preprint on surface-coil intensity correction states that the sensitivity maps estimated by the usual methods differ from the true maps, so that reconstructed images show undesired spatial variation in intensity, and that this can be particularly problematic in thoracic and abdominal imaging of larger patients, where the central area of the body can appear excessively dark compared with tissue closer to the receive coils; the preprint proposes a correction from pre-scan data collected with the body coil [24]. In the 1.5 T patient study of Section 6 the authors attribute a central signal loss on eight image stacks to imperfect intensity normalisation in post-processing [17]. The preprint is not peer reviewed in the extract read.
9.2 Parallel Imaging and the Coil
The review for non-physicists gives the relation SNRaccelerated = SNRunaccelerated / (g·√R), with R the reduction factor and g the geometry factor, and states that more channels reduce the g-factor and so allow higher acceleration [10]. The 2013 review adds that increasing the channel count helps accelerated imaging most at the centre, where the g-factor is highest, and that acceleration above tenfold is possible with two-dimensional acceleration on highly parallel arrays [11]. The measured g-factor reductions are in Section 6. The reconstruction methods and the reference scans that use the sensitivities are on 9588 and 9564.
9.3 Quantitative Values and Comparability
The knee study in Section 6 found that cartilage T2 values depended on the coil used and on the cartilage plate, with higher SNR giving longer T2 values, and the authors call their figures a lower-bound estimate because of the small sample (10 knees, two sites) [20]. (Reasoning) For follow-up examinations and for comparison between sites, the coil is therefore one more item to keep constant, together with the sequence parameters on 9501.
9.4 Fit, Comfort and Motion
The overview of 2014 lists coil weight, dimensions, flexibility, cable position, the number and type of connectors, easy positioning without repositioning during the examination, and openness (so that patients can see out and feel less claustrophobic, which may cost some SNR) among the requirements; it adds that a tight fit can impair cooling and cause perspiration and that patient discomfort increases motion, which reduces image quality [12]. The review for non-physicists states that close-fitting arrays maximise SNR but must be balanced against comfort and workflow [10].
10. Handling and Safety of Receive Coils in Outline
The mechanisms of RF burns, the role of loops and contact and the rules for padding are on 9804; the table below gives only what the sources read state about the receive coil and its cable.
| Topic | Statement in the source | Source |
|---|---|---|
| Unused coils | The ACR Manual (2020) states that all unnecessary or unused electrically conductive materials external to the patient should be removed before imaging and that it is insufficient to disconnect and leave unused conductive devices "such as surface coils" in the scanner during imaging | [9] |
| Connectors and insulation | All electrical connections, for example those used for surface coils, are to be checked for insulation integrity, and the technologist is to check them visually before each use | [9] |
| Leads and loops | Large conductive loops, including those formed by patient tissue, should not form inside the scanner; leads should be placed as far as possible from the inner wall when the body coil transmits | [9] |
| Padding | The ACR texts address insulating pads between the skin and the transmit coil: a single-layer bed sheet is insufficient insulation or spacing, and the distances are manufacturer-specific. They do not give padding rules for receive coils; skin-to-skin contact within the transmit volume should be prevented | [9]; [7] |
| Burns despite normal SAR | The manual notes that burns have occurred even when the system was operating within the guidelines for RF power deposition, so SAR limits alone do not prevent them | [9] |
| Manufacturer's coil warnings (one example) | The operator's manual of an 8-channel wrist array (March 2022, as shown on a manual-hosting page) states: remove the coil from the system when it is not plugged in (burn hazard); do not cross or loop cables (arcing and burns); route cables out of the magnet so that they do not touch the patient; check cable insulation, strain reliefs and connector boxes before each use and stop using the coil if the insulation is broken or the cable frayed; use only approved accessories; do not lift the coil by its latch; prevent the patient's body from forming a loop and use pads to keep the hands away from metal parts of the table or cradle | [26] |
| Direct contact burn (case) | A figure caption in a RadioGraphics review shows a third-degree burn on a patient's calf that was pressed against the cable of a phased-array coil (caption only; no more detail read) | [25] |
| Built-in protection | Preamplifier protection by active trap, passive detuning and fuse (Section 4) | [10] |
The ACR text and the manufacturer's manual agree on removing unused coils, checking connectors and cables before use and avoiding loops; the manual is one example and the wording of other makers' manuals was not read.
11. Checking Coils: Acceptance, Annual and Routine Tests
| Check | Who and when | Content as stated | Source |
|---|---|---|---|
| Acceptance testing | Before patient imaging; the medical physicist reports the findings | Evaluation of all coils, "including all breast imaging and biopsy coils" (ACR); AAPM Report 100 asks for an inventory of components with model, serial and version numbers and notes that missing items such as surface coils are a common acceptance problem | [5]; [6] |
| Annual performance evaluation | Medical physicist or MR scientist; up to 14 months between surveys | Coil tests "on all coils used clinically (including breast coils)": frequency and gain or attenuator values, image SNR, intensity uniformity for volume coils, phase stability and image-artefact assessment; the test checks the trade-off between uniformity and SNR inherent to the coil types. Breast accreditation requires a statement that the bilateral breast coil was checked and acceptable; the survey need not be repeated if the breast coils were checked in a general survey within the past year | [5] |
| Weekly technologist tests | Technologist | Transmitter gain and centre frequency are the weekly tests named; the page does not mention ghosting or element failure for coils | [5] |
| SNR of coils (AAPM Report 100) | Acceptance and ongoing QC | Head and body coils, and in three orthogonal planes for at least one coil; "reference SNR tests for all coils are preferable"; for surface and phased-array coils an image showing the region-of-interest locations is required and maximum-SNR measures are recommended; reconstructing images from each array element is "preferable" because it allows SNR characterisation of each element but typically needs a service or research mode; acceptance means meeting or exceeding the manufacturer's values, if provided | [6] |
| Uniformity (AAPM Report 100) | Acceptance | Percent image uniformity of a volume head coil should meet or exceed 90 % for scanners at 2 T or below; above 2 T water phantoms are expected to give lower values because of dielectric and penetration effects | [6] |
| Breast coil element check | Weekly or before use (UK technical guidelines, mandatory) | Verify that each element is functional, preferably with the manufacturer's QC, otherwise by comparing the image of each element with a baseline set; a baseline spin-echo example is TR 500 ms, TE under 20 ms, 128 × 128 matrix, 10 mm slices, field of view 300–340 mm, uncombined element images, each showing high signal next to the element; a coil that fails the manufacturer's action level is taken out of service until the fault is rectified; most failures of the SNR test come from positioning, region-of-interest errors or phantom-filler movement, so the test is repeated before a fault is assumed; breast coils may not be tested by default in the manufacturer's service schedule, and a repaired or replaced coil must pass the manufacturer's image-quality checks before clinical use | [2] |
| Breast coil, physicist testing | Qualified medical physicist or scientist | The ACR practice parameter states that equipment monitoring follows the ACR–AAPM technical standard and includes testing of the breast coil(s) | [1] |
The tests for the system as a whole (field homogeneity, RF shielding, ghosting) are outside this page. No test in the sources read looks for a failed element in a clinical array other than the breast coil check and the AAPM suggestion of element-by-element images in a service mode (Section 15).
12. The Technologist's View (Reasoning)
This section is the editor's inference from Sections 3 to 11 and is not a procedure.
- The choice of coil is a choice of distance and filling factor: the smallest array that covers the region of interest and fits the patient comfortably gives the best SNR [10], [12]; a coil that is too large or too far loses signal, and a coil that is too tight is less comfortable and can increase motion [12].
- Centring and covering the region matters more with an array than the number of channels printed on it: sensitivity falls with depth and the gain from more elements is at the periphery [10], [11].
- A band of low signal, a localised region of noise or an unexpected shading that follows a coil element may be a coil or connection fault rather than a patient effect; the breast-coil guideline expects each element image to show high signal next to that element and asks for the test to be repeated before a fault is assumed [2].
- Seating the connector, looking at the cable and keeping unused coils out of the bore are checks required by the sources before every examination [9], [26].
- After a change of coil (replacement, repair, a different model for the same region), a quantitative or follow-up comparison needs a note that the coil changed (Section 9.3).
13. Discrepancies Between Sources
| Topic | What differs | How the page handles it |
|---|---|---|
| Date of the phased-array paper | The 2014 overview dates the simultaneous array acquisition of Roemer to 1991 in its text, while its own reference list and the other sources give 1990 | The date 1990 is used [12], [23] |
| Gain at the centre of an array | The review for non-physicists states that arrays give no SNR gain at the centre over a size-matched volume coil; the 2013 review describes ideal averaging as giving a central gain above tenfold for 128 channels and reports a 1.2-fold central gain of a 64-channel over a 32-channel array at acceleration 4 and a central SNR equal to that of a 32-channel coil for a 96-channel array | The baselines differ (a volume coil, a single element, a smaller array); the page reports each with its baseline (Sections 5.4 and 6) [10], [11] |
| Overlapped or gapped elements at high acceleration | In one simulation the text puts the advantage of the gapped design at acceleration "R>5" while the figure caption says "R>4" | The threshold is not used [11] |
| Role of preamplifier decoupling | The review for non-physicists states that it does not directly reduce mutual inductance but reduces the current in the loop; the overview and the syllabus list it among the decoupling methods | Both are reported; the sources use "decoupling" for the effect on coupled current, not always for the mutual inductance (Section 4) [10], [12], [14] |
| Endorectal coil | The PI-RADS recommendation as quoted in PI-QUAL v2 discourages endorectal coils; the same text and other studies report an SNR advantage, mainly at 1.5 T, and a small 3 T study found no significant difference in lesion detection | The page reports the recommendation and the studies without ranking them (Section 7) [8], [22] |
| Channel counts over time | 2013: 16–32 channels on most new scanners; 2014: 32 commercial maximum; 2018: hardware limit 128 channels | Counts are given with the year of the source; none is stated as current (Section 2) |
| SNR gain and diagnostic value | Large SNR gains in volunteers (up to threefold; nearly tenfold in the distal cortex) against a patient study in which the 32-channel coil was preferred for image quality but only 4 of 41 lesions were rated as better in diagnostic value | Reported side by side; the page draws no conclusion on diagnostic benefit (Section 6) [11], [17], [18] |
14. Open Questions
- Which receive-coil channel counts and element layouts are in current clinical use for each region, and which coils each major manufacturer offers; the sources read describe published and research coils and counts of 2013–2018.
- How often elements fail in clinical arrays and which routine test detects a failed element outside breast imaging; AAPM Report 100 suggests per-element images in a service mode and the UK breast guideline gives a weekly test, but no study of failure frequency was read.
- How the console chooses which elements are active for a given position, and how a technologist should check it; the sources read do not describe it.
- Whether receive-coil choice changes diagnostic accuracy for the common indications; the only patient comparison read (12- against 32-channel head coil at 1.5 T) rated only 4 of 41 lesions as better in diagnostic value.
- Receive coils for low-field and open systems: not covered by the sources read (see 9705 for the magnets).
15. Evidence Gaps and the Attempts Made
| Gap | What was tried | Result |
|---|---|---|
| IEC 60601-2-33 clauses on coils (RF burns, coil labelling, receive-only coils) | Catalogue entries only (3.4.18, for 9710); the standard is sold | Not read; no statement of this page is taken from it |
| Textbook treatment of coil design (for example McRobbie, Haacke) | Not obtained | No textbook was read; the structure relies on one review for non-physicists, one review of massively parallel arrays, three ISMRM syllabi and experimental studies |
| Original paper of Roemer 1990 | Search for the paper; its abstract read on an aggregator page through a summarising tool (a paraphrase, not verbatim); the publisher page and the PubMed page were not reached | Abstract-level only; the theory of combination is cited through [10], [11], [13] |
| Sodickson 2007 and Wright 2011 encyclopedia chapters (named as references in a syllabus) | Not searched | Not read |
| Manufacturer coil specifications and service documents | Search returned one manufacturer product page for a 15-channel knee coil (no specification in the extract) and an operator's manual page of an 8-channel wrist coil, read through a summarising tool | One manual (the warnings of Section 10) only; no element count, geometry or test procedure of any current product |
| ACR Manual on MR Safety, coil statements | The 2020 copy was read in two ranges (0–100,000 and 100,000–164,784 of 164,784 characters) with a question about coils; the first range ended partway through the RF section and the second contained no coil statement; the 2024-cover copy used for 9709 and 9710 was not queried for coils | The 2020 text gives connectors and unused coils; it gives no rule on coil cables, damaged coils or manufacturer-only coils |
| Spine, shoulder, wrist, foot and ankle, TMJ, orbit, cardiac and paediatric coils | No guideline or study read | Only the channel ranges of one review table (Section 7) |
| Full text of the PI-RADS technical requirements and of the PI-QUAL v2 authors | The text of PI-QUAL v2 was read through a summarising tool; its Table 1 and author list were not captured | PI-RADS wording is taken from its quotation in PI-QUAL v2 |
| Reading depth | Review for non-physicists read to 100,000 of 122,782 characters, the rest being the reference list; the 2013 review of massively parallel arrays read to 100,000 of 150,044 characters, the rest being references; figures in both were not available to the reader | Figure-only values (for example the optimal radius of loops by depth) are not reported |
16. Where to Go Next
- Transmit side, body coil and the decoupling of transmit and receive: 9708.
- Where the receive coils sit in the signal chain: 9703; physics of receive coils and the phased array: 9208.
- What the coil sensitivities are used for: 9588 (parallel imaging) and 9564 (reference scans, simultaneous multi-slice); SNR and the number of averages: 9589.
- Burns, loops and padding: 9804; SAR: 9210.
- Other applications that need dedicated coils (nuclei other than hydrogen, fetal and interventional imaging): 9215.
17. Evidence-Based References
No guideline or systematic review addresses receive-coil structure. Sources are ordered by type and quality (guidelines, consensus and accreditation documents first; reviews and syllabi, then single studies and extracts); the reading level of each is stated. Search extracts are cited only for what the extract states.
A. Guidelines, Consensus Statements, Accreditation, Technical Reports and Safety Guidance
B. Reviews and Educational Syllabi
C. Experimental and Clinical Studies
Related Protocols
Recent PubMed search for this protocol