Other Applications of MRI: Survey and Status

MRIninja Knowledge Base | Orientation — Survey and Status Parent page: MRI Physics — Fundamentals and Principles (9201) Version 1.0 — November 2026

Orientation — Survey and Status

Other Applications of MRI: Survey and Status

What MRI is used for beyond the routine hydrogen examination of a patient on a diagnostic scanner: nuclei other than hydrogen, spectroscopy, MRI in radiotherapy, MR-guided focused ultrasound, MR-guided interventions, fetal imaging, functional MRI and tractography for neurosurgical planning, elastography, and preclinical research. Each application is described at survey level, with what the sources state about its status: in routine use, approved for a defined indication, or investigational.


1. Scope, Boundary and Information Status

This page follows the introduction What Is MRI in Clinical Diagnostics? (page 9214), which describes the ordinary examination, and names the applications that lie outside it. It does not teach any of them in depth: each deserves its own page, and none has one yet unless a pointer in Section 14 says so.

This page is the home of the survey of non-routine MRI applications and of the status wording used for them. It is not the home of the physics of nuclear magnetic resonance (9202), of the hardware (9208, 9703), of MR safety (9802 and the other safety pages), of pregnancy and MRI safety (9812), of diffusion physics (9207) or of the brain tumour protocol, which holds the interpretation of spectroscopy in glioma (1185).

Information status. Forty-one sources were read at different depths; each reference states its reading level. Sixteen are guidelines, consensus statements, technical reports or regulatory documents (references [1]–[16]); the others are reviews, single studies, search extracts and news items (references [17]–[41]). No single guideline covers the whole subject, and the evidence differs sharply between applications: for some there is a consensus statement or a profile that sets requirements, for others only narrative reviews, and for hyperpolarised and non-hydrogen imaging mostly early-phase studies. The page therefore states, for each application, only what the sources state about its status and says where the sources are old.

Attempts that failed and were not worked around: the Thieme-Connect pages of the MR-HIFU review and of the interventional MRI white paper (a security challenge page was returned); three PMC articles (a 2015 review of MR-guided focused ultrasound, a 2019 review of HIFU applications and a German guideline on MRI simulation for stereotactic radiotherapy; a browser check was returned); the full text of the 2024 resting-state fMRI recommendations at two addresses of the journal and of a 2021 RadioGraphics article (client error). For these, only the abstract or the search extract is cited, and the reading level is stated.

2. How the Status Words Are Used

  • Established or in routine use: used only where a source says so in those terms, for example the consensus statement on proton spectroscopy [3] or the white paper on interventional MRI [30].
  • Approved for an indication: a regulator has approved a named device or agent for a named indication, with a date. An approval is not a guideline recommendation and does not say how widely the application is used (Reasoning).
  • Investigational: the source describes the application as research, a trial, feasibility work, an animal study or "potential".

Status is a property of a source at a date. A reader should check the date of the source before relying on a status, and the dates are given throughout.

3. Summary of the Applications and Their Stated Status

ApplicationWhat is imaged or doneStatus as stated in the sources readSources
Proton spectroscopy of the brainSignal from hydrogen read as a spectrum of metabolitesClinical usefulness established for brain neoplasms, neonatal and paediatric disorders, demyelinating disorders and infectious lesions; may contribute in neurodegenerative disease, epilepsy and stroke; some organisations still regard it as investigational (2014)[3]
Phosphorus-31 spectroscopyEnergy metabolism (phosphocreatine, ATP), pH, magnesiumLongest clinical history of the non-hydrogen nuclei in one review, in cardiac ischaemia; no regulatory status found[18]
Sodium-23 imagingTissue sodium concentrationNot a routine clinical tool; cartilage the most developed area (2021 review)[17]
Hyperpolarised carbon-13 (mainly pyruvate)Metabolic conversion in real timeInvestigational: more than 200 human subjects in 17 clinical studies by 2021; more than 20 ongoing trials[19, 20]
Hyperpolarised xenon-129Lung ventilationFDA-approved on 23 December 2022 for evaluation of lung ventilation in adults and in paediatric patients aged 12 years and older; not evaluated for lung perfusion[14, 15]
MRI in radiotherapy planningDelineation of target and organs at risk, registered to CTUse increasing; dedicated MR simulators in few departments (2 of 71 UK centres in 2018)[9, 10, 11]
MR-linac (online MR-guided radiotherapy)Daily imaging and adaptation of the planTwo systems in clinical use; implementation guidance exists; evidence on benefit by tumour site scarce (2021)[2]
MR-guided focused ultrasoundThermal ablation guided and monitored by MRIFDA-approved for named neurological indications (essential tremor from 2016, Parkinson disease tremor 2018, pallidotomy 2021, supplement of 2025) and, earlier, for uterine fibroids (2004); other indications investigational[13, 24, 25, 26, 27, 28, 40]
MR-guided biopsy and interventionsNeedle and device placement under MRIBreast biopsy of otherwise undetectable lesions established in routine; prostate biopsy being introduced (2017)[29, 30, 31]
Fetal MRIFetal brain, thorax, abdomen and placentaComplement to an expert ultrasound examination, not a screening test; standard method only in selected high-risk cases[1]
Functional MRI and tractography for neurosurgical planningLanguage, motor and visual cortex; white-matter tractsWeak (Level C) evidence for some uses in epilepsy surgery; consensus recommendations for language mapping; tractography described as feasible in a small protocol study[4, 5, 6, 34, 35]
MR elastography (liver)Liver stiffness from shear wavesQuantitative biomarker with a QIBA profile; stated to give the highest classification accuracy for fibrosis staging in a 2024 review[8, 33]
Preclinical MRIAnimal models on small-bore systemsResearch[37, 38]

4. Nuclei Other Than Hydrogen

4.1 Why they are harder

One review states that in theory any nucleus that can be detected by nuclear magnetic resonance can be imaged [18]. Hydrogen is used because of its abundance: a lecture abstract notes that hydrogen is about 62 % of the atoms of the body and that the other nuclei, such as sodium-23, phosphorus-31, oxygen-17, potassium-39 and chlorine-35, are present in very low concentration in comparison [23]. Two reviews and a lecture state the consequences: a relatively poor signal-to-noise ratio, and limits in acquisition time and resolution [17, 22, 23]. For sodium the review of Zaric and colleagues gives a sensitivity of 9.2 % of that of hydrogen, a concentration of about 15–30 mM in muscle and 250–350 mM in articular cartilage, relaxation times about two orders of magnitude shorter than those of protons, and a biexponential T2 in which most of the sodium signal is lost within a few milliseconds [17]. A 2013 review adds that sodium imaging can be done in around 10 to 15 minutes with a resolution of a few millimetres [21].

Because the gyromagnetic ratio of these nuclei is lower, their resonance frequency at a given field is lower: a review of hyperpolarised imaging gives 32 MHz for carbon-13 against 128 MHz for hydrogen at 3 T and notes that gradient demands are proportionally higher [20].

4.2 Nucleus by nucleus

NucleusWhat it is used to measure, as statedMaturity, as statedSources
Phosphorus-31Cell-membrane phospholipid metabolism in relation to ATP energy metabolism; phosphocreatine and ATP and their ratio; also tissue pH and magnesiumThe ratio has been useful for identifying ischaemia in animals and in humans with coronary stenosis; the review describes this as the longest clinical history of the nuclei it covers[18]
Sodium-23Tissue sodium concentration, which reflects intracellular and extracellular sodium and the volume fractions; intra- and extracellular signals cannot yet be reliably separated in vivoMost developed in cartilage; heart, kidney, breast and brain work is mostly feasibility or small cohorts; "not yet a routine clinical tool"; the authors say it is meant to give early metabolic information and not to replace hydrogen MRI[17]
Carbon-13, hyperpolarisedConversion of [1-13C]pyruvate to lactate, bicarbonate and alanine, which reflects enzyme activity and metabolic fluxFirst clinical trials in prostate cancer; early human studies in brain tumours and mild traumatic brain injury; in the 2021 review oncology had the highest translation state and brain, kidney and liver the lowest[18, 19]
Xenon-129, hyperpolarisedLung ventilation and gas exchange; diffusion reflecting lung structureApproved in the United States for ventilation imaging (Section 4.4); other uses described as research or potential[14, 18]
Fluorine-19Strong signal but no natural occurrence in biological tissue, so used to label drugs (one search extract); hyperpolarised 19F gases have been reported for lung studies (another review)Little detail given in the reviews read[18, 39]
Deuterium (2H)Quantification of 2H-labelled glucose metabolism in tumoursEarly-stage research; animal studies[18]
Oxygen-17, potassium-39, chlorine-35Named as nuclei that play a role in metabolism, ion homeostasis and action potentialsNamed only; no status given in the sources read[23]

4.3 Hyperpolarisation

Hyperpolarisation raises the signal by creating a nuclear polarisation far above thermal equilibrium. A review of detection strategies names spin-exchange optical pumping for noble gases, dissolution dynamic nuclear polarisation (dDNP), which is the method most applied in vivo and mainly with carbon-13, and parahydrogen-induced polarisation [20]. The magnetisation is not renewable and decays with T1, which the review gives as a few tens of seconds in vivo; each radiofrequency excitation uses up part of it, so sequences are designed around that limit, and for dDNP doses there are intervals of more than 30 minutes between dissolutions [20]. For carbon-13 a 2021 review describes polarisation at 0.8 K and 5 T taking about 30–180 minutes, followed by rapid dissolution to an injectable solution, a signal that is "typically completely relaxed within minutes (1–2 min)", and a workflow that the authors name as the limiting factor of the method, not the MR methods [19]. The same review counts 50 polarisers prepared for human injection installed worldwide and more than 10 sites performing clinical trials [19].

4.4 Hyperpolarised xenon-129: an approved use

The FDA approved xenon Xe 129 hyperpolarised (Xenoview) on 23 December 2022 as a hyperpolarised contrast agent for use with MRI for evaluation of lung ventilation in adults and paediatric patients aged 12 years and older [14]. The approval rested on two prospective, multicentre, open-label crossover trials in 83 patients with lung disorders, comparing the agent with xenon-133 scintigraphy [14]. The review document states the limitation of use: it has not been evaluated for lung perfusion imaging [15]. The dose is inhaled during a breath-hold of up to 15 seconds [15]. Pooled data of 123 adults from the published literature reported transient adverse reactions: numbness, tingling, euphoria and dizziness [14]. A later efficacy supplement document proposes use from age 6 years; the outcome of that submission was not read [16].

4.5 Hardware

The sources agree on what is needed. Dual-resonance coils exist for hydrogen with carbon-13, sodium-23 or phosphorus-31, triple-resonance coils have been used to calibrate carbon-13 transmit fields from the sodium signal, and cryogenic coils have given a threefold gain for a carbon-13 coil at 3 T [20]. Sodium imaging asks for low operating frequencies and, for a combined hydrogen and sodium coil, a double-tuned multichannel array whose design can degrade performance at one or both frequencies; multichannel sodium systems are said to be available on only a small number of scanners [17]. A review of multinuclear methods notes that a second heteronuclear transmitter and a range of radiofrequency coils allow imaging and spectroscopy of carbon-13, fluorine-19, sodium-23 or phosphorus-31 [39]. The generic coil and transmit hardware is on 9208 and 9708. What follows for a technologist is that an examination of another nucleus is not a variant of a routine protocol: it needs a coil and a transmit chain that most clinical systems do not have (Reasoning, drawn from [17, 20]).

5. Proton MR Spectroscopy of the Brain

Spectroscopy is not a different nucleus: it reads the signal of hydrogen as a spectrum in which each metabolite appears at its own frequency, instead of forming an image. The consensus statement of the MRS Consensus Group (Radiology, 2014) is the most substantial source on its clinical use [3].

5.1 Status

The consensus states that the clinical usefulness of proton MR spectroscopy is established for brain neoplasms (diagnosis, grading, prognosis, treatment planning, and differentiating recurrence from radiation injury), neonatal and paediatric disorders (hypoxia-ischaemia, inherited metabolic diseases, traumatic brain injury), demyelinating disorders and infectious focal lesions (for example pyogenic versus tuberculous abscess) [3]. It lists neurodegenerative disease, epilepsy (mainly temporal lobe epilepsy) and stroke after the hyperacute phase as conditions where it may contribute to management [3]. It adds that some medical organisations still regard it as investigational and that reimbursement is a frequently cited barrier [3].

5.2 What the main metabolites are said to indicate

MetaboliteStated interpretation
N-acetylaspartate (NAA, total NAA)Marker of neuronal and axonal integrity and viability; reduced in neurodegeneration, tumours and infarcts; grossly elevated in Canavan disease
Choline (total choline)Marker of cellular proliferation, membrane turnover or inflammation; elevated in tumours
Creatine (total creatine)Often used as the denominator of ratios; absent or severely reduced in creatine synthesis or transporter defects
LactateAnaerobic glycolysis; a non-specific marker elevated in neonatal hypoxia-ischaemia, the infarct core and some tumours
Lipids (near 1.3 ppm)Linked to necrosis or hypoxia; associated with higher-grade tumours; also seen in some abscesses
Myo-inositolMarker of gliosis or glial activation; elevated early in dementia and in some demyelinating disease

All rows are from the consensus statement [3]. Other compounds (glutamate and glutamine, GABA, glutathione, succinate, acetate, amino acids and others) are named there for particular conditions, and some are detectable only at 3 T or higher or with a short echo time [3].

5.3 Acquisition guidance in the consensus

Clinical spectroscopy can be performed successfully at 1.5 T or 3 T; the statement says that spectral quality matters more than field strength alone, and that a good 1.5 T spectrum can be better than a poor 3 T one [3]. Single-voxel acquisition is preferred for a well-defined lesion or a single clinical question (about 2–5 minutes for voxels of 6–8 mL), and spectroscopic imaging (chemical-shift imaging) for multiple, heterogeneous or undefined lesions, with voxels of about 1 mL or less and about 6–30 minutes depending on resolution [3]. A short echo time is 25–35 ms for PRESS and 20 ms or less for STEAM, and a long echo time is 135–270 ms; lactate inversion uses 144 ms, with a caution about chemical-shift displacement errors at 3 T [3]. Shimming and lipid and water suppression are harder over large spectroscopic-imaging volumes than in a single voxel [3]. The statement defines minimum water linewidth and signal-to-noise criteria for interpretation in a figure whose numerical thresholds were not captured in the text read; they are (to be sourced) from the article itself. Quantification with LCModel or similar tools uses Cramér–Rao lower bounds as error estimates; test–retest coefficients of variation of 6 % or less are reported as achievable in single-voxel studies for total NAA, total creatine, total choline, myo-inositol and glutamate [3].

5.4 Limits stated by the consensus

Lack of standardisation and quality assurance is called "a current impediment"; the evidence on impact on diagnosis, treatment and outcome is limited, and no study had measured societal impact; changes in neurodegeneration and epilepsy are subtle and hard to detect in individual patients; and paediatric normative data are scarce [3]. The authors recommend standardising acquisition, analysis and reporting across vendors, publishing normative data and running large prospective multicentre trials with assessment of outcome and economic benefit [3]. The statement is from 2014; no more recent consensus was read.

The interpretation of spectroscopy in glioma and its place in a tumour protocol are on the brain tumour page (1185). Spectroscopy of organs other than the brain was not covered by the sources read.

6. MRI in Radiotherapy

Three situations are described by the sources, and they ask different things of the MR service.

SettingWhat MRI is used forScannerWhat the sources ask forSources
MRI for planning in a CT-based workflowBetter delineation of target and organs at risk; images registered to the planning CTA diagnostic scanner adapted for the purpose, or a dedicated MR simulatorPositioning as for treatment (flat indexed couch, immobilisation, coil supports, lasers), low distortion, RT-specific quality assurance, MR-trained radiographers[9, 10]
MR simulator in the radiotherapy departmentPlanning on dedicated MR-simulation platforms, whose use "has expanded rapidly" according to the AAPM descriptionDedicated wide-bore system (nominal 0.7 m)Equipment selection, siting, acceptance testing, quality assurance, workflow optimisation, safety and staffing, as stated by the AAPM report (page description read only)[10, 11]
MR-linacImaging before and during delivery, with daily adaptation of the planA hybrid system: 0.35 T with a split magnet or 1.5 T, each with a linacWritten MR safety procedures, training of all staff in the MR environment, patient selection, adaptive workflow with checks, backup arrangements[2]

6.1 MRI for planning in a CT-based workflow

The IPEM report states that MRI provides superior soft-tissue contrast compared with CT, and limits its scope to MRI acquired for external-beam planning in a CT-based workflow; MRI-only planning, brachytherapy and gamma radiosurgery are excluded [9]. It records that only 2 of 71 UK centres had dedicated MR scanners in radiotherapy departments in 2018 [9]. A British Journal of Radiology guidance (2020) says that the use of MRI within the radiotherapy pathway is becoming more prevalent as access to MRI increases [10].

The sources describe what changes in the examination. A flat indexed couch top greatly improves the reproducibility of patient set-up and is called essential for extracranial scans, coil supports or bridges should avoid compressing the patient contour, and conductive materials in an immobilisation device (carbon fibre, for example) may cause signal loss [9]. The flat overlay has "unintended consequences": it can reduce comfort and raise the patient within the bore, which increases motion artefacts, and standard head and neck coils may not fit with immobilisation devices [10]. Geometric distortion and chemical shift depend on the receive and slice-selection bandwidth, and a higher bandwidth reduces distortion but lowers signal-to-noise [10]. The IPEM report gives parameter guidance for planning sequences (fat–water chemical-shift displacement of 1 mm or less, zero slice gap, three-dimensional acquisition and three-dimensional distortion correction where possible, and at least two-dimensional correction) and works one example in which a bandwidth of 450 Hz/pixel keeps displacement below 1 mm for up to 3.5 ppm of inhomogeneity on a 3 T system with 1 mm voxels [9]. How bandwidth, chemical shift and distortion work is on the parameter pages (9501, 9597, 9607).

The quality-assurance values of the IPEM report are given in the following table.

TestAction value or tolerance statedFrequency or commentSection of the report
External laser alignment, basic and full±2 mmBasic check on every day the lasers are used; full check monthly8.5
Gradient-related distortion2 mm within the clinically relevant volume; no more than 1 mm for intracranial radiosurgerySuggested volumes: 20 cm sphere for brain or prostate, 40 cm sphere for full-field head and neck or pelvis; at least annually, more often when introducing MRI8.6.1
Field-inhomogeneity distortion1–2 mm for most applicationsIf sequences are optimised to a 3.5 ppm fat–water shift, inhomogeneity not exceeding 3.5 ppm gives clinically acceptable errors8.6.2
Couch flatness and deflection0.2° lateral, 0.5° longitudinal, ±2 mm deflectionAt acceptance, then annually8.7
End-to-end test2 mm for most applications; tighter for radiosurgeryTable 10 assigns frequencies; the column alignment in the extract was unclear8.8

The report recommends commissioning the MRI-to-CT registration workflow on at least ten patients and a qualitative check by two independent people for each patient, using a graded uncertainty scale [9]. The BJR guidance adds that MR-only planning and synthetic CT need individual verification before clinical use [10]. A German guideline on MRI simulation for cranial stereotactic radiotherapy defines minimum quality requirements and advanced options; only the search extract was read, because the full text was refused to automatic readers [12].

6.2 The MR-linac

The ESTRO–ACROP recommendation of 2021 describes online MR-guided radiotherapy as one of the most innovative applications of image-guided radiotherapy and names two hybrid systems in clinical use: one with a 0.35 T MR scanner with a split magnet and a 6 MV flattening-filter-free linac, and one with a 1.5 T MR system and a 7 MV flattening-filter-free linac [2]. Daily imaging allows the plan to be adapted to the anatomy of the day, from simple shifts of the leaves of the collimator to full re-optimisation [2]. Indications preferred for adaptation are moving targets: lung lesions, locally advanced pancreatic cancer, and primary or secondary liver tumours; others named are head and neck, prostate, breast, pelvic nodes, oligometastases, and kidney and adrenal metastases [2].

The recommendation sets conditions that concern the MR service directly. Patients are classed as physically incompatible (for example with a pacemaker that is not MR conditional), clinically incompatible, borderline or fully compatible; borderline patients may need psychological support, anaesthesia or music; severe obesity or cachexia (body mass index above 40 or weight under 40 kg) is reviewed case by case in view of artefacts, heating and a bore under 70 cm [2]. Treatment slots with full online re-planning are significantly longer than on a conventional linac and may last up to 60 minutes [2]. Centres should write MR safety procedures before implementation, train all professionals who work in the MR environment, including cleaning staff, in access restrictions, noise, radiofrequency and gradient fields, quench and emergencies, and have approved procedures for emergencies; screening includes metal, implants and pacemakers, and ferromagnetic detectors may be used only after conventional screening [2]. The team typically comprises radiation therapists, radiation oncologists and medical physicists, whose roles "may cross traditional boundaries" [2]. The authors state limits: the benefit of daily adaptation varies between tumour sites; the scarcity of published evidence does not allow them to say which approach performs best; national education of the professionals varies greatly between countries; and implementation is "a resource intensive investment" [2].

6.3 Roles and safety in the radiotherapy setting

Staffing rules differ with the source and the country. The IPEM report requires at least two staff classed as MR Authorised Persons when patients are scanned, at least one of whom is an MR Authorised Person (Supervisor); its safety content is based on United Kingdom guidance and is stated to be UK-specific [9]. The BJR guidance recommends that at least one MR-trained radiographer is present at all times and that radiotherapy radiographers have MR safety competence [10]. Neither rule is a European standard (Reasoning). The MR safety pages of the library are 9802 and the pages linked from it.

7. MR-Guided Focused Ultrasound

7.1 What it is and what MR does

Focused ultrasound directs acoustic waves at a target; the effect depends on intensity and duty cycle, and tissue devitalisation is generally achieved at temperatures above 55 °C [24]. The intracranial application became reliable after CT-derived skull data were registered to the MR images, and the phased array can steer up to 1,024 elements to correct for distortion by the skull with a resolution of about 1 mm [24]. A review in JAMA Neurology reports that a 650 kHz transducer can create lesions 4 to 6 mm in diameter with an accuracy under 2 mm, and that treatment is monitored by real-time thermal imaging and clinical feedback [25]. The roles of MR in the sources are three: anatomical targeting with high soft-tissue contrast, near-real-time MR thermometry during sonication, and assessment of the treated volume afterwards (a 2015 review states that the non-perfused volume of the target can be determined and can indicate the success of treatment) [24, 26].

7.2 Approvals and status

The FDA summary of the 2025 supplement gives the history of one intracranial system: original approval on 11 July 2016 (for medication-refractory essential tremor, according to a trade report [40]), a supplement on 16 December 2018 for unilateral thalamotomy for tremor-dominant Parkinson disease with medication-refractory tremor in patients aged at least 30, a supplement on 29 October 2021 for unilateral pallidotomy in advanced idiopathic Parkinson disease with medication-refractory moderate to severe motor complications, and a supplement approved on 3 July 2025 [13]. Trade reports state that approval for the second side of essential tremor, at least nine months after the first, was announced in December 2022 and that a staged bilateral pallidothalamic tractotomy for advanced Parkinson disease was also approved; the extracts do not show the date of the last [40]. For the body, the sponsor protocol of a transcranial study states that a body system received FDA approval for uterine fibroids in October 2004 [28]; a 2015 review states a CE mark and FDA approval for thermal ablation of uterine fibroids and for pain from bone metastases, and no approval for prostate or breast cancer [26]; a 2019 review states that MR-HIFU is CE approved for uterine fibroids, alleviation of bone pain, prostate tissue ablation and, as a first neurological application, essential tremor [27]. Other neurological uses (dystonia, neuropathic pain, obsessive–compulsive disorder, epilepsy, brain tumours, opening of the blood–brain barrier with microbubbles for drug delivery, neuromodulation) are described as under investigation [24, 25].

7.3 Safety and selection points in the sources

The 2017 AJNR review lists adverse effects from the glioma trial (a fatal intracranial haemorrhage a few days after treatment, possibly related to coagulopathy; incomplete ablation in three patients), states that lower-frequency application seems associated with an increased risk of bleeding, and says that patient selection is "based on heat generation at the calvaria" [24]; it does not report the essential tremor adverse effects or a skull density ratio. A US Medicare local coverage determination, quoted in a provider's reconsideration request, lists a skull density ratio below 0.45 among its non-covered limitations; this is a coverage rule, not a clinical guideline [41]. The infiltrative nature of primary gliomas makes ablation uniquely challenging, and acoustic access limits superficial lesions [24].

The MR safety dimension of an examination on a focused-ultrasound system (device compatibility, the patient's implants, the table) is on the safety pages; this page does not repeat it.

8. MR-Guided Biopsy and Interventions

A RadioGraphics article states that clinical use of MRI for guidance of interventions emerged shortly after the introduction of clinical diagnostic MRI in the late 1980s, and that MRI guidance is particularly advantageous when the disease is best visualised with MRI or when superior soft-tissue contrast is needed to monitor treatment (search extract only) [31]. A 2017 white paper states that biopsy of breast lesions not detectable by other modalities has been performed under MRI guidance for decades and that prostate biopsy under MRI guidance is being introduced in a similar fashion; it describes endovascular and cardiac MR-guided procedures as having advantages for a number of indications (search extract only) [30]. A review of MRI-guided robots lists neurosurgery, breast biopsy, cardiovascular intervention, prostate biopsy and radiotherapy among current applications, and notes that "MR Conditional" is different from "MR Compatible", an obsolete term first defined in a 1997 FDA draft guidance (search extract only) [32].

8.1 MRI-guided breast biopsy and localisation as a worked example

The most detailed source read is a 2020 review of MRI-guided breast interventions [29]. Indications are lesions that are "mammographically and sonographically occult" and suspicious lesions that cannot be re-identified on ultrasound or mammography [29]. The procedure is done prone, with a dedicated breast array coil at 1.5 T or 3 T, after intravenous gadolinium and with pre- and post-contrast fat-saturated T1 sequences; a grid guides needle entry with a fiducial marker in a neighbouring square, vacuum-assisted large-gauge systems (8–12 G) are recommended with typically 12 samples and a metallic clip, and localisation uses a hook wire confirmed by two-view mammography [29]. Reported figures: complications of vacuum biopsy 0–6 %, of localisation 0–3.3 %, neoplastic seeding 2 of 1,644 and 8 of 4,010 biopsies, and non-visualisation (procedure cancelled) in about 12–13 % of cases; a biopsy takes 20–90 minutes and localisation about 40 minutes [29]. The review states limits: prone and supine positions differ, the presumed sonographic correlate was not the true correlate in 26 % of 38 cases in one study, background enhancement can cause non-visualisation, wire migration is more likely than with mammographic guidance, and MRI-guided procedures are longer and more costly than other image-guided procedures [29]. The review was read to 100,000 of 129,537 characters.

The contrast agent used is on 9101 and its safety on 9806; the zones and access rules that govern an MR suite in which instruments and a physician work are on 9802.

9. Fetal MRI

The updated ISUOG practice guideline (2023) states that fetal MRI complements an expert ultrasound examination, either by confirming ultrasound findings or by adding information, and is not a primary screening tool; it is a standard method only in selected high-risk cases, for example affected first-degree relatives [1]. It is indicated when ultrasound leaves the diagnosis incomplete; for sonographically isolated findings (ventriculomegaly, agenesis of the corpus callosum, absent septum pellucidum, cerebellar or vermian anomalies) MRI may be recommended for parental reassurance, and other reasons include a monochorionic twin pregnancy after the death of a co-twin and abnormalities of the face, neck, chest and abdomen [1]. Before 18 weeks MRI usually adds nothing beyond transvaginal ultrasound, its value grows after about 22 weeks, and organs are generally best seen at 26–32 weeks [1].

9.1 Technique in the guideline

The guideline accepts 1.5 T or 3 T: 1.5 T is most common and gives acceptable resolution from about 18 weeks; 3 T offers higher resolution but 1.5 T is preferred in polyhydramnios because it is less prone to fluid-wave artefacts [1]. T2-weighted fast (turbo) spin-echo is the mainstay, steady-state free precession distinguishes vessels from solid tissue, T1-weighted 2D gradient echo is used at 1.5 T, and single-shot gradient-echo echo-planar imaging shows bone, calcification and blood products; diffusion-weighted imaging, diffusion tensor imaging and dynamic SSFP are optional [1]. The stated slice thickness is 2–5 mm with a 10–15 % gap, T2 is needed in three orthogonal planes of the brain and body, and "this minimum protocol should be executable in under 30 min"; the minimum protocol includes an SSFP sequence covering the whole uterus to document placental position and allow volumetry [1]. The woman is positioned supine or left lateral, contraindications and claustrophobia are screened, sedation may be considered, and informed consent is obtained [1]. Performance and interpretation should be limited to people with specific training; desirable attributes of a training centre include at least 500 fetal MRI examinations with at least two per week [1].

9.2 Safety statements

Without contrast media, MRI "is not associated with known adverse fetal effects at any point in pregnancy"; whole-body radiofrequency transmission is limited by the IEC to a specific absorption rate not exceeding 2 W/kg, and postnatal studies showed no impact on hearing or growth after prenatal 3 T MRI [1]. The guideline does not address acoustic noise or contrast beyond the no-contrast statement [1]. The full discussion of the safety evidence, of gadolinium in pregnancy and of the pregnant worker is on 9812 and 9806. The limits stated by ISUOG are maternal obesity, fetal movement, fluid-wave artefacts at 3 T and, after 32–35 weeks, digits that may not be assessable; the evidence on MRI for isolated findings is described as inconclusive and heterogeneous, and the guideline is not intended to set a legal standard of care [1].

10. Functional MRI and Tractography for Neurosurgical Planning

The main clinical applications of functional MRI in an individual patient are preoperative risk assessment and the mapping of "critical" cortex for presurgical planning [5]. A 2010 review states that BOLD fMRI made a major transition in the preceding decade from a research technique to a clinical one used primarily for presurgical planning in patients with brain tumours and other resectable lesions (abstract only) [34].

10.1 Guideline-level evidence

The American Academy of Neurology guideline on fMRI in the presurgical evaluation of patients with epilepsy (published in January 2017) grades its conclusions: fMRI may be considered instead of the intracarotid amobarbital procedure (Wada test) for language lateralisation in temporal and extratemporal epilepsy (weak evidence, Level C), with insufficient evidence for temporal neocortical epilepsy and temporal tumours; presurgical fMRI of verbal memory or language encoding should be considered to predict verbal memory outcome after left medial temporal lobe surgery (moderate evidence, Level B); and fMRI of language may be considered to predict postsurgical language outcome after anterior temporal lobectomy (Level C) [4]. The guideline lists its limits: small, heterogeneous samples, no multicentre data or data on generalisability across centres, imperfect concordance between fMRI and the Wada test (which has its own limitations), and little paediatric data [4].

10.2 Consensus recommendations for language mapping

A consensus of the Organization for Human Brain Mapping working group, published in 2025 as a shortened form of the ratified document, recommends choosing tasks from the clinical question, using a combination of tasks for lateralisation with at least one sentence-level task, preferring active non-linguistic control conditions to rest, practising before scanning (a simulated session for children), acquiring at 3 T where possible with a repetition time of 1–3 s and voxels no larger than 3.0 mm isotropic, minimising smoothing (ideally 4 mm, at most 5 mm), and inspecting a range of thresholds and unthresholded maps [5]. It states that task-based mapping remains most reliable and that resting-state data alone are insufficient when a task can be acquired [5]. The ASFNR task algorithm for adult language lists sentence completion, silent word generation, rhyming, object naming and passive story as tasks (search extract only) [7]. Recommendations for resting-state acquisition from three neuroradiology societies give at least 6 minutes, eyes open with fixation, acquisition before task-based fMRI and contrast administration, a temporal resolution printed as "#2 seconds" (probably 2 s or better, Reasoning), and 3 T or higher (abstract only) [6].

The consensus states its limits plainly. Lateralised activation does not show that an area is dispensable; false negatives are the greatest concern because a missed language area could be resected; EPI distortion and susceptibility signal loss can misplace or hide activation; pathological vessels in high-grade tumours may restrict the use of fMRI and reduce BOLD signal; more areas are typically activated by fMRI than are associated with language errors during direct electrical stimulation, giving only modest specificity (approximately 55 %–71 %); the Wada test is itself not infallible; and the literature "consists of mainly uncontrolled observations" [5]. The same consensus states that if the interpreter does not believe a report to be valid no images should be generated, that fixed distance limits from activations should not set surgical margins, and that fMRI should be interpreted by the operating neurosurgeon after interdisciplinary consultation [5].

10.3 Diffusion tensor imaging and tractography

Diffusion tensor imaging measures the directional diffusion of water along myelinated fibres and allows white-matter tracts to be reconstructed by tractography [36]. The only primary source read on its presurgical use is a protocol study of five language-related bundles with regions of interest placed on a T1-weighted MPRAGE image, acquired at 3 T with 2 mm isotropic voxels, b = 1300 s/mm² and 60 directions in 5 min 47 s; in 150 region pairs placed by three raters, 132 (88 %) tracked successfully, success was 78.7 % in tumour patients and 97.3 % in patients with myelopathy, and inter-rater agreement was fair (kappa 0.300) [35]. The authors state that deterministic tractography cannot account for uncertainty or distributed connectivity, that tracking stops at thresholds and "does not mean that there are no fibers", and that the study compared no tract with direct stimulation or histology [35]. The study is one centre's feasibility work with 20 subjects, not a guideline (Reasoning, from the counts in [35]). Diffusion physics is on 9207 and the diffusion sequences on 9012.

11. MR Elastography

MR elastography uses an external driver that sends low-frequency mechanical vibrations into the body, phase-contrast imaging with motion-encoding gradients records the shear waves, and an inversion algorithm converts the wave propagation into maps of shear stiffness in kilopascals [33]. A 2024 Radiology review states that liver stiffness measured by MRE provides the highest classification accuracy for non-invasive staging of fibrosis and that the QIBA provides guidance to standardise acquisition and interpretation [33]. It recommends thresholds of 3.0 kPa for stage F1 or higher, 3.5 kPa for F2 or higher, 4.0 kPa for F3 or higher and 5.0 kPa for F4, from the LI-RADS Quantitative Imaging Working Group and a Society of Abdominal Radiology panel, applied regardless of cause [33]. Acquisition is during an end-expiration breath-hold; regions of interest avoid artefacts, large vessels, lesions and the outer centimetre below the capsule, and the mean of four sections is reported [33]. The article names obese patients as technically difficult [33].

The QIBA profile (November 2023) is a document that sets requirements to reach a claim, not a standard of care: a measured change in hepatic stiffness of 19 % or larger indicates a true change with 95 % confidence, based on a within-subject coefficient of variation of 7 % in normal liver [8]. It requires a driver frequency of 60 Hz, four slices at the widest transverse extent of the liver, a field of view of 420 mm, EPI-based MRE at 3 T if available (otherwise gradient-echo MRE), at least 4 hours of fasting, the driver over the right lower chest wall, confirmation of signal loss in the subcutaneous fat under the driver and of shear waves in the phase images, and rejection and repetition of examinations with iron overload, interposed colon or incoherent waves; the physicist confirms the driver frequency and field of view and linearity, and the phantom stability difference should be 10 % or less, assessed about every 6 months [8]. Which of these steps belong to the technologist and which to the physicist is stated in the profile; its own checklist says "coronal" where its text and appendix specify axial sections [8]. The liver protocol is on 5101. Applications of elastography outside the liver were not covered by the sources read, apart from the statement that flexible drivers may allow stiffness measurements of multiple organs [33].

12. Preclinical and Research MRI

A 2024 review defines preclinical MRI as animal imaging with small-bore systems used for disease pathophysiology and translational research, different from whole-body systems used for larger animals [37]. It lists installed systems of 1.0 T to 11.7 T, a 16 cm bore for mice and rats at 7 T and a 30 cm bore at 7 T for mice, rats and rabbits, and both superconducting and permanent-magnet systems, the latter with a limited number of sequences [37]. Applications described are brain (diffusion tensor imaging, fMRI, spectroscopy, quantitative susceptibility mapping, CEST), heart (function and strain after infarction), liver (relaxation times, dynamic contrast imaging) and multinuclear spectroscopy with phosphorus-31 and carbon-13 coils [37]. An earlier review of small-animal imaging of central nervous system models (year not shown) states that most small-animal systems operate at 4.7–11.7 T (search extract only) [38].

The review states several practical differences from clinical work: control of movement is the most important factor, isoflurane and sevoflurane are the most common anaesthetics, mice become hypothermic and need temperature control, monitoring covers respiration, temperature and ECG, and rodents require anaesthesia unlike clinical subjects [37]. It also states that standardised acquisition and analysis protocols are needed for reproducibility, and that translational claims are hedged and not validated quantitatively against human data [37]. Two citation errors were found in the review (an isoflurane statement citing an unrelated paper and a CEST discussion citing a fMRI protocol), so its statements should be checked against primary sources before they are used (Reasoning, from the check of the reference list in [37]).

13. What These Applications Have in Common for the MR Service

The following points are inferences from the sources read, marked as such, and not statements of the sources (Reasoning).

  • Most of these applications change the examination set-up or hardware, not only the sequence: a flat couch and radiotherapy immobilisation [9, 10], a coil and a transmit chain for another nucleus [17, 20], a driver for elastography [8], a focused-ultrasound phased-array transducer [24], a grid and MR-conditional instruments for biopsy [29, 32].
  • Several have their own quality assurance and written requirements: geometric distortion and laser checks in radiotherapy [9], a conformance profile and a phantom for elastography [8], quality criteria for spectra [3].
  • Several are team activities in which the technologist is one role among others: radiation therapists and physicists on the MR-linac [2], an obstetrician or radiologist with a trained fetal MRI team [1], a neurosurgeon and neuroradiologist in functional mapping [5].
  • For most of them the evidence on outcome is limited, and the sources say so: spectroscopy [3], the MR-linac [2], functional MRI [4, 5], sodium and hyperpolarised imaging [17, 19].

14. Where to Go Next

If the reader wantsPage
The ordinary examination and the introduction to MRIWhat Is MRI in Clinical Diagnostics?
The physics of nuclear magnetic resonance and chemical shiftFundamentals of Nuclear Magnetic Resonance, Chemical Shift
Coils and the transmit chainRF hardware, RF transmit system
Spectroscopy in brain tumoursCerebral gliomas and glial tumours
DiffusionFlow, motion and diffusion physics, DWI sequence
Perfusion in brain tumoursDSC perfusion
Pregnancy and fetal safetyMRI in pregnancy and the pregnant worker
The contrast agent used in guided proceduresContrast media, gadolinium safety
The safety of an MR suite and of implantsZones, active implants
The liver protocolLiver — generic standard protocol
Uterine fibroids and the female pelvisFemale pelvis — generic standard protocol

There is no dedicated page yet for proton spectroscopy as a technique, for functional MRI, for elastography, for MRI in radiotherapy or for focused ultrasound; they are candidates for later pages and would be added only on request.

15. Discrepancies Between Sources

TopicOne sourceAnother sourceHandling on this page
Stage of the liver MRE profileA summarising-tool answer gave "Stage 3: Clinically Feasible" for the cover, and the file name contains "CLINICALLY-FEASIBLE" [8]A search extract of the same document shows "Stage: 3: Technically Confirmed" on the cover [8]Both stated; not resolved; the stage is not used for any claim
Change in stiffness that is a true change19 % or larger, 95 % confidence [8]"A change exceeding 20 %" citing QIBA and a second study [33]19 % given as the profile's claim; the 20 % is a rounded or different statement; the question uses 19 %
Non-visualisation in MRI-guided breast biopsyAbout 12–13 % of cases [29]8–13 % in another section of the same review [29]Both ranges given
MR-guided focused ultrasound for prostate2015 review: no CE or FDA approval had been obtained [26]2019 review: CE approved for prostate tissue ablation [27]Different dates and regulators; not a conflict (Reasoning); no current regulatory status checked
Date of the first approval of the intracranial system11 July 2016 [13]16 July 2016 in a provider's request quoting a coverage document [41]The FDA document is used; the second source was seen only as a search extract
Age limit of hyperpolarised xenonApproved from age 12 [14]A 2025 supplement document proposes age 6 and above [16]12 years stated as approved in 2022; the outcome of the supplement was not read
Language fMRI recommendationsThe 2025 text is a shortened form of the ratified document and "slightly differs" from the official version [5]The official version was not readThe text read is cited as such

16. Open Questions

  • How the benefit of daily adaptation on the MR-linac compares with conventional radiotherapy, by tumour site: the guideline states the evidence is scarce [2].
  • Whether fMRI changes surgical outcome in tumour patients: the consensus states that the literature consists of mainly uncontrolled observations and that early postsurgical outcomes alone are insufficient [5].
  • Whether proton spectroscopy changes management and outcome, and whether it can be standardised across vendors [3].
  • How and when hyperpolarised carbon-13 imaging will move from trials to practice, given its workflow limits [19].
  • Which focused-ultrasound indications beyond tremor and fibroids will be confirmed by trials [24, 25].

17. Evidence Gaps

  • No guideline for the whole subject exists in the sources read; each section stands on its own sources, and some are old: the proton spectroscopy consensus is from 2014, the reviews of focused ultrasound from 2015 to 2019, the interventional white paper from 2017.
  • The numerical water-linewidth and signal-to-noise thresholds of the spectroscopy consensus were not captured; the supplementary appendix of that article was not available in the text read. (to be sourced)
  • The current regulatory status of focused ultrasound in Europe and elsewhere, and of the 2025 supplement, was not read from regulators; the extracts are partial. The Thieme and PMC reviews of focused ultrasound were refused to automatic readers and were not worked around. (to be sourced)
  • The AAPM Task Group 284 report and the German guideline on MRI simulation for stereotactic radiotherapy were not read in full text (the AAPM page description and a search extract only).
  • Radiotherapy planning with MRI alone (synthetic CT) was not covered beyond the statements of the BJR guidance and the exclusion in the IPEM report.
  • Elastography beyond the liver, spectroscopy beyond the brain, MRI in cardiac or vascular intervention, and MRI-guided prostate biopsy accuracy were not read in detail; the RadioGraphics 2021 article and the interventional white paper were available only as search extracts.
  • The roles and titles of technologists in these services in Italy, France and other countries were not sourced. (to be sourced)
  • Reading levels: Öz 2014 was read to 100,000 of 115,662 characters and its remainder, which is the reference list; the ISUOG guideline to 100,000 of 101,790 characters and its remainder, which is page boilerplate; the language fMRI consensus to 100,000 of 168,588 characters and then from 100,000 onward; the IPEM report in two ranges covering 184,942 characters; the hyperpolarised carbon-13 review to 100,000 of 198,755; the hyperpolarised detection review to 100,000 of 179,183; and the breast intervention review to 100,000 of 129,537 characters.

18. Evidence-Based References

No systematic review or meta-analysis forms the basis of this page. Sources are ordered by type and quality (guidelines, consensus statements and regulatory documents first; reviews, single studies and extracts after); the reading level of each is stated. Search extracts are cited only for what the extract states.

A. Guidelines, Consensus Statements, Technical Reports and Regulatory Documents

Guideline
[1] Prayer D, Malinger G, De Catte L, et al. ISUOG Practice Guidelines (updated): performance of fetal magnetic resonance imaging. Ultrasound Obstet Gynecol. 2023;61(2):278–287.
PMC copy through a summarising tool, characters 0–100,000 of 101,790 (the remainder was page boilerplate).
Recommendation
[2] Corradini S, Alongi F, Andratschke N, et al. ESTRO-ACROP recommendations on the clinical implementation of hybrid MR-linac systems in radiation oncology. Radiother Oncol. 2021;159:146–154.
Accepted manuscript (ICR repository) through a summarising tool; no truncation notice. Several authors declare relations with a manufacturer.
Consensus
[3] Öz G, Alger JR, Barker PB, et al; MRS Consensus Group. Clinical proton MR spectroscopy in central nervous system disorders. Radiology. 2014;270(3):658–679.
Open copy (UCSF eScholarship) through a summarising tool, characters 0–100,000 of 115,662, and the remainder (reference list). Figure 9 thresholds and Appendix E1 not captured.
Guideline summary
[4] American Academy of Neurology. Summary of practice guideline for clinicians: use of fMRI in the presurgical evaluation of patients with epilepsy. Published in Neurology, 11 January 2017 (per the summary page; authors not shown in the text read).
Summary page through a summarising tool; the full guideline was not read.
Consensus
[5] Voets NL, Ashtari M, Beckmann CF, et al. Consensus recommendations for clinical functional MRI applied to language mapping. Aperture Neuro. 2025;5. doi:10.52294/001c.128149.
Open copy (Oxford repository) through a summarising tool, characters 0–100,000 and then from 100,000 of 168,588; the text is a shortened form of the ratified OHBM document.
Consensus (abstract)
[6] Recommended resting-state fMRI acquisition and preprocessing steps for preoperative mapping of language, motor and visual areas in adult and paediatric patients with brain tumours and epilepsy. ASFNR, ASPNR and ASNR task force. AJNR. 2024;45(2):139 (first author shown only as "Sair" in a search extract).
Abstract on a university profile page through a summarising tool; the full text was refused at two addresses and not read.
Recommendation (extract)
[7] Black DF, Vachha B, Mian A, et al. American Society of Functional Neuroradiology-recommended fMRI paradigm algorithms for presurgical language assessment. AJNR. 2017;38(10):E65–E73.
Search extracts only (abstract and article information).
Technical profile
[8] QIBA MR Biomarker Committee. QIBA Profile: Magnetic Resonance Elastography of the Liver. Radiological Society of North America; maintenance version of 7 November 2023. doi:10.1148/QIBA/20231107.
PDF through a summarising tool; no truncation notice; a search extract of the cover was also read (see Section 15 for the stage wording).
Technical report
[9] Speight R, et al. IPEM topical report: guidance on the use of MRI for external beam radiotherapy treatment planning. Phys Med Biol. Accepted manuscript (year 2021 from the file name).
Accepted manuscript (UCL repository) through a summarising tool in two ranges (0–100,000 and from 100,000 of 184,942 characters). The safety content is UK-specific.
Guidance
[10] Guidance on the use of MRI for treatment planning in radiotherapy clinical trials. Br J Radiol. 2020;93(1105):20190161 (authors not shown in the text read).
PMC copy through a summarising tool; no truncation notice. The authors state that it draws on literature and their experience, not a formal consensus standard.
Technical report (description)
[11] Glide-Hurst CK, Paulson ES, McGee KP, Tyagi N, Hu Y, Balter JM, et al. AAPM Task Group 284: Magnetic resonance imaging simulation in radiotherapy: considerations for clinical implementation, optimization, and quality assurance. AAPM Report No. 284, 2021. doi:10.1002/mp.14695.
Description page of the AAPM through a summarising tool; the report itself was not read.
Guideline (extract)
[12] Quality requirements for MRI simulation in cranial stereotactic radiotherapy: a guideline from the German Taskforce "Imaging in Stereotactic Radiotherapy". Strahlenther Onkol (journal not shown in the extract; DOI 10.1007/s00066-023-02183-6 appears in the address of the publisher page).
Search extracts only; the PMC copy returned a browser check and was not worked around.
Regulatory document (extract)
[13] US Food and Drug Administration. Summary of Safety and Effectiveness Data, PMA P150038/S037, MR-guided focused ultrasound system; notice of approval 3 July 2025.
Search extract only (history of the original approval and two supplements).
Regulatory document
[14] US Food and Drug Administration. Drug Trials Snapshot: Xenoview (xenon Xe 129 hyperpolarized), approval date 23 December 2022.
Search extracts only.
Regulatory document (extract)
[15] US Food and Drug Administration. NDA 214375, xenon Xe 129 hyperpolarized (Xenoview) and HPX hyperpolarization system: multi-disciplinary review and evaluation; completed 22 December 2022.
Search extract only (indication, limitation of use, dose volume and breath-hold).
Regulatory document (extract)
[16] US Food and Drug Administration. Xenoview efficacy supplement review documents (clinical pharmacology review and recommended indication; review completion 30 May 2025).
Search extracts only; the outcome of the submission was not read.

B. Reviews, Studies and Other Sources

Review
[17] Zaric O, Juras V, Szomolanyi P, Schreiner M, Raudner M, Giraudo C, Trattnig S. Frontiers of sodium MRI revisited: from cartilage to brain imaging. J Magn Reson Imaging. 2021;54(1):58–75 (online 26 August 2020).
Open repository copy through a summarising tool; no truncation notice.
Review
[18] Clinical application of multi-nuclear magnetic resonance spectroscopy in the various biologic processes (review). Insights Imaging. 2022. doi:10.1186/s13244-022-01262-z (authors not shown in the text read).
Through a summarising tool; no truncation notice. The statements on 23Na and 19F are minimal in this review.
Review
[19] Comprehensive literature review of hyperpolarized carbon-13 MRI: the road to clinical application. Metabolites. 2021;11(4):219 (authors not shown in the text read).
PMC copy through a summarising tool, characters 0–100,000 of 198,755.
Review
[20] Review of spatially resolved detection of hyperpolarized nuclei (PMC7109201; authors and journal not shown in the text read).
PMC copy through a summarising tool, characters 0–100,000 of 179,183.
Review (extract)
[21] Madelin G, Regatte RR. Biomedical applications of sodium MRI in vivo. J Magn Reson Imaging. 2013 (PMC3759542).
Search extract only (abstract).
Review (abstract)
[22] Non-proton nuclei (X-nuclei) MRI review. Magnetic Resonance Letters, November 2023 (authors not shown).
Abstract in a search extract only.
Educational abstract
[23] Basic principles for imaging non-proton nuclei. ISMRM 2018 educational course abstract E1303 (author not shown in the text read).
Page through a summarising tool; gives no Larmor frequencies, no resolution values and no role for each nucleus.
Review
[24] Intracranial applications of MR imaging–guided focused ultrasound. AJNR. 2017;38(3):426–431 (authors not shown in the text read).
PMC copy through a summarising tool; no truncation notice.
Review (abstract)
[25] A review of the current therapies, challenges, and future directions of transcranial focused ultrasound technology: advances in diagnosis and treatment. JAMA Neurol. 2018;75(2):246–254.
Abstract on PubMed through a summarising tool (86 full-text articles reviewed qualitatively, from 2,855 records).
Review (extract)
[26] Kobus T, McDannold N. Update on clinical MR-guided focused ultrasound applications. Magn Reson Imaging Clin N Am. 2015;23(4):657–667.
Search extract only; the PMC copy returned a browser check and was not worked around.
Review (extract)
[27] Siedek F, Yeo S, Heijman E, et al. Magnetic resonance-guided high-intensity focused ultrasound (MR-HIFU): technical background and overview of current clinical applications (Part 1). Fortschr Röntgenstr. 2019;191:522–530.
Search extracts only; the publisher page returned a security challenge and was not worked around.
Sponsor document (extract)
[28] ExAblate transcranial MR guided focused ultrasound for the treatment of essential tremors: study protocol (ClinicalTrials.gov NCT01827904), section 1.4.
Search extract only; a manufacturer's protocol, cited only for the date of the fibroid approval.
Review
[29] Gao P, et al. Recent progress for the techniques of MRI-guided breast interventions and their applications on surgical strategy. J Cancer. 2020;11(16):4671–4682.
PMC copy through a summarising tool, characters 0–100,000 of 129,537.
White paper (extract)
[30] White paper: interventional MRI: current status and potential for development considering economic perspectives, part 1: general application. Rofo. 2017;189(7):611–623.
Search extracts only; the publisher page returned a security challenge and was not worked around.
Review (extract)
[31] MRI-guided interventional procedures (RadioGraphics, 2021; DOI 10.1148/rg.2021210040; authors not shown in the extract).
Search extracts only; the page returned a client error.
Review (extract)
[32] MRI-guided robot intervention: current state-of-the-art and new challenges. Med-X. 2023;1:4.
Search extracts only.
Review
[33] Interpretation, reporting, and clinical applications of liver MR elastography. Radiology. 2024;310(3):e231220 (PMC10982829; authors not shown in the text read).
PMC copy through a summarising tool, characters 0–100,000 of 114,483; Appendix S1 and Table 1 (an image) not captured.
Review (abstract)
[34] Pillai JJ. The evolution of clinical functional imaging during the past 2 decades and its current impact on neurosurgical planning. AJNR. 2010;31(2):219–225.
Abstract in a search extract only.
Study
[35] A user-friendly protocol for tractography of the major language fibre bundles in presurgical work-up (PMC6525736; authors and journal not shown in the text read).
PMC copy through a summarising tool; the authors' text gives inconsistent failure counts for the tumour and myelopathy groups.
Book chapter (extract)
[36] Chapter on presurgical functional neuroimaging (DOI 10.1007/978-3-030-83343-5_1; title and authors not shown in the extract).
Search extract only; cited for the definition of diffusion tensor imaging and tractography.
Review
[37] Saito S, Ueda J. Preclinical magnetic resonance imaging and spectroscopy in the fields of radiological technology, medical physics, and radiology. Radiol Phys Technol. 2024;17(1):47–59.
PMC copy through a summarising tool, characters 0–100,000 of 102,481. The tool flagged two citation errors in the reference list.
Review (extract)
[38] Small animal magnetic resonance imaging: translational research on central nervous system disease models (PMC3075741; authors not shown in the extract).
Search extract only.
Review (extract)
[39] Review of the use of multinuclear MRI for investigations in drug delivery (PMC9572840; authors not shown in the extract).
Search extract only; cited for the fluorine-19 statement and the second heteronuclear transmitter.
News items (extracts)
[40] MassDevice, 20 December 2022; NeurologyLive, 21 December 2022; AuntMinnie (date not shown): reports on FDA approvals of a focused-ultrasound system for the second side of essential tremor and for staged bilateral pallidothalamic tractotomy.
Search extracts only; trade reports, used only where the FDA document does not give the information.
Coverage document (extract)
[41] Centers for Medicare & Medicaid Services. Local coverage determination L37761 (MR-guided focused ultrasound for essential tremor), as quoted in a provider's reconsideration request (November 2023).
Search extract only; a coverage rule, not a clinical guideline.