Active Implants Other Than Cardiac Devices in MRI — Deep Brain, Spinal Cord, Vagus and Sacral Stimulators, Cochlear and Other Auditory Implants, Drug Infusion Pumps and Glucose Monitors: What the Evidence Shows and Where It Stops

up to this point verified by human experts

Active Implants Other Than Cardiac Devices in MRI — Deep Brain, Spinal Cord, Vagus and Sacral Stimulators, Cochlear and Other Auditory Implants, Drug Infusion Pumps and Glucose Monitors: What the Evidence Shows and Where It Stops

MRIninja Knowledge Base | Child Page — MRI Safety (deep dive S7) Version 1.0 — October 2026

Parent page: MRI Safety — Master Page (hazard table in Section 3). Builds on: Implant and Device MR Safety Framework (test methods, label content, matching a label to the scanner) · Cardiac Implantable Electronic Devices in MRI (cardiac devices) · RF Heating and Burns (RF heating of leads)


1. Scope, Status and Boundaries

This page gathers the evidence on active implanted devices that are not cardiac: neurostimulators (deep brain, spinal cord, peripheral nerve, vagus and sacral nerve), cochlear and other active auditory implants, implanted drug infusion pumps, and body-worn or implanted glucose monitors and insulin delivery devices. It states what the sources read report about adverse events and about the conditions under which scanning has been done, and where those sources stop. Three neighbouring topics are held elsewhere and are not repeated: how a label is read and matched to the scanner, and the test standards behind it (Implant and Device MR Safety Framework); pacemakers, defibrillators and other cardiac devices (Cardiac Implantable Electronic Devices in MRI); and the mechanism of heating of leads and conductors (RF Heating and Burns). Passive implants and foreign bodies are the subject of Passive Implants and Retained Foreign Bodies in MRI.

Reading status. The page rests on the fifth edition (July 2026) of the UK MHRA guidelines (general sections read verbatim; Section 5.5 on implanted devices was not available in the text read), one national guideline module on hearing implants (Netherlands, 2024, read in Dutch), one systematic review and one overview on auditory implants (abstract level), a narrative review of neuromodulation, and clinical studies, case reports, a phantom study and manufacturer or institutional documents, nearly all at abstract level. No systematic review of MRI in deep brain, spinal cord or vagus nerve stimulation was read; a systematic review of MRI in vagus nerve stimulation appeared in search results but could not be accessed. No current manufacturer conditions are reproduced: they are specific to each model, they change, and they are held on the label and the manufacturer’s MRI guidance for the device in front of the operator.

What this page does not do. It gives no scanning conditions to be applied to a patient, because the label of the exact implanted system is the authority (Implant and Device MR Safety Framework). Statements that are the page’s own inference are marked Reasoning; where a source needed to complete a statement was not found they are marked (to be sourced).


2. Sources Read and What They State

Table 1. Documents used and how they were read.

DocumentUsed forLimits of the reading
MHRA guidelines, fifth edition 2026 [1]Hazard mechanisms for active implants; definitions; adherence to manufacturer instructionsSections 1.5, 2.4.2, 2.5 and 3.2 verbatim; Section 5.5 not available; no statement on the devices below
Dutch national guideline module on hearing implants, 2024 [2]Recommendations by implant category; summary of the evidence and its limitsRead in Dutch; last 1,343 characters and the table of manufacturer conditions not read
Systematic review, cochlear implants and MRI, 2026 [3]Pooled counts of adverse eventsAbstract only
Overview of active auditory implants, 2021 [4]Statement on labelling and adverse eventsAbstract only
Narrative review of neuromodulation and MRI [5]Cross-device picture; labelling history to about 2016No search method; snapshot of about 2016
Deep brain stimulation: prospective series [6], spine protocol study [7], phantom study [8], review [9], case report [10]Clinical and technical evidenceAbstract level; small numbers; single devices
Spinal cord and peripheral nerve stimulation: manufacturer compilation [11], impedance case series [12]Scope of labelled conditions; failure of the conditional modeAbstract level; not clinical outcome studies
Cochlear implant series [13], [14]Adverse events and complication rates in single centresAbstract level; 1.5 T mainly
Drug pumps: manufacturer guide [15], case report [16]Pump stall, steps before and after, baclofen riskManufacturer document; one case
Glucose monitors: institutional web page [17]Classification by one UK departmentNo published evidence cited; undated content

3. Hazards Common to Active Implants

The MHRA guidelines state that the effects of the MR environment can adversely affect medical equipment, including passive and active implantable medical devices, and name stimulation systems (deep brain, sacral nerve or vagus nerve stimulation) as examples; the effects listed are device malfunction (possible effects include device reset, battery drain and permanent damage, with serious consequences for patients who are dependent on the device), movement of the implant within the body with possible tissue damage, heating (gradient-induced eddy currents in implants), and unintended stimulation from voltages induced in leads or circuitry [1]. For devices with leads or electrodes the guidelines state that there is an increased potential for localised burns or SAR hot spots, and that the B1+rms limit is most often found in the labelling of MR Conditional implants [1]. They also state that the operating modes do not in isolation consider the risk from implants, and that for MR Conditional implants exposure risks are minimised by adhering to the manufacturer’s instructions [1]. The standard behind such labelling and the meaning of the label content are in Implant and Device MR Safety Framework.

Table 2. Hazard by device group and where the evidence is reviewed below.

Device groupDominant concern in the reports readSection
Deep brain stimulationElectrode heating; labelled conditions that are restrictive4
Spinal cord and peripheral nerve stimulationLabelled conditions that vary by manufacturer; failure of the conditional mode5
Vagus and sacral nerve stimulationLabelled conditions; little published evidence in the sources read6
Cochlear and other auditory implantsPain and displacement of the internal magnet7
Implanted drug infusion pumpsRotor stall and delayed restart; withdrawal with baclofen8
Glucose monitors and insulin deliveryClassification of each device; removal or off-label retention9

Reasoning. The hazards are the same four for every device in this page; what differs is which one dominates in the published reports. For stimulators with leads the evidence is mostly about heating and about whether a device can be put into its MR-conditional state; for cochlear implants it is about the internal magnet; for pumps it is about the mechanism of the motor; for glucose monitors it is about classification. This is the page’s own summary of the sections that follow.


4. Deep Brain Stimulation

A narrative review describes the labelling history: in 2002 the manufacturer required head-only transmit and a SAR of 0.4 W/kg, revised in 2005 to 0.1 W/kg, with extended labelling in 2015 for eligible devices that allows a B1+rms-based condition with body-coil transmit [5]. The same review reports that patient series described few or no adverse incidents attributable to the MR environment, and cites two published cases of harm when scanning was outside the prescribed conditions: a permanent neurological deficit and a transient dystonia [5]; the original reports were not read. An earlier review stated that manufacturer restrictions had been little tested and that very few detailed investigations of heating had been published [9].

Table 3. Evidence on deep brain stimulation read for this page.

SourceDesignResult as reportedLimits
Boutet 2019 [6]Prospective, 102 participants with active systems; 1.5 T or 3 T; T1-weighted and gradient-echo echo-planar imagingNo short- or long-term MRI-related adverse events, no acute changes; electrode-contact artefact about 9 mm; coil artefact obscured about 1.9 per cent of the brain, more at 3 T than at 1.5 T; authors: safe “with a priori local safety testing”Abstract only; number of scans, devices and testing not given; one research programme
Kahan 2015 [8]Phantom, bilateral system, fibre-optic thermometryElectrode-tip heating 0.45 to 0.79 °C (1.5 T) and 1.25 to 1.44 °C (3 T) from head to body transmit; below 1 °C at 1.5 T in every position tested; at 3 T cranial MRI can readily exceed guidelinesOne device and protocol; in vitro; no tissue measurement
Spine protocol study [7]Phantom, then 67 sequences in 9 participants at 1.5 THeating greatest for cervical, lowest for lumbar spine; sequences modified to keep rise below 2 °C; no acute or long-term adverse effects; protocol exceeded vendor slice limitsNine participants; one protocol; authors not shown on the record read
Hayley 2022 [10]Case report, four electrodes and two generatorsOff-label spinal MRI without adverse effectsOne patient
Review of heating [9]; narrative review [5]ReviewsRestrictions little tested in 2011; labelled conditions evolved to 2015; two reported harms outside prescribed conditionsAbstract level or snapshot of 2016

Reasoning. Zero events in 102 participants is compatible with a true event rate of up to about 3 per cent (rule of three applied by this page to participants, not to scans; the abstract gives neither the number of scans nor the devices). The series show that events were not seen under local safety testing; they do not show that scanning outside the labelled conditions is safe, and the phantom study shows that heating depends on field strength and transmit mode.


5. Spinal Cord and Peripheral Nerve Stimulation

A compilation of data collected by direct communication with ten neurostimulation companies up to December 2024 reports that full-body 1.5 T is standard across most spinal cord stimulation systems, that some systems permit 3 T, and that limits vary by manufacturer: spatial gradient limits of 1,000 to 4,000 gauss/cm, SAR limits and active scan times of 15 to 60 minutes; peripheral nerve stimulation systems show similar patterns [11]. The authors note that no central source of this information existed and that it will need continued updating [11]. The narrative review of about 2016 described several systems as labelled for head-transmit only at 1.5 T and others for whole-body transmit at 1.5 T, and reported safe MRI in two patient series of spinal cord stimulation [5]; the difference from the later compilation illustrates how fast labelling changes.

A label can be unusable in an individual patient. In a case series of three patients the spinal cord stimulator could not be placed in its MRI-conditional mode because of high impedance; all three needed alternative imaging (including CT) and two had the system explanted [12]. The authors state that the cause of high impedance is incompletely understood and that impedance changes can have device-related and physiological causes [12].

Reasoning. Because the conditional mode has to be enabled in the device before the scan, the label cannot be applied by the scanner operator alone: the check that the device can reach its MRI mode belongs to the pathway before the examination. The sources read do not describe who performs this check or how it is documented (to be sourced).


6. Vagus and Sacral Nerve Stimulation

The only source read that covers these devices is the narrative review of about 2016: vagus nerve stimulation labelled for 1.5 T or 3 T with a head-only transmit coil, an early report of 25 patients at 1.5 T without permanent adverse effects, and a sacral nerve system permitting head-only transmit at 1.5 T [5]. No counts or rates for sacral nerve stimulation are reported in that source, and the labelling is a snapshot that may have changed [5]. The evidence on these devices in the sources read is therefore limited to a labelling description and a single early series (to be sourced: a current review or the manufacturer conditions).


7. Cochlear and Other Auditory Implants

7.1 Pooled and Single-Centre Evidence

A systematic review of 14 observational studies (10 retrospective, 4 prospective; searches to March 2024) reports 814 MRIs in 668 implanted ears with 167 adverse events (20 per cent); pain accounted for 51 per cent of the events, magnet displacement for 22 per cent and polarity reversal for 2 per cent; no adverse events were reported with diametric magnets, events were more frequent with axial magnets, and most occurred at 1.5 T, a bias of the literature that limits comparison [3]. The abstract does not itemise the remaining quarter of events. Mitigating measures (head wrapping, anaesthesia, anxiolytics) were common and not uniformly effective, and paediatric patients were under-represented [3]. Figure 1 shows the composition.

Figure 1 — Adverse events after MRI in cochlear implant recipients: composition reported by a systematic review

Horizontal bars of the share of 167 adverse events reported in 814 MRIs of 668 cochlear implant ears: pain 51 per cent, magnet displacement 22 per cent, polarity reversal 2 per cent and a remaining 25 per cent not itemised in the abstract.

A prospective review of 14.5 years at one centre (97 patients, 428 sessions at 1.5 T) reports that magnets were removed in 28 sessions (6.5 per cent), that there were 15 complications (3.5 per cent of sessions), mostly discomfort, with five magnet dislocations and no implant failure or excessive demagnetisation, and the authors conclude that MRI was generally safe and well tolerated without magnet removal [14]. An earlier retrospective record review of 18 patients and 30 examinations reported that five patients with head bandages could not finish their scans because of pain, that one of them had magnet displacement, that another needed surgery to remove and reinsert the magnet, and that one patient had a polarity reversal [13]. An overview of the topic states that adverse events still occur although more MRI-conditional active hearing implants are available, and that older devices often have more restrictive labelling, which can cause confusion [4].

7.2 The Dutch National Guideline

The Dutch module (2024) recommends following the manufacturer’s MR conditions; for cochlear implants it distinguishes devices that can be scanned without a bandage (scheduled at 1.5 T or 3 T according to the question and the manufacturer’s conditions) from those that are scanned after a bandage and splint (local procedure), refers to the implanting centre when the magnet must be removed and no alternative imaging is possible, and recommends staff training and a written local procedure before bandaging; for auditory brainstem implants it refers fixed-magnet devices to the implanting centre; for percutaneous bone conduction implants it allows scanning at 1.5 T and 3 T without further conditions; for transcutaneous bone conduction and active middle ear implants it requires identification of supplier and model and adherence to the supplier’s conditions; and for passive middle ear implants it allows 1.5 T or 3 T without restrictions [2]. The guideline states that it found no comparative patient studies, so GRADE could not be applied; that the evidence is expert consensus, a systematic review of 38 studies (pain the most frequent adverse event, magnet dislocation the most frequently reported device problem) and four small non-comparative studies in which dislocation rates varied strongly by implant type (about 30 per cent with one implant series in two studies, none in about 40 procedures in a third); that the incidence of risks cannot be estimated because the number of MRIs performed is unknown; and that the bandage recommendations rest on expert experience [2].

Table 4. Cochlear implant evidence side by side.

SourceDenominatorResult as reportedLimits
Systematic review 2026 [3]814 MRIs, 668 ears, 14 studies167 events (20 per cent); pain 51 per cent, displacement 22 per cent, reversal 2 per cent; none with diametric magnetsAbstract only; mostly 1.5 T; observational
Tam 2020 [14]428 sessions, 97 patients (cochlear and brainstem implants)15 complications (3.5 per cent of sessions); five dislocations; no device failureOne centre; 1.5 T only; abstract only
Kim 2015 [13]18 patients, 30 examinationsFive patients with bandages could not finish (pain); one displacement; surgery to reinsert a magnet in another; one reversalVery small; older devices; abstract only
Dutch guideline 2024 [2]Reported evidence of other authorsDislocation about 30 per cent in two studies of one implant series; none in about 40 procedures in a third; incidence cannot be estimatedTable of conditions not read; guideline reports the studies at second hand

Reasoning. The sources do not agree on a single rate because they count differently (events per implanted ear, complications per session, patients unable to finish), cover different devices and eras, and differ in whether a bandage was used. What they agree on is the ranking: pain and magnet displacement dominate, axial magnets are involved more than diametric ones in the pooled data, and harm to hearing was not reported in the series read. The review of single-centre series cannot say how many displacements went unreported.


8. Implanted Drug Infusion Pumps

A manufacturer guide for one family of programmable intrathecal pumps states conditions for 1.5 T and 3 T cylindrical scanners, a maximum spatial field gradient of 19 T/m, a slew rate of 200 T/m/s or less per axis, the First Level Controlled Operating Mode, and that the risk of heating increases for active torso scanning over 30 minutes [15]. The guide describes motor stall during the scan, with recovery typically detected within 20 minutes of the end of the scan and, for pumps delivering at least 0.048 mL/day, restart of infusion possibly delayed by up to 24 hours; states that a pump aligned 90° to the scanner’s z-axis carries a risk of permanent motor stall; advises deciding before the scan whether the patient can go without drug delivery during it, that stopping the pump is not recommended, and that the current infusion parameters be recorded; requires that all pumps be interrogated after the scan, with a second interrogation possibly needed to detect stall recovery; and warns that baclofen withdrawal can be life-threatening [15].

A case report adds a patient with several MRIs and a pump of the same family who had two memory failures with withdrawal symptoms (one resolved by a memory reset, the second needing pump replacement 29 months later); the authors state that MRI can temporarily stop the rotor and that every patient with an implantable drug delivery system should have the pump interrogated after MRI, with particular attention to baclofen [16].

Reasoning. The guide applies only to the models it names; other pumps (including those used for other drugs or with other mechanisms) have their own conditions. The consequence of a stalled or reset pump depends on the drug, which is why the same label-based pathway needs a drug-specific step. No independent study of pump function after MRI was read (to be sourced), and the case report cannot show how common memory errors are.


9. Glucose Monitors and Insulin Delivery Devices

The only source read on these devices is a web page of one UK department: one implanted-sensor system is classified as MR Conditional in the UK with its external transmitter MR Unsafe and to be removed; a second sensor system is not MR Conditional in the UK but may stay in place off-label under a local risk assessment and signed consent, citing the manufacturer’s US conditions; all other glucose monitors are classified MR Unsafe and are to be removed before scanning [17]. The page states that the risk of keeping the second system in place “has been demonstrated to be low” without giving the data, cites no published evidence, and does not mention insulin pumps [17].

No published study, guideline or manufacturer document on insulin pumps was read for this page (to be sourced). The MHRA text read does not mention them [1]. Because these devices are worn on the body and are normally removable, the usual question is whether and how they are removed and reapplied, which the sources read do not settle.


10. Practice: the Pathway the Sources Support

Reasoning. The following arrangement is the page’s synthesis of statements in the sources and is not itself a recommendation of any one source. (1) Identify every implanted component and its model, including leads, extensions and any abandoned parts, and read the label for the complete system rather than for the generator alone (Implant and Device MR Safety Framework; the MHRA states that the manufacturer’s instructions are the basis of risk minimisation [1]). (2) Confirm that the device can be put into its MRI mode and that the conditions of the label can be met for the intended anatomy; the label may be unusable in the individual patient [12]. (3) Apply the device-specific steps required before and after the scan: for drug pumps the recording of infusion settings and the interrogation afterwards [15]; for cochlear implants the local bandaging procedure and the patient information on discomfort [2]. (4) Treat any scan outside the label as an exceptional, documented risk-benefit decision, noting that published off-label scanning without harm exists but cannot establish safety for others [7, 10]. The sources read do not say who performs each step or how it is recorded (to be sourced).


11. Where the Documents Differ and Open Questions

Table 5. Differences and unresolved questions.

TopicPosition APosition BStatus
Deep brain stimulation at 3 TCranial MRI at 3 T in a phantom can readily exceed heating guidelines [8]102 participants scanned at 1.5 T or 3 T without events under local safety testing [6]Not directly comparable (phantom, one device and protocol versus clinical series under local testing); device and protocol specific
Scanning beyond vendor limitsProtocol exceeded vendor slice limits in nine participants without events [7]Manufacturer limits remain the label conditions [1]Small data against a label; no consensus read
Cochlear implant displacementAbout 30 per cent in two studies of one implant series, none in about 40 procedures in a third (as reported in the guideline) [2]Five dislocations in 428 sessions [14]; 22 per cent of events in a pooled review [3]Different denominators and devices; cannot be reconciled with the sources read
Magnet removalMagnets were removed in 6.5 per cent of sessions and complications were few without removal [14]Bandaging with pain and displacement despite bandages [13]Practice differs by centre; the guideline defers to local procedure [2]
Labelling over timeMany neurostimulators labelled head-transmit only around 2016 [5]Full-body 1.5 T standard for most spinal cord systems by 2024 [11]Labelling changes; compare only against the current label

12. Limits of the Evidence and What Was Not Read

The evidence is mostly single-centre series, case reports, phantom studies and manufacturer documents; one systematic review (cochlear implants) was read at abstract level, and no systematic review for the other device groups was read. MHRA Section 5.5, the ACR Manual on MR Safety, ISO/TS 10974 and the manufacturers’ current conditions were not read. Events in small series cannot establish the absence of risk, passive reports and unreported displacements are not captured, and the evidence is concentrated at 1.5 T. The statements on labelling are snapshots and must not be applied without the current label of the exact system.


13. Topics Not Covered

Label content, test standards and matching a label to the scanner (Implant and Device MR Safety Framework); cardiac devices (Cardiac Implantable Electronic Devices in MRI); heating of conductors and burns (RF Heating and Burns); passive implants, aneurysm clips and foreign bodies (Passive Implants and Retained Foreign Bodies in MRI); pregnancy and staff exposure (MRI in Pregnancy and the Pregnant Worker); sedation, monitoring and the care of a patient with an implant during the scan (Claustrophobia, Sedation, Monitoring, Equipment and Emergencies in the MR Environment); deep brain stimulation research at ultra-high field; programming and stimulation-related effects of the scan on symptoms; and the rules of individual jurisdictions (Incident Learning, Governance, Quality and Jurisdictions in MR Safety).


14. Evidence-Based References

One systematic review (cochlear implants) and one overview were read at abstract level; no systematic review of MRI in deep brain, spinal cord or vagus nerve stimulation was read. The sections below follow the hierarchy of the sources actually read.

A. Guidelines, Consensus, Standards and Regulation

Guideline (regulatory, UK; Sections 1.5, 2.4.2, 2.5 and 3.2 read verbatim)
[1] Medicines and Healthcare products Regulatory Agency. Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use. Fifth edition, July 2026 (revision history on the pages read lists versions 5.0.1 to 5.0.4).
Relevance: Cited in Sections 2, 3, 9, 10, 11. Definitions of MR Safe, MR Conditional and MR Unsafe (Section 1.5), the statement on leads and electrodes and on B1+rms in labelling (Section 2.4.2), the effects on active and passive implants including stimulation systems (Section 2.5 and 2.5.1 to 2.5.4) and the statement that exposure risks are minimised by adhering to the manufacturer’s instructions (Section 3.2). The text available ended inside Section 4.1.4, so Section 5.5 (implanted medical devices and other potential contraindications) was not read; the text read contains no statement on cochlear implants, drug pumps or insulin pumps.
Guideline module (national, Netherlands; page read nearly in full, in Dutch)
[2] Nederlandse Vereniging voor Keel-Neus-Oorheelkunde en Heelkunde van het Hoofd-Halsgebied (NVKF). Gebruik MRI bij patiënten met implantaten: module MRI bij gehoorimplantaat. Richtlijnendatabase, published 1 May 2024.
Relevance: Cited in Sections 2, 7, 10, 11. Recommendations for cochlear, auditory brainstem, bone conduction and middle ear implants, the evidence summary (no comparative patient studies; GRADE not applied; a systematic review of 38 studies; four small non-comparative studies; variable dislocation rates) and the considerations stated. The last 1,343 characters of the page were not read, and the table of manufacturer conditions (Table 4) was embedded as images and could not be read; the guideline’s own limits are reported on the page. A national guideline of one specialty society; the recommendations on bandaging rest on expert experience.

B. Systematic Reviews, Overviews and Narrative Reviews

Systematic review of observational studies (abstract read)
[3] Mak LE, Pai V, Heyn C, Symons S, Maralani PJ. Systematic review of the adverse events in patients with cochlear implants following MRI. AJNR Am J Neuroradiol. 2026;47(6):1643-1647. doi:10.3174/ajnr.A9142.
Relevance: Cited in Sections 2, 7, 11. The most recent systematic review of cochlear implants and MRI in the sources read: 14 observational studies (10 retrospective, 4 prospective), 814 MRIs in 668 implanted ears, 167 adverse events (20 per cent); pain 51 per cent of events, magnet displacement 22 per cent, polarity reversal 2 per cent; no events with diametric magnets; events more frequent with axial magnets and mostly at 1.5 T. Abstract only; the full text, with device detail, was not accessible. The abstract does not itemise the remaining quarter of events, and the authors state that the literature’s bias toward 1.5 T limits comparison.
Overview with a systematic review of reported adverse events (abstract read)
[4] Fierens G, Standaert N, Peeters R, Glorieux C, Verhaert N. Safety of active auditory implants in magnetic resonance imaging. J Otol. 2021;16(3):185-198.
Relevance: Cited in Sections 2, 7. Used only for what the abstract states: adverse events are still reported although more MRI-conditional active hearing implants are available, and older devices often have more restrictive labelling, which can cause confusion. The abstract gives no counts; the figures of the review (38 studies) are quoted here only as reported in the Dutch guideline.
Narrative review (read through a summarising tool; journal details from the university repository record)
[5] Thornton JS. Technical challenges and safety of magnetic resonance imaging with in situ neuromodulation from spine to brain. Eur J Paediatr Neurol. 2016;21(1):232-241. doi:10.1016/j.ejpn.2016.06.001.
Relevance: Cited in Sections 2, 4, 5, 6, 11. The only source read that covers deep brain, spinal cord, vagus nerve and sacral nerve stimulation together: the history of the deep brain stimulation labelling (head-only transmit; SAR limits of 0.4 and then 0.1 W/kg; extended labelling in 2015), labelled conditions by device type as they stood around 2016, qualitative statements on adverse events including two published cases outside prescribed conditions, and the author’s conclusion. No search strategy or inclusion criteria are described; no rates are given; the labelling is a snapshot of about 2016 and may have changed; one value in the summarised text (a B1+rms limit given in tesla) was evidently a unit error and is not used.
Review article (abstract read)
[9] Gupte AA, Shrivastava D, Spaniol MA, Abosch A. MRI-related heating near deep brain stimulation electrodes: more data are needed. Stereotact Funct Neurosurg. 2011;89(3):131-140.
Relevance: Cited in Sections 2, 4. Used only for the authors’ statement that manufacturer restrictions on MRI in deep brain stimulation had been little tested at the time and that very few detailed investigations had been published. Abstract only; the full text was behind a paywall; published in 2011 and not an indication of the current evidence.

C. Clinical Studies, Case Series and Case Reports (read at abstract level)

Prospective clinical study (abstract read)
[6] Boutet A, Rashid T, Hancu I, Elias GJB, Gramer RM, Germann J, Dimarzio M, Li B, Paramanandam V, Prasad S, Ranjan M, Coblentz A, Gwun D, Chow CT, Maciel R, Soh D, Fiveland E, Hodaie M, Kalia SK, Fasano A, Kucharczyk W, Pilitsis J, Lozano AM. Functional MRI safety and artifacts during deep brain stimulation: experience in 102 patients. Radiology. 2019;293(1):174-183. doi:10.1148/radiol.2019190546.
Relevance: Cited in Sections 2, 4, 11. The largest prospective clinical series on deep brain stimulation in the sources read: 102 participants with active systems scanned at 1.5 T or 3 T (T1-weighted and gradient-echo echo-planar imaging) between June 2017 and October 2018; no short- or long-term MRI-related adverse events and no acute changes; artefact size by field strength; the authors’ conclusion that deep brain stimulation patients may safely undergo 1.5-T and 3-T MRI “with a priori local safety testing”. Abstract only; the number of scans, the devices and the local testing are not given in the abstract; single research programme; a series without events cannot show that an event rate is zero.
Prospective pilot study with a phantom stage (abstract read; authors not shown on the page read)
[7] Safety assessment of spine MRI in deep brain stimulation patients. J Neurosurg Spine. 2020 (published online 14 February 2020; print 1 June 2020). ClinicalTrials.gov NCT03753945. (The authors were not shown on the page read and are not given here; volume and pages were not shown.)
Relevance: Cited in Sections 2, 4, 10, 11. Heating of a deep brain stimulation device during 1.5-T spine imaging in a phantom (greatest for cervical, lowest for lumbar); sequences raising the temperature by more than 2 °C were modified (3D to 2D, fewer slices); the modified protocol was used for 67 sequences in 9 participants with no acute or long-term adverse effects, stable impedances and no acute brain MRI findings, although it still exceeded the vendor slice limits. Abstract only; nine participants; a single protocol and device system.
Case report (abstract read)
[10] Hayley J, Hart MG, Mostofi A, Morgante F, Pereira EA. No adverse effects following off-label magnetic resonance imaging in a patient with two deep brain stimulation systems: a case report. Stereotact Funct Neurosurg. 2022;100(4):253-258. doi:10.1159/000525538.
Relevance: Cited in Sections 2, 4, 10. One patient with four electrodes and two pulse generators, scanned (spine) off-label without regard for manufacturer guidance with no adverse effects; the authors suggest the guidance might be overly restrictive. Used only to show that off-label scanning has been reported without harm and cannot establish safety for others. Abstract only.
Case series of three patients with a literature review (abstract read)
[12] Jotwani R, Abd-Elsayed A, Villegas K, Shakil A, Gulati A, Sayed D, Lam C, Mehta N. Failure of SCS MR-conditional modes due to high impedance: a review of literature and case series. Pain Ther. 2021;10:729-737 (published online 21 November 2020).
Relevance: Cited in Sections 2, 5, 10. Three patients in whom the spinal cord stimulator could not be placed in its MRI-conditional mode because of high impedance; all needed alternative imaging, two had systems explanted. Used to show that an MR Conditional label can be unusable in an individual patient. Three patients, retrospective; the authors state that the cause of high impedance is incompletely understood.
Retrospective single-centre record review (abstract read)
[13] Kim BG, Kim JW, Park JJ, Kim SH, Kim HN, Choi JY. Adverse events and discomfort during magnetic resonance imaging in cochlear implant recipients. JAMA Otolaryngol Head Neck Surg. 2015;141(1):45-52. doi:10.1001/jamaoto.2014.2926.
Relevance: Cited in Sections 2, 7, 11. 18 patients and 30 examinations between September 2003 and February 2014 (16 patients at 1.5 T, 2 at 3.0 T); 13 patients (25 examinations) completed their scans; five patients with head bandages could not finish because of pain, one of these with magnet displacement, another needing surgery to remove and reinsert the magnet, and one patient with polarity reversal; hearing performance unaffected in those with major events; magnet artefacts compromised diagnosis of ipsilateral brain lesions at 1.5 T. Abstract only; very small, retrospective, older devices; the figures on pain, displacement and reversal could not be separated from the abstract wording.
Prospective review of a clinical series (abstract read)
[14] Tam YC, Lee JWY, Gair J, Jackson C, Donnelly NP, Tysome JR, Axon PR, Bance ML. Performing MRI scans on cochlear implant and auditory brainstem implant recipients: review of 14.5 years experience. Otol Neurotol. 2020 (issue 5):e556-e562.
Relevance: Cited in Sections 2, 7, 11. 97 patients (21 auditory brainstem implant, 76 cochlear implant, 23 bilateral), 428 MRI sessions at 1.5 T over 14.5 years; magnets removed in 28 sessions (6.5 per cent) and left in place in 400; 15 complications (3.5 per cent of sessions), mostly discomfort, five magnet dislocations; no implant failure or excessive demagnetisation; artefact sizes with and without the magnet. Abstract only; a single tertiary centre, 1.5 T only; the volume number of the journal was not shown on the record read and is not given here.
Case report (abstract read)
[16] Kosturakis A, Gebhardt R. SynchroMed II intrathecal pump memory errors due to repeated magnetic resonance imaging. Pain Physician. 2012;15(6):475-477.
Relevance: Cited in Sections 2, 8. One patient with a SynchroMed II pump and several MRI examinations who had two memory failures with withdrawal symptoms; the authors state that MRI can temporarily stop the rotor and that all implantable drug delivery system patients should have the pump interrogated after MRI, with particular attention to baclofen. A single case of one device in 2012; causation by repeated MRI is the authors’ attribution.

D. Technical Studies and Manufacturer or Institutional Documents

In vitro (phantom) heating study (abstract read)
[8] Kahan J, Papadaki A, White M, Mancini L, Yousry T, Zrinzo L, Limousin P, Hariz M, Foltynie T, Thornton J. The safety of using body-transmit MRI in patients with implanted deep brain stimulation devices. PLoS One. 2015;10(6):e0129077.
Relevance: Cited in Sections 2, 4, 11. Bilateral deep brain stimulation phantom with a specific pulse generator and electrodes, fibre-optic thermometry; electrode-tip heating rose from 0.45 to 0.79 °C at 1.5 T and from 1.25 to 1.44 °C at 3 T when changing from head to body transmit; the greatest heating at 1.5 T in any position remained below 1 °C; the authors conclude that body-transmit cranial MRI at 1.5 T with their protocol does not produce heating above international guidelines and that cranial MRI at 3 T can readily do so. Abstract only; one device and one protocol; the study is in vitro and does not measure heating in tissue.
Compilation of manufacturer data (abstract read)
[11] Vu PD, Slitzky M, Miller K, Yong RJ, Robinson CL, Chung M. Provision of up-to-date magnetic resonance imaging conditionality for current peripheral nerve stimulation and spinal cord stimulation systems in pain management. Pain Med. 2025;26(12):924-929. doi:10.1093/pm/pnaf076.
Relevance: Cited in Sections 2, 5, 11. Data from 10 neurostimulation companies up to December 2024: full-body 1.5 T is standard across most spinal cord stimulation systems and some allow 3 T; limits vary by manufacturer, with spatial gradient limits of 1,000 to 4,000 gauss/cm, SAR limits and active scan times of 15 to 60 minutes; peripheral nerve stimulation systems show similar patterns. Abstract only; not a clinical study; the conditions printed in the paper change as devices change and are not reproduced here.
Manufacturer MRI safety guide (read through a summarising tool)
[15] Medtronic. SynchroMed II and III Infusion System MRI Safety Guide (document code 2025-synchromed-3-tesla-mri-en-gb-emea-17476373; dated only by a 2025 copyright notice).
Relevance: Cited in Sections 2, 8, 10. Conditions for the named pumps (1.5 T and 3 T cylindrical scanners, maximum spatial field gradient 19 T/m, slew rate 200 T/m/s or less per axis, First Level Controlled Operating Mode, heating risk increasing for active torso scanning over 30 minutes), motor stall and its recovery, steps before and after the scan, orientation risk, the warning on baclofen withdrawal. Manufacturer document: it applies only to the named products and models; no independent evaluation; the full document was read only through a summary.
Institutional web page (NHS Greater Glasgow and Clyde MRI Physics; read through a summarising tool)
[17] NHS Greater Glasgow and Clyde MRI Physics. Continuous glucose monitors and MRI (web page, copyright notice 2016, undated content; contact address of the MR safety expert).
Relevance: Cited in Sections 2, 9. Used only to show how one UK department classifies continuous glucose monitors: one sensor system MR Conditional in the UK with its transmitter MR Unsafe, one system not MR Conditional in the UK but allowed to stay in place off-label under a local risk assessment and signed consent, and all others MR Unsafe and to be removed. The page cites no published evidence (internal documents and a manufacturer web page only) and does not mention insulin pumps; the conditions may have changed since the page was read.

End of document — Active Implants Other Than Cardiac Devices (S7) — MRIninja v1.0 — October 2026. Parent: MRI Safety master.