Incident Learning, Governance, Quality and Jurisdictions in MR Safety — What the Surveys Show, What Each Framework Requires of Reporting, Roles and Training, and Where the Documents Differ

up to this point verified by human experts

Incident Learning, Governance, Quality and Jurisdictions in MR Safety — What the Surveys Show, What Each Framework Requires of Reporting, Roles and Training, and Where the Documents Differ

MRIninja Knowledge Base | Child Page — MRI Safety (deep dive S13) Version 1.0 — October 2026

Parent page: MRI Safety — Master Page (governance and personnel in Section 5; occupational exposure in Section 8.3). Builds on: MR Site Zones and Access Control (responsibility for access policy) · Projectile and Torque Hazards in the Static Field (incident evidence on projectiles) · Claustrophobia, Sedation, Monitoring, Equipment and Emergencies (emergencies) · Cryogens, Quench and Oxygen Depletion in MRI (quench)


1. Scope, Status and Boundaries

This page is the last of the deep dives under the MRI Safety master. It gathers what the sources read state on four questions: what is known about the frequency and the under-reporting of MR safety incidents; what is reported, to whom and when under the frameworks that could be read; which roles, training and audit the frameworks describe beyond the ACR and MHRA role comparison already held in MRI Safety — Master Page; and where the frameworks differ. The roles of the ACR and the MHRA (MR Medical Director, MR Safety Officer, MR Safety Expert, MR Responsible Person and others) are not repeated; the comparison of the ACR and MHRA positions is in MRI Safety — Master Page (Section 5), and Directive 2013/35/EU (Article 10 and its conditions for MRI) is in MRI Safety — Master Page (Section 8.3).

Reading status. The page rests on the full text of the 2017 RANZCR guideline (through a summarising tool of a PDF copy), a review tool that restates the Joint Commission Sentinel Event Alert #38 and the 2013 ACR guidance, the web pages of the US FDA (device reporting), the UK HSE (exemption for MRI), the NHS (reporting to the MHRA) and Health Canada (archived Safety Code 26), a poster of a UK professional society, two European surveys (abstracts), one Swedish national survey (abstract and text) and one analysis of US adverse-event reports. The regulations themselves (21 CFR 803, the UK Control of Electromagnetic Fields at Work Regulations 2016, national laws transposing Directive 2013/35/EU), the MHRA guideline chapters on reporting and training, the ACR Manual on MR Safety, the 2021 (third) version of the RANZCR guideline, the legislation or guidance of Japan, Germany, France, Italy and other countries, and the content of the four-tier European training curriculum were not read. No incident database with denominators, no study of the effect of a reporting system on harm, and no audit study of MR safety programmes were found.

What this page does not do. It is not a legal analysis and does not say what the law of any country requires; a unit applies the law and the guidance that bind it. No jurisdiction is treated as the default. Statements that are the page’s own inference are marked Reasoning; where a source needed to complete a statement was not found they are marked (to be sourced).


2. Sources Read and What They State

Table 1. Documents used, what they were used for and how they were read.

DocumentUsed forLimits of the reading
RANZCR MRI Safety Guidelines, Version 2.0, 2017 [1]Roles, training, screening, reporting within 24 hours, auditFull text through a summarising tool; garbled passages; Version 3 (2021) not opened
Review tool on Sentinel Event Alert #38 and the ACR guidance 2013 [2]Objectives on safety officer, written policies, annual education; reporting to the vendor and the FDASummary; co-authored by an author of the guidance; no counts of events
FDA page on mandatory device reporting [3]Reporting duties of user facilities and manufacturers (United States)Web page; definitions and the regulation not read
NHS page on reporting to the MHRA [4]; professional-society poster [5]Reportability criteria, routes and categories (United Kingdom)Written for adopters of devices generally; poster; MHRA MRI chapters not read
HSE page on the MRI exemption [6]Occupational exposure limits and MRI (United Kingdom)Web page, not the legal text
Health Canada Safety Code 26 [7]Legal status and levels of an archived national guidelineArchived; nothing on management or reporting
Swedish national survey [8]; European surveys [9][10]; analysis of US reports [11]Frequency, under-reporting, training, regulationSelf-reported surveys; abstracts for the European surveys; voluntary reports without denominator

3. What the Incident Evidence Shows

A Swedish national web survey covering 12 months collected reports from 529 eligible participants of 200 MR safety incidents and 156 CT safety incidents (human injury, material damage and close calls); 16% of the MR incidents received the highest potential severity score, and the authors concluded that incidents occur in clinical MR environments and that the assessed risk level is high [8]. A European survey distributed through ESR member societies in 2023 received 793 responses from 44 countries: 77% said their institution had a clear MR safety guideline, 52% said nobody there had received specific MR safety training, about 12% reported MR-related accidents causing injury to patients or personnel, MR Conditional cardiac devices were scanned at 75% of respondents’ university hospitals and 42% of outpatient facilities, and MR Unsafe cardiac devices were scanned off-label at 27% of university hospitals [9]. The analysis of US adverse-event reports is a set of spontaneous voluntary reports without a denominator [11]; its narratives on projectiles are on Projectile and Torque Hazards in the Static Field. Figure 1 shows some of the percentages.

Figure 1 — Selected self-reported results of two surveys (not comparable with each other)

Horizontal bars of percentages. Swedish survey: 16 percent of MR incidents had the highest potential severity score; 73 percent of MR workers (50 percent of CT workers) were confident they would be aware of any incident; 69 percent of MR workers (83 percent of CT workers) were not aware of reported incidents at their hospitals. European survey: 77 percent had a clear MR safety guideline; 52 percent said nobody had specific MR safety training; about 12 percent reported MR-related accidents with injury.

Reasoning. None of these sources gives a rate of incidents per examination, a rate by type of incident, or a trend over time. The two surveys differ in country, year, respondents and wording, so their percentages cannot be compared or combined; each respondent counted incidents or accidents by its own definition (the definitions were not retrieved). The sources support the statements that incidents with injury and with high potential severity are reported by respondents in two settings, and that reported training and guidelines were incomplete in the European sample; they do not support an estimate of how frequent incidents are.


4. Under-Reporting and Confidence in Reporting

In the Swedish survey, 73% of MR workers (50% of CT workers) were confident they would be aware of any incident at their workplace, yet 69% of MR workers (83% of CT workers) were not aware of reported incidents at their hospitals; the authors describe a false sense of security [8]. A poster of a UK professional society states that MR incidents are widely under-reported across Europe, citing a multicentre interview survey and a three-year review of a UK independent-sector provider; the poster gives no figures and the two cited studies were not read [5]. The US adverse-event analysis is limited by voluntary reporting and gives no estimate of under-reporting [11].

Reasoning. The Swedish figures are a measure of what staff believe about their reporting system and not of how many incidents go unreported: they show that many respondents did not know of incidents that had been reported at their own hospitals, which suggests that learning from reports was not reaching them. The sources read do not state the proportion of incidents that are reported, and none evaluates a reporting system by its effect on later incidents (to be sourced).


5. What Is Reported, to Whom and When

The documents read name different recipients and different deadlines. They are of different kinds (a regulator’s requirement, a guidance page for adopters, a professional guideline), so the table gives each in the terms of its source and does not equate them.

Table 2. Reporting of MR-related incidents, by source.

SourceWho reports whatTo whom and whenStatus as statedLimits
United States, FDA [3]User facility: device-related death; device-related serious injury; malfunctions optionalDeath: FDA and manufacturer, within 10 work days; serious injury: manufacturer (FDA only if the manufacturer is unknown), within 10 work days; annual summary by January 1 (Form FDA 3419)Described as mandatory reportingDefinitions and the regulation not read
United States, review tool [2]All MR accidents and near misses, in the 2013 ACR guidance table (Section A.4)Vendor and FDA; no deadline stated in the summaryGuidanceSummary; the table itself not seen
United Kingdom, NHS page [4]Adverse incident that occurred, with the device suspected to be a contributory cause, and death or serious deterioration in health resulted or might have resultedAs soon as possible; Yellow Card (England and Wales); Health Facilities Scotland; Northern Ireland: contact the MHRADevelopers have a legal duty; for adopters described as best practice, not legally requiredNot MRI-specific; MHRA MRI chapters not read
United Kingdom, professional-society poster [5]MR incident categories: burns, projectiles, mechanical failures, cryogen and quench, injector failures, noise, major artefacts, device function changes, defective medicines, adverse drug reactionsInternally by local policy; to the MHRA by Yellow CardPosterDoes not say which incidents go internally only
Australia and New Zealand, RANZCR 2017 [1]Incidents, through a reporting system involving the Responsible Person or Safety OfficerWithin 24 hours; monitored; documented responses; committee reviews root causes where one existsNon-statutory professional guidelineExternal recipients not stated in the summary; Version 3 not read
Canada, Safety Code 26 [7]Nothing statedNot applicableArchived non-binding guidanceNo content on reporting
Europe, survey [9]Not applicableNot applicableRespondents from five countries said MR safety is mandated by law; Italy and Austria confirmedNo information on reporting duties in the abstract

Reasoning. The sources do not say, for any of these systems, which MR incidents fall within a legal reporting duty: the FDA page defines the duty for “device-related” deaths and serious injuries without saying how an MR accident such as a projectile or a burn is classified, and the NHS page gives criteria that require suspicion of the device as a contributory cause. A unit should establish with its regulator and its manufacturer how each type of incident is classified (to be sourced). The existence of an internal reporting system is stated by the Australian and New Zealand guideline; the other sources read describe external routes and do not state whether an internal system must exist.


6. Roles, Training and Screening Beyond the ACR and MHRA Comparison

The 2017 RANZCR guideline states that responsibility for the safe operation of the MRI site must be explicitly assigned to a nominated medical practitioner (the Responsible Person, the MRI Medical Director), that an MRI Safety Officer handles day-to-day implementation and need not be a medical practitioner, that an MRI Safety Expert, often from another practice, provides third-party advice and audits, and that a safety committee is optional for larger sites [1]. It describes three levels of training (Level I: controlled-area rules, hearing protection, emergencies, quench procedures and magnet screening; Level II: electrical, static, gradient and RF safety, exposure limits, cryogens and implant conditions; Level III: bio-effects and contrast agent risks), states that junior personnel need Level I and senior personnel Levels I and II, and that all training must be documented with at least an annual refresher [1]. It recommends three safety screenings, of which at least two on site (at booking, on arrival and immediately before the scan), reviewed with the patient by a senior staff member who initials the checklist, with annual screening of MRI staff [1].

The review tool of the Joint Commission alert lists as objectives that all MR personnel receive safety education at least annually, that non-emergent patients are screened twice by trained personnel, that a safety officer is appointed, and that written MR safety policies are kept and their compliance reviewed periodically [2]. The European survey proposes a four-tier curriculum for MR safety training; its content was not read [9]. Health Canada’s archived Safety Code 26 contains no content on safety management [7].

Reasoning. The frameworks agree on a named accountable function, annual training and repeated screening, and differ on the qualification of the accountable person, on the number of screenings and on how training is graded. The sources give no evidence that any of these arrangements reduces incidents; the guidance is consensus without stated evidence grading, and the European survey reports that in its sample 52% had no specific training [9]. The responsibility for access policy is on MR Site Zones and Access Control (Section 12.3).


7. Occupational Exposure Rules

The conditions of Directive 2013/35/EU for MRI are in MRI Safety — Master Page (Section 8.3). The UK Health and Safety Executive page states that the exemption from the exposure limits of the Control of Electromagnetic Fields at Work Regulations covers work with MRI equipment for patients in the health sector, including development, testing, installation, use, maintenance and related research, on condition that exposure is kept as low as reasonably practicable and employees are protected against the related health effects and safety risks; employers must comply with all other requirements except the action plan, and MRI not used for patients in the health sector falls outside the exemption [6]. The archived Canadian Safety Code 26 stated that operators should not be continuously exposed to more than 0.01 T during the working day, with higher levels allowed for about 10 minutes per hour; it described its levels as not strict limits, and Health Canada archived it as no longer reflecting current MRI technologies [7].

Reasoning. The exemption described by the HSE is conditional on a duty to reduce exposure, not an absence of duty, and its condition is stated in general terms; the sources read do not give a numerical limit for staff in either the HSE page or the Directive summary held on MRI Safety — Master Page. The Canadian values are those of an archived guideline and should not be read as a current limit. National laws transposing the Directive, and the rules of countries outside the United Kingdom and Canada, were not read (to be sourced).


8. Quality Assurance and Audit

The RANZCR guideline states that the safe practice manual is reviewed at least annually and after hardware or major software changes, that periodic quality assurance and screening audits are required, that records are reviewed annually and that maintenance records should comply with accreditation standards [1]. The review tool lists periodic review of compliance with written policies [2]. A European survey on MRI quality assurance and quality control (269 responses from 37 countries, 2024) found that only a few countries mandate national QA/QC legislation, that many rely on voluntary guidelines or lack formal protocols, and that respondents agreed on the need for harmonised guidelines; it concerns image and system quality rather than safety incidents [10].

Reasoning. The documents read describe audit as a periodic review of policy, screening records and maintenance; they do not give an audit method, an indicator, or a standard against which an audit is scored, and none reports an audit result (to be sourced). The survey of quality assurance shows heterogeneity, not an association with safety outcomes.


9. The Frameworks Side by Side

No framework is treated as the default. Each cell states what the source read says; “not read” means the source exists but was not available in the text used.

Table 3. Frameworks side by side (as read).

ItemUnited States (alert, ACR 2013, FDA)United Kingdom (MHRA, HSE, NHS)Australia and New Zealand (RANZCR 2017)Canada (Safety Code 26, archived)
Legal statusGuidance (ACR); FDA device reporting described as mandatory [2][3]MHRA guideline: roles in MRI Safety — Master Page; occupational regulation with an MRI exemption [6]Non-statutory professional guideline [1]Non-binding; archived [7]
Accountable rolesSee MRI Safety — Master Page (ACR)See MRI Safety — Master Page (MHRA)Responsible Person (medical practitioner), Safety Officer, Safety Expert, optional committee [1]None stated
TrainingAnnual education for MR personnel [2]Appendix not readLevels I to III, documented, annual refresher [1]None stated
Screening of patientsTwo screenings for non-emergent patients [2]See MRI Safety — Master PageThree recommended, at least two on site [1]None stated
Incident reportingDevice-related deaths and serious injuries to manufacturer and FDA within 10 work days; accidents and near misses to vendor and FDA in the 2013 table [3][2]Yellow Card; “as soon as possible”; legal duty on developers; adopters “not legally required” per the NHS page [4]Internal reporting within 24 hours [1]None stated
AuditPeriodic compliance review [2]Chapters not read (the quench page lists items an audit may check)Annual manual review; periodic QA and screening audits [1]None stated
Staff exposureSee MRI Safety — Master PageExemption from limits on conditions [6]Section on occupational exposure not summarised in the text read0.01 T continuous; higher for short periods (archived) [7]

Reasoning. The rows are not equivalent in weight: two are professional guidelines, one is a regulator’s requirement for device events, and one is an archived guideline. Japan, Germany, France, Italy and most other countries are absent because their rules were not read; the European survey states only that respondents from five countries said MR safety was mandated by law, of which two could be confirmed [9].


10. Where the Documents Differ and Open Questions

Table 4. Differences and unresolved questions.

TopicPosition APosition BStatus
Number of screeningsTwo for non-emergent patients [2]Three recommended, at least two on site [1]Different documents; no comparative data
Time to reportWithin 24 hours to the Responsible Person or Safety Officer [1]Within 10 work days to the manufacturer (and FDA for death) [3]; “as soon as possible” in the UK [4]Different recipients and purposes; internal and external deadlines are not alternatives
Qualification of the safety officerSafety Officer need not be a medical practitioner; the Responsible Person must be [1]Roles of the ACR and MHRA in MRI Safety — Master PageDifferent structures; no outcome data
Duty to report for usersMandatory for user facilities for deaths and serious injuries [3]Described as not legally required for adopters [4]Different legal systems; the MHRA MRI text was not read
Legal mandate for MR safetyRespondents from five countries said mandated; Italy and Austria confirmed [9]Few countries mandate QA/QC [10]Survey data; legal texts not read
Staff field limit0.01 T continuous in an archived guideline [7]Conditional exemption from limits in the UK [6]Different instruments; neither is a current measured limit here

11. Practice: the Arrangements the Sources Support

Reasoning. The following arrangements are the page’s synthesis of statements in the sources and are not a recommendation of any one source, and a unit applies the law and the guidance that bind it. (1) Name the accountable person and the day-to-day safety officer in writing, and record who provides independent advice [1][2]. (2) Keep written policies with a periodic review and an audit of screening and training records [1][2]. (3) Document training and repeat it at least annually for staff who enter the controlled areas [1][2]. (4) Establish an internal reporting system that includes near misses and gives feedback to staff, because the Swedish survey found that most MR workers did not know of incidents reported at their own hospital [8]. (5) Establish with the regulator and the manufacturer which incidents must be notified externally and by when, and keep the notification separate from the internal report [3][4]. (6) Record the legal basis of the unit’s occupational exposure arrangement (see MRI Safety — Master Page, Section 8.3, and the HSE page for the UK) [6]. (7) Review incident reports for root causes at a committee or an equivalent forum where the site is large enough [1].


12. Limits of the Evidence and What Was Not Read

The evidence on frequency is two self-reported surveys and a set of voluntary reports; the evidence on governance is consensus guidance and regulators’ web pages. No source read gives an incident rate with a denominator, quantifies under-reporting, evaluates the effect of a reporting system, a safety officer, a training programme or an audit on incidents, or reports audit results. The regulations, the MHRA chapters on reporting and training, the ACR Manual, the 2021 version of the RANZCR guideline, the content of the European four-tier curriculum and the rules of most countries were not read, so the page cannot say what is currently required in any country. The 2017 RANZCR text may differ from the 2021 version. All statements are those of the documents as returned by summarising tools and were not checked page by page.


13. Topics Not Covered

National legislation and its transposition of Directive 2013/35/EU; the legal text of any reporting duty; the rules of Japan, Germany, France, Italy and other countries; accreditation schemes and their requirements; root-cause analysis methods and the content of an incident form; just culture and legal protection of reporters; contrast-agent reaction treatment and its governance (not covered on the contrast page either, because the ACR Manual on Contrast Media could not be read); research-volunteer governance; MR-guided radiotherapy and interventional suites; and insurance and liability.


14. Evidence-Based References

No systematic review, no incident database with denominators and no evaluation of a reporting or governance system was read; the sections below follow the hierarchy of the sources actually read, and the guidance documents rest on consensus without stated evidence grading.

A. Guidelines, Regulators’ Pages and Guidance

Professional guideline (Australia and New Zealand; Version 2.0 of 2017, full text read through a summarising tool of a PDF copy; Version 3 of 2021 not opened)
[1] Royal Australian and New Zealand College of Radiologists, Faculty of Clinical Radiology. MRI Safety Guidelines. Version 2.0, approved 8 December 2017 (Version 1.0, April 2007). The College web page lists Version 3 (2021) as current; its PDF was not opened.
Relevance: Cited in Sections 2, 5, 6, 8, 9, 10, 11. Roles (Responsible Person, MRI Safety Officer, MRI Safety Expert, optional committee), three levels of training with documented training and annual refresher, three screenings, a reporting system with incidents reported within 24 hours, annual review of the safe practice manual and periodic audits. Non-statutory professional guideline. Some passages of the text read were garbled or incomplete; the document of 2017 may differ from Version 3 (2021), which was not read.
Review tool relating a Joint Commission alert to the ACR guidance (summary of a PDF copy read)
[2] Gilk T, Kanal E. Interrelating Sentinel Event Alert #38 with the ACR guidance document on MR safe practices: 2013. An MRI accreditation safety review tool. J Magn Reson Imaging. 2013;37(3):531-543.
Relevance: Cited in Sections 2, 5, 6, 8, 9, 10, 11. Lists the ten objectives derived from Sentinel Event Alert #38 (February 2008: four-zone access, two screenings, annual education, no resuscitation in Zone IV and others) and three additional objectives (a safety officer, written policies reviewed periodically, only MR Safe or MR Conditional devices), and records that the 2013 ACR guidance table lists reporting of all MR accidents and near misses to the vendor and the FDA (Section A.4). It gives no counts of events. Co-authored by an author of the guidance it interrelates; not independent.
Regulator’s web page (US Food and Drug Administration; page read in full)
[3] US Food and Drug Administration. Mandatory Reporting Requirements: Manufacturers, Importers and Device User Facilities. FDA web page (Medical Device Reporting).
Relevance: Cited in Sections 2, 5, 9, 10, 11. Device user facilities (hospitals, imaging centres): a device-related death is reported to the FDA and the manufacturer, and a serious injury to the manufacturer (to the FDA only if the manufacturer is unknown), within 10 work days of becoming aware on Form FDA 3500A; malfunctions are not required but may be reported voluntarily (MedWatch); an annual summary on Form FDA 3419 is due by January 1; manufacturers report within 30 calendar days, or 5 work days when remedial action is needed. The page does not define “device-related” or “serious injury” for an MR accident, and the regulation (21 CFR 803) was not read.
Guidance web page (NHS England Digital Regulations Innovation Service; page read in full)
[4] NHS England. Reporting safety issues about medical devices to the MHRA. Digital Regulations Innovation Service guidance page for adopters.
Relevance: Cited in Sections 2, 5, 9, 10, 11. Defines an adverse incident as an event that caused or almost caused injury to a patient or other person; sets three criteria of reportability (an event occurred; the device is suspected to be a contributory cause; death or serious deterioration in health resulted or might have resulted); gives “as soon as possible” with no deadline; routes (Yellow Card in England and Wales; Health Facilities Scotland; Northern Ireland to contact the MHRA); states that developers have a legal requirement to report and that for adopters reporting is “not legally required” but essential. Written for adopters of digital health technology generally, not MRI; the MHRA MRI guideline sections on reporting were not read.
Regulator’s web page (UK Health and Safety Executive; page read in full)
[6] Health and Safety Executive. Exemptions from exposure limits in the Control of Electromagnetic Fields at Work Regulations: MRI exemption for medical purposes. HSE web page (hse.gov.uk/radiation/nonionising/emf-exemptions.htm).
Relevance: Cited in Sections 2, 7, 9, 10, 11. States that the exemption from the exposure limits covers work with MRI equipment for patients in the health sector (development, testing, installation, use, maintenance and related research), on condition that exposure is as low as reasonably practicable and employees are protected against the related health effects and safety risks; employers must comply with all other requirements except the action plan; MRI not for patients in the health sector falls outside the exemption. The page does not give the regulation number; the Regulations themselves were not read. A web page is not the legal text.
Archived national guideline (Canada; non-binding; web page read)
[7] Health Canada. Guidelines on Exposure to Electromagnetic Fields from Magnetic Resonance Clinical Systems — Safety Code 26. First published 1987; web page with an issue date of 2018-06-29 and a modification date of 2024-04-18; archived by Health Canada as no longer reflecting current MRI technologies and radiation safety practices.
Relevance: Cited in Sections 2, 5, 6, 7, 9, 10. Used only for its legal status (non-binding guidance), its exposure levels for patients (2 T; 3 T/s rms; SAR limits) and operators (0.01 T continuously during the working day, higher for about 10 minutes per hour) and for the fact that it contains nothing on safety management or incident reporting and excludes projectiles and quench. Archived: it is not current guidance and the page does not say what replaced it.

B. Surveys and Analysis of Reports

Cross-sectional national survey (Sweden; abstract, methods, results and discussion read)
[8] Hansson B, Olsrud J, Wilén J, Owman T, Höglund P, Björkman-Burtscher IM. Swedish national survey on MR safety compared with CT: a false sense of security? Eur Radiol. 2020;30(4):1918-1926. doi:10.1007/s00330-019-06465-5.
Relevance: Cited in Sections 2, 3, 4, 11. 529 eligible participants reported 200 MR safety incidents and 156 CT safety incidents over 12 months; 16% of the MR incidents received the highest potential severity score; 73% of MR workers (50% of CT workers) were confident they would be aware of any incident at their workplace, but 69% of MR workers (83% of CT workers) were not aware of reported incidents at their hospitals. Self-reported web survey; recall and selection bias are possible; the response rate and the definition of severity were not retrieved; one country.
Cross-sectional international survey (Europe; abstract and key points read through a database record)
[9] European Society of Radiology (ESR). The European MR safety landscape. Insights Imaging. 2024;15(1). doi:10.1186/s13244-024-01813-6. (The record read did not show the author list or the article number.)
Relevance: Cited in Sections 2, 3, 5, 6, 9, 10. 793 responses from 44 countries (April to June 2023); respondents from five countries said MR safety is legally mandated, only two (Italy and Austria) could be confirmed; 77% said their institution had a clear MR safety guideline; 52% said nobody there had received specific MR safety training; MR Conditional cardiac devices were scanned at university hospitals (75% of respondents from those sites) and at 42% of outpatient facilities; MR Unsafe cardiac devices were scanned off-label at 27% of university hospitals; about 12% reported MR-related accidents with injury; a four-tier training curriculum is proposed. Self-selected respondents distributed through ESR member societies; the full text, the denominators and the content of the curriculum were not read.
Cross-sectional international survey on quality assurance (Europe; abstract read)
[10] Busoni S, Pichiecchio A, European Society of Radiology (ESR), ESR MR Safety and Quality Subcommittee, Cristiano L, England A, Oei EHG, Lundberg P, Williams MC, Santini F, Neri E. The MR quality landscape in Europe. Insights Imaging. 2026;17(1):125. doi:10.1186/s13244-026-02280-x.
Relevance: Cited in Sections 2, 8, 10. 269 responses from 37 countries (October to December 2024); only a few countries mandate national MRI QA/QC legislation, many rely on voluntary guidelines or lack formal protocols; low awareness and implementation of quality guidelines; strong agreement on the need for harmonised guidelines. Concerns image and system quality rather than safety incidents. Only the abstract was read; self-selected respondents.
Descriptive analysis of regulatory adverse-event reports (full text read through a summarising tool)
[11] Delfino JG, Krainak DM, Flesher SA, Miller DL. MRI-related FDA adverse event reports: a 10-yr review. Med Phys. 2019;46(12):5562-5571. doi:10.1002/mp.13768.
Relevance: Cited in Sections 2, 3, 4. Used only to state that the available US evidence on MR incidents is a set of spontaneous, voluntary reports without denominator or rate; the incident narratives are on the projectile and quench pages. Under-reporting is not quantified in the article.

C. Professional-Society Poster

Professional-society poster (UK; page read in full)
[5] Society and College of Radiographers, MR Advisory Group. Incident Reporting (poster for MR Safety Week 2022).
Relevance: Cited in Sections 2, 4, 5. States that MR incidents are widely under-reported across Europe (citing a multicentre interview survey and a three-year review of a UK independent-sector provider, neither of which was read), lists categories of MR incidents reportable to the MHRA (burns and overheating, projectiles, mechanical failures, cryogen and quench problems, contrast injector failures, noise problems, major image artefacts, changes in device function caused by MR exposure, defective medicines, adverse drug reactions) and refers to the online Yellow Card system. A poster, not a study; it does not state which incidents are reported internally only.

End of document — Incident Learning, Governance, Quality and Jurisdictions (S13) — MRIninja v1.0 — October 2026. Parent: MRI Safety master.

Recent PubMed search for this protocol

Last updated: October 2026
MRI.ninja has no commercial vendor support. Donations help cover maintenance and hosting costs. Donate & Request