Permanent, Resistive and Low-Field MRI Magnets — Construction, Open and Dedicated Systems, Comparison with Superconducting Magnets

MRIninja Knowledge Base | Foundations — Scanner Hardware Parent page: MRI Physics — Fundamentals and Principles (9201) Builds on: MRI Scanner Structure and Components — Overview (9703) Version 1.0 — November 2026

Foundations — Scanner Hardware

Permanent, Resistive and Low-Field MRI Magnets

Construction and behaviour of the non-superconducting magnets used in open and dedicated (for example extremity) systems, what limits each type, and how they differ from the superconducting magnet.


1. Scope, Boundary and Information Status

The superconducting magnet is on Superconducting MRI Magnet — Construction and is not repeated. The overview of all components is on Scanner Structure and Components, and the general description of the magnet types and the physics of homogeneity is on Magnetic Field, Hardware and Homogeneity. This page states what is specific to permanent magnets, resistive electromagnets and low-field systems, including the open and dedicated scanners built on them. The imaging consequences of field strength (relaxation times, SNR scaling) are on Relaxation Phenomena and are only summarised here. Clinical performance of low-field systems is not a subject of this page; a few reported comparisons are listed only to show the evidence level.

Information status. Six sources were read, all through a tool that returns a model-written summary of the page, not the raw text. The most useful is a 2019 review of low-field MR physics [1]; resistive magnets rest on a 2021 review whose subject is cooling of low-field research and low-resource systems, not clinical dedicated scanners [2]; permanent-magnet geometry on a 2023 review of point-of-care hardware [3]; clinical context on a 2024 narrative review [4]. Two further sources are secondary: the 2013 narrative review used on the overview page [5] and an undated teaching page [6]. No source was found that lists which commercial dedicated or open scanners use a permanent magnet and which a resistive one, and none describes the magnets of the dedicated systems in common use in Italy; these points are marked (to be sourced). No standard, manufacturer manual or textbook was read.

SourceTypeReading level and weight
[1] Marques, Simonis, Webb 2019, J Magn Reson ImagingReviewOpen full-text PDF through a summarising tool; strongest source; the summary flagged unclear table-to-vendor mapping and a value that conflicts with the text
[2] Natukunda et al. 2021, BMC Biomed EngReviewOpen full-text PDF through a summarising tool; resistive coils below 200 mT for low-resource settings; applicability to clinical dedicated scanners limited
[3] Webb, O’Reilly 2023, MAGMAReviewPage through a summarising tool; permanent-magnet geometries for point-of-care systems only
[4] Pogarell et al. 2024, Skeletal RadiolNarrative reviewFull text through a summarising tool; low-field musculoskeletal MRI; evidence mostly old
[5] Kathiravan, Kanakaraj 2013Narrative reviewFull text through a summarising tool; two statements on resistive and permanent limits
[6] King’s College London teaching pageTeaching page, undatedLow weight; one statement on field range

2. How the Field Strength Ranges Are Named

RangeDefinitionSource
Mid-field0.1–1 T[3]
Low-field0.01–0.1 T[3]
Ultra-low-field (whole-body)Below 0.01 T[3]
"Low field" in the resistive-coil reviewStudies below 200 mT (0.2 T)[2]
Point-of-care systemsAbout 45–80 mT (examples at 47, 50.9, 55, 70 and 80 mT)[3]
Early musculoskeletal low-field systems0.25–1.0 T permanent magnets (1980s); the share of systems below 1.5 T fell from about 30% in 2000 to about 5% (statement of [4])[4]

The sources define the classes differently (the 0.2 T dedicated systems are "mid-field" in [3] and "low-field" in [2]); this page uses the field value in tesla wherever it matters.


3. Permanent Magnets

TopicStatementSource
MaterialNeodymium-iron-boron (NdFeB), remanence 1.2–1.425 T; grades from N35 to N52; harder grades (for example N48M, N48H) have a larger demagnetisation margin but are weaker mechanically and dearer because they contain dysprosium[1]
GeometriesH-shaped design (two ferromagnetic yokes) and C-shaped design (one yoke), with shaped ferromagnetic pole pieces; for point-of-care systems the two designs are the yoked C-/H-shaped magnet and the Halbach array[1]; [3]
Field rangeFor 0.25–0.5 T the options are a permanent magnet alone or a permanent magnet combined with an electromagnet; the review states that superconductors are normally used above 0.5 T, but it also lists a 1 T hand/wrist permanent-magnet imager (an internal conflict)[1]
Homogeneity targetppm-level over an ellipsoidal volume; field fluctuation during a scan below 100 nT[1]
ShimmingShaping the magnet, adding movable magnetic pieces (passive) or electrical shim coils; shimming is "substantial" because the imaging volume is a large fraction of the magnet volume[1]; [3]
Halbach arraysStrongest field per unit mass of magnetic material; discretising into polygonal elements increases inhomogeneity by about 20–30% compared with a continuous ring and square or cubic elements can be up to an order of magnitude worse; flux reduction about 50% for square and about 25% for octagonal or circular cross-sections[3]
Yoked C-/H-shaped magnetsMuch easier to construct than Halbach arrays[3]
Temperature dependence and driftThe review gives only a reference to a hysteresis model and no numbers; the existing statement that the field depends on temperature and needs thermal management is on 9208[1] (gap)
Weight and floor loadA 1 T hand/wrist imager needs a floor able to support almost 2,000 kg; the weights in the review table are unclear; weight is named as a disadvantage in [4][1]; [4]
Stray field"Almost complete absence of fringe fields"; the 5 gauss line is within the magnet cover in small extremity scanners; the table lists 5 gauss distances of 1.8, 0.6 and 1.75 m (vendor mapping unclear)[1]
FootprintWhole-body low-field scanners about 20–30 m²; extremity scanners as small as 9 m²; a 1 T hand/wrist imager can be sited in about 3 × 4 m[1]
Power and coolingNo cooling system and low power consumption; a standard 220 V supply in some cases; longevity[1]; [4]

A permanent magnet cannot be switched off (Reasoning) (the field is a property of the material); the sources read do not describe emergency procedures for these magnets (to be sourced).


4. Resistive (Normal-Conducting) Electromagnets

TopicStatementSource
Definition and historyEarly commercial MRI magnets, later overtaken by superconducting magnets; often not above 0.3 T before power demands made them uneconomical[2]
Power at whole-body scaleContinuous power about 40 kW; impractical for whole-body imaging above about 0.18 T[5]; [2]
Heating lawJoule losses; the power deposited in a coil increases with the square of the field strength, so the maximum field is limited by cooling capacity[2]
Temperature effectsHeat ages the winding insulation (the weak link), increases image noise and affects patient comfort; coil expansion shifts coil positions and reduces the field (drift); no numerical drift value is given[2] (gap)
Field only with currentThe field exists only while the coil is powered (Reasoning)physics; no source
Fringe fieldSmaller than that of superconducting magnets[2]
Weight"Lighter than permanent magnets" (qualitative, no figure); one 0.08 T water-cooled magnet was 1.9 m high and 2 tons[2]
CostLess expensive than superconducting systems; no figure; no liquid helium[2]
CombinationA permanent magnet can be combined with an electromagnet in the 0.25–0.5 T range[1]
HomogeneitySensitive to coil expansion and spacers; Helmholtz, Lee-Whiting and Merritt coil layouts give uniform fields; no numeric values[2]

4.1 Cooling of resistive coils (low-field research and low-resource systems)

MethodReported figuresSource
AirLimited to current densities of about 2 A/mm²; heat-transfer coefficients 2.5–10 W/m²K (natural) and 10–500 W/m²K (forced); examples: about 4 kW with coil surfaces up to 70 °C at 6.5 mT; air cooling suited to up to about 500 W at 23 mT[2]
WaterHeat-transfer coefficients 100–15,000 W/m²K (all liquids); examples: 25 L/min for a 0.08 T 2-ton magnet; 17.5 kW at 0.2 T with a rise of at most 40 °C at 15.14 L/min; 10 kW at 47.5 mT with the coil 5 °C above the coolant[2]
Oil and other fluidsTransformer oil at 78.4 kW for a 1.6 T, 200 A coil; perfluoropolyether at 0.95 T and 800 A with an average conductor temperature of 26 °C[2]
ConductionSuited to low heat loads[2]
DrawbacksAir cooling is bulky; liquid cooling risks leakage, corrosion and pumping cost; closed loops carry an environmental risk on leakage[2]

These examples come from research and low-resource prototypes, mostly below 0.2 T, not from commercial dedicated scanners; they show the principle and orders of magnitude of heat removal, not the specifications of any clinical system (to be sourced).


5. Open and Dedicated Systems

TopicStatementSource
Open geometryMost low-field scanners are open designs; bore openings of 35.1–58 cm are quoted; intended for claustrophobia, obesity, patient positioning and access during scanning; upright imaging uses a vertical bore or a rotatable magnet[1]
Typical musculoskeletal designPermanent-magnet systems are open and typically C-shaped, with longevity and low power consumption as advantages and complex shimming and weight as disadvantages[4]
Extremity scannersHand/wrist systems with gradients up to 215 mT/m; footprint as small as 9 m²[1]
Other dedicated systemsNeonatal (compact, low noise, bed temperature control), radiotherapy-integrated (Co-60 source or linear accelerator), interventional (no vendor listed a low-field interventional scanner in the review)[1]
Older general statementResistive and permanent magnets reach about 0.2–0.3 T with lower image quality; open designs suit interventional MRI and obese or paediatric patients[6]
Which systems use which magnetNot stated in the sources read; the review says only that resistive magnets are the "most notable exception" to permanent magnets in low-field scanners[1] (to be sourced)

Dedicated extremity and open scanners in widespread local use are not described in the sources read; the technology of each model (permanent or resistive magnet, field strength, geometry) must be taken from its documentation (to be sourced).

5.1 Reported clinical comparisons (secondary reports in [4], not appraised)

Report as given in [4]Finding
0.2 T versus 1.5 T, knee, 90 patients (2000)No significant difference in sensitivity or specificity for meniscal and ACL tears
1.0 T whole-body versus 0.2 T extremity scanner, ankle (1998)Higher SNR at 1.0 T; 96% of lesions found at 1.0 T were also found at 0.2 T
0.2 T permanent-magnet extremity scanner, wrist, 195 patients (2003)Scaphoid fractures in 19%, distal radius in 14%; MRI changed treatment in 92%; not compared with CT
Overall statement of the authorsMost supporting evidence is several decades old and new high-quality research is needed; modern low-field MRI is an extension of the high-field portfolio, not a replacement

These are individual older studies quoted by a narrative review; they are not an assessment of current dedicated scanners and the level of evidence is low.


6. Comparison of Magnet Types

FeaturePermanentResistiveSuperconducting (see 9704)
Typical field in the sourcesAbout 0.2–0.5 T; a 1 T example [1]Up to about 0.3 T historically; 0.18 T limit at whole-body scale [2, 5]0.15–4 T common; 1.5 and 3 T dominant [5]; 0.55 T also exists [4]
CryogenNone [1]None [2]Helium, in bath or sealed form
PowerLow [1, 4]Up to about 40 kW at whole-body scale; cooling needed [2, 5]Cryocooler and ancillary power only (persistent mode)
Cooling of the magnetNone [1]Air, water or oil [2]Cryogenic
Field when unpoweredAlways present (Reasoning)Absent (Reasoning)Persistent unless ramped down
Stray fieldAlmost absent [1]Smaller than superconducting [2]Large; see 9803
WeightHeavy [4]; floor load about 2,000 kg for a 1 T example [1]Lighter than permanent (qualitative) [2]Several tons; see 9704
Homogeneity and stabilityHard shimming; temperature sensitivity [1, 4]Coil expansion and drift with heating [2]Persistent mode, decay below 0.1 ppm/h [see 9704]
SitingSmall footprint, open geometry [1]Flexible siting [2]Dedicated room, shielding

Cells without a citation are derived from the cited rows of this page or of 9704 (Reasoning).


7. Consequences of the Lower Field (Summary of [1])

QuantityStatementSource
SNRSignal scales with B0² and overall SNR was assumed to scale as B0^3/2; fitted exponents of 1.04 (T2*-weighted), 0.90 (T1-weighted) and 0.92 (PD) are lower than postulated, so the loss at low field is smaller than predicted (fits depend on noise and sequence assumptions)[1]
TimeExample: a 9-minute T1-weighted acquisition at 1.2 mm would take 91 or 243 minutes at 0.5 T depending on the assumed law; matching the time needs about 1.7 mm resolution[1]
T1 and T2*T1 increases with field; apparent T2* decreases with field (low-field T2 and T2* are longer)[1]; 9203
SARScales with B0²; fewer SAR limits at low field[1]
Susceptibility artefactScales with B0; imaging nearer implants is possible[1]
Other limitsLower polarisation, g-factor amplification in parallel imaging, higher gradient and magnet inhomogeneity needing advanced reconstruction[1]

8. Discrepancies Between Sources and Evidence Limits

TopicDifference or gapHandling
Permanent versus superconducting boundarySuperconductors "normally above 0.5 T" but a 1 T permanent-magnet imager is listed [1]Both stated
Resistive field limitAbout 0.3 T [2] versus 0.18 T for whole-body imaging [5]Different scope (historical versus whole-body continuous power); not reconciled
Definition of low-fieldBelow 0.1 T [3] versus below 0.2 T [2]Tesla value used
Resistive cooling dataFrom research and low-resource prototypes, mostly below 0.2 TNot applicable to clinical dedicated scanners without confirmation
Drift and temperature dependenceNo numeric values for permanent or resistive magnetsAll (to be sourced)
Dedicated scanners in local useMagnet type of each model not in the sourcesAll (to be sourced)
Source levelReviews read through summarising tools; no standard, manual or textbookConfirm against equipment documentation
Not coveredSafety and emergency procedures of these magnets, installation, shielding of the room, drift correction, specific commercial systemsAll (to be sourced)

9. Related Pages and Next Steps

The cooling architectures of superconducting magnets, including the sealed low-helium magnets, are compared on Superconducting MRI Magnet — Construction (Section 5). Planned: shim hardware, gradient system, RF transmit system, patient table and bore, RF coils by type (topic D). A page on dedicated extremity and open systems as a group (siting, coils, protocols) needs documentation of the systems in use (to be sourced).


10. Evidence-Based References

No guideline or study with an EBM level applies to this descriptive engineering page. Sources are ordered by publication quality; the reading level of each is stated.

A. Peer-Reviewed Reviews

Review
[1] Marques JP, Simonis FFJ, Webb AG. Low-field MRI: an MR physics perspective. J Magn Reson Imaging. 2019;49(6):1528–1542. doi:10.1002/jmri.26637.
Open full-text PDF through a summarising tool; table-to-vendor mapping unclear in the summary.
Review
[3] Webb A, O’Reilly T. Tackling SNR at low-field: a review of hardware approaches for point-of-care systems. MAGMA. 2023;36:375–393.
Page through a summarising tool; section labels inconsistent; permanent-magnet designs only.
Review
[2] Natukunda F, Twongyirwe TM, Schiff SJ, Obungoloch J. Approaches in cooling of resistive coil-based low-field magnetic resonance imaging (MRI) systems for application in low resource settings. BMC Biomed Eng. 2021;3:3. doi:10.1186/s42490-021-00048-6.
Open full-text PDF through a summarising tool; low-field research systems; one helium figure in it comes from a weak source and is not used.
Narrative review
[4] Pogarell T, Heiss R, Janka R, Nagel AM, Uder M, Roemer FW. Modern low-field MRI. Skeletal Radiol. 2024;53:1751–1760. doi:10.1007/s00256-024-04597-4.
Full text through a summarising tool; musculoskeletal focus; earlier studies are quoted second-hand.
Narrative review
[5] Kathiravan S, Kanakaraj J. A review on potential issues and challenges in MR imaging. Sci World J. 2013;2013:783715. doi:10.1155/2013/783715.
Full text through a summarising tool (also used on 9703 and 9704); two statements.

B. Teaching Page

Teaching page
[6] King’s College London. Magnetic Resonance Imaging — MRI Instrumentation (e-health teaching page). Undated, no author.
Page through a summarising tool; low weight; one statement.

Recent PubMed search for this protocol

Last updated: November 2026

Report an error on this page

The report stores the page address, the text above and, if you give it, your e-mail. It does not store your IP address, browser type or any cookie. See the Disclaimer, section 7.