MRI Patient Table and Bore — Table Parts and Docking, Table Motion and Field Change, Emergency Removal, Reported Bore Dimensions and Evidence on Bore Size

MRIninja Knowledge Base | Foundations — Scanner Hardware Parent page: MRI Physics — Fundamentals and Principles (9201) Builds on: MRI Scanner Structure and Components — Overview (9703); Laser Positioning, Landmarking, Isocentre and the Scanner Coordinate System (9702) Version 1.2 — November 2026

Foundations — Scanner Hardware

MRI Patient Table and Bore

The parts of the scanner that the patient touches and lies in: the docking patient table, its motion through the fringe field and the symptoms that follow, the reported bore dimensions, what the evidence says about bore size and patient outcomes, and the emergency removal of a person from the bore as written in site procedures.


1. Scope, Boundary and Information Status

The position of the table relative to the isocentre, the lasers and the coordinate system is on Laser Positioning, Landmarking, Isocentre and the Scanner Coordinate System. The place of the table and bore among the other blocks, and the cantilevered table that isolates the patient from gradient vibration, are on 9703 (Section 7); neither is repeated. Claustrophobia management, sedation and the full discussion of the scanner-design trial are on Claustrophobia, Sedation, Monitoring, MR-Conditional Equipment and Emergencies. The static-field contours, the exposure limits for staff and the movement-induced effects as regulated are on Static Magnetic Field Contours. Acoustic noise is on 9808. Quench, zones and screening are on the safety pages.

This page is the home of the hardware description of the patient table and of the bore, of the reported table speeds and bore dimensions, and of the evidence that links bore geometry to patient motion and tolerance.

Information status. Twenty-three sources were read, nearly all through a tool that returns a model-written answer about the page, not the raw text, and a number of them only as search extracts. Two are editions of the ACR Manual on MR Safety, read in parts only [1, 2]. Six are peer-reviewed articles, of which one was read in nearly its full text [3], one in full [4] and four only as abstracts or search extracts [5, 6, 7, 8]; one further journal article is known through the conference abstract of the same cohort [9]. Four are conference abstracts [10, 11, 12, 13]. Five are written procedures of research or hospital sites [14, 15, 16, 17, 18]. Three are patents [19, 20, 21]. Two are a manufacturer article and a news report [22, 23]. No standard (including the patient-support requirements of IEC 60601-2-33), textbook or manufacturer service documentation was read. The copy of the ACR Manual with a 2024 cover was read from its start to its end in consecutive ranges through a summarising tool that answered questions on the table, the bore, claustrophobia, transfer and emergencies; passages outside those topics were not examined, and the 2020 edition was read only in part. The page is therefore thinner on engineering than on procedure. The page is therefore thinner than the other component pages; the gaps are listed in Section 11. Statements marked (Reasoning) are derived by the editor; gaps are marked (to be sourced).

SourceTypeReading level and weight
[1] ACR Manual on MR Safety, copy with a 2024 coverGuidance document of a professional societyWhole document in consecutive ranges (characters 0–385,941) through a summarising tool answering topic questions; Chapter 14 (pp. 103–107), Chapter 15 (pp. 108–111), Chapter 16 (7 T, p. 118) and Appendices 1, 2, 3 and 6 read; hosted on a third-party site; edition not confirmed (see reference list)
[2] ACR Manual on MR Safety, version 1.0, 2020Guidance document of a professional societyFirst 100,000 of 164,784 characters (zones, emergency response, implant discovered during an examination) and a second extract with Appendices 2 and 3, through a summarising tool; superseded edition
[3] Hansson et al., Bioelectromagnetics 2019Peer-reviewed descriptive study, 154 examinations at 7 TFirst 100,000 of 117,512 characters through a summarising tool; reference list cut; single site, no control group, self-reported symptoms
[4] Mian et al., PLoS ONE 2013Peer-reviewed experimental study, 25 volunteers at 7 TFull text through a summarising tool; figures read as captions only; one table partly garbled
[5] Iwan et al., Eur Radiol 2021Peer-reviewed survey within a randomised trialSearch extracts only; design and one figure; results not read in full
[6] Koh et al., Proc Singapore Healthc 2017Peer-reviewed retrospective comparisonCatalogue abstract extract only; denominators not read
[7] Brunnquell et al., Top Magn Reson Imaging 2020Peer-reviewed review of patient comfortSearch extract only; title not shown; used for the dimensions of the common design
[8] Low-field 0.55 T against 1.5 T comparison, PMC9692643Peer-reviewed prospective comparisonAbstract extract only; journal and authors not shown in the extract
[9] Dewey et al., JMRI 2007Peer-reviewed cohortNot read; figures taken from the RSNA 2007 abstract of the same cohort, read in full
[10] Lane et al., ISMRM 2008Conference abstract, 32 patientsFull abstract page through a summarising tool; chart not legible
[11] Ahn et al., RSNA 2010Conference abstract, 15 volunteersSearch extract of the abstract; units not shown
[12] ISMRM 2014 abstract 4870 (7 T sensations)Conference abstractSearch extract only; title and authors not shown
[13] ISMRM 2018 abstract 2636 (compact 3 T)Conference abstract, research scannerSearch extract only
[14] Queen's University fMRI Facility SOP #10Written site procedure, 3 T research scannerDocument read through a summarising tool; undated, manufacturer not named, excerpt ends before its last step
[15] Emergency Subject Evacuation (3T), January 2023Written site procedureRead through a summarising tool; organisation not named in the file; the referenced diagram is not included
[16] Emergency Subject Evacuation (7T), January 2023Written site procedureSearch extract only
[17] Duke BIAC SOP104, [18] UAB MRI safety policyWritten site proceduresSearch extracts only
[19, 20, 21] Three US patentsPatent descriptionsSearch extracts only; background statements by the applicants, not evidence that a design is in use
[22] GE HealthCare article on bore sizes, [23] Radiology Business news itemManufacturer article; news reportSearch extracts only; the news item reports a journal study that was not read

2. The Patient Table: Parts Named in Site Procedures

No description of the table by a manufacturer, a standard or a textbook was found. What is known comes from two written site procedures for 3 T research scanners [14, 15] and from the background statements of patents [19, 20, 21]. The procedures do not name the manufacturer, and the two tables may not be the same design, so the table below lists what each procedure names and is not a general description (Reasoning).

ElementStatementSource
Removable tabletop on a drive unitThe tabletop is mechanically decoupled from the motorised drive unit by an emergency release handle at a 7 T site, and can then be pulled out by the handle at the foot end[16] (extract only)
Docking to the scannerThe table is docked to the system and can be undocked only when it is fully out in the home position; a left pedal undocks it, and an emergency undocking handle between the two foot-end pedals does the same if the pedal fails[15]
Carrier plate and trolleyThe table unit locks onto a trolley through docking rods entering a guide rail; the unit is lowered until the tabletop rests on the trolley's mounting device, the carrier plate is lowered and the table-trolley unit is wheeled out of the magnet room[14]
Brakes and guiding wheelBrake position 1 disengages the brakes and lifts or deactivates the guiding wheel; position 2 is said to hinder extraction to the preparation room; position 3 engages the brakes. The extract gives the same parenthetical for positions 1 and 2, so the difference is unclear[15]
Side railsRails may be up, horizontal or down but must be down for the tabletop to move[15]
Coil sockets and accessories on the tableAt undocking the console asks for all coil connectors and hoses (squeeze ball, vacuum cushion, headphones) to be disconnected and the socket covers to be closed[14]
Single vertical positionFor cylindrical patient receiving regions the table is described as having a single vertical position inside the bore, which is optimal for an average patient; the patent proposes a way to change this[20]
Isocentre positioningA patent describes adjusting the table position when the geometric centre of the region lies within a set tolerance of the isocentre; the tolerance is not given in the extract[21]

Table speed, positioning accuracy, load limit, vertical range, lowest height, travel and the connector layout were not found in any source read (to be sourced).

3. Table Motion Through the Fringe Field

3.1 Reported speeds and rates of field change

Reported speeds are few and come from 7 T systems and from one vendor's range. A patent states that table speed in MRI is limited to avoid exposing the patient to an overly large rate of field change, that some conventional approaches give an unduly slow speed, and proposes a speed that varies along the direction of movement according to the spatial gradient of the field [19]. The patent is an applicant's description; it does not show that any commercial table works that way.

SystemTable motionRate of field change reportedSource
Actively shielded 7 T system (Philips Achieva)Motorised, 88 mm/sMaximum about 0.72 T/s on entry[3]
Passively shielded 7 T system, same vendorMotorised, 105 mm/sNot given in the extract[3]
Clinical 1.5 T system, same vendorMotorised, 180 mm/sNot given in the extract[3]
7 T research scanner (Philips Achieva)Bed pushed by hand, mean entry speed about 0.069 m/s; entry about 4 m in 50–60 sPeak |dBz/dt| about 0.63 T/s[4]
Another 7 T scanner, Japanese volunteersAutomatic, constant 2 cm/s, described as slowNot given in the extract[12]

If the maximum rate of 0.72 T/s in [3] occurred at the table speed of 88 mm/s, the implied steepest spatial gradient on entry is about 8 T/m; the corresponding calculation for [4] gives about 9 T/m (Reasoning). Both values assume that peak rate and mean speed coincide, which the sources do not state, and the two scanners differ. The calculation only illustrates that the rate of change equals speed multiplied by spatial gradient; it is not a specification. Speeds of 1.5 T and 3 T tables from other vendors, and the field-change rate at those field strengths, were not found (to be sourced).

3.2 Symptoms in the 7 T studies

In the single-site study of 154 examinations in 124 volunteers and research patients at an actively shielded 7 T system, dizziness was reported in 84 % (130), inconsistent movement in 70 % (108), nausea in 52 % (81), headache in 52 % (81), metallic taste in 43 % (66), peripheral nerve stimulation in 67 % (103) and magnetophosphenes in 23 % (35) [3]. Dizziness, inconsistent movement and nausea were significantly more frequent during movement into or out of the scanner than when stationary (P < 0.01), while headache and metallic taste did not differ; dizziness was reported on entry in 119 of 154 examinations (mean 40 on a 0–100 scale), against 81 when stationary in the bore [3]. Head-first examinations produced stronger dizziness and inconsistent movement than feet-first examinations, but 118 examinations were head-first brain studies and 25 were feet-first knee studies, so position and anatomy are confounded [3]. The authors suggest that exposure may be similar to a passively shielded system because the table speed was reduced, and state that dB/dt cannot be excluded as a cause of the higher symptom rates [3]. Limits: descriptive design, no control group, self-report, symptoms depending on information and expectation, and predicted nerve-stimulation values for only 83 of 154 examinations [3].

In a controlled volunteer study at 7 T without imaging, 25 healthy adults were moved without vision into the bore. Nystagmus began at about 1.67 ± 0.58 T and perceived vertigo at about 5.14 ± 1.08 T, values the authors state are not thresholds [4]. Horizontal rotation was the dominant percept (about 84–86 % incidence), nearly always in the same direction, lasting about 50 s on average with about 42 s after the bed stopped; the percept reversed on exit and lasted about 30 s [4]. The authors favour continuous stimulation of the semicircular canals, likely by Lorentz forces on ionic currents in the endolymph, over a transient effect of movement or of a time-varying field, which they consider less likely but not excluded [4]. Limits: small subgroups of 6–19, manual bed speed that differed between entry and exit, onset fields likely overestimated because measured during movement, and several results shown as "data not shown" [4].

The two studies differ on what the table can change. If the effect is continuous and depends on the field itself, a slower table alters the time course but not the presence of the effect; if it depends on dB/dt, the table speed is the lever (Section 10). No study that varied table speed within one scanner and measured symptoms was found (to be sourced); the ACR Manual states that limiting motion during table movement decreases vertigo at 7 T, without naming a study in the extract [1]. How the limits for staff and the guidance on slow movement are formulated is on 9803; no data on 1.5 T and 3 T patients were found in the sources read (to be sourced).

4. Emergency Removal From the Bore as Written in Site Procedures

The two procedures that could be read in full are for 3 T research scanners [14, 15]; a 7 T procedure was seen only as an extract [16]. They are local documents, not guidelines; they do not name the manufacturer and one has no date [14]. They describe the mechanics of the table, not the clinical response or the quench.

Situation or stepQueen's University fMRI Facility [14]3 T procedure dated January 2023 [15]
Power and motorised drive intactPress the Home Position button on the front face of the magnet, to the right or left of the boreThe jog wheel is called the safest method for moving the tabletop out; the Home button moves the tabletop completely out but the table moves at a faster speed, so it is to be used with caution
Power failure or defective driveActivate the emergency table release under the support frame of the table and pull the tabletop out by hand; without the release the table can still be moved by hand but with considerably more forcePull the emergency release, whose location is marked by a label, fully outward, raise the pull-out handle and pull the tabletop out of the magnet
Before undockingTable fully withdrawn; position the trolley under the table until the docking rods lock; set the brakes at the foot endTable fully out in the home position; side rails down for the tabletop to move; set brake position 1
UndockingHold the remote's Lower button until the tabletop rests on the trolley; step on the left interlock pedal until it clicks; disconnect all coil connectors and hoses when the console requests it; acknowledge with the Speed key; lower the carrier platePress the left pedal fully down; if it fails and the table is fully out, release the brakes and pull the emergency undocking handle between the pedals
ResultThe table releases from the carrier plate and stays on the trolley, to be wheeled out of the magnet roomPull the table away from the scanner by the handle at the foot end until clear of the docking mechanism

The 7 T extract [16] adds a table-to-trolley transfer: slide the participant down the bed and out of the head coil, park the trolley beside the table with its sides down and brakes on, bridge the gap with a sliding board, slide the participant over, then release the brakes and raise the sides before leaving. What must be unplugged, the control layout and the steps after leaving the magnet room are not stated in [15] and its diagram is missing. The introduction of [14] reminds the reader that the static field is always on.

The two other site documents seen as extracts differ on what comes next. One states that if the subject can be safely removed from the bore, the subject is taken out to the console room, but that an unconscious or severely injured subject should not be moved out of the magnet room without qualified medical assistance [17]. The other states that, for an emergency requiring assistance or a Code Blue call, the patient must be removed from the magnet room before emergency care is attempted, and for one vendor's scanner lists undocking the table and unplugging the RF coil if a tabletop coil is used [18]. Both are single-site rules known only from extracts. The ACR Manual on MR Safety is read below.

4.1 What the ACR Manual on MR Safety says

The manual does not describe the mechanics of any table; it treats the table as part of the emergency response. In Chapter 14 of the copy with a 2024 cover (pp. 102–107, page numbers inferred by the reading tool from running headers) the Emergency Table Stop is described as designed to stop scanning and table motion at once and as meant for situations such as something caught on the table; the operator is expected to be familiar with the manual process of moving or undocking the table when a patient must be removed quickly; the Emergency Power Off is to be used with caution and only in emergencies, and may need an electrician to reset [1]. For a medical code or cardiac arrest the patient is to be removed immediately from Zone IV to a predetermined, magnetically safe location; facilities are to have immediate access to an MR Conditional gurney when the table cannot be undocked; if transfer is delayed, MR personnel are to prioritise patient safety and start cardiorespiratory resuscitation while still in Zone IV; MR Unsafe items such as automated external defibrillators are not to be brought into Zone IV; and quenching is not routinely recommended for these cases [1]. A medical code is stated not to be a quench situation, whereas entrapment of a patient or staff member against the bore by a sizable ferromagnetic object without timely extraction is a situation in which quenching is recommended [1]. The 2020 edition gives the same direction for a code (basic life support or CPR while removing the patient, then moving the patient out of Zone IV) and adds that a quench could take more than a minute [2]. In the section on an implant discovered during an examination, the 2020 edition advises extracting the patient slowly along a straight line parallel to the centre of the magnet and not letting the patient sit up as soon as the body is out of the bore [2]. The 2024-cover copy also lists claustrophobia and large body habitus among special populations and asks for a communication system or emergency call button in Zone IV and for a Zone IV door that can be unlocked from inside [1].

The same copy adds statements that touch the table and the bore. Chapter 16 on alternative MR environments, in its 7 T section (p. 118), states that limiting patient motion during table movement has been shown to decrease vertigo, that moving the patient with a slow table velocity in and out of the scanner may also help, and that staff should check with the patient before full removal from the table [1]; the study behind the first statement is not identified in the extract (to be sourced). Chapter 12 (pp. 97–98) says that slow table speed and slow patient movement should be beneficial for patients with an implant, and that, if the table is dockable, it is detached and the patient moved into Zone III before sitting up, or transferred slowly and horizontally to an MR Conditional stretcher in Zone IV [1]; Chapter 8 (p. 60) gives the reason for slow movement with a metallic device, the Lenz effect, and states that translational forces are greatest near the openings of the bore (pp. 57, 59) [1]. Chapter 11 (p. 86) requires transport into the scanner on an MR Conditional wheelchair or stretcher and transfer to the MR table in Zone III [1]. Appendix 1 (p. 125) expects a technologist notification device (squeeze ball) and an open two-way intercom channel for conscious patients, and hearing protection in Zone IV [1]; Chapter 15 (p. 110) recommends SOPs for claustrophobic or anxious patients, assessment of padding in sedated patients, and warns that contact or excessive proximity to the bore can cause near-field RF burns [1]. Where the heating mechanism and the positioning rules are concerned, see 9804.

Not found in the whole copy read or in the parts read of the 2020 edition: the emergency release of the table, table speed values, a weight limit (large body habitus appears only as a risk factor for heating, p. 64) and bore dimensions [1, 2]. The 2020 edition leaves the content of the procedure for claustrophobic or anxious patients to local policy [2]. Because the manual refers the operator to the manual process of the table, the release and undocking steps of Section 4 are presumably the manufacturer-specific part of the procedure and must come from the documentation of the system in use (Reasoning).

5. The Bore: Reported Dimensions

The clinical bore is the cylindrical patient space inside the body coil and its cover. The construction layers around it are on 9703. The dimensions found are few, come from clinical reports and a review, and are not manufacturer specifications.

Standard60 cm × 1.9 m [10] Shorter, wider0.6 m × 1.6 m [9] Wide short bore70 cm × 1.25 m [10]
Figure 1. Reported bore diameter (height) and length (width) drawn at the same scale (140 px per metre). The 0.6 m × 1.6 m bore is described as conical, so a rectangle overstates its narrowest part. Constructed schematic from published figures; not a specification.
System describedDiameter and lengthField and contextSource
Standard design60 cm; length 1.5 to 2.25 m, called the common designReview of patient comfort[7] (extract only)
Standard bore, hospital60 cm, 1.9 m long1.5 T[10]
Wide short bore70 cm, 1.25 m long (Siemens MAGNETOM Espree named)1.5 T[10]
Newer shorter, wider conical bore0.6 m wide, 1.6 m long; 97 % less acoustic noise (about 30 dB(A))Cohort of 12,736 patients; field strength of the comparison not stated in the abstract[9]
Wide bore, manufacturer article70 cm, about 10 cm wider than the traditional 60 cm; available at 1.5 T and 3 T; the article states that some image quality is lost in wide-bore against narrow-bore systemsManufacturer statement; in Portuguese; extract only[22]
Low-field wide bore80 cm against 60 cm for a 1.5 T system0.55 T, same-day comparison of patients[8]
Compact research 3 T37 cm for the head and 60 cm for the shoulders, with asymmetric gradient design to allow accessResearch system, compared with 60 cm whole-body scanners[13]

Bore diameter and length are not independent of the other blocks: a larger bore changes the radius of the gradient and body coils and the volume over which the field must be homogeneous (Reasoning). No source read quantifies the cost of a wider bore in gradient performance, field homogeneity or coil efficiency; the only statement found is the manufacturer's remark about image quality [22] (to be sourced). The relation of bore geometry to the open and permanent designs is on 9705.

6. Bore Size, Patient Motion and Tolerance

Every comparison found confounds bore diameter with other properties of the scanner. No randomised comparison of diameter alone was found.

StudyDesign and sizeResultLimits
Dewey 2007 [9]Prospective, non-randomised cohort, 55,734 consecutive patients, 1997–2005Claustrophobic reactions 0.7 % (95 % CI 0.6–0.9) on the newer shorter, wider scanner against 2.1 % (2.0–2.3) on the conventional one; adjusted odds ratio 3.1 (2.5–3.9); head-first, female sex and middle age also associatedNewer scanner also much quieter; conventional bore not described; abstract only (the cohort is discussed on 9813)
Lane 2008 [10]32 patients whose head or spine MRI had failed on a 60 cm bore, rescheduled on a 70 cm × 1.25 m bore, 11 months, single institution27 of 32 (84 %) completed a diagnostic examination; 20 of 25 head or head-containing examinations (80 %); all five failures involved head imagingNo control group; selected previous failures; protocols and durations differed between scanners; spine-only results not reported separately
Koh 2017 [6]Retrospective comparison of 60 cm and 70 cm scanners, Singapore General HospitalThe abstract states that 74 % of the claustrophobia occurred at the 60 cm scanners and 26 % at the 70 cm scanners, a reduction by a factor of 2.95Denominators not read; the shares may reflect different numbers of examinations; extract only
Ahn 2010 [11]15 volunteers, 100 scans (as printed), 70 cm 3 T against two 60 cm 1.5 T systems of two vendors; realignment by SPMX translation 2.2558 (1.08263) at 60 cm against 1.0424 (0.37330) at 70 cm, p = 0.03; Z rotation 0.0341 (0.00742) against 0.0203 (0.0069), p = 0.007Field strength, vendor and coil differ together with bore; units and the meaning of the bracketed values not shown in the extract; abstract only
0.55 T against 1.5 T [8]Prospective, same-day patientsPatients perceived the 0.55 T system, with an 80 cm bore and lower noise, as more comfortableBore and noise change together; extract only
News report of a journal study [23]Early terminations; known claustrophobia analysed separately0.8 % of all patients terminated early; in patients with known claustrophobia 6.2 % at 70 cm against 8.5 % at 60 cm, a difference reported as not significant; prior claustrophobia associated with 33-fold higher early terminationNews item only; the journal article was not read; the cohort and the definitions are not shown
Iwan 2021 [5]160 high-risk patients from a randomised trial of an open 1 T scanner against a short-bore 1.5 T scanner completed a design questionnaire44 (27.5 %) had a claustrophobic event, defined as inability to complete the examination; patients visually favoured open designs over short- and closed-bore designsPreference and visual impression, not outcome of the examination; extract only

The direction is the same in most reports, with fewer failures and less motion on wider or shorter bores, but the sizes range from no significant difference to a threefold lower rate, and each comparison differs in noise, length, field strength or vendor (Reasoning). The sources do not allow a statement that diameter alone reduces claustrophobia. The ACR Manual lists claustrophobia among special populations and recommends SOPs for claustrophobic or anxious patients in both editions, without specifics; its pre-appointment guide notes that some scanners are wider and shorter [1, 2].

7. Differences by System Type

SystemWhat was foundSource
Standard cylindrical 1.5 T and 3 T60 cm diameter, 1.5–2.25 m length is the common design; 70 cm wide-bore variants exist at both field strengths; table speed of 180 mm/s reported for one vendor's 1.5 T system[7, 3, 22]
7 TAll symptom and table-motion studies found are at 7 T; table speeds of 88 mm/s (actively shielded) and 105 mm/s (passively shielded) in one vendor's pair; emergency release and trolley transfer described for a 7 T site[3, 4, 12, 16]
Low-field wide bore80 cm bore at 0.55 T judged more comfortable than a 60 cm 1.5 T bore; table, speed and emergency removal for these systems not found[8]
Open scannersPatients at risk of claustrophobia prefer the appearance of open designs; table and bore details not found[5]
Compact research systemsReduced bore for the head with a wider region for the shoulders[13]

8. Technologist View (Reasoning)

The statements below connect the sources to what is seen at the scanner. They are derived by the editor from the sources named, are limited to the systems described, and are not recommendations.

What the technologist sees or doesComponent behind itBasis and limit
The table will not move into the boreSide rails not down; table brakes engaged[15] for one 3 T system; other systems not documented
Table moves faster when sent to the home positionHome button against jog wheel[15]; the speed values are not given
A patient reports dizziness or a feeling of movement on enteringMovement through the steepest part of the field[3, 4] at 7 T only; the proportion at 1.5 T and 3 T was not found
The console asks to disconnect coil connectors, squeeze ball or headphonesUndocking the table onto a trolley[14] for one system
Head-first patients more often anxious or dizzyHead position inside the boreHead-first examinations had higher claustrophobia in [9] and stronger dizziness in [3]; both confounded by anatomy and scanner
A 7 T patient reports dizziness on enteringTable movement and field change[1] states that limiting motion and a slow table velocity may help; [3, 4] describe the symptoms; no value of speed is given
Something is caught on the table during movementEmergency Table Stop[1]: designed to stop scanning and table motion; its effect on a given system is not documented here
A patient needs resuscitationZone IV emergency response[1, 2]: remove from Zone IV to a predetermined safe location, CPR meanwhile if delayed; an MR Conditional gurney is needed if the table cannot be undocked; no quench for a code
Patients who failed on a standard bore may complete on a wide short boreDiameter and length of the bore[10] only; selected, uncontrolled

9. Discrepancies Between Sources

ItemValues or statementsTreatment
Table speed88 and 105 mm/s at two 7 T systems and 180 mm/s at a 1.5 T system of one vendor [3]; about 69 mm/s by hand [4]; 20 mm/s at another 7 T site [12]All stated; they are different systems and methods; no general value can be given
Cause of movement-induced vertigodB/dt cannot be ruled out [3]; continuous stimulation by the static field favoured, transient mechanisms less likely [4]; limiting motion during table movement stated to decrease vertigo at 7 T, and a slow table velocity to be helpful [1]Views held by the authors and by the manual; not resolved; the study behind the manual statement is not identified; the speed of the table is not shown here to be the controlling variable
Where to take a person in a medical emergencyRemove from the magnet room before care [18]; remove to the console room if it can be done safely, but not out of the room if unconscious or severely injured without qualified assistance [17]; remove immediately from Zone IV to a predetermined, magnetically safe location, with CPR started in Zone IV if transfer is delayed [1, 2]The ACR text read agrees with the UAB extract and does not support the restriction in the Duke extract; the Duke rule is a research-site procedure known only from an extract and may rest on local arrangements for medical assistance (Reasoning); neither edition of the manual gives table mechanics
Effect of bore width on toleranceThreefold lower rate on a newer design [9]; reduction by 2.95 [6]; no significant difference in a news report [23]Confounded, different definitions and populations; the news item is not verified
Brake positionsPositions 1 and 2 have the same description in the extract [15]Unresolved; diagram missing
Length of the standard bore1.5 to 2.25 m [7]; 1.9 m [10]; 1.6 m for a shorter newer design [9]Consistent as a range; length depends on the design and field strength

10. Open Questions

Whether the table speed can reduce the symptoms at 7 T is open, although the ACR Manual states that limiting motion during table movement decreases vertigo [1]: the authors of [3] suggest that exposure may be similar to a passively shielded system when the table is slower, and those of [4] consider a continuous mechanism more likely. Whether a wider bore reduces claustrophobia by itself, apart from noise and length, is open (Section 6). Whether the restriction on moving an unconscious or severely injured subject, found in one research-site extract, has any basis beyond local arrangements is open; the ACR text read directs removal from Zone IV for a code (Section 4.1).

11. Evidence Gaps

GapStatus
Patient-support requirements of IEC 60601-2-33 (including mechanical safety, speeds and emergency release)(to be sourced): standard not read
Table specifications: load limit, travel, vertical range, positioning accuracy, speed ranges at 1.5 T and 3 T(to be sourced): manufacturer documentation needed
Bore covers, lighting, ventilation, intercom, alarm and other comfort features(to be sourced)
Study behind the ACR statement that limiting motion during table movement decreases vertigo at 7 T; edition status of the ACR copy; national guidance outside the United States(to be sourced): the whole copy with a 2024 cover was read through a summarising tool; table release, speed, weight limit and bore dimensions not found in either edition
Symptoms and table speed at 1.5 T and 3 T in patients; a study varying table speed on one scannerNot found
Randomised comparison of bore diameter with noise and length controlledNot found
Effect of a wider bore on gradient performance, homogeneity volume and body-coil efficiency(to be sourced)
Patients of high weight or size and the limits of table and bore for them(to be sourced)
Cleaning, maintenance and quality assurance of the table and of the docking mechanism(to be sourced)
Journal article behind the news report [23] and the full journal text of [5, 6, 7, 8, 9]Not read

12. Evidence-Based References

No guideline or study with an EBM level applies to this descriptive engineering page. Sources are ordered by publication quality; the reading level of each is stated.

A. Guidance Documents

Guidance document
[1] American College of Radiology, Committee on MR Safety. ACR Manual on MR Safety. Cover dated 2024; copy hosted on a third-party site (static.seram.es).
Read through a summarising tool in consecutive ranges covering characters 0 to 385,941 (Chapters 1 to 16 and Appendices 1 to 6), each range answering questions on the table, bore, claustrophobia, transfer and emergencies; passages outside those topics not examined. Edition not confirmed: one extract reported a copyright line suggesting a 2026 update, another that references were accessed in April 2024. Page numbers inferred by the tool; wording paraphrased here.
Guidance document
[2] American College of Radiology, Committee on MR Safety. ACR Manual on MR Safety, version 1.0, 2020 (revisions of 15 April and 15 May 2020); copy hosted by a university department (radiology.wisc.edu).
First 100,000 of 164,784 characters (to the electrically conductive clothing subsection, about p. 35) and a second extract containing Appendices 2 and 3 (pp. 49–56), through a summarising tool; superseded edition; the end of the main text and the rest not read.

B. Peer-Reviewed Articles

Descriptive study
[3] Hansson B, Höglund P, Markenroth Bloch K, Nilsson M, Olsrud J, Wilén J, Björkman-Burtscher IM. Short-term effects experienced during examinations in an actively shielded 7 T MR scanner. Bioelectromagnetics. 2019;40(4):234–249. doi:10.1002/bem.22189.
First 100,000 of 117,512 characters through a summarising tool; single site, no control group, self-reported, unbalanced head-first and feet-first groups.
Experimental study
[4] Mian OS, Li Y, Antunes A, Glover PM, Day BL. On the vertigo due to static magnetic fields. PLoS ONE. 2013;8(10):e78748.
Full text through a summarising tool; 25 volunteers; figures read as captions; one table partly garbled.
Survey in a randomised trial
[5] Iwan E, Yang J, Enders J, Napp AE, Rief M, Dewey M. Patient preferences for development in MRI scanner design: a survey of claustrophobic patients in a randomized study. Eur Radiol. 2021;31(3):1325–1335. doi:10.1007/s00330-020-07060-9.
Search extracts only; trial results not read.
Retrospective comparison
[6] Koh SAS, Lee W, Rahmat R, Salkade PR, Li H. Interethnic variation in the prevalence of claustrophobia during MRI at Singapore General Hospital: does a wider bore MR scanner help? Proc Singapore Healthc. 2017;26(4):241–245.
Catalogue abstract extract only; denominators not read.
Review
[7] Brunnquell CL, et al. Patient comfort in MR scanner design (title not shown in the extract). Top Magn Reson Imaging. 2020;29:167–174.
Search extract only; used only for the dimensions of the common design.
Prospective comparison
[8] More space, less noise — new-generation low-field magnetic resonance imaging systems can improve patient comfort: a prospective 0.55T–1.5T-scanner comparison. PMC9692643 (authors and journal not shown in the extract).
Abstract extract only.
Cohort
[9] Dewey M, Schink T, Dewey CF. Claustrophobia during magnetic resonance imaging: cohort study in over 55,000 patients. J Magn Reson Imaging. 2007;26:1322–1327. doi:10.1002/jmri.21147.
Journal text not read; figures from the RSNA 2007 abstract SSC22-09 of the same cohort, read in full through a summarising tool; conventional bore not described.

C. Conference Abstracts

Conference abstract
[10] Lane JI, Wood CP, Bolster BD Jr, Johnson KJ, Bernstein MA, Witte RJ. Wide short bore MR at 1.5T: reducing the failure rate in severe claustrophobics. Proc Intl Soc Mag Reson Med. 2008;16:2713.
Full page through a summarising tool; chart not legible; two authors affiliated with the manufacturer.
Conference abstract
[11] Ahn K, Kim I, Kim B. Benefit of wide-bore magnet in reducing patient's motion: comparison with narrow-bore magnet. RSNA 2010, SSJ22-05.
Search extract of the abstract; units and spread statistic not shown.
Conference abstract
[12] Sensations of Japanese subjects during 7 T examination with a slow table-feed speed (title and authors not shown in the extract). Proc Intl Soc Mag Reson Med. 2014;22:abstract 4870.
Search extract only.
Conference abstract
[13] Patient experience on a compact 3T scanner (C3T) compared with 60-cm whole-body systems (authors not shown in the extract). Proc Intl Soc Mag Reson Med. 2018;26:abstract 2636.
Search extract only; research system.

D. Written Site Procedures

Site procedure
[14] Queen's University Centre for Neuroscience Studies, fMRI Facility. Standard Operating Procedure #10: Exiting a person from the magnet in an emergency. Undated.
Read through a summarising tool; 3 T, manufacturer not named; the excerpt ends before its last step.
Site procedure
[15] Emergency subject evacuation (3T). January 2023. Published on the oxcin.ox.ac.uk site; the file names no organisation.
Read through a summarising tool; diagram missing; what must be unplugged and the transfer to a trolley are not described.
Site procedure
[16] Emergency subject evacuation (7T). January 2023. Same site as [15].
Search extract only.
Site procedure
[17] Duke University Brain Imaging and Analysis Center. SOP104: Medical emergencies.
Search extract only.
Site procedure
[18] University of Alabama at Birmingham Radiology. MRI safety policy R31r6.
Search extract only; the vendor-specific lines were cut in the extract.

E. Patents

Patent
[19] US 10739421. Systems and methods for table movement control.
Search extract; applicant's background statement; assignee not shown.
Patent
[20] US 11786181. Magnetic resonance apparatus with a patient positioning apparatus.
Search extract; applicant's background statement; assignee not shown.
Patent
[21] US 10747412. Method and apparatus for adjustment of a table position in a medical data acquisition scanner.
Search extract; tolerance value not shown.

F. Manufacturer Article and News

Manufacturer article
[22] GE HealthCare. MRI bore sizes and benefits (Portuguese edition).
Search extract only; promotional context; no data or references shown.
News report
[23] Radiology Business. News item on a European Journal of Radiology study of early termination of MRI examinations (journal article not identified in the extract).
Search extract only; the study was not read.

Recent PubMed search for this protocol

Last updated: November 2026

Report an error on this page

The report stores the page address, the text above and, if you give it, your e-mail. It does not store your IP address, browser type or any cookie. See the Disclaimer, section 7.