Laser Positioning, Landmarking, Isocentre and the Scanner Coordinate System in MRI — Table Movement, Offsets, Relation to the Patient Frame and Positioning Quality Control
MRIninja Knowledge Base | Foundations — Geometry and Positioning Parent page: MRI Physics — Fundamentals and Principles (9201) Builds on: Patient Planes, Orientation and Positions in MRI (9701) Version 1.1 — November 2026
Foundations — Geometry and Positioning
Laser Positioning, Landmarking, Isocentre and the Scanner Coordinate System
Reference page on how a point of the patient is brought to the magnet isocentre, how slice and field-of-view offsets are realised, how the scanner frame relates to the patient frame, and how positioning accuracy is checked.
1. Scope, Boundary and Information Status
The patient frame, planes and position codes are defined on Patient Planes, Orientation and Positions and are not repeated here. This page covers the equipment side: the positioning laser light, the landmark, the isocentre, table travel, the physical axes of the scanner, the way slice and field-of-view offsets are produced, how these map to the DICOM image position, and how positioning accuracy is tested. The magnet, gradient system, table and coils as hardware are planned as separate pages (Section 11).
Information status. Seven sources were consulted, all online and none as a full peer-reviewed article. The core is a conference educational poster on geometry in MR reconstruction (Atkinson et al., ISMRM 2018) [2], the DICOM Image Plane Module [1], ACR phantom test instructions [3] and two undated educational pages by one author [4, 5]. Every page was read through a tool that returns a model-written summary of the page, not the raw text; this is a limit on the exactness of the details quoted. The IEC 60601-2-33 safety standard was available as a preview only and defines neither coordinate system nor isocentre in the part seen [6]. Statements not supported by these sources are marked (Reasoning) (derived) or (to be sourced) (no source read).
| Source | What it supports | Reading level / weight |
|---|---|---|
| [1] DICOM PS3.3 C.7.6.2 | Patient-based coordinate system, Image Position and Orientation (Patient) | Summarised page of the standard; high authority, summarised reading |
| [2] Atkinson et al., ISMRM 2018, abstract 5652 (poster, UCL) | Magnet XYZ vs patient axes, laser point vs isocentre, offsets, gradient angulation, IPP computation | Abstract page with figure captions, summarised reading; conference educational poster, not peer-reviewed article |
| [3] ACR Large/Medium Phantom Testing: MRI (revised 2026-02-06) | Phantom positioning with the indicator light and levelling; geometric slice layout | Summarised web page; no numeric tolerance given on the page |
| [4, 5] Elster, mriquestions.com: Slice position and accuracy; Routine quality control | Tolerances, causes of slice-position error, test frequency | Expert educational site, undated, no author-stated sources on the pages read; low-to-moderate weight |
| [6] IEC 60601-2-33 ed. 4.0 (2022-08), preview | Scope; B0 stated at MR isocentre with ±5% accuracy | Front matter only |
| [7] ISMRM 2012 abstract 1451 (automatic landmarking) | Existence of automated landmark approaches; knee must be at best homogeneity | Text garbled in extraction; lowest weight |
2. Vocabulary
| Term | Meaning used on this page | Status |
|---|---|---|
| Positioning (laser) light | A light line or cross projected on the patient to mark where the patient is relative to the bore; set before the table moves | Described as "initial laser positioning" that sets a point in the body [2]; ACR calls it the positioning indicator light [3] |
| Landmark | The point or plane of the patient that the operator marks with the light (for example the centre of the coil or of the region of interest) | Operator-chosen reference (Reasoning); the word is used for the automatic detection in [7] |
| Isocentre (MR isocentre) | The reference point of the magnet at which the nominal B0 is specified (IEC: B0 given at MR isocentre with ±5% accuracy) and about which the imaging and shim gradients vary the field | [6] (B0 specification only); [2] (gradients vary field about the isocentre); no definition read (to be sourced) |
| Table movement | Translation of the patient support to bring a chosen point to the isocentre or to image different regions | [2] |
| FOV centre / offset | The position of the centre of the field of view or slab relative to the isocentre, set by the operator on the localiser | Definition (Reasoning); implementation in Section 5 |
3. Frames and Their Relation
Three frames matter (see also Section 2 of Patient Planes, Orientation and Positions): the patient-based frame of DICOM (x toward the patient’s left, y posterior, z toward the head) [1]; the magnet frame with the physical X, Y and Z gradient axes [2]; and the table, whose travel moves the patient through the magnet (Reasoning).
| Statement | Source |
|---|---|
| DICOM uses a right-handed patient system with Left-Posterior-Head axes; its origin is not defined by the standard but stays consistent between scans until the patient is repositioned | [2]; the axis directions in [1] |
| One choice of origin is the magnet isocentre plus the table movements | [2] |
| The orientation of the patient axes relative to the magnet XYZ axes depends on head-first or feet-first entry and on supine or prone position | [2] |
| In-plane gradient directions and the slice-select gradient can have any angulation and are formed by combinations of the physical X, Y and Z gradients | [2] |
| Shim fields may be defined along X, Y and Z | [2] |
| The poster does not define a gantry or device frame, table axes or explicit transformation matrices between magnet, table and patient frames | [2] (limit stated by the summary of the page) |
Consequence (Reasoning): the mapping between patient axes and magnet axes is different for each of the entry directions and body positions listed on Patient Planes, Orientation and Positions, so the same operator action (for example “move the table in”) does not correspond to the same patient direction for head-first and feet-first entry. The numerical form of the mapping for each combination, the naming and sign of the magnet axes by system, and the direction of the B0 field relative to the axes were not found in the sources read (to be sourced).
4. Laser Positioning, Landmarking and Table Movement
4.1 What the Sources State
The poster describes the sequence as follows: the initial laser positioning sets a point in the body; after planning, table movements can leave that point away from the magnet isocentre [2]. The ACR phantom instructions give the version of the procedure used in quality control: place the phantom’s centre notch in the centre of the head coil, align it with the positioning indicator light so that it is at isocentre, check levelling in all three axes with a non-metallic bubble level, keep the phantom unchanged for the whole series, and, after moving it to magnet centre, verify the position with a sagittal and optionally a coronal localiser, repeating until correct [3].
4.2 Generic Workflow (derived)
The table below is a logical sequence built from the definitions above, not a documented protocol (Reasoning). Site and manufacturer procedures differ and none was read (to be sourced).
| Step | Action | Why it matters (derived) |
|---|---|---|
| 1 | Position the patient and the receive coil; confirm the entered patient position (see 9701 Section 7) | The entered position governs orientation labels |
| 2 | Set the landmark with the positioning light on the structure that defines the coil or region centre | Fixes which point will be brought to the isocentre |
| 3 | Move the table so the landmark reaches the isocentre (distance d in Figure 1) | Places the region of interest where the gradients and B0 specification refer to [2, 6] |
| 4 | Acquire localisers and plan the slices or slab on them | Planning is done in the frame of the localiser |
| 5 | If the patient moves or the table is moved after planning, check or repeat the localiser | The planned offsets refer to the table position at the time of the localiser (Reasoning) |
5. How Offsets and Slice Position Are Realised
| Quantity | How it is produced | Source |
|---|---|---|
| Slice or slab position | Set by the RF frequency together with the slice-select gradient | [2] |
| Slice orientation (angulation) | The slice-select gradient can have any angulation, made by combining the physical X, Y and Z gradients | [2] |
| Frequency-encode (readout) offset | Change of the centre frequency | [2] |
| Phase-encode offset | A linear change in the acquired phase | [2] |
| Difference between requested and realised offset | Can arise when a half-pixel shift is added during acquisition, or when in-plane offsets are applied by post-processing shifts (for example with readout ramp sampling) | [2] |
| k-space origin | The point acquired when the net gradient area is zero | [2] |
The poster validates geometry by comparing scans acquired at several offsets, angulations and phase-encode directions [2]. Whether and how this is done in routine clinical quality control is not stated in the sources read (to be sourced).
6. Relation to the DICOM Image Position and Orientation
Each image is located by Image Position (Patient), the position of the upper-left voxel centre in millimetres, and by the row and column direction cosines of Image Orientation (Patient) [1]. The slice order of a multi-slice volume follows from the scalar product of the position with the slice normal, which is the cross product of the row and column vectors [2]. These relations tie the offsets of Section 5 to the patient frame of Patient Planes, Orientation and Positions (Reasoning): the position of each slice is the isocentre-referenced centre plus the table offset, expressed in the patient system.
| Point | Statement | Source |
|---|---|---|
| Origin | Not defined by the standard; one practical choice is the isocentre plus table movements | [1, 2] |
| Reference point of the image position | DICOM: centre of the first transmitted voxel; the poster uses the centre of the top-left pixel, whereas some implementations use the edge of that corner pixel | [1, 2] |
| Transformations (reflection, rotation by 90°, zero-padding, cropping) | Keep the voxel size; the spatial reference is updated through the position and orientation vectors; for a 90° counter-clockwise rotation the new row vector is the old column vector and the new column vector is the negative of the old row vector | [2] |
| Software with different corner conventions | Positions of the same voxel can differ by half a voxel between implementations; checking by phantom is advisable | (Reasoning) |
7. Why the Landmark Position Matters
The sources give three links between position and image quality, none of them quantified in the sources read (to be sourced):
| Link | Statement | Source / status |
|---|---|---|
| Gradient non-linearity | The poster recommends imaging near the isocentre when validating geometry to reduce gradient non-linearity effects | [2] |
| Field homogeneity | B0 is specified at the isocentre; an automatic landmark abstract notes that the knee must be positioned at the best field homogeneity | [6]; [7] (garbled abstract, lowest weight) |
| Slice position accuracy | Listed causes of slice-position error include operator set-up, laser–table misalignment, table-positioning inaccuracy, gradient miscalibration, slice-select gradient problems and poor B0 homogeneity | [4] (educational page) |
Region-specific effects (fat suppression failure and geometric distortion for off-centre regions, shoulder and extremity imaging off the isocentre) were not found in the sources read (to be sourced). They belong on regional pages once sourced.
8. Quality Control of Positioning
| Item | Content | Source / weight |
|---|---|---|
| Phantom set-up | Centre notch of the phantom in the coil centre, aligned with the positioning light; bubble-level check on the z-axis and the chin bar; verification with sagittal (and optionally coronal) localisers, repeated until correct | [3] |
| Slice-position test layout | Sagittal localiser spin echo through the phantom centre, 10 mm preferred (20 mm acceptable); axial series of 11 slices of 5 mm with a 5 mm gap, slice centres spanning 100 mm; crossed-wedge vertices 100 mm apart; for 3D site protocols 21 slices of 5 mm with odd-numbered images analysed | [3]; the page gives no numeric tolerance |
| Slice-position tolerance | Generally ±10% of the prescribed position (slice position, via two opposing ramps) | [4] educational page |
| Laser–table alignment check | Sagittal localiser; the central alignment bar of the ACR phantom expected within about 2 mm of the laser-established centre | [4] educational page |
| Table position accuracy | Measured with the weekly phantom set; an example tolerance of ±1 mm is given as an example only, thresholds being site-established | [5] educational page; [5] cites AAPM Report 100 (2010), not read |
| Frequency | Daily visual inspection of hardware and general image quality; at least weekly phantom measurements including table position; detailed assessment by a trained physicist or engineer at delivery and every 6–12 months | [5] educational page |
Limit: the numeric tolerances above come from educational pages without stated primary sources; the ACR page itself gives none. They should not be used as acceptance criteria. The original ACR phantom test guidance, AAPM Report 100 and the IEC standard were not read (to be sourced).
9. Failure Modes and Checks (Derived)
| Failure | Consequence | Check | Status |
|---|---|---|---|
| Landmark set on a structure other than the intended centre | Region of interest off the isocentre | Verify on the localiser that the region is centred (Section 4.2) | (Reasoning) |
| Table moved or patient moved after planning | Planned offsets no longer match the anatomy | Repeat the localiser and re-plan | (Reasoning) |
| Wrong patient position entered | Wrong orientation labels and image position mapping | Compare entered and physical position before scanning (9701 Section 7) | (Reasoning) |
| Different corner convention between software | Half-voxel mismatch of positions between tools | Compare a phantom across tools | (Reasoning); [2] |
| Requested and realised offset differ | Half-pixel shift or post-processing shift added | Check the realised position in the image header; not only the requested value | [2] |
| Laser or table out of tolerance | Systematic position error | Periodic phantom test (Section 8) | [3, 4, 5] |
10. Evidence Gaps and Limits
(a) No peer-reviewed article, guideline or manufacturer-independent textbook was read for any statement on this page; the strongest sources are a conference poster [2] and the DICOM standard [1]. (b) All pages were read through a summarising tool and one abstract was garbled [7]. (c) No definition of isocentre or of the magnet axes was read; the IEC preview contains none [6]. (d) Numeric position tolerances come from undated educational pages [4, 5]. (e) Laser safety, laser class and the technical design of the positioning light were not read (to be sourced). (f) Practice by region (where to place the landmark for each coil and anatomy) is not given here and was not sourced (to be sourced). (g) Open-bore, low-field and mobile-table systems may use different procedures; none was read (to be sourced). (h) The figure is a constructed schematic.
11. Planned Related Pages (Not Yet Published)
Published: MRI Scanner Structure and Components — Overview. Planned: the component pages (magnet, gradient system, radiofrequency transmit system, table, bore, cabinets and console) and radiofrequency coils by type. Until they exist, the hardware terms are used here only as far as the positioning workflow needs them.
12. Evidence-Based References
No guideline or study with an EBM level applies to this page. Sources are ordered by authority for the claims they support; every source is listed with the way it was read.
A. Standards and Technical Specifications
B. Technical and Educational Sources
Related Protocols
Recent PubMed search for this protocol