Parallel Acquisition Options: Reference Scans, Simultaneous Multi-Slice and CAIPIRINHA

MRIninja Knowledge Base | MRI Parameter Deep Dive Parent page: MRI Parameters — Overview and Classification (9501), Section 6.3 Builds on: Parallel Imaging (9588) Version 1.0 — November 2026

MRI Parameter Deep Dive

Parallel Acquisition Options: Reference Scans, Simultaneous Multi-Slice and CAIPIRINHA

What the calibration data of an accelerated scan are and what they cost, how controlled aliasing (CAIPIRINHA) and simultaneous multi-slice (SMS) acquisition change the acquisition rather than the reconstruction, how the slice-direction FOV shift is chosen, what is published about their g-factor and SNR, and which labels a Siemens console shows for them.


1. Scope, Boundary and Information Status

The principles of SENSE-type and GRAPPA-type reconstruction, the g-factor, the acceleration factor R, the effective acceleration after autocalibration lines and a short description of simultaneous multi-slice imaging are on Parallel Imaging and are not repeated. This page adds what a console exposes and what each option changes: reference lines and reference scans, controlled aliasing in 3D, SMS factor and FOV shift, and the label mapping. Partial Fourier, which the console offers in the same tab, is on 9563; echo-train arithmetic is on 9561. Compressed sensing, wave-CAIPI, deep-learning reconstruction and diffusion-specific SMS protocols are not covered.

Information status. No guideline, consensus statement or systematic review on these options was found. The substance rests on two original papers read as full text through a tool that returns a model-written answer about the page, not the raw text [1, 4]; one vendor article [3]; one methods paper read up to 100,000 of 139,627 characters [2]; and abstract or excerpt-level items [5, 6, 7, 8, 9, 11]. The vendor article was written by the technique's originators and published by a manufacturer; its statements are not independent evidence. Statements marked (Reasoning) are derived by the editor; gaps are marked (to be sourced). Console values come from screenshots of one 3 T head-and-neck examination supplied by the site owner [10].

SourceTypeReading level and weight
[1] Griswold et al., MRM 2002 (GRAPPA)Landmark original articleFull text through a summarising tool; some equations garbled
[2] Brau et al., MRM 2008Methods comparison of autocalibrating reconstructionsFirst 100,000 of 139,627 characters through a summarising tool; 1D-accelerated Cartesian data only
[3] Breuer, Blaimer, Griswold, Jakob, MAGNETOM Flash 2015, "CAIPIRINHA – Revisited"Vendor article by the technique's authorsFull text through a summarising tool; layout-garbled passages
[4] Setsompop et al., NeuroImage 2012Original article (diffusion SMS)Full text through a summarising tool; one 3 T and one Connectom system, research setting
[5] Setsompop et al., MRM 2012 (blipped-CAIPI)Original articleAbstract only (library records)
[6] Breuer et al., MRM 2005 (CAIPIRINHA)Landmark original articleAbstract excerpt only
[7] ACS-mismatch artefacts at 11.7 T, MRM 2025Research articleAbstract and record excerpt only; authors not retrieved; ultra-high field
[8] Bertle, TU Wien thesis 2024 (CAIPIRINHA in susceptibility imaging)Academic thesisAbstract excerpt only; not peer reviewed; lowest weight
[9] Patent documents US 8,354,844, US 10,330,759 and US 9,989,610PatentsExcerpts only; they describe implementations, not performance; lowest weight
[10] Console screenshotsUnpublished observationsRead visually; not vendor documentation
[11] Sheltraw et al., arXiv 1208.0972 (autocalibration errors in GRAPPA EPI)PreprintAbstract and excerpt only; not peer reviewed

2. Calibration Data: Reference Lines and Reference Scans

Data-driven (GRAPPA-type) methods need fully sampled autocalibration signal (ACS) lines from the central k-space to fit the reconstruction weights and do not need coil sensitivity maps; physically based (SENSE-type) methods need explicit coil sensitivity information [2]. In the original GRAPPA article the weights are fitted from acquired lines to an ACS line, each coil is reconstructed separately into an uncombined image and the images are then combined, in that study by sum of squares [1]. The calibration lines can be acquired within the accelerated scan, or before or after it [2].

OptionWhat the sources say
Integrated (embedded) reference lines"Commonly used because it minimizes artifacts from misregistration" [2]. A 3D GRE study at 11.7 T reports that artefacts arise from a mismatch between the calibration data and the accelerated data, because the calibration lines have an intrinsically lower resolution, and that an externally acquired reference with a short echo time gave fewer artefacts than an integrated one [7]. A thesis excerpt reports that integrated reference lines roughly doubled the acquisition time and caused strong artefacts at later echoes in a multi-echo susceptibility acquisition with CAIPIRINHA [8]. These two findings come from ultra-high-field or specialised multi-echo 3D settings and do not by themselves transfer to clinical protocols.
Separate reference scanUsed in the original article for single-shot HASTE of the lung at 3× acceleration, where the lines were acquired before the reduced acquisition and discarded, and for a cine acquisition where eight extra lines were acquired once and reused in other phases [1]. The vendor article shows an "extra reference scan" for a breath-hold liver VIBE with a matrix of 32 × 24 × 24 and for a brain 3D FLASH with 32 × 32 × 32, and does not compare separate with integrated references [3].
Number of reference linesMore lines add training examples and improve the fit but reduce efficiency [2]; the methods paper used 20 lines in both experiments [2]. The original article used 16 reference lines in a 3D GRE example [1]. A typical range of 16 to 32 lines for 1D acceleration appears in a patent excerpt [9], which is not performance evidence.

Common sources of calibration error stated for physically based methods are low SNR, Gibbs ringing, motion and a tight field-of-view prescription [2]. For EPI, a preprint excerpt states that interleaved calibration acquired over several shots can be corrupted by perturbations between the shots, producing artefacts in the reconstructed images [11]; echo-planar calibration is not otherwise developed here (to be sourced).

2.1 What the reference lines cost (Reasoning, using the effective-acceleration relation of 9588)

With integrated lines the number of acquired lines is Nacq = Nref + (Nlines − Nref) / R. For a hypothetical 190 phase-encoding lines, 32 reference lines and R = 2: 32 + (190 − 32) / 2 = 111 lines, an effective acceleration of 190 / 111 = 1.71 rather than 2. With a separate reference scan the full count of the accelerated scan is unchanged but the reference acquisition adds its own time, which a console may or may not include in the displayed duration (to be sourced). The original article states that for single-shot imaging the calibration lines add acquisition time and does not quantify it [1].

3. Controlled Aliasing in 3D and Total Acceleration

CAIPIRINHA modifies the appearance of the aliasing during the acquisition so that the following parallel reconstruction is better conditioned [6]. In 3D, sampling positions are shifted in the partition direction by an offset Δ as additional gradient offsets in the phase-encoding tables; an elementary cell of PATy × PATz positions defines the pattern, with the total factor the product, for example 2 × 3 = 6 [3]. For a total factor of 4 the vendor article shows 1 × 4, 2 × 2 and 4 × 1 patterns with shifts from 0 to 3; Δ = 0 is conventional 2D acceleration and the others are CAIPIRINHA-type [3]. The same article states that selecting the optimal 2D pattern automatically is "a challenging task" that is not solved, and that sensitivity variations depend on coil geometry, field of view and object position, so results vary [3].

Published performance. The vendor article reports lower g-factors for 2D-CAIPIRINHA than for conventional 2D acceleration in an abdominal example (breath-hold VIBE, total factor 4, 9 s at 1.5 T) and shows that a 1 × 5 pattern at a total factor of 5 compares well with 2 × 2 at 4 in a brain 3D FLASH example at 3 T, with scan times of 1 min 16 s against 1 min 40 s [3]. These are single examples with qualitative or map-based comparison, not a trial. The effect depends on the receiver array; the console values of Section 6 do not state the coil used for the series.

4. Simultaneous Multi-Slice (SMS): Factor, FOV Shift and Calibration

In SMS acquisition several slices are excited together with summed frequency-modulated pulses and separated with the coil sensitivities [4]. For two-dimensional multi-slice acquisition with RF phase cycling, alternating multiband pulses give each slice a phase cycle along the phase-encoding direction: alternating two pulses shifts every second slice by FOV/2, a four-pulse set shifts by FOV/4 [3]. That alternation does not work for single-shot EPI, which needs blipped-CAIPIRINHA, where the shift is produced with gradient blips [3, 5].

StatementValueSource and level
g-factor penalty of 3-fold slice acceleration at 3 T, 32-channel head array, with inter-slice shift versus withoutBelow 1 % versus 32 %[5] abstract (Monte-Carlo simulation); also quoted in [4] as prior work
3-fold slice acceleration combined with 2-fold in-plane accelerationg-factor penalty 19 % at 3 T and 10 % at 7 T, versus 41 % and 23 % without slice shift[5] abstract (simulation)
Measured SNR ratio (proxy for 1/g) in [4], 3 T-class research system3× slice acceleration alone: 1.04 ± 0.07 average, minimum 0.89; 3× combined with 2× in-plane: 0.9 ± 0.08 average, minimum 0.73[4] full text
Net time saving for 3× SMS after the SNR costThe SNR cost corresponds to 1.38 averaged acquisitions, a net saving of about 2.2-fold (3 / 1.38)[4], diffusion protocol

The FOV shift. In [4] the field of view in the slice direction is small compared with the in-plane field of view in the example given (12 cm against 21 cm), so closely spaced slices alias close to each other and produce a high g-factor penalty unless the shift separates them. In that paper a shift of FOV/2 was used without in-plane acceleration; with 2× in-plane acceleration a FOV/2 shift would give no net shift, so FOV/4 was used [4]. The rule is the following (Reasoning): with in-plane acceleration R the aliased in-plane replicas are separated by FOV/R, and the inter-slice shift must be chosen so that it does not coincide with them.

Calibration. The slice-separation kernel is fitted to each slice of a pre-scan calibration dataset acquired one slice at a time, with separate kernel sets per simultaneously excited slice, and separate even-line and odd-line kernels to reduce inter-slice ghosting [4]. Patent excerpts describe a multiband TSE sequence with one-half or one-third FOV shift and a gradient-echo reference scan, and a TSE-based reference scan for SMS [9]; which reference type a given product uses and its consequence for image quality are (to be sourced).

Costs and limits stated. Multi-slice excitation deposits significantly more RF energy, limiting the number of slices or requiring low flip angles [3]. In [4] pulse design (SLR, VERSE) was used to limit peak voltage and SAR, and no numerical SAR value is given. Slice cross-talk: with interleaved SMS, when the number of slices per subgroup is even, adjacent slices at subgroup boundaries are excited back to back and produce signal loss at edge slices; an odd number of slices per subgroup was recommended [4]. This concerns the slice ordering of the interleaved multi-slice scheme; see 9592 for scan order in general.

5. Console View (Siemens, One 3 T Head-and-Neck Examination)

Values read from screenshots [10]; labels are the console's. For single-tab captures the series is not stated and its assignment is inferred (Reasoning).

Console label and modeValues observedSeries (inferred)
SMSAcceleration factor PE 2; reference lines PE 64; SMS factor 2; FOV shift factor 4; reference scans TSE / SeparateA TSE series (probable)
GRAPPAAcceleration factor PE 2; reference lines 32; self-calibrated; phase partial Fourier 6/8A TSE series (probable)
CAIPIRINHATotal factor 2; mode Free; reference scans GRE with Separate or Integrated; reference lines PE 24–30; acceleration factor 3D 1; phase and slice partial Fourier Off or 7/8; asymmetric echo Weak3D VIBE-type series

Observations (Reasoning, arithmetic and comparison only). (a) SMS factor 2 with PE acceleration 2 and a FOV shift factor of 4 is consistent with the rule of Section 4 (FOV/4 when in-plane acceleration is 2) if "FOV shift factor 4" means a shift of FOV/4; the console documentation of the label is (to be sourced). (b) The total acceleration of such a scan would be the product of the SMS factor and the in-plane factor, as in the convention of 9588, but the console capture does not show a combined figure. (c) For the CAIPIRINHA series the total factor of 2 with a 3D acceleration factor of 1 corresponds to an in-plane factor of 2 and no acceleration in the partition direction, so the series is in effect a one-dimensional acceleration; what "Free" selects is (to be sourced). (d) The reference-line counts differ between modes (64 for SMS, 32 for GRAPPA, 24–30 for CAIPIRINHA); no source read explains how a vendor sets them, and the original article's examples of 16 and 8 lines concern different sequences [1]. (e) The reference scan of the SMS capture is labelled as TSE and Separate, which agrees with the existence of TSE-based reference scans described in a patent excerpt [9]; whether the time of a separate reference scan is included in the displayed scan time is (to be sourced).

6. Technologist Decision Points (Reasoning)

  1. Treat the reference lines as part of the scan time. With integrated lines the effective factor is lower than the nominal one (Section 2.1); with a separate reference scan check where the time is added.
  2. Use separate references when a mismatch between calibration and imaging data is likely (very short or multiple echo times, ultra-high field) and integrated references when misregistration between scans is the larger risk [2, 7] — a decision rule built from two limited findings, not a recommendation of a guideline.
  3. Do not combine more acceleration than the coil can support. The published g-factor figures come from 32-channel head arrays at 3 T and 7 T [4, 5]; for other coils they are not transferable (Reasoning).
  4. When using SMS, check RF energy first (flip angles, SAR display) because it is the stated limit on the number of simultaneous slices [3].
  5. Interpret residual aliasing or shading as a calibration problem first: motion between the reference and the accelerated data, a tight field of view and low SNR are the stated causes [2].

7. Discrepancies, Controversies and Limits of the Evidence

  • SNR of SMS. The vendor article states that SMS gives an SNR gain of about the square root of the number of slices at a cost in RF energy [3]; page 9588 writes SNRSMS = SNRsingle / (gz × √MB). Reasoning: the two statements use different baselines (the same total scan time against the same number of slices per excitation); the baseline is not stated in either. The paper [4] reports the net time saving of 2.2-fold for 3× acceleration after the SNR cost.
  • Integrated against separate reference lines. Integrated lines are described as commonly used because they minimise misregistration [2]; two excerpts at ultra-high field and in multi-echo imaging report that integrated lines caused artefacts or doubled the acquisition time [7, 8].
  • g-factor values. The values of Section 4 are simulation results [5] or research systems [4], and the vendor article states that results vary with coil, field of view and object position [3].
  • Weight of the sources. Two original papers are read in full, the methods comparison only up to 100,000 of 139,627 characters; Breuer 2005 and Setsompop MRM 2012 only as abstract. The vendor article is by the technique's authors.
  • Not covered: clinical outcome studies, vendor reconstruction details, the wave-CAIPI and non-Cartesian variants, SMS in diffusion beyond one example, the meaning of console labels not captured (for example the available modes of CAIPIRINHA beyond "Free").

8. Evidence-Based References

No guideline, consensus statement or systematic review on parallel acquisition options was found. Sources are ordered by type and recency; the reading level of each is stated.

A. Guidelines / Consensus / Systematic Reviews

None found.

B. Original Articles and Methods Papers

Original article
[7] On the impact of artifacts induced by mismatches between auto-calibration signal and accelerated 3D GRE data at 11.7 T. Magn Reson Med. 2025;95(3). doi:10.1002/mrm.70127 (authors not retrieved).
Abstract and record excerpt only; ultra-high field.
Original article
[4] Setsompop K, Cohen-Adad J, Gagoski BA, Raij T, Yendiki A, Keil B, Wedeen VJ, Wald LL. Improving diffusion MRI using simultaneous multi-slice echo planar imaging. NeuroImage. 2012;63:569–580.
Full text through a summarising tool; research systems.
Original article
[5] Setsompop K, Gagoski BA, Polimeni JR, Witzel T, Wedeen VJ, Wald LL. Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty. Magn Reson Med. 2012;67(5):1210–1224. doi:10.1002/mrm.23097.
Abstract only; simulation figures.
Methods paper
[2] Brau ACS, Beatty PJ, Skare S, Bammer R. Comparison of reconstruction accuracy and efficiency among autocalibrating data-driven parallel imaging methods. Magn Reson Med. 2008;59(2):382–395. doi:10.1002/mrm.21481.
First 100,000 of 139,627 characters through a summarising tool; 1D-accelerated Cartesian data.
Preprint
[11] Sheltraw D, Deshpande V, Trumpis M, Inglis B. Simultaneous reduction of two common autocalibration errors in GRAPPA EPI time series data. arXiv:1208.0972.
Abstract and excerpt only; not peer reviewed.

C. Vendor Article, Thesis and Patents

Vendor article
[3] Breuer F, Blaimer M, Griswold M, Jakob P. CAIPIRINHA – Revisited. MAGNETOM Flash. 2015;(63) 3/2015. Siemens Healthineers.
Full text through a summarising tool; written by the technique's authors.
Thesis
[8] Bertle J. Susceptibility imaging using CAIPIRINHA acceleration. Master's thesis (level not stated in the excerpt), TU Wien, 2024.
Abstract excerpt only; not peer reviewed.
Patents
[9] US 8,354,844 (data acquisition acceleration with two or more receiver coil arrays); US 10,330,759 (turbo spin echo-based reference scan for simultaneous multislice acquisitions); US 9,989,610 (multiband slice-accelerated imaging with balanced slice-selective gradients).
Excerpts only; describe implementations, not performance.

D. Landmark Historical References

Landmark
[1] Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202–1210. doi:10.1002/mrm.10171.
Full text through a summarising tool; some equations garbled.
Landmark
[6] Breuer FA et al. Controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi-slice imaging. Magn Reson Med. 2005;53(3):684–691. PMID 15723404.
Abstract excerpt only; author list not verified in this session; volume and pages as cited in a patent excerpt.

E. Console Observations (Unpublished)

Observation
[10] Screenshots of the Resolution (Acceleration) tab of one Siemens 3 T head-and-neck examination, supplied by the site owner.
Read visually; not vendor documentation; no patient data used.

End of document — Parallel Acquisition Options — MRIninja v1.0 — November 2026