Partial Fourier and Asymmetric Echo: Phase, Slice and Readout Direction

MRIninja Knowledge Base | MRI Parameter Deep Dive Parent page: MRI Parameters — Overview and Classification (9501), Section 6.2 Builds on: Signal Localisation and Image Formation — k-Space (9205), Section 6.3 Version 1.0 — November 2026

MRI Parameter Deep Dive

Partial Fourier and Asymmetric Echo: Phase, Slice and Readout Direction

What is saved and what is paid when only part of k-space or only part of an echo is acquired: the conjugate-symmetry principle and why phase errors limit it, the reconstruction methods, the arithmetic of lines, scan time and SNR for partial Fourier in the phase direction, the different purpose of the asymmetric (partial) echo in the readout direction, and the labels a Siemens console shows for both.


1. Scope, Boundary and Information Status

The principle of conjugate symmetry and a short statement of partial Fourier acquisition are on Signal Localisation and Image Formation — k-Space (Section 6.3) and in 9501 (Section 6.2). This page adds the parameter view: the three directions in which an asymmetry can be applied (phase, slice, readout), how each changes lines, time, echo time and SNR, the reconstruction methods and where each fails, and the console labels. Echo trains and scan-time arithmetic of turbo spin echo are on 9561; parallel imaging is on 9588; the Dixon acquisition is on 9562. Compressed sensing and deep-learning reconstruction are not covered.

Information status. Eleven items were used. No guideline, consensus statement or systematic review on partial Fourier parameters was found. Most of the substance rests on educational web pages and a book-chapter-style text, read through a tool that returns a model-written answer about the page, not the raw text [1, 2, 3, 4]; one research article on partial echo was read the same way [5]; one 1996 technical note [6]; and four items known only as abstract or search-result excerpt [7, 9, 10, 11]. Equations of [1] were lost in extraction. Statements marked (Reasoning) are derived by the editor; gaps are marked (to be sourced). Console values come from screenshots of one 3 T head-and-neck examination supplied by the site owner [8].

SourceTypeReading level and weight
[1] Chapter 2, "Partial k-Space Reconstruction" (hosted file "Recon_Pauly_read"; author not named in the retrieved text)Textbook-style chapterFull chapter through a summarising tool; OCR errors, equations lost; 0.5 T example images
[2] Elster, mriquestions.com, "Partial Fourier"Educational web pageFull page through a summarising tool; no SNR figures on this page
[3] Elster, mriquestions.com, "Phase conjugate symmetry"Educational web pageFull page through a summarising tool; source of the SNR and scan-time figures; no primary reference for them was shown
[4] Elster, mriquestions.com, "Read conjugate symmetry"Educational web pageFull page through a summarising tool
[5] Bydder et al., NMR Biomed 2021Research article (one 3 T system, 2-channel coil)Essentially all text through a summarising tool; used only for the partial-echo statements
[6] Semelka et al., JMRI 1996Technical note (HASTE)Full text through a summarising tool; OCR errors
[7] Noll, Nishimura, Macovski, IEEE Trans Med Imaging 1991Landmark (homodyne)Abstract only (library record)
[8] Console screenshotsUnpublished observationsRead visually; not vendor documentation
[9] mriquestions.com, "Why not use both"Educational web pageSearch-result excerpt only
[10] Single-shot fast spin echo with variable refocusing flip angle, 3 T abdomen (PMC4684814)Research articleAbstract and introduction excerpt only; authors and journal not retrieved
[11] HASTE-based VASO study (PMC2797844)Research articleMethods excerpt only; authors and journal not retrieved; used for one acquisition-order detail

2. Principle and Its Limits

Images are ideally real-valued, so their k-space has conjugate symmetry: a point on one side of the centre is the complex conjugate of the point on the opposite side, and the uncollected data can be synthesised by reflection through the origin [1, 2]. A little more than half of k-space then suffices, which in the text of [1] reduces the scan time by "almost a factor of two".

The assumption fails because real images are complex. Resonance-frequency variation, flow and motion add phase [1]. Susceptibility differences at air–tissue interfaces produce local shifts of several ppm over short distances, which sets how much resolution the phase estimate needs, how much central k-space must be acquired symmetrically and how "partial" the acquisition can be [1]. For this reason scanners acquire more than half: about 5/8, 6/8 or 7/8 of the lines in practice [2, 3]. The phase is estimated from a low-resolution image built from the symmetrically sampled central strip of k-space [1, 2]; the example in [1] uses a strip of ±1/16 of k-space for a 9/16 acquisition and gives no general rule for the number of overlap lines.

The method most associated with the first practical use of conjugate-symmetric synthesis is the homodyne approach of Noll et al., whose abstract describes synchronous detection used in place of magnitude detection, with applications including the reconstruction of partial k-space images [7]. Feinberg (1986), MacFall (1988) and McGibney (1993) are cited by [2] for the origins and for phase correction; they were not read.

3. Reconstruction Methods and Where They Fail

MethodPrincipleStrength and failure stated in the sources
Zero fillingMissing k-space set to zero before the transformAcceptable near a fraction of 1; blurs in the phase direction as the fraction approaches 0.5 (example of 144 of 256 lines) [1]
Phase-corrected conjugate synthesisPhase corrected in image space, then missing data synthesisedErrors near the boundary of acquired data; weighting with homodyne-style ramps reduces them, but a 9/16 acquisition still ghosts, so 5/8 is used in the example [1]
HomodyneSingle pass: weighting of k-space (high-pass step or ramp), low-resolution phase map, phase removed, real part kept [1, 2]A step weighting ghosts at 9/16; a linear ramp from 0 to 2 removes one ghost but adds an artefact where phase changes rapidly; requires small phase gradients [1]
POCSIteration between image domain (phase constraint) and k-space (measured data)About four to five iterations in [1]; fewer artefacts than step- or ramp-weighted homodyne at 9/16; both fail near rapid local phase change [1, 2]

All methods work when image phase varies smoothly [1]. Which method a given scanner uses is (to be sourced); one educational page states that homodyne is a currently popular method [2]. Taking the magnitude after reconstruction also suppresses the noise component in quadrature with the signal when signal greatly exceeds noise [1].

4. Phase Direction: Lines, Time and SNR

4.1 Arithmetic of lines and time (Reasoning, from the fractions in [2, 3])

FractionLines not acquiredExample: 256 lines → lines acquiredFactor on SNR if SNR ∝ √(samples)
5/837.5 %1600.79
6/825 %1920.87
7/812.5 %2240.94
Exactly 1/2 (theoretical limit)50 %1280.71

The educational page [3] states that the SNR is multiplied by about 0.707 for exactly half of k-space and by about 0.87 for 6/8, before other reconstruction effects, which agrees with √(fraction); the values for 5/8 and 7/8 are computed here, not quoted. It also states a theoretical scan-time saving of up to 50 % and a practical saving of about 25 % to 40 % depending on the factor [3]. A different page states that in practice common fractions are 5/8, 6/8 or 7/8, and that EPI often uses about 6/8 to 7/8 [2]. The primary source of the √(fraction) relation and of the practical time saving was not shown on [3] (to be sourced).

4.2 In turbo spin echo (Reasoning)

In a multi-shot turbo spin echo each echo fills one line, so the number of echo trains per slice is the number of acquired lines divided by the turbo factor, rounded up (see 9561). A hypothetical case: 190 lines, partial Fourier 6/8, turbo factor 21. Acquired lines ≈ 190 × 0.75 = 142.5, rounded here to 143 (the rounding rule and the number of extra central lines used by a vendor are to be sourced); trains = ⌈143 / 21⌉ = 7 against ⌈190 / 21⌉ = 10, so the repetition count falls by about 30 %. The effect on a console scan time also depends on the fixed overhead per slice and on concatenation, which are not modelled here.

The same step can be used to shorten the echo train instead of the scan: for a given number of trains, fewer lines allow a lower turbo factor and therefore a shorter echo-train duration and less T2 blurring (Reasoning, from the train-duration relation of 9561). In single-shot sequences fewer lines shorten the single train: HASTE is described as a single-section T2-weighted sequence acquiring images in less than 1 second, with half-Fourier acquisition and reconstruction; in the technical note an echo train length of 104, an effective TE of 60 ms and a 192 × 256 matrix are given, the exact k-space fraction is not stated, and low SNR is called the major drawback [6]. Blurring in the phase direction from T2 decay along a single-shot train is described in the introduction excerpt of [10], and a HASTE-based study excerpt describes an ascending acquisition of 36 echoes (56 % partial Fourier, eight central lines used for phase correction) that starts on the undersampled side [11]. That order places the central lines early in the train (Reasoning); its consequence for effective TE is on 9561.

4.3 What goes wrong

Imperfect phase estimation can leave residual blurring, ringing or phase-related artefacts with homodyne, and POCS can leave artefacts where phase varies rapidly [2, 3]. In EPI, lines acquired late in the train accumulate different phase, which complicates the reconstruction [2]; the chapter [1] states that a partial acquisition reduces the echo time of EPI, which can be long for a fully symmetric acquisition [1]. The statement that partial Fourier reduces geometric distortion by shortening the train is plausible (Reasoning) but no source read states it (to be sourced). The sources read do not use the term "Gibbs" for the artefact of partial Fourier [1].

5. Readout Direction: Asymmetric (Partial, Fractional) Echo

In the readout direction only part of each echo is sampled, and the missing early portion is estimated with conjugate symmetry and phase correction; usually slightly more than half of the echo is acquired [4]. Vendor names given by [4]: "Partial Echo" (GE, Philips), "Half Echo" (Fujifilm), "Asymmetric Echo" (Siemens), "Matched Bandwidth" (Canon).

The purpose is different from the phase direction. The full number of phase-encoding steps is still acquired, so there is no direct reduction in total scan time [4]. The purpose is to shorten the minimum echo time, because the echo centre can move closer to the start of the readout [4]; the chapter [1] adds that the readout dephasor is reduced so the echo arrives earlier, which helps against flow-dependent dephasing and through-plane susceptibility signal loss. Applications named in [4] are gradient-echo imaging, MR angiography, EPI and short-TE spin echo.

PropertyPhase-direction partial FourierAsymmetric echo
Main useFewer lines: shorter scan or shorter echo train [3, 4]Shorter minimum TE [4]
Direct scan-time savingYes, about 25–40 % in practice according to [3]No [4]
Missing dataLines on one side of the centre in ky [3]Early part of each echo in kx [4]
SNRFalls with fewer samples [3]Missing data must be estimated, which can reduce SNR; with a fixed short TE a longer readout window may allow lower receiver bandwidth and possibly higher SNR [4]
Typical failureBlurring, ringing, residual phase artefact [2, 3]The same family; more likely when the acquired fraction is too small or phase varies rapidly [4]

5.1 How far can the echo be shortened?

One educational page gives no explicit minimum fraction [4]. A research article states that sampling asymmetrically in the readout direction requires an overlap region around the centre of k-space as a calibration region, which limits the minimum fractional echo to about 60 % before artefacts are observed, and that the image is unusable at 50 % in its first figure [5]. That article proposes a different method, two half-echo acquisitions of opposite readout polarity combined by low-rank reconstruction, and reports for a balanced steady-state free precession sequence at 3 T (192 points, 220 mm field of view, 200 Hz/Px):

Echo fractionMinimum TE (ms)Minimum echo spacing ΔTE (ms)Minimum TR (ms)
100 %3.175.796.67
66 %1.534.965.03
50 %0.753.044.25

Limits stated by the authors of [5]: two readouts lengthen the acquisition (offset in that study by 5/8 partial Fourier in the phase direction and by the shorter TR); the combined images gave no SNR gain over a single image; chemical shift artefacts are not modelled and may remain at low bandwidth; reconstruction is slow; testing used one 3 T system with a 2-channel coil. These values describe a research method and one sequence, not a clinical protocol, and the 60 % limit is the authors' statement for conventional partial-echo reconstruction.

Reasoning. Because the shortened part is the pre-echo readout, the effect on the minimum TE of a given sequence is largest where the readout window dominates the TE, as in short-TE gradient-echo imaging with narrow bandwidth, and smaller where TE is set by other elements, as in a long-TE spin echo. No source read quantifies this for clinical sequences (to be sourced). The interaction with receiver bandwidth and chemical shift is on 9597.

6. Slice Direction and Combinations

Slice direction (3D). The console shows a separate slice partial Fourier setting for 3D acquisitions (Section 7). No source read explains this option. That the principle applies along any encoded axis of a 3D k-space is a Reasoning, as is the expectation that the lines saved are slice-encoding steps; its practical limits are (to be sourced).

Phase and readout together. An educational page argues that the read and phase conjugate-symmetry techniques cannot be used together, because the symmetry is diagonal and a single quadrant (one quarter of k-space) is not enough to reconstruct the image accurately [9]. The console of one examination nevertheless offers a phase partial Fourier option and an asymmetric echo option at the same time (Section 7). Reasoning: the two statements are compatible if the acquired fractions are well above one half in each direction, so that the combined fraction stays far above one quarter; how a vendor combines them, and the limit it enforces, are to be sourced.

With parallel imaging. The console shows partial Fourier together with GRAPPA and CAIPIRINHA settings (Section 7). The two reduce different things: partial Fourier removes data that are redundant by symmetry, whereas parallel imaging replaces data using coil sensitivity, so their SNR penalties are separate factors (Reasoning). No source read quantifies the combination (to be sourced); see 9588 for the parallel-imaging side and 9564 for reference scans, CAIPIRINHA and SMS.

7. Console View (Siemens, One 3 T Head-and-Neck Examination)

Values read from screenshots [8]; labels are the console's. The series for some single-tab captures is not stated and its assignment is inferred (Reasoning).

Label on the consoleValue observedWhere seen
Phase Partial Fourier (acceleration tab, GRAPPA)6/8, with GRAPPA factor PE 2 and 32 reference linesCapture of the acceleration tab; series not stated (a TSE series is probable, Reasoning)
Phase and Slice Partial Fourier (CAIPIRINHA)Off, or 7/8 in the same tab; the 3D Dixon VIBE series also shows 7/8 for bothAcceleration tab of a VIBE-type series; t1_vibe_dixon_tra_post iso hr
Asymmetric EchoWeak (3D series with CAIPIRINHA and the Dixon VIBE); Off (a capture of a TSE sequence with turbo factor 19 and bandwidth 525 Hz/Px, probably the T2 Dixon series, Reasoning)Sequence tab

Observations (Reasoning, arithmetic only). (a) The educational page [3] gives "Half Fourier" as the Siemens term; the console labels read "Phase Partial Fourier", "Slice Partial Fourier" and "Asymmetric Echo". The page [4] also gives "Asymmetric Echo" as the Siemens term for the readout direction, which agrees. (b) Only the settings Off and Weak were captured for asymmetric echo; whether further settings exist and what fraction of the echo each corresponds to is (to be sourced). (c) For the VIBE Dixon series the console shows a first echo of 2.46 ms and a bandwidth of 780 Hz/Px; the full-echo readout time at that bandwidth, 1/780 Hz ≈ 1.28 ms, cannot by itself show how much of the echo is acquired, because the echo fraction and the readout window are not displayed together in the captures. (d) The interplay of asymmetric echo with the Dixon echo times is not described by any source read (to be sourced); the Dixon echo-time arithmetic is on 9562.

8. Technologist Decision Points (Reasoning)

  1. Know which direction you are changing. Phase partial Fourier trades SNR for lines; asymmetric echo trades reconstruction robustness for a shorter TE and does not shorten the scan [3, 4].
  2. Expect an SNR cost in proportion to the data removed (about 13 % for 6/8 according to [3]) and do not combine it with another SNR-lowering measure (parallel imaging, narrow slices, higher bandwidth) without checking the SNR of the result.
  3. Be suspicious where phase varies quickly: air–tissue interfaces, flowing blood, metal and the skull base [1]. Residual blurring or ringing at such places can come from the reconstruction rather than from motion.
  4. Use asymmetric echo when the TE is the limit (short-TE gradient echo, susceptibility- or T2*-sensitive imaging) and accept that the scan time does not change [4].
  5. Read the time shown on the console after each change. Part of the saving of the phase direction can be offset by fixed overhead and rounding (Reasoning).

9. Discrepancies, Controversies and Limits of the Evidence

  • Practical time saving. Theoretical saving of up to 50 % against a practical 25–40 % [3]; fractions of 5/8 to 7/8 imply 12.5–37.5 % fewer lines by arithmetic. The difference is not explained by any source read.
  • Typical fractions. 5/8, 6/8 and 7/8 [2, 3]; EPI often 6/8 to 7/8 [2]; 9/16 and 5/8 in the chapter example [1]; 56 % in one HASTE-type study excerpt [11].
  • Minimum echo fraction. About 60 % according to a research article [5]; no explicit minimum on an educational page [4]; the chapter gives no general rule for overlap [1].
  • Combination of directions. An educational page says phase and read symmetry cannot be combined [9]; the console offers both settings (Section 6).
  • Weight of the sources. The educational pages carry no primary reference for the SNR and scan-time figures; the original papers (Feinberg 1986, MacFall 1988, McGibney 1993, Noll 1991 in full) and a comparison of reconstruction methods on clinical images were not read.
  • Not covered: vendor algorithms and their parameters, partial Fourier in diffusion-weighted EPI and its effect on distortion, quantitative SNR measurements on clinical systems, interaction with Dixon and with compressed sensing, and the full list of console settings.

10. Evidence-Based References

No guideline, consensus statement or systematic review on partial Fourier or asymmetric echo parameters was found. Sources are ordered by type and recency; the reading level of each is stated.

A. Guidelines / Consensus / Systematic Reviews

None found.

B. Research Articles and Technical Notes

Research article
[5] Bydder M, Ali F, Ghodrati V, Hu P, Yao J, Ellingson BM. Minimizing echo and repetition times in magnetic resonance imaging using a double half-echo k-space acquisition and low-rank reconstruction. NMR Biomed. 2021;34(4). doi:10.1002/nbm.4458.
Essentially all text through a summarising tool; one 3 T system, 2-channel coil; used only for partial-echo statements.
Research article
[10] Single-shot fast spin echo with variable-flip-angle refocusing at 3 T, abdomen (PMC4684814; authors and journal not retrieved).
Abstract and introduction excerpt only; used for one statement on T2-decay blurring in single-shot sequences.
Research article
[11] HASTE-based vasoactive-stimulus study (PMC2797844; authors and journal not retrieved).
Methods excerpt only; used for one acquisition-order detail.
Technical note
[6] Semelka RC, Kelekis NL, Thomasson D, Brown MA, Laub GA. HASTE MR imaging: description of technique and preliminary results in the abdomen. J Magn Reson Imaging. 1996;6:698–699.
Full text through a summarising tool; OCR errors.

C. Educational Texts and Web Resources

Educational
[1] Chapter 2, "Partial k-Space Reconstruction" (hosted file "Recon_Pauly_read"; author and book not named in the retrieved text).
Full chapter through a summarising tool; OCR errors, equations lost.
Educational
[2] Elster AD. Partial Fourier. Questions and Answers in MRI (mriquestions.com).
Full page through a summarising tool.
Educational
[3] Elster AD. Phase conjugate symmetry. Questions and Answers in MRI (mriquestions.com).
Full page through a summarising tool; SNR and scan-time figures without primary reference.
Educational
[4] Elster AD. Read conjugate symmetry. Questions and Answers in MRI (mriquestions.com).
Full page through a summarising tool.
Educational
[9] Elster AD. Why not use both read and phase conjugate symmetry? Questions and Answers in MRI (mriquestions.com).
Search-result excerpt only.

D. Landmark Historical References

Landmark
[7] Noll DC, Nishimura DG, Macovski A. Homodyne detection in magnetic resonance imaging. IEEE Trans Med Imaging. 1991;10(2):154–163.
Abstract only.
Landmark
Feinberg DA et al., 1986; MacFall JR, Pelc NJ, Vavrek RM. Magn Reson Imaging. 1988;6:143–145; McGibney G et al., 1993.
Cited by [2]; not read; titles and journals of the first and third not verified.

E. Console Observations (Unpublished)

Observation
[8] Screenshots of the Resolution (Acceleration) and Sequence tabs of one Siemens 3 T head-and-neck examination, supplied by the site owner.
Read visually; not vendor documentation; no patient data used.

End of document — Partial Fourier and Asymmetric Echo — MRIninja v1.0 — November 2026