Dixon Water–Fat Separation: Echo Times, Outputs and Console Settings
MRIninja Knowledge Base | MRI Parameter Deep Dive Parent page: MRI Parameters — Overview and Classification (9501), Section 7.3 Builds on: Chemical Shift and Fat/Water Physics (9206); Chemical Shift — Theoretical Foundations (9596) Version 1.0 — November 2026
MRI Parameter Deep Dive
Dixon Water–Fat Separation: Echo Times, Outputs and Console Settings
How a Dixon acquisition produces water-only, fat-only, in-phase and opposed-phase images: the two-point model, the echo times and where published values differ, why more than two echoes are used, what each output is for, what the technique costs in scan time and what it does not fix, and which labels a Siemens console shows for it.
1. Scope, Boundary and Information Status
The origin of the chemical shift and the physical comparison of Dixon with STIR and spectral fat suppression are on Chemical Shift and Fat/Water Physics (Sections 6.1–6.3); the console parameters of chemical shift are on 9596; the echo-time tables for opposed-phase imaging and the India-ink artefact are on 9608; the selection of fat-suppression methods by anatomical region is in 9501 (Section 7.3) and on the protocol pages. This page adds the acquisition side of the Dixon technique: the signal model and the two-point solution, where published echo-time values differ, the step to three or more echoes, the output images and their uses, the consequences for scan time and contrast use, and the console labels. The water–fat swap of Dixon reconstructions (page 9632) and the failure of fat suppression (page 9609) are planned separate pages; they are named here only at the point where they arise. Quantification of the proton-density fat fraction is not developed here.
Information status. Eight items were used. Two narrative reviews were read as full text through a tool that returns a model-written answer about the page, not the raw text [1, 2]; the equations of one lecture were partly garbled in extraction [6]; three reviews were available as abstract or abstract excerpt only [3, 4, 5]; the 1984 original [7] was not read and is known only through the citations of [5] and [6]. No guideline, consensus statement or systematic review on the acquisition parameters was found; one review states that "there are no validated data regarding the best parameters" for quantitative Dixon sequences [1]. Statements marked (Reasoning) are derived by the editor; gaps are marked (to be sourced). Console values come from screenshots of one 3 T head-and-neck examination supplied by the site owner [8].
| Source | Type | Reading level and weight |
|---|---|---|
| [1] Lins et al., Radiol Bras 2021 (Dixon in the musculoskeletal system) | Narrative review | Full text through a summarising tool (two readings); figure images not seen; musculoskeletal focus |
| [2] Bray et al., Br J Radiol 2017 | Narrative review (fat-fraction mapping) | Full text through a summarising tool, stated complete; used for multi-echo reasoning and echo spacing |
| [3] Eggers and Börnert, JMRI 2014 | Technical review | Abstract excerpt only |
| [4] Bley et al., JMRI 2010 | Review of fat-suppression and water–fat methods | Abstract only |
| [5] Ma, JMRI 2008 | Technical review of Dixon methods | Abstract only |
| [6] Wu, UCLA lecture 2018 | Educational lecture | Full text through a summarising tool; some equations garbled; numbers marked as given by the lecture |
| [7] Dixon, Radiology 1984 | Landmark | Not read; cited as in [5] and [6] |
| [8] Console screenshots | Unpublished observations | Read visually; not vendor documentation |
2. Principle and the Two-Point Method
Water and fat protons resonate at slightly different frequencies, so the relative phase of their signals changes with the echo time; sampling at several echo times allows the two signals to be told apart [2]. The original technique of 1984 acquired two images with a modified spin-echo sequence: a conventional one with water and fat in phase, and one with the readout gradient shifted so that the two signals are 180° out of phase; the two images give a water-only and a fat-only image, and the water-only image can serve as fat suppression [5, 7].
In the notation of a lecture [6], with sW and sF the water and fat signals and ψ the local field offset, the signal at echo time TEn is [sW + sF ei2π·Δf·TEn] ei2π·ψ·TEn. For an in-phase echo s0 = sW + sF, for an opposed-phase echo s1 = sW − sF, and with ψ = 0:
sW = ½ (s0 + s1), sF = ½ (s0 − s1) [6]. In the words of a review: adding the in-phase and opposed-phase images gives a water image, subtracting the opposed-phase from the in-phase image gives a fat image [1].
Two limits of the original are stated in the sources. First, the original method discarded phase data, so on in-phase and opposed-phase magnitude images pure fat and pure water look alike, and fat-dominant tissue can be given a fat fraction near 0 %; extended two-point methods used the phase to resolve this, but they were sensitive to B0 inhomogeneity [2]. Second, if the field offset ψ is not zero, the separation suffers crosstalk and signal loss [6]. Both are the reason for the step to more echoes (Section 4).
3. Echo Times
The frequency difference between the main fat peak and water is given as 3.4 ppm, about 217.6 Hz at 1.5 T and about 435 Hz at 3.0 T, with the first opposed-phase echo at about 2.3 ms at 1.5 T and about 1.15 ms at 3.0 T, and the in-phase echo at twice the first opposed-phase echo [1]. A lecture gives the main fat peak at −3.5 ppm, quoted as −210 Hz at 1.5 T and −420 Hz at 3.0 T, with these echo times [6]:
| Field strength | In-phase TE (ms) [6] | Opposed-phase TE (ms) [6] |
|---|---|---|
| 1.5 T | 0, 4.6, 9.2, 13.8 … | 2.3, 6.9, 11.5 … |
| 3.0 T | 0, 2.3, 4.6, 6.9 … | 1.2, 3.5, 5.8 … |
The same lecture gives a practical abdominal example at 3 T with an in-phase TE of 2.6 ms and an opposed-phase TE of 1.3 ms [6], that is, not at the nominal first opposed-phase value, because the shortest achievable echo time and the echo spacing depend on the number of readout points, the bandwidth, the field of view and the gradient and slew-rate limits [6].
The nominal values follow from the frequency difference (Reasoning: opposed-phase TE = 1 / (2 Δf), in-phase TE = 1 / Δf):
| Basis | Δf (Hz) | Opposed-phase TE (ms) | In-phase TE (ms) |
|---|---|---|---|
| 217.6 Hz (1.5 T, review [1]) | 217.6 | 2.30 | 4.60 |
| 435 Hz (3.0 T, review [1]) | 435 | 1.15 | 2.30 |
| 3.4 ppm at 123.26 MHz (console of [8]) | 419 | 1.19 | 2.39 |
| 3.5 ppm at 123.26 MHz | 431 | 1.16 | 2.32 |
The published values are therefore not identical: 3.4 ppm [1] against 3.5 ppm [6], and the lecture's 210 Hz at 1.5 T corresponds to about 3.3 ppm at 63.9 MHz while its tabulated echo time of 2.3 ms corresponds to about 217 Hz (Reasoning, arithmetic). The differences are of a few per cent in echo time; the shift value used by each manufacturer is (to be sourced). The consequence for copying echo-time pairs between field strengths is on 9608.
4. Beyond Two Points
| Variant | What it adds | Limit stated in the sources |
|---|---|---|
| Two-point | Water and fat from an in-phase and an opposed-phase echo [1, 6] | Sensitive to B0 inhomogeneity; crosstalk if the field offset is not zero [1, 6] |
| Three-point | A third echo allows the field offset to be estimated together with water and fat; the usual scheme is one in-phase and two opposed-phase echoes [1, 2, 6] | Phase unwrapping or region growing can fail where signal cancels; water–fat swaps "remain difficult to prevent entirely" [2]; the lecture calls unwrapping "not solved yet" [6] |
| Multi-point with iterative estimation (IDEAL, mDIXON and equivalents) | Arbitrary echo times, several coil channels and several fat peaks; can be combined with several sequences and weightings [1]; the lecture adds multiple coils [6] | A two-species model is inaccurate because fat has several spectral components; the lecture states that peaks near water account for about 8 % of the fat signal and that a multi-peak model needs at least four measurements [2, 6] |
| Echo-count for quantification | Six echoes are recommended for fat-fraction work; simpler two- or three-point methods suffice for some applications [2, 6] | T2* decay is not accounted for by two- and three-point methods; a low flip angle (typically 5° or less) and a long TR are used against T1 bias [2] |
One review adds that the shorter the echo spacing, the lower the sensitivity to B0 inhomogeneity, but that higher spatial resolution requires longer readouts and therefore a larger echo spacing; the authors of that review use a first echo of 0.5–1.5 ms and an echo spacing of 0.8–1.5 ms [2]. Neither review nor lecture compares bipolar with monopolar readouts or gives a preference [2, 6]. The statement that "two- or three-point methods are enough for qualitative work" is supported by the lecture's "N = 2 or 3 is common for qualitative separation" [6] and by [2]; no study comparing the two for a defined diagnostic task was found (to be sourced).
5. Output Images and What Each Is For
| Output | Content | Use stated or implied by the sources |
|---|---|---|
| Water-only | Fat-suppressed image | Fat suppression performed in post-processing, described as homogeneous [1] |
| Fat-only | Water-suppressed image | Direct fat visualisation; with water-only gives a signal fat fraction [1, 6] |
| In-phase | Sum of water and fat signal | Image in which fat and water add, without fat suppression (Reasoning) |
| Opposed-phase | Difference of water and fat signal | Signal loss where fat and water share a voxel; its artefact at fat–water boundaries is on 9608; use for detecting microscopic fat is (to be sourced) here |
| Fat fraction and R2* maps | Quantitative variants only | Not developed here [1, 2] |
A single acquisition therefore yields both fat-suppressed and non-suppressed images [1]. Which of them are stored is a console choice (Section 7); storing fewer outputs reduces the number of images to archive (Reasoning).
6. Consequences for Protocols
| Property | Chemical fat saturation | STIR | Dixon |
|---|---|---|---|
| B0 heterogeneity | Prone to incomplete suppression from B0 and B1 inhomogeneity [1] | Not stated in the passages checked (to be sourced) | Multi-point variants (three or more echoes) are described as not significantly affected by field heterogeneity; the two-point version can give incomplete or incorrect separation [1]; local swaps cannot be fully excluded [2] |
| After intravenous contrast | Described as reliable for contrast-enhanced T1 imaging [1] | Unsuitable, because it also suppresses other short-T1 tissue [1] | Usually acquired as T1 after contrast [1] |
| SNR | High [1] | Multi-point Dixon is described as having better SNR than STIR [1] | Better SNR efficiency than conventional fat-suppressed sequences, because long spectral pulses are avoided [1] |
| Scan time | Not quantified in [1] | Not quantified in [1]; see 9006 | About three times the minimum time of a conventional sequence; typically 4–5 min for IDEAL, Dixon and mDIXON, 16–70 s for the quantitative variants [1] |
Further limits stated: for quantification, imaging of extremely large regions should be avoided and the technique is not recommended in regions strongly affected by B0 heterogeneity; reconstruction is longer; the technique cannot distinguish different lipid molecules; and availability on older scanners is limited [1]. The scan-time figures come from musculoskeletal reports and are not a protocol (to be sourced for the body and neck at 3 T).
7. Console View (Siemens, One 3 T Head-and-Neck Examination)
Three series of the examination show the fat–water settings side by side (values read from screenshots [8]; labels are the console's):
| Series | Fat-Water Contrast | Other values read |
|---|---|---|
| t2_tse_dixon_tra | Dixon, Fat Saturation Strong | TR 5060 ms, TE 116 ms, 50 slices of 3.5 mm, 3:30; Dixon pop-up: Original Echoes off; Water, Fat, In Phase, Opposed Phase on |
| t1_vibe_tra_dynamic | Standard | TR 2.91 ms, TE 1.15 ms, flip angle 10° |
| t1_vibe_dixon_tra_post iso hr | Dixon | TR 5.67 ms, TE 1 = 2.46 ms, Contrasts 2, flip angle 10°, 180 slices of 1.0 mm; Sequence tab: bandwidth 780 Hz/Px, 3D, optimisation In Phase, readout mode bipolar, asymmetric echo Weak (see Partial Fourier and Asymmetric Echo); Dixon pop-up: only Water on |
Observations (Reasoning, arithmetic only). (a) The transmit frequency of the same examination is 123.26 MHz, so the opposed-phase echo of the non-Dixon VIBE at 1.15 ms is close to the 1.16 ms computed for 3.5 ppm and not to the 1.19 ms for 3.4 ppm (Section 3). (b) The first echo of the Dixon VIBE, 2.46 ms, equals neither the nominal in-phase value for 3.4 ppm (2.39 ms) nor for 3.5 ppm (2.32 ms); the second echo time was not captured, and the reason, possibly a flexible echo-time choice [6], is (to be sourced). (c) The "Contrasts" field of the Dixon series is 2 for the Dixon series; the standard series value was not recorded, and 2 agrees with two echoes being acquired. (d) The meaning of "Fat Saturation Strong" together with Dixon in the T2 TSE series, of "Lines per Shot" and of the "In Phase" optimisation is (to be sourced); no source read explains these labels. (e) The post-contrast series saves only the water image; the T2 series saves four outputs, which multiplies the number of images by four for the same acquisition (Reasoning).
8. Technologist Decision Points (Reasoning)
- Choose Dixon where B0 homogeneity or large coverage defeats fat saturation, and where STIR is not allowed (after contrast). This follows the comparison in Section 6 [1].
- Do not copy an echo-time pair between field strengths. The nominal opposed-phase TE is about 2.3 ms at 1.5 T and 1.15 ms at 3 T [1] (Section 3).
- Accept that scan time will be longer than for the corresponding fat-saturated series and check the time shown before starting.
- Select the outputs on purpose: water for suppression; add in-phase and opposed-phase only when they are going to be read.
- Look for swaps. Where the field is poor (neck, near metal, large field of view) check that fat and water are not locally interchanged; swaps cannot be fully prevented [2].
9. Discrepancies, Controversies and Limits of the Evidence
- Chemical-shift value. 3.4 ppm [1] against 3.5 ppm [6], and a lecture whose Hz values (210/420) correspond to 3.3 ppm; the difference is a few per cent in echo time.
- Qualitative scope of two- and three-point methods. The sources say that two or three points are common for qualitative work and enough for some applications [2, 6]; none compares them in a diagnostic task.
- Acquisition parameters. One review states that no validated data exist for the best parameters of quantitative Dixon sequences [1]; the same applies, as far as the sources read show, to bandwidth, bipolar against monopolar readout and flip angle for qualitative sequences (to be sourced).
- Weight of the sources. The two reviews read in full are narrative; three technical reviews, including the main review of Dixon methods [5], were read as abstract only; the 1984 original was not read. A reading of [3], [4] and [5] in full text would add the algorithms (phase correction, unwrapping, multi-peak modelling) that are only named here.
- Not covered: vendor algorithms and their limits, fat-fraction accuracy by organ, the effect of contrast agents on quantitative variants, Dixon in diffusion imaging, the time-saving single-point variants, and the console series whose labels are unexplained (Section 7).
10. Evidence-Based References
No guideline, consensus statement or systematic review on Dixon acquisition parameters was found. Sources are ordered by type and recency; the reading level of each is stated.
A. Guidelines / Consensus / Systematic Reviews
None found.
B. Narrative and Technical Reviews
C. Educational Lecture
D. Landmark Historical Reference
E. Console Observations (Unpublished)
End of document — Dixon Water–Fat Separation — MRIninja v1.0 — November 2026