MRI Heart – Generic Standard Protocol

up to this point verified by human experts

MRIninja Knowledge Base | Master / General Protocol Page
Related pages: bSSFP · Phase-Contrast MRI · MRI Parameters Overview (§14.2 cardiac triggering) · Cardiac Implantable Devices in MRI · Gadolinium Contrast Agent Safety (technique and safety pages, not parents)
Scope: adult cardiac MRI (CMR), technical focus. Congenital, paediatric and fetal imaging are outside this page. Indication-specific child pages are planned (Section 2.3).
Version 1.0 — October 2026

1. Executive Summary

Cardiac magnetic resonance (CMR) is a multi-sequence examination in which every sequence is synchronised to the heartbeat and most are acquired in a breath-hold. The examination is built from a small number of modules: localisers and view planning, cine imaging for ventricular function, tissue characterisation (T2-weighted imaging, T1/T2/T2* mapping), first-pass perfusion with or without stress, late gadolinium enhancement (LGE) and, when needed, velocity-encoded flow imaging [1]. The clinical question decides which modules are used and in what order; the technique of each module is largely the same whatever the disease [1].

This page describes the generic technique: electrocardiographic (ECG) set-up and gating, breath-hold management, view planning, the acquisition parameters recommended by the Society for Cardiovascular Magnetic Resonance (SCMR) for each module, contrast dosing and timing, the conventions for contouring and reporting, and the errors that most often degrade the examination [1, 2, 3]. It deliberately leaves disease-specific decision making, diagnostic criteria and differential diagnosis to child pages. Physical principles and generic gating theory are not repeated: see bSSFP (9015), MRI Parameters Overview (9501) and Phase-Contrast MRI (9022).

Four technical features distinguish CMR from other body MRI. First, the quality of the ECG signal determines the quality of nearly every image, and the ECG is distorted inside the magnet by the flow of blood in the static field, an effect that grows with field strength [4]. Second, the examination depends on repeated, reproducible breath-holds, so patient instruction is part of the protocol [1, 5]. Third, the sequence parameters are tied to heart rate and rhythm: the LGE readout, the mapping scheme and the cine window all change with the R–R interval [1, 3]. Fourth, quantitative values (volumes, mass, T1, T2, ECV) are valid only when contouring conventions, sequence parameters and reference ranges are the same as those used to define the normal range [2, 3].

1.1 Core Strengths

Standardised building blocks: the SCMR protocol document defines a core cine protocol, LGE, mapping, perfusion and flow modules, and lists which modules are required for each clinical question [1]. This allows one technical page to serve many indications.

Tissue characterisation without contrast: native T1 and T2 mapping and T2* imaging need no gadolinium [1, 3]. T2* is the sequence listed for myocardial iron overload in the documents read [1, 3].

Quantification with stated conventions: the SCMR post-processing document fixes how volumes, mass, LGE extent, mapping regions and flow are measured, and states where no consensus exists [2].

1.2 Intrinsic Limitations of the Generic Protocol

Rhythm and heart rate: arrhythmia, tachycardia and an unreliable ECG degrade cine, LGE and mapping; the mitigations are partial (Section 5.1) [1, 3].

Devices: patients with cardiac implantable electronic devices need a dedicated pathway Cardiac Implantable Devices in MRI, and parametric mapping is stated to be unreliable in the presence of a pacemaker or implantable cardioverter-defibrillator unless special shimming or sequences are used [3].

Missing numbers: the SCMR protocol document does not state the in-plane spatial resolution for cine, the number of short-axis slices, or a numerical PSIR inversion time (Section 11) [1].

1.3 Evidence Base and Its Reading Level

The page is built on three society consensus documents: the SCMR standardised protocols of 2020 [1], the SCMR standardised image interpretation and post-processing document of 2020 [2], and the consensus recommendations on mapping of T1, T2, T2* and extracellular volume of 2017 [3]. All three were read in full text but as rendered by a summarising tool, which may have omitted or condensed passages; for [3] the final 261 characters of the text were not read. Numerical values in this page are those the tool reported and should be checked against the original before being adopted as a local protocol. Two further sources are of lower level: a 7 T feasibility study in 10 healthy volunteers, used only for the description of the magnetohydrodynamic effect and of vectorcardiographic triggering [4], and a practical guide for technologists whose publication date is not stated in the document read, used for view planning [5]. The American Heart Association segmentation statement [6] is cited as the origin of the 17-segment model; its content was not read (only the bibliographic record).

Limits to state at the outset. First, no systematic review or meta-analysis was used: the page rests on expert consensus, and the recommendations are graded as consensus, not as outcome-tested. Second, disease-specific guidelines are not summarised on this page; myocarditis, including the updated Lake Louise criteria, is covered in the child page MRI Heart for Suspected Myocarditis (10102), and the other diseases are planned child pages. Third, vendor-specific implementations and sequence names are not described. Fourth, safety content (devices, gadolinium) is delegated to the safety pages Cardiac Implantable Devices in MRI and Gadolinium Contrast Agent Safety.

2. Main Clinical Indications

This section lists which modules the SCMR document includes in the core protocol for each clinical question. It is a technical map, not a statement of appropriateness: appropriateness criteria and diagnostic algorithms were not read for this page (to be sourced) and are the task of the child pages.

2.1 Core Protocol by Clinical Question

Each protocol in [1] starts with left-ventricular (LV) structure and function by cine imaging, followed by the items below. The source presents these as lists in the text, not as a single table; the table below is a rearrangement of those lists.

Clinical questionModules added to LV structure and functionSource
Acute myocardial infarction / acute coronary syndromeTissue characterisation (optional), rest first-pass perfusion (optional), early gadolinium enhancement (optional), LGE[1]
Chronic ischaemia and viabilityTissue characterisation (optional), low-dose dobutamine or vasodilator / high-dose dobutamine (optional), LGE[1]
Hypertrophic cardiomyopathyTissue characterisation, LVOT flow with 3-chamber cine and velocity mapping if obstructed, LGE; vasodilator stress optional[1]
Hypertensive, dilated, non-compaction, cancer-related, drug-induced, post-transplantTissue characterisation (mostly optional), optional stress perfusion, LGE; T1/T2 mapping listed for post-transplant and cancer-related[1]
Arrhythmogenic cardiomyopathy (right ventricle)Transaxial or oblique bSSFP covering the right ventricle and RVOT, optional black-blood or fat-suppressed sequences, LGE with optional RV nulling[1]
Siderotic (iron overload)T2* mapping, optional stress, optional LGE[1]
Restrictive disease, sarcoidosis, myocarditisTissue characterisation and LGE; early gadolinium enhancement may be added in myocarditis; real-time mid-LV short-axis cine may be added in restrictive disease[1]
Valvular diseaseValve-specific velocity-encoded imaging planes[1]
Pericardial diseaseReal-time imaging with LGE of the pericardium[1]
Masses and thrombiT1- and T2-weighted fast spin echo, first-pass perfusion, post-contrast fat-suppressed T1, LGE with the inversion time set to null thrombus[1]
Angiography (peripheral, thoracic aorta, coronary, pulmonary veins)Dedicated MRA protocols with volume, timing and resolution details given in the source[1]

2.2 Red Flags and Escalation

The SCMR document states that vasodilator and dobutamine stress require monitoring of blood pressure, at least a single-lead ECG, an intercom, and pulse oximetry for patients with devices, with a resuscitation policy, a crash cart outside the scanner room and immediately accessible drugs (beta-blocker, nitroglycerin, aminophylline, bronchodilators, oxygen) [1]. A stress examination therefore requires a physician-led pathway defined by the department. Escalation criteria for the non-stress examination (for example haemodynamic instability on the table) were not found in the sources read (to be sourced).

2.3 Planned Child Pages

Published: MRI Heart for Suspected Myocarditis (10102); MRI Heart for Ischaemic Heart Disease (10104) (stress perfusion, viability, acute infarction); MRI Heart for Non-Ischaemic Cardiomyopathies (10105); MRI Heart for Pericardial Disease (10107); MRI Heart for Valvular Disease and Shunts (10110); MRI Heart for Cardiac Masses and Thrombi (10111); MRI Heart for Cardio-Oncology (10112); MRI Heart for the Right Ventricle and Pulmonary Hypertension (10113). All planned cardiac child pages are now published. Published deep dives: Late Gadolinium Enhancement: Technical Deep Dive (10103); T1 and T2 Mapping and ECV: Technical Deep Dive (10106); Stress Perfusion in Cardiac MRI: Technical Deep Dive (10108); Cine and Real-Time Cardiac MRI: Technical Deep Dive (10109). All planned technical deep dives of this cluster are now published.

3. Preparation Reference

General preparation, safety screening and consent are on the universal preparation page (Universal MRI Preparation) and are not repeated. The table lists the cardiac-specific items.

3.1 Preparation Table

ItemStatementSource
ECG leadsPlace the leads securely to limit motion artefact and allow accurate gating; use contact gel if needed and check the signal strength before starting[5]
Gating hardwareECG gating is required; vector-cardiographic (VCG) gating is preferred; prospective, retrospective and triggered gating should be supported; peripheral pulse gating may work for retrospectively gated cine when the ECG is unreliable, for example at 3 T[1]
CoilCardiac-specific multi-element surface coil, typically at least 8 elements, highly recommended and required for parallel imaging[1]
Breath-hold instructionExplain the breath-hold instructions and the scan duration before starting; expiration gives more consistent positioning, inspiration is more comfortable for some patients[1, 5]
Power injectorA compatible power injector is required for perfusion and contrast-enhanced angiography[1]
Intravenous accessAdenosine: two lines preferred (one for gadolinium, one for adenosine, one in each arm); regadenoson: one line. For stress perfusion place the blood-pressure cuff on the arm connected to the power injector[1, 5]
Caffeine, theophylline, dipyridamoleAvoid for 12 to 24 h before vasodilator stress; for dobutamine avoid beta-blockers and nitrates for 12 to 24 h. Caffeine and nicotine effects may be overcome with higher adenosine or regadenoson doses[1]
FastingNot mandatory, but often advised because of the risk of nausea and vomiting during stress[1]
Haematocrit (for ECV)Measured immediately before the scan if possible, otherwise within 24 h; if not available, synthetic ECV can be used when the relationship between native blood T1 and haematocrit is known for the sequence and field strength[3, 2]
Contraindications to gadoliniumSevere renal dysfunction, pregnancy and known hypersensitivity are to be checked before injection; details on the safety page[5]; Gadolinium Contrast Agent Safety
Implanted cardiac devicesDedicated pathway; see the safety pageCardiac Implantable Devices in MRI

3.2 ECG Set-Up and the Magnetohydrodynamic Effect

Blood is a conductive fluid moving through the static field; it induces a voltage perpendicular to both the field and the direction of flow, which is superimposed on the ECG. The ST segment and the T wave are most affected, which can lead to wrong QRS detection and mistimed triggering; the effect grows with field strength and is greatest in systole, when aortic flow is highest, and the time-varying gradient fields add further distortion [4]. These statements come from the introduction and discussion of a 7 T study in 10 healthy volunteers [4]; the principle is the one that makes ECG gating less reliable at 3 T than at 1.5 T, and the SCMR document states that peripheral pulse gating may work for retrospectively gated cine when the ECG is unreliable, such as at 3 T [1].

In that 7 T study, vector-cardiographic triggering, with a learning phase performed outside the magnet, succeeded in all 10 volunteers; the sensitivity and specificity across the R-waves were 97.6% and 98.4% in the results section (the abstract reports different figures, 98.7% specificity on a different number of R-waves, an internal discrepancy of the source) [4]. Patients with arrhythmia were not studied, so these figures do not apply to them [4]. A pulse sensor was used as a backup trigger [4].

Reasoning: no source read gives a step-by-step electrode layout that suppresses the magnetohydrodynamic voltage at 1.5 T or 3 T (to be sourced). A published study of electrode placement for suppression of the effect exists in the literature found in the search but could not be opened and was not read; the practical advice remains to follow the vendor's lead placement, to check the trace after the patient is on the table and in the bore, and to reposition the leads when the T wave is taller than the R wave or the trigger indicator is irregular.

3.3 Breath-Hold Management

Cine imaging is acquired in breath-hold; expiration gives a more consistent position from one acquisition to the next, while inspiration is more comfortable for some patients [1]. Localisers are acquired as free-breathing low-resolution series in the transaxial, sagittal and coronal planes [5]. The number of breath-holds needed for mapping is stated for several diseases (from about 6 for diffuse fibrosis or iron overload to 16–25 for regional disease such as myocarditis or acute infarction), and motion correction is not a substitute for breath-holding in sequences without a navigator: only in-plane motion is corrected [3]. Reasoning: because the whole examination can require several tens of breath-holds, the order of the sequences should place the diagnostic modules before the optional ones, so that a tired patient does not compromise the essential images.

4. Standard Protocol Design

4.1 Modules and Typical Order

StepModuleNotesSource
1LocalisersFree-breathing, low resolution, transaxial, sagittal and coronal[5]; [1]
2View planningPseudo two- and four-chamber views, then true short axis (Section 4.2)[5]
3Cine (LV structure and function)bSSFP, breath-hold; short-axis stack and long-axis views[1]
4Tissue characterisation before contrastT2-weighted imaging, native T1 and T2 mapping; T2* acquired before contrast[1]
5Stress or rest perfusion (when indicated)Vasodilator stress, gadolinium during first pass[1]
6Post-contrast T1 for ECV (when indicated)10 to 30 min after the bolus[1, 3]
7LGEAt least 10 min after contrast, shorter after low doses; about 5 min after a second stress or rest injection[1]
8Flow and angiography (when indicated)Perpendicular to the vessel or valve[1, 5]

The order above reflects the timing constraints stated in the sources (T2* before contrast, post-contrast T1 10 to 30 min, LGE at least 10 min after contrast); the exact sequence in a department depends on whether stress perfusion is performed and is a matter of local protocol [1, 3]. The SCMR document states that rest perfusion adds little and should be omitted when possible, and that cine can be acquired during the 10-minute wait between stress and rest perfusion [1].

4.2 View Planning

TargetViews and planningSource
Left ventriclePseudo two-chamber and pseudo four-chamber views, then a true short-axis stack covering the LV; true four-, two- and three-chamber views planned orthogonally to one another. Simplified planning can use the pseudo two- and four-chamber views[5]
Short-axis stackBase to apex; the two-chamber and four-chamber long-axis views pass through the mitral valve and apex; the three-chamber view is aligned with the LVOT[1]
Right ventricleTransaxial stack from the diaphragm to the pulmonary bifurcation, sagittal and coronal RVOT views, and an RV inflow-outflow view planned with three points[5]
Pulmonary arteriesLeft and right pulmonary artery views, angulation adjusted for proximal or distal segments[5]
Aortic valve and aortaCoronal LVOT / aortic valve view, in-plane aortic valve view (4 to 5 slices for an en face view), aortic arch view planned with three points; axial images always reviewed for the aorta[5]
FlowAortic and main pulmonary artery flow planes perpendicular to the vessel, plus left and right pulmonary artery planes; planned from coronal and axial localisers[5]

Reasoning: the short-axis plane is perpendicular to the long axis of the LV, the line from the centre of the mitral valve to the apex; the sources read state the views to acquire but not the landmark logic, which is anatomical and is given here as an aid, not as a recommendation of the sources. Automatic view-planning tools exist from vendors; no source read evaluates their reliability (to be sourced).

4.3 Cine Imaging

bSSFP is the method of choice; at 3 T spoiled gradient echo can be considered because bSSFP is more prone to banding [1]. The short-axis stack is acquired from base to apex with a slice thickness of 6–8 mm with or without inter-slice gaps of 2–4 mm (total 10 mm) and a temporal resolution of 45 ms or better; parallel imaging or compressed sensing shortens the breath-hold [1]. Long-axis views are two-chamber and four-chamber (orthogonal, through the mitral valve and apex) and a three-chamber view aligned with the LVOT; more than three long-axis views is optional [1]. Real-time cine is optional, with a preferred temporal resolution of 60 ms or better, and absolute volumes are less accurate [1]. The sequence physics are on the bSSFP page (9015).

4.4 Tissue Characterisation Before Contrast

T2-weighted imaging: black-blood short-tau inversion recovery (STIR), or bright-blood T2-prepared bSSFP or a fast spin echo–bSSFP hybrid [1]. T2 mapping: T2-prepared bSSFP, gradient- and spin-echo (GraSE) or fast spin echo; short-axis maps are always recommended; slice thickness 6–8 mm, in-plane resolution about 1.6–2.0 mm [1]. Native T1 mapping: MOLLI, ShMOLLI or equivalent, diastolic acquisition when the rhythm is regular, systolic in atrial fibrillation; at least one short-axis map is required [1, 3]. T2*: acquired before contrast, single breath-hold multi-echo gradient echo, single mid-ventricular short-axis slice (Section 10.4) [1, 3]. Slice position and number for mapping depend on whether the disease is diffuse, patchy or focal (Section 10.3) [3].

4.5 First-Pass and Stress Perfusion

First-pass perfusion uses saturation-recovery bSSFP, gradient echo or gradient echo–EPI, with at least three short-axis slices per heartbeat; ischaemia evaluation should acquire every heartbeat if possible [1]. The vasodilator agents and their doses are in Section 10.5. Adenosine can be started outside the bore, with gadolinium injected during the last minute of the infusion; with regadenoson, gadolinium is injected about 45 to 60 s after the bolus [1]. Dobutamine stress for function starts at 10 µg/kg/min and increases by 10 µg/kg/min every 3 min up to a target heart rate of 85% of (220 − age), with atropine in 0.5 mg increments if needed and a typical maximum of 40 µg/kg/min; it is stopped for a new wall-motion abnormality, a serious side effect or the target heart rate [1]. A dedicated page on stress perfusion is planned.

4.6 Late Gadolinium Enhancement

LGE uses a 2D segmented inversion-recovery gradient echo or bSSFP sequence, phase-sensitive inversion recovery (PSIR), or a 3D sequence if breath-holding and signal allow; a single-shot bSSFP readout is an optional back-up [1]. The views are the same as for cine, with the same slice thickness [1]. The inversion time (TI) is set to null normal myocardium; a TI scout helps but may differ by up to 50 ms from the segmented sequence, and PSIR avoids the need for precise TI setting [1]. A slightly long TI is preferred on magnitude images, because an "etched" appearance of normal myocardium indicates a TI that is too short [2]. Enhancement should be confirmed in an orthogonal plane or by re-acquiring with a changed readout direction [2]. The readout frequency depends on heart rate (Section 5.1) [1].

4.7 Velocity-Encoded Flow

The physics and the choice of velocity encoding are on the phase-contrast page (Phase-Contrast MRI); only the cardiac-specific items are given here. The planes are perpendicular to the vessel or valve (Section 4.2) [5]. One practical guide recommends at least 30 phases per cardiac cycle and encoding velocities of 150–200 cm/s for the aorta and 100–150 cm/s for the pulmonary arteries to avoid aliasing [5]; the SCMR post-processing document states that 8 to 16 pixels should span the vessel and that at least 11 to 16 frames per cardiac cycle are needed, and that a velocity encoding that is too high gives poor jet visualisation and signal, while one that is too low gives a mosaic pattern [2]. The two sources therefore differ on the required temporal sampling (Section 11).

5. Optimisation Strategy

5.1 Heart Rate and Rhythm

ModuleAdjustment stated in the sourcesSource
LGEReadout usually every other heartbeat; every heartbeat if the heart rate is below 60 bpm; every third heartbeat if above 100 bpm or in arrhythmia. Acquisition window under 200 ms per R–R interval, less in tachycardia[1]
T1 / T2 mappingAcquire in diastole if the rhythm is regular; use sequences designed for higher heart rates in tachycardia; in atrial fibrillation repeat acquisitions and average; systolic readout is robust in tachyarrhythmia but gives lower values and needs its own normal ranges; acquisition windows must be short enough to avoid blurring at the patient's heart rate[3]
Mapping with pacemaker or ICDUnreliable unless special shimming or sequences are used[3]
CinePeripheral pulse gating may work for retrospectively gated cine when the ECG is unreliable; real-time cine is an optional acquisition with a preferred temporal resolution of 60 ms or better. Reasoning: real-time cine is the usual fallback when segmented acquisition fails because of rhythm or breath-holding; the sources read do not state this use explicitly (to be sourced)[1]

Generic definitions of prospective and retrospective gating, trigger delay and trigger window are on the parameters page (MRI Parameters Overview, Section 14.2) and are not repeated.

5.2 Field Strength and Shimming

A field strength of at least 1.0 T is required and most examinations use 1.5 T; at 3 T bSSFP is more prone to banding and shimming is important [1]. Volume-selective B0 shimming on the heart is highly recommended at 1.5 T and essential at 3 T, and B1 shimming is recommended at 3 T [3]. Native T1 values are not directly comparable between 1.5 T and 3 T, and T2* is currently recommended at 1.5 T [3]. Manual shimming should be considered, especially at 3 T [1].

5.3 Artefacts and Countermeasures

ProblemCauseCountermeasureSource
Mis-triggering, wrong cardiac phaseT wave or magnetohydrodynamic voltage read as R wave; poor electrode contactSecure leads and check the signal [5]; VCG gating preferred [1]; peripheral pulse as back-up [1]; repositioning the leads is Reasoning[1, 4, 5]
Banding in cineOff-resonance in bSSFP, worse at 3 TShimming; spoiled gradient echo can be considered at 3 T[1]
"Etched" normal myocardium in LGETI too shortUse a slightly longer TI or PSIR; confirm in an orthogonal plane[2]
Skipped beats or motion in mappingArrhythmia, breath-hold failureCheck raw images, error maps and motion-corrected source images immediately; repeat if suboptimal[3]
Mapping region of interest too smallPartial-volume and noiseAvoid regions under 20 pixels[2, 3]
Aliasing or mosaic pattern in flowVelocity encoding too lowRaise the encoding velocity and re-acquire (Reasoning, implied by the cause)[2, 5]

General artefact physics is on Physics of MRI Artefacts.

6. Contrast Use: Principles Specific to Cardiac MRI

Safety, renal function, retention and hypersensitivity are on Gadolinium Contrast Agent Safety and the general principles on Contrast Media in MRI. The dosing in the SCMR document is expressed per module and the document states that the lowest dose giving adequate image quality should be used [1].

IndicationDose (mmol/kg)Injection rateSaline chaser
Perfusion0.05–0.13–7 ml/s30 ml at 3–7 ml/s
LGE0.1–0.2Not given20 ml
Angiography (carotid, renal, aorta)0.1–0.22–3 ml/s20 ml at 2–3 ml/s
Time-resolved angiography0.053–5 ml/s30 ml at 3–5 ml/s
Peripheral angiography0.2First 10 ml at 1.5 ml/s, then 0.4–0.8 ml/s20 ml at 0.4–0.8 ml/s

Source of the table: [1]. Injection rates halve for 1 mmol/ml agents compared with 0.5 mmol/ml agents; the total gadolinium dose for a stress and rest study is 0.1–0.2 mmol/kg, with extra given for LGE as needed [1]. For ECV the consensus recommends an extracellular, non-protein-bound gadolinium agent at 0.1–0.2 mmol/kg and a post-contrast T1 acquisition 10 to 30 min after administration (10 min is now regarded as sufficient); in split-dose protocols the timing is taken from the last dose [3]. Native T1 and T2 can be measured without contrast, at least 24 h after the last dose if renal function is normal [3]. LGE should wait at least 10 min after the contrast, possibly less after lower doses, and about 5 min after a second stress or rest injection [1]. The dose table applies to the agents and concentrations for which it was written; the SCMR document notes that dose depends on agent and protocol [1].

7. Reporting Essentials

Disease-specific reporting items and diagnostic criteria belong to the child pages. The items below are the technical conventions that apply to every report and that the SCMR document fixes [2].

7.1 Items to Report

ItemConventionSource
Papillary muscles and trabeculaeMyocardial tissue, ideally in LV mass; clinical practice often includes them in the blood-pool volume, which the document calls acceptable; the reference ranges must match the chosen approach[2]
End-diastole and end-systoleEnd-diastole is the frame with the largest LV blood volume and end-systole the frame with the smallest; one phase is identified across all short-axis slices[2]
Basal sliceEither track wall thickening or use a slice with at least 50% of its blood volume surrounded by myocardium; the document states that there is no expert consensus on which method to use[2]
Epicardial borderIn the LV volume analysis drawn at the middle of the chemical shift artefact line when present[2]
LV massMyocardial density 1.05 g/ml; for a basal slice with only a myocardial crescent or an apical slice with only a ring, an epicardial-only contour is drawn[2]
LGE location and extentReported on the American Heart Association 17-segment model; visual transmural extent per segment in five bands: 0%, 1–25%, 26–50%, 51–75%, 76–100%[2]; model origin [6]
LGE quantification (research)No single method endorsed; options are manual planimetry, n-SD and full width at half maximum; "5-SD for infarction" suggested as a starting point; no cut-off given for non-ischaemic LGE[2]
No-reflow in acute infarctionSubendocardial and mid-myocardial hypo-enhanced zones count toward infarct size[2]
T1 / T2 mapping and ECVReport the absolute value, the Z-score if available and the local normal range (mean ± 2 SD); septal region of interest on mid-cavity short-axis maps for diffuse disease; blood ROI in the centre of the blood pool excluding papillary muscles and trabeculae; same-day haematocrit or synthetic ECV[2, 3]
FlowNet volume = antegrade minus retrograde; regurgitant fraction = retrograde / antegrade × 100[2]

7.2 Structured Reporting

The sources read describe how to measure, not a report template. A structured reporting template for CMR was not found in the sources read (to be sourced). Reasoning: a report should state the field strength, the sequence set used, the reference range applied for each quantitative value, and any technical limitation (rhythm, breath-hold quality, device), because the interpretation of mapping and volumes depends on them [2, 3].

8. MRI Technologist Pearls

This section is partly Reasoning: statements with a citation are supported by a source read; the others are practical inferences from the supported statements and are marked.

8.1 Before the Examination

Instruct the patient about breath-holds and scan duration before the leads are applied [1, 5]. Secure the leads and check the ECG signal strength before moving the patient into the bore [5]. For stress examinations check 24 h of caffeine abstinence and the other exclusion criteria, and place the blood-pressure cuff on the arm connected to the power injector [5]. Obtain a haematocrit on the day if ECV is planned [3]. Reasoning: ask the patient to practise the breath-hold on the table and to say the length they can sustain, because the sequence acquisition windows are fixed by the heart rate and not by the patient's comfort.

8.2 During the Examination

Check the ECG trace after the patient is inside the magnet: the trace seen outside the bore can differ [4]. Review each series for mis-triggered or skipped beats, motion and off-resonance; review raw images, error maps and parametric maps immediately and repeat suboptimal series [3]. Keep slice prescription and cardiac phase identical for native and post-contrast T1 mapping [3]. When using a TI scout, verify the result on the first LGE images, because the scout can differ from the segmented sequence by up to 50 ms [1].

8.3 Common Errors (Reasoning)

Reasoning: (1) accepting a trigger indicator without looking at the trace, so that the T wave triggers acquisition; (2) changing the slice prescription between native and post-contrast mapping, which invalidates ECV; (3) using one reference range for results acquired with different sequences, field strengths or contrast agents, which the consensus excludes [3]; (4) comparing native T1 values between 1.5 T and 3 T [3]; (5) starting LGE at a fixed time without regard to the dose and a second injection [1]; (6) a velocity encoding chosen without a quick look at the expected peak velocity [2].

9. Quality Control Checklist

CheckBasis
ECG signal checked inside the bore; gating method recorded[1, 4, 5]
Cardiac coil with at least 8 elements; parallel imaging active[1]
Short-axis stack covers base to apex; two-, three- and four-chamber views acquired[1]
Cine slice thickness 6–8 mm (gap 2–4 mm as the department chooses), temporal resolution ≤45 ms[1]
LGE: waiting time recorded, TI or PSIR documented, enhancement confirmed in a second plane[1, 2]
Mapping: same slice prescription and cardiac phase for native and post-contrast; ROI ≥ 20 pixels; local reference range available[2, 3]
ECV: haematocrit within 24 h (or synthetic ECV), post-contrast T1 at 10–30 min[3]
Mapping phantom quality control every 3 months and after hardware or software changes[3]
Stress: monitoring, resuscitation equipment and drugs available; caffeine abstinence checked[1, 5]
Contrast dose and agent recorded; same technique at follow-up[1]; Reasoning

10. Advanced Technical Parameters

This section is intended for MRI technologists and protocol optimisation specialists. The values are those the SCMR documents give, as rendered by a summarising tool, and must be verified against the originals before use. Where a document gives no value, the table states so. They are not vendor settings.

10.1 Cine

ParameterValue statedSource
SequencebSSFP; spoiled gradient echo can be considered at 3 T[1]
Slice thickness6–8 mm, with or without gaps of 2–4 mm (total 10 mm)[1]
Temporal resolution≤45 ms (segmented); ≤60 ms preferred for real-time[1]
In-plane spatial resolutionNot stated (to be sourced)[1]
Number of short-axis slicesNot stated; a 9-slice stack is shown as an example[1]
Breath-holdExpiration for reproducibility; inspiration if more comfortable[1]

10.2 LGE

ParameterValue statedSource
Waiting time after contrastAt least 10 min (shorter with lower doses); about 5 min after a second stress or rest injection[1]
In-plane resolutionAbout 1.4–1.8 mm[1]
Acquisition windowUnder 200 ms per R–R interval; less in tachycardia[1]
ReadoutEvery other heartbeat; every heartbeat if HR below 60 bpm; every third heartbeat if HR above 100 bpm or arrhythmia[1]
TITo null normal myocardium; TI scout may differ by up to 50 ms from the segmented sequence[1]
Long-TI variant (no-reflow, thrombus)About 550 ms at 1.5 T and 850 ms at 3 T as rendered by the tool[1]
Short-TI variant with PSIR (subendocardial scar)About 200 ms as rendered by the tool[1]
Dark-blood LGEOptional[1]

10.3 T1 and T2 Mapping and ECV

ParameterValue statedSource
Native T1MOLLI, ShMOLLI (and the 5(3)3 MOLLI variant) or equivalent; saturation-based schemes (SASHA, SAP-T1, AIR) and SAPPHIRE as alternatives; slice 6–8 mm, in-plane about 1.6–2.0 mm[1, 3]
Trade-off of schemesMOLLI: excellent precision but sensitive to magnetisation transfer, T2, flip angle, inversion efficiency, heart rate and off-resonance; saturation schemes avoid most of these at the cost of precision[3]
T2 mappingT2-prepared bSSFP or gradient-echo with at least 3 source images; two-parameter fit; example echo times 0, 25 and 50 ms; slice 6–8 mm, in-plane about 1.6–2.0 mm[1, 3]
Slices: diffuse diseaseBasal and mid short-axis, optional single long-axis map[3]
Slices: patchy diseaseAt least one long-axis map plus basal and mid short-axis; four-chamber for amyloid, three-chamber for Anderson-Fabry[3]
Slices: focal or acute diseaseShort-axis maps covering the most abnormal area and an area of apparent minimal abnormality, or the whole LV if cine and T2-weighted data are unavailable[3]
Breath-holds (T1 mapping)About 7 for amyloid (repeat post-contrast); 4–7 Anderson-Fabry; 6 iron overload (native only); 6 diffuse fibrosis; 16–25 myocarditis, acute infarction or regional disease[3]
ECV formulaECV = [(1/T1 myocardium post − 1/T1 myocardium native) / (1/T1 blood post − 1/T1 blood native)] × (100 − haematocrit), expressed as a percentage[3]
Contrast and timing for ECVExtracellular non-protein-bound agent at 0.1–0.2 mmol/kg; post-contrast T1 10–30 min after administration[3]
TemperatureT1 rises by about 1% per 1 °C of body temperature; usually negligible in patients[3]
Reference rangesLocal ranges (mean ± 2 SD) from data acquired and analysed in the same way; if no local native T1 or T2 range exists, quantitative results should not be reported clinically; 15–20 healthy subjects may suffice for large-magnitude changes, 50 or more with sex and age adjustment for diffuse fibrosis[3]

10.4 T2*

ParameterValue statedSource
SequenceSingle breath-hold multi-echo gradient echo, acquired before contrast; delay 0 ms after the R wave[1]
Echoes6–9 echoes from about 2 ms to about 18 ms, spaced about 2 ms [1]; 8 equally spaced echoes from 2 to 18 ms [3]. For severe iron: 6–9 echoes from about 1 ms to about 12 ms, spaced about 1 ms [1][1, 3]
SliceSingle mid-ventricular short-axis, thickness 8–10 mm, in-plane about 1.6–3.0 mm; septal ROI[1, 3]
Field strength1.5 T recommended[3]
Black-blood approachRecommended if available (less bias and observer variability)[3]
Risk tiers at 1.5 T with at least 8-point gradient echoLow risk >20 ms; intermediate 10–20 ms; high risk <10 ms[3]
Liver T2* (optional)Non-gated axial through the mid-liver; fat saturation and the shortest available TE (consensus)[1, 3]

10.5 Perfusion and Stress

ParameterValue statedSource
Sequence and coverageSaturation recovery with bSSFP, gradient echo or gradient echo–EPI; at least 3 short-axis slices per heartbeat[1]
Slice and resolutionSlice 8–10 mm; in-plane under 3 mm; readout about 100–125 ms or shorter[1]
Acquisition lengthAt least 50–60 heartbeats (more in low output); breath-hold starts before contrast reaches the LV; free-breathing with motion correction optional[1]
Adenosine140 µg/kg/min for 2–4 min; consider up to 210 µg/kg/min if HR rises <10 bpm or systolic pressure drops <10 mmHg after 2–3 min[1]
ATP140 µg/kg/min for 3–5 min, same escalation option[1]
Regadenoson0.4 mg single injection; gadolinium about 45–60 s later[1]
Dipyridamole0.142 µg/kg/min over 4 min[1]
Rest perfusionAt least 10 min after stress; adds little and should be omitted when possible[1]
Dobutamine (function)Start 10 µg/kg/min, increase by 10 µg/kg/min every 3 min, target 85% of (220 − age), atropine 0.5 mg increments, typical maximum 40 µg/kg/min[1]
Dobutamine (viability)2.5–10 µg/kg/min for 5–10 min[1]
ReversalMany sites give aminophylline 100 mg IV after stress images; the source cautions that it is arrhythmogenic[1]

Quantitative perfusion (optional) uses a dual-bolus or dual-sequence approach with proton-density images before contrast [1]. Drug doses are quoted from a consensus document for information; the use of any stress agent follows the departmental protocol and the supervising physician.

10.6 Flow

ParameterValue statedSource
Pixels across the vessel8–16[2]
Frames per cardiac cycleAt least 11–16 [2]; at least 30 phases [5] (the two sources differ)[2, 5]
Velocity encoding150–200 cm/s aorta; 100–150 cm/s pulmonary arteries (practical guide)[5]
Peak velocitySingle-pixel peaks are noisy; averaging adjacent pixels may underestimate the true peak[2]
Baseline correctionNo recommendation in the document read[2]

11. Evidence Gaps and Ongoing Debate

Missing numerical values: the SCMR protocol document, as rendered by the tool, does not state the in-plane resolution of cine, the number of short-axis slices, a numerical PSIR inversion time, or a consolidated table of core protocols per indication [1]. The post-processing document endorses no single LGE quantification method and no cut-off for non-ischaemic LGE, no T2-mapping oedema cut-off, no apex-specific rule for the 17-segment model in LGE, and no basal-slice method for LV volumes [2].

Discrepancy on flow sampling: at least 11–16 frames per cardiac cycle [2] against at least 30 phases [5]; the second source is a practical guide of undated publication and is of lower level.

Gating at 3 T and above: the only source read on the magnetohydrodynamic effect is a 7 T feasibility study in 10 healthy volunteers without arrhythmia, whose abstract and results differ on the number of R-waves and on specificity [4]. Evidence on gating performance at 3 T and in arrhythmia is missing from the sources read (to be sourced).

Tool-rendered reading: all three consensus documents were read through a summarising tool; omissions and condensations are possible. No systematic review, randomised trial or outcome study was read for this page.

Not read: ESC guidelines on cardiomyopathies, the SCMR consensus on stress CMR, CMR in patients with devices beyond the safety page, and vendor application guides. Their content is deliberately not paraphrased here. The ESC 2025 guideline on myocarditis and pericarditis and the updated Lake Louise criteria were read, in part, for the child page MRI Heart for Suspected Myocarditis (10102).

12. Evidence-Based References

References are ordered by level of evidence and then by recency. The reading level of each source is stated in its line; a source read only through a summarising tool or at record level supports only what the tool reported.

A. Guidelines / Consensus / Society Recommendations

High
[1] Kramer CM, Barkhausen J, Bucciarelli-Ducci C, Flamm SD, Kim RJ, Nagel E. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson. 2020;22:17. DOI: 10.1186/s12968-020-00607-1.
(High — Society consensus document; full text read as rendered by a summarising tool; numerical values to be checked against the original) Cited throughout.
High
[2] Schulz-Menger J, Bluemke DA, Bremerich J, Flamm SD, Fogel MA, Friedrich MG, Kim RJ, von Knobelsdorff-Brenkenhoff F, Kramer CM, Pennell DJ, Plein S, Nagel E. Standardized image interpretation and post-processing in cardiovascular magnetic resonance – 2020 update. J Cardiovasc Magn Reson. 2020;22:19. DOI: 10.1186/s12968-020-00610-6.
(High — Society consensus document of the Task Force on Standardized Post-Processing; full text read as rendered by a summarising tool) Cited in Sections 1, 3, 4, 5, 7, 8, 9, 10, 11.
High
[3] Messroghli DR, Moon JC, Ferreira VM, Grosse-Wortmann L, He T, Kellman P, Mascherbauer J, Nezafat R, Salerno M, Schelbert EB, Taylor AJ, Thompson R, Ugander M, van Heeswijk RB, Friedrich MG. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: A consensus statement. J Cardiovasc Magn Reson. 2017;19:75. DOI: 10.1186/s12968-017-0389-8.
(High — Consensus statement; full text read as rendered by a summarising tool, the final 261 characters not read) Cited in Sections 1, 3, 4, 5, 6, 7, 8, 9, 10, 11.

B. Systematic Reviews / Meta-analyses

No systematic review or meta-analysis was read for this page (to be sourced).

C. Important Prospective / Original Studies

No prospective or outcome study was read for this page (to be sourced).

D. Technical MRI Papers

Technical
[4] Hamilton-Craig C, Stäb D, Al Najjar A, O'Brien K, Crawford W, Fletcher S, Barth M, Galloway G. 7-Tesla functional cardiovascular MR using vectorcardiographic triggering – overcoming the magnetohydrodynamic effect. Tomography. 2021 (published August 2021). PMC8396263. DOI not verified.
(Technical — Feasibility study in 10 healthy volunteers at 7 T; read as rendered by a summarising tool; journal name inferred by the tool; abstract and results differ on the number of R-waves) Cited in Sections 1, 3, 5, 8, 9, 11; used for the mechanism of the effect, not for 1.5 T or 3 T performance.
Technical
[5] Venkateshvaran A. Optimized image planning in CMR: a practical guide for technologists. Educational article hosted at dam-assets.escardio.org (European Society of Cardiology); publication date and affiliation not stated in the document read.
(Technical — Practitioner guide; read as rendered by a summarising tool; cites EACVI and SCMR standards and references from 2017 and 2020) Cited in Sections 1, 3, 4, 8, 9, 10; publication details to be verified.

E. Landmark Historical References

Expert
[6] Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. J Nucl Cardiol. 2002;9(2):240–245. DOI: 10.1067/mnc.2002.123122. PMID: 11986572.
(Expert — Society statement; bibliographic record only, content not read) Cited in Section 7 as the origin of the 17-segment model, which is applied in [2].

Child Protocols

Clinical pages derived from this master protocol. These pages document what changes for specific indications.

Recent PubMed search for this protocol

Last updated: October 2026

Report an error on this page

The report stores the page address, the text above and, if you give it, your e-mail. It does not store your IP address, browser type or any cookie. See the Disclaimer, section 7.