MRI Heart for Valvular Disease and Shunts

up to this point verified by human experts

MRIninja Knowledge Base | Child Page — Pathology-Specific Protocol
Parent page: Heart master (10101); related: 9022 (phase-contrast physics and VENC), 10109 (cine)
Version 1.0 — November 2026

Prerequisite: this page assumes the Heart master (ECG set-up, cine in the valve-specific planes, breath-hold management) and the phase-contrast page for VENC physics, and does not repeat them. It documents what changes when the clinical question concerns a valve or an intracardiac shunt: the valve-specific flow planes, the direct and indirect methods for regurgitant volume and fraction, the heterogeneous thresholds proposed in the literature, stenosis assessment, the Qp/Qs calculation and the first quantitative studies of shunts.

1. Executive Summary

1.1 Added Value over the Master Protocol

The master gives valvular disease one row: valve-specific velocity-encoded imaging planes [1]. This page states them valve by valve (Section 2), together with the quantification rules of the SCMR post-processing document (stroke volume, regurgitant fraction, two methods for atrioventricular valve regurgitation, Qp/Qs from pulmonary and sinotubular stroke volumes) [2] and a review that tabulates the direct and indirect methods and the CMR thresholds proposed in single studies [3]. For shunts, the SCMR protocol document has no content and refers to a separate congenital document [1]; the page therefore rests on the post-processing document [2] and two original studies of 4D flow, in 29 adults with atrial septal defect and in 24 children with ventricular septal defect [4, 5].

1.2 Limits of the Dedicated Protocol

No consensus severity thresholds exist in the sources read. The SCMR post-processing document states none for regurgitation or stenosis [2]; the values in Section 4.3 are single-study cut-offs collected by a narrative review, whose text was truncated before the pulmonary regurgitation thresholds [3]. The review states that echocardiography remains first-line and that CMR is a second tool when echocardiography is inconclusive [3]. Not read: the ESC/EACTS and AHA/ACC valvular guidelines, the SCMR congenital protocol (Fratz 2013), the primary studies behind the thresholds, and any study of prosthetic valves or endocarditis (all to be sourced). Points without a source are marked (Reasoning).

1.3 Evidence Base and Its Reading Level

SourceTypeReading levelUsed for
Kramer et al., SCMR protocols 2020 [1]Society consensus documentValvular and flow sections as rendered by a summarising toolValve-specific planes, flow parameters
Schulz-Menger et al., SCMR post-processing 2020 [2]Society consensus documentFlow section as rendered by a summarising tool (repository copy)Definitions, Qp/Qs, contouring, offsets
Vermes et al., Front Cardiovasc Med 2022 [3]Narrative reviewArticle as rendered by a summarising tool; text truncated in the pulmonary regurgitation section; primary studies not readMethods, thresholds, pitfalls
Chelu et al., 2018 [4]Multicentre original study, 29 patientsArticle text as rendered by a summarising tool; journal not named on the page read4D flow in ASD
Zamani-Aliabadi et al., Front Cardiovasc Med 2024 [5]Prospective single-centre, 24 childrenArticle page as rendered by a summarising tool4D and 2D flow in VSD against catheterisation

2. Valve-Specific Acquisition

The protocol document assesses valve morphology with bSSFP cine in the plane of the valve in question and uses velocity-encoded imaging for velocity and direction; the velocity encoding is adapted to the actual velocity, using the lowest value without aliasing, and the shortest possible echo time is used for high-velocity jets [1]. Gradient echo or hybrid gradient echo–echo planar sequences may visualise regurgitant jets with higher sensitivity, for qualitative purposes only [1].

ValveRegurgitation (SCMR protocol)Stenosis (SCMR protocol)
MitralLV structure and function; plane perpendicular to the aortic valve at the sinotubular junction, end diastole, retrospective gating [1]Through-plane encoding parallel to the valve at peak flow disturbance on a long-axis cine; in-plane alternative along the jet [1]
AorticLV structure and function; plane perpendicular to the valve about 5 mm above the valve plane, end diastole, retrospective gating; plane perpendicular to the descending aorta at the level of the main pulmonary artery for diastolic flow reversal [1]Through-plane encoding parallel to the valve at peak flow disturbance, or in-plane along the jet [1]
TricuspidRV structure and function; plane perpendicular to the pulmonic valve about 5 mm above the valve plane (the document names the pulmonic valve in the tricuspid row, which may be an error in the source) [1]Through-plane encoding parallel to the valve on a long-axis cine, or in-plane along the jet [1]
PulmonicRV structure and function, planned from RVOT and coronal views; plane perpendicular to the valve about 5 mm above the valve plane, end diastole, retrospective gating [1]Through-plane encoding parallel to the valve at peak flow disturbance, or in-plane along the jet [1]

3. Flow Technique for Quantification

ItemStatementSource
Plane positionCentred in the vessel, orthogonal to the main flow direction, at the scanner isocentre[1]
Slice thickness / in-plane5–8 mm; in-plane resolution at least one tenth of the vessel diameter (8–16 pixels across the vessel)[1, 2]
Temporal resolution50 ms or better on acquired, not interpolated, frames; at least 11–16 frames per cycle[1, 2]
VENCToo low gives a mosaic pattern; too high loses the jet and SNR; one review advises a value within 25% of the true maximal velocity for the aortic forward flow used in mitral regurgitation[2, 3]
Aliasing checkReview the phase-difference images after each scan and raise VENC if aliased[1]
Background phase offsetVolumes can be dramatically affected by a small offset; the task force makes no dedicated statement on baseline correction[2]
ContoursTrace the vessel cavity on each phase and magnitude image; the phase images carry the encoded information[2]
4D flowConsider if available; no valve-specific parameters given[1]

4. Regurgitation: Quantification and Severity

4.1 Definitions and Internal Checks

Net volume is antegrade minus retrograde volume; regurgitant fraction is retrograde volume divided by antegrade volume, multiplied by 100 [2]. Regurgitant volume of an atrioventricular valve is either diastolic flow across the valve minus systolic forward flow across the semilunar valve, or ventricular stroke volume from cine minus forward flow across the semilunar valve [2]. Without shunts or valve regurgitation, right and left stroke volumes should be nearly equal, branch pulmonary artery flows should sum to the main pulmonary artery flow, and flow-based stroke volume can be compared with cine volumetry [2].

4.2 Direct and Indirect Methods

ValveMethod named in the reviewSource
MitralIndirect, preferred: LV stroke volume from short-axis cine minus aortic forward flow, measured at the sinotubular junction or aortic valve and not in the mid-ascending aorta; average two or three acquisitions. Alternative: LV minus RV stroke volume. The direct method is not recommended for clinical practice[3]
AorticDirect, preferred: phase-contrast about 5 mm above the valve; diastolic flow reversal in the descending aorta. Indirect: aortic minus pulmonary forward flow, or LV minus RV stroke volume. A large dilated aorta may make the direct measurement inaccurate, in which case net pulmonary artery flow is subtracted [2][2, 3]
TricuspidIndirect: RV stroke volume minus pulmonary forward flow (or aortic forward flow if there is no aortic regurgitation)[3]
PulmonaryDirect (reference): phase-contrast 5 mm to 1 cm above the valve. Indirect: volumetric, or forward-flow differences[3]

Indirect methods are invalid when pulmonary regurgitation or a shunt coexists [3].

4.3 Proposed Thresholds and Their Status

ValveEcho guideline thresholds citedCMR-derived proposals cited
MitralRegurgitant fraction 50% or more; regurgitant volume 60 mL or more (ASE)Fraction above 42% for severe (vs Doppler echo); above 35% for moderate to severe; fraction 40% and volume 55 mL for surgical indication; fraction above 41% and volume above 64 mL. The review concludes severe is "very likely" at fraction 40% or more, pending validation
AorticFraction 50% or more; volume 60 mL or more; moderate–severe band 45 to under 60 mLFraction cut-offs between 30% and 37% in five studies (above 30%, 30% or more, above 33%, above 32%, 37% or more); volume above 40 mL or 50 mL in two studies. The review concludes "very likely" significant at fraction 30% or more, pending larger trials

All values from [3]; they come from different small cohorts and reference methods and are not a consensus. For the mitral valve, proximal isovelocity surface area (PISA) yields higher regurgitant volumes than CMR and the two sets of thresholds are not interchangeable; for the aortic valve the correlation with echocardiography is modest, especially with eccentric jets; for the tricuspid valve correlations between echo parameters and CMR were 0.3 to 0.49 [3].

4.4 Association With Surgery and Outcome

Study cited in [3]Valve (n)EndpointCMR cut-off (sensitivity / specificity)
MyersonMitral (109)Surgical indicationFraction above 40% (76% / 74%); volume above 55 mL (72% / 87%)
CavalcanteIschaemic mitral (578)Death or transplant at 1 yearFraction 35% or more (specificity 84%)
MyersonAortic (113)Surgical indicationFraction above 33% (85% / 92%)
HarrisAortic (29)Valve surgery or heart failureFraction 37% or more (100% / 75%)

Most endpoints are an indication for surgery rather than hard outcomes, and the review notes that no reference method exists [3].

5. Stenosis

The protocol document gives the acquisition (through-plane encoding parallel to the valve at the point of peak flow disturbance, or in-plane along the jet) [1]. The post-processing document warns that peak velocity reported in a single pixel is noise-sensitive, that through-plane peak velocity may underestimate the true peak along the vena contracta, and that peak velocity (unlike volume flow) depends on the angle between plane and flow; it does not mention the Bernoulli gradient, valve area planimetry or grading thresholds [2]. (to be sourced): grading thresholds, planimetry and comparison with echocardiography for stenosis were not found in the sources read.

6. Intracardiac Shunts

6.1 What the Consensus Documents Say

The protocol document has no shunt content and states that congenital protocols are covered in a separate document [1]. The post-processing document estimates Qp/Qs from stroke volumes obtained by flow measurement in the pulmonary artery and at the sinotubular junction, and states that shunts can also be quantified by direct measurement of the flow through the defect; it gives no numerical cut-offs [2].

6.2 Original Studies of 4D Flow

ItemASD, 4D flow [4]VSD in children, 4D and 2D flow against catheterisation [5]
DesignMulticentre (four sites; three retrospective, one prospective); 29 patients analysed; 1.5 T (69%) and 3 T (31%); gadolinium at all sitesProspective single centre; 24 children analysed of 31 recruited (7 excluded or cancelled); no contrast or sedation for MRI; free breathing with compressed sensing
Qp/QsMedian 1.6 at valve, main artery and peripheral levels; 1.8 by stroke volumes4D 1.403 ± 0.344, 2D 1.326 ± 0.283, catheter 1.499 ± 0.396
AgreementSpearman 0.95 (valve, peripheral) and 0.93 (stroke volume) against the main artery level; interobserver ICC for Qp/Qs 0.968–0.9834D versus catheter r = 0.962, bias 0.096 (limits −0.127 to 0.319); 2D versus catheter r = 0.894, bias 0.172 (limits −0.201 to 0.546)
Direct shunt volumeMeasurable in 20 of 29; underestimated by 0.57 L/min (limits −2.71 to 1.57)Not applicable
AcquisitionScan time median 10.75 min; VENC 150–500 cm/sNot stated in the extraction

Limits stated or evident: modest samples, single vendor in [4] and single centre, scanner and software in [5]; no direct comparison with 2D phase-contrast in [4]; the invasive reference is itself imperfect and the MRI was done about an hour before catheterisation under different conditions in [5]; the abstract and results of [4] disagree on the direction of one bias, and [5] reports two sets of 4D results (r = 0.936 and 0.962) without a full explanation [4, 5]. Three patients straddled the 1.5 threshold in [4], which shows how classification near the threshold depends on the measurement level.

7. Reporting Framework

ItemContent
ValveMorphology on valve-plane cine; leaflet motion; jet direction and visualised extent (qualitative)
RegurgitationMethod (direct or indirect), flow level, regurgitant volume and fraction, consistency check of the right and left stroke volumes [2, 3]
ThresholdsState which cut-offs were used and that CMR thresholds are not consensus-based [3]
StenosisPeak velocity (pixel or averaged), plane angle, aliasing check [2]
ShuntQp/Qs from pulmonary and aortic stroke volumes, level of measurement, direct shunt flow if measurable [2, 4]
Ventricular consequencesVolumes, mass and function by cine (see 10109)

8. Technical Pitfalls

PitfallConsequenceSource
Mitral regurgitation in arrhythmia or poor breath-holdingLess accurate quantification[3]
Basal slice selection with prolapse or mitral annular disjunctionError in the stroke volume used for the indirect method[3]
Eccentric or non-laminar aortic jet, or aortic stenosisDirect measurement less accurate[3]
Indirect method with pulmonary regurgitation or shuntInvalid[3]
Aliasing, VENC misestimation, phase offsetWrong velocities and volumes[2, 3]
Large slice thicknessHigh velocities averaged out[2]
Pacemaker or defibrillator artefactReduces tricuspid quantification[3]
Turbulent pulmonary flow in repaired tetralogy or dilated arteriesRegurgitant volume may be underestimated[3]
Jet imaging with bSSFPCan underestimate the degree of regurgitation[3]

9. MRI Technologist Pearls and Quality Checklist

(Reasoning) Plan the flow plane from two orthogonal cine views of the valve and check on the magnitude image that the vessel lies in the middle of the plane before acquiring; write the VENC and the heart rate on the exam sheet, and repeat a series whose phase image is aliased rather than leaving the correction to post-processing.

CheckPass criterion
PlanePerpendicular to flow, vessel centred, about 5 mm above the semilunar valve [1]
Resolution8–16 pixels across the vessel; slice 5–8 mm; at least 11–16 frames [1, 2]
AliasingPhase-difference images checked after each series [1]
GatingRetrospective gating over the whole cycle for regurgitation planes [1]
RepeatTwo or three acquisitions for the aortic forward flow of the indirect mitral method [3]
ConsistencyStroke volumes cross-checked with cine [2]

10. Advanced Technical Parameters

ParameterValue statedSource
Flow plane distance above semilunar valveAbout 5 mm (pulmonary 5 mm to 1 cm in the review)[1, 3]
Slice thickness5–8 mm[1]
Temporal resolution50 ms or better; 11–16 frames minimum[1, 2]
Pixels across the vessel8–16[2]
VENC example of aliasing shift175 cm/s peak with VENC 150 cm/s shifted to −100 to +200 cm/s by software[2]
Bronchial flowAbout 5% of aortic output (overestimation in indirect aortic regurgitation methods)[2]
4D flow in ASDVENC 150–500 cm/s; median scan time 10.75 min[4]
4D flow aliasing limit in the reviewAliasing above 5 m/s; long scan times[3]

11. Evidence Gaps and Ongoing Debate

Not read: ESC/EACTS 2021 and AHA/ACC 2020 valvular guidelines, the SCMR congenital protocol (Fratz 2013), the SCMR clinical indications document, primary studies behind thresholds, prosthetic valves, endocarditis and perivalvular leak, mitral annular disjunction, tricuspid regurgitation outcome data, stenosis grading and planimetry, 4D flow consensus parameters, and the pulmonary regurgitation thresholds truncated in the review (all to be sourced). Debate: CMR thresholds differ from echo thresholds and among studies; the review concludes values of about 40% (mitral) and 30% (aortic) are very likely significant, pending validation, and notes no reference method exists [3]. Source inconsistencies: the tricuspid row of the protocol document names the pulmonic valve [1]; the 4D flow papers contain internal inconsistencies noted in Section 6.2 [4, 5].

12. Evidence-Based References

References are ordered by level of evidence and then by recency. The reading level of each source is stated in its line; a source read only at abstract level supports only what its abstract states.

A. Guidelines / Consensus / Society Recommendations

High
[1] Kramer CM, Barkhausen J, Bucciarelli-Ducci C, Flamm SD, Kim RJ, Nagel E. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson. 2020;22:17. DOI: 10.1186/s12968-020-00607-1.
(High — Society consensus document; valvular and flow sections as rendered by a summarising tool; no shunt, Qp/Qs or congenital content; tricuspid row names the pulmonic valve) Cited in Sections 1, 2, 3, 5, 6, 8, 9, 10, 11.
High
[2] Schulz-Menger J, Bluemke DA, Bremerich J, et al. Standardized image interpretation and post-processing in cardiovascular magnetic resonance – 2020 update: Society for Cardiovascular Magnetic Resonance (SCMR) Board of Trustees Task Force on Standardized Post-Processing. J Cardiovasc Magn Reson. 2020;22:19. DOI: 10.1186/s12968-020-00610-6.
(High — Society consensus document; flow quantification section as rendered by a summarising tool from a repository copy; no dedicated valvular or congenital section; no regurgitation or stenosis thresholds; no Bernoulli or planimetry statement) Cited in Sections 1, 3, 4, 5, 6, 7, 8, 9, 10.

B. Systematic Reviews / Meta-analyses

Moderate
[3] Vermes E, Iacuzio L, Levy F, Bohbot Y, Renard C, Gerber B, Maréchaux S, Tribouilloy C. Role of cardiovascular magnetic resonance in native valvular regurgitation: a comprehensive review of protocols, grading of severity, and prediction of valve surgery. Front Cardiovasc Med. 2022;9:881141. DOI: 10.3389/fcvm.2022.881141.
(Moderate — Narrative review, not a systematic review; article as rendered by a summarising tool, text truncated in the pulmonary regurgitation section; primary studies behind thresholds not read; thresholds from small cohorts without consensus) Cited in Sections 1, 3, 4, 7, 8, 9, 10, 11.

C. Important Prospective / Original Studies

Moderate
[4] Chelu RG, Horowitz M, Sucha D, Kardys I, Ingremeau D, Vasanawala S, Nieman K, Paul JF, Hsiao A. Evaluation of atrial septal defects with 4D flow MRI: multilevel and inter-reader reproducibility for quantification of shunt severity. 2018. DOI: 10.1007/s10334-018-0702-z.
(Moderate — Multicentre original study, 29 patients, single vendor, gadolinium at all sites; article text as rendered by a summarising tool; journal not named on the page read; no comparison with 2D phase-contrast; internal inconsistencies in the direction of bias) Cited in Sections 1, 6, 7, 10, 11.
Moderate
[5] Zamani-Aliabadi SM, Qanadli SD, Fatehi-Feyzabad SH, Ghasemnezhad M, Ghaemi H, Azarine A, Mohammadzadeh A, Bitarafan-Rajabi A, Mortezaeian H, Rezaei-Kalantari K. Assessment of 4D flow MRI for quantification of left-to-right shunt in pediatric patients with ventricular septal defect: comparison with right heart catheterization. Front Cardiovasc Med. 2024;11:1399110. DOI: 10.3389/fcvm.2024.1399110.
(Moderate — Prospective single-centre study, 24 children (3–15 years), one scanner and software; article page as rendered by a summarising tool; catheter reference imperfect; MRI about one hour before catheterisation; 7 of 31 recruited excluded; two sets of 4D results reported) Cited in Sections 1, 6, 7, 11.

D. Technical MRI Papers

None used on this page.

E. Landmark Historical References

None used on this page.

Recent PubMed search for this protocol

Last updated: November 2026

Report an error on this page