MRI Heart for Valvular Disease and Shunts
MRIninja Knowledge Base | Child Page — Pathology-Specific Protocol
Parent page: Heart master (10101); related: 9022 (phase-contrast physics and VENC), 10109 (cine)
Version 1.0 — November 2026
Prerequisite: this page assumes the Heart master (ECG set-up, cine in the valve-specific planes, breath-hold management) and the phase-contrast page for VENC physics, and does not repeat them. It documents what changes when the clinical question concerns a valve or an intracardiac shunt: the valve-specific flow planes, the direct and indirect methods for regurgitant volume and fraction, the heterogeneous thresholds proposed in the literature, stenosis assessment, the Qp/Qs calculation and the first quantitative studies of shunts.
1. Executive Summary
1.1 Added Value over the Master Protocol
The master gives valvular disease one row: valve-specific velocity-encoded imaging planes [1]. This page states them valve by valve (Section 2), together with the quantification rules of the SCMR post-processing document (stroke volume, regurgitant fraction, two methods for atrioventricular valve regurgitation, Qp/Qs from pulmonary and sinotubular stroke volumes) [2] and a review that tabulates the direct and indirect methods and the CMR thresholds proposed in single studies [3]. For shunts, the SCMR protocol document has no content and refers to a separate congenital document [1]; the page therefore rests on the post-processing document [2] and two original studies of 4D flow, in 29 adults with atrial septal defect and in 24 children with ventricular septal defect [4, 5].
1.2 Limits of the Dedicated Protocol
No consensus severity thresholds exist in the sources read. The SCMR post-processing document states none for regurgitation or stenosis [2]; the values in Section 4.3 are single-study cut-offs collected by a narrative review, whose text was truncated before the pulmonary regurgitation thresholds [3]. The review states that echocardiography remains first-line and that CMR is a second tool when echocardiography is inconclusive [3]. Not read: the ESC/EACTS and AHA/ACC valvular guidelines, the SCMR congenital protocol (Fratz 2013), the primary studies behind the thresholds, and any study of prosthetic valves or endocarditis (all to be sourced). Points without a source are marked (Reasoning).
1.3 Evidence Base and Its Reading Level
| Source | Type | Reading level | Used for |
|---|---|---|---|
| Kramer et al., SCMR protocols 2020 [1] | Society consensus document | Valvular and flow sections as rendered by a summarising tool | Valve-specific planes, flow parameters |
| Schulz-Menger et al., SCMR post-processing 2020 [2] | Society consensus document | Flow section as rendered by a summarising tool (repository copy) | Definitions, Qp/Qs, contouring, offsets |
| Vermes et al., Front Cardiovasc Med 2022 [3] | Narrative review | Article as rendered by a summarising tool; text truncated in the pulmonary regurgitation section; primary studies not read | Methods, thresholds, pitfalls |
| Chelu et al., 2018 [4] | Multicentre original study, 29 patients | Article text as rendered by a summarising tool; journal not named on the page read | 4D flow in ASD |
| Zamani-Aliabadi et al., Front Cardiovasc Med 2024 [5] | Prospective single-centre, 24 children | Article page as rendered by a summarising tool | 4D and 2D flow in VSD against catheterisation |
2. Valve-Specific Acquisition
The protocol document assesses valve morphology with bSSFP cine in the plane of the valve in question and uses velocity-encoded imaging for velocity and direction; the velocity encoding is adapted to the actual velocity, using the lowest value without aliasing, and the shortest possible echo time is used for high-velocity jets [1]. Gradient echo or hybrid gradient echo–echo planar sequences may visualise regurgitant jets with higher sensitivity, for qualitative purposes only [1].
| Valve | Regurgitation (SCMR protocol) | Stenosis (SCMR protocol) |
|---|---|---|
| Mitral | LV structure and function; plane perpendicular to the aortic valve at the sinotubular junction, end diastole, retrospective gating [1] | Through-plane encoding parallel to the valve at peak flow disturbance on a long-axis cine; in-plane alternative along the jet [1] |
| Aortic | LV structure and function; plane perpendicular to the valve about 5 mm above the valve plane, end diastole, retrospective gating; plane perpendicular to the descending aorta at the level of the main pulmonary artery for diastolic flow reversal [1] | Through-plane encoding parallel to the valve at peak flow disturbance, or in-plane along the jet [1] |
| Tricuspid | RV structure and function; plane perpendicular to the pulmonic valve about 5 mm above the valve plane (the document names the pulmonic valve in the tricuspid row, which may be an error in the source) [1] | Through-plane encoding parallel to the valve on a long-axis cine, or in-plane along the jet [1] |
| Pulmonic | RV structure and function, planned from RVOT and coronal views; plane perpendicular to the valve about 5 mm above the valve plane, end diastole, retrospective gating [1] | Through-plane encoding parallel to the valve at peak flow disturbance, or in-plane along the jet [1] |
3. Flow Technique for Quantification
| Item | Statement | Source |
|---|---|---|
| Plane position | Centred in the vessel, orthogonal to the main flow direction, at the scanner isocentre | [1] |
| Slice thickness / in-plane | 5–8 mm; in-plane resolution at least one tenth of the vessel diameter (8–16 pixels across the vessel) | [1, 2] |
| Temporal resolution | 50 ms or better on acquired, not interpolated, frames; at least 11–16 frames per cycle | [1, 2] |
| VENC | Too low gives a mosaic pattern; too high loses the jet and SNR; one review advises a value within 25% of the true maximal velocity for the aortic forward flow used in mitral regurgitation | [2, 3] |
| Aliasing check | Review the phase-difference images after each scan and raise VENC if aliased | [1] |
| Background phase offset | Volumes can be dramatically affected by a small offset; the task force makes no dedicated statement on baseline correction | [2] |
| Contours | Trace the vessel cavity on each phase and magnitude image; the phase images carry the encoded information | [2] |
| 4D flow | Consider if available; no valve-specific parameters given | [1] |
4. Regurgitation: Quantification and Severity
4.1 Definitions and Internal Checks
Net volume is antegrade minus retrograde volume; regurgitant fraction is retrograde volume divided by antegrade volume, multiplied by 100 [2]. Regurgitant volume of an atrioventricular valve is either diastolic flow across the valve minus systolic forward flow across the semilunar valve, or ventricular stroke volume from cine minus forward flow across the semilunar valve [2]. Without shunts or valve regurgitation, right and left stroke volumes should be nearly equal, branch pulmonary artery flows should sum to the main pulmonary artery flow, and flow-based stroke volume can be compared with cine volumetry [2].
4.2 Direct and Indirect Methods
| Valve | Method named in the review | Source |
|---|---|---|
| Mitral | Indirect, preferred: LV stroke volume from short-axis cine minus aortic forward flow, measured at the sinotubular junction or aortic valve and not in the mid-ascending aorta; average two or three acquisitions. Alternative: LV minus RV stroke volume. The direct method is not recommended for clinical practice | [3] |
| Aortic | Direct, preferred: phase-contrast about 5 mm above the valve; diastolic flow reversal in the descending aorta. Indirect: aortic minus pulmonary forward flow, or LV minus RV stroke volume. A large dilated aorta may make the direct measurement inaccurate, in which case net pulmonary artery flow is subtracted [2] | [2, 3] |
| Tricuspid | Indirect: RV stroke volume minus pulmonary forward flow (or aortic forward flow if there is no aortic regurgitation) | [3] |
| Pulmonary | Direct (reference): phase-contrast 5 mm to 1 cm above the valve. Indirect: volumetric, or forward-flow differences | [3] |
Indirect methods are invalid when pulmonary regurgitation or a shunt coexists [3].
4.3 Proposed Thresholds and Their Status
| Valve | Echo guideline thresholds cited | CMR-derived proposals cited |
|---|---|---|
| Mitral | Regurgitant fraction 50% or more; regurgitant volume 60 mL or more (ASE) | Fraction above 42% for severe (vs Doppler echo); above 35% for moderate to severe; fraction 40% and volume 55 mL for surgical indication; fraction above 41% and volume above 64 mL. The review concludes severe is "very likely" at fraction 40% or more, pending validation |
| Aortic | Fraction 50% or more; volume 60 mL or more; moderate–severe band 45 to under 60 mL | Fraction cut-offs between 30% and 37% in five studies (above 30%, 30% or more, above 33%, above 32%, 37% or more); volume above 40 mL or 50 mL in two studies. The review concludes "very likely" significant at fraction 30% or more, pending larger trials |
All values from [3]; they come from different small cohorts and reference methods and are not a consensus. For the mitral valve, proximal isovelocity surface area (PISA) yields higher regurgitant volumes than CMR and the two sets of thresholds are not interchangeable; for the aortic valve the correlation with echocardiography is modest, especially with eccentric jets; for the tricuspid valve correlations between echo parameters and CMR were 0.3 to 0.49 [3].
4.4 Association With Surgery and Outcome
| Study cited in [3] | Valve (n) | Endpoint | CMR cut-off (sensitivity / specificity) |
|---|---|---|---|
| Myerson | Mitral (109) | Surgical indication | Fraction above 40% (76% / 74%); volume above 55 mL (72% / 87%) |
| Cavalcante | Ischaemic mitral (578) | Death or transplant at 1 year | Fraction 35% or more (specificity 84%) |
| Myerson | Aortic (113) | Surgical indication | Fraction above 33% (85% / 92%) |
| Harris | Aortic (29) | Valve surgery or heart failure | Fraction 37% or more (100% / 75%) |
Most endpoints are an indication for surgery rather than hard outcomes, and the review notes that no reference method exists [3].
5. Stenosis
The protocol document gives the acquisition (through-plane encoding parallel to the valve at the point of peak flow disturbance, or in-plane along the jet) [1]. The post-processing document warns that peak velocity reported in a single pixel is noise-sensitive, that through-plane peak velocity may underestimate the true peak along the vena contracta, and that peak velocity (unlike volume flow) depends on the angle between plane and flow; it does not mention the Bernoulli gradient, valve area planimetry or grading thresholds [2]. (to be sourced): grading thresholds, planimetry and comparison with echocardiography for stenosis were not found in the sources read.
6. Intracardiac Shunts
6.1 What the Consensus Documents Say
The protocol document has no shunt content and states that congenital protocols are covered in a separate document [1]. The post-processing document estimates Qp/Qs from stroke volumes obtained by flow measurement in the pulmonary artery and at the sinotubular junction, and states that shunts can also be quantified by direct measurement of the flow through the defect; it gives no numerical cut-offs [2].
6.2 Original Studies of 4D Flow
| Item | ASD, 4D flow [4] | VSD in children, 4D and 2D flow against catheterisation [5] |
|---|---|---|
| Design | Multicentre (four sites; three retrospective, one prospective); 29 patients analysed; 1.5 T (69%) and 3 T (31%); gadolinium at all sites | Prospective single centre; 24 children analysed of 31 recruited (7 excluded or cancelled); no contrast or sedation for MRI; free breathing with compressed sensing |
| Qp/Qs | Median 1.6 at valve, main artery and peripheral levels; 1.8 by stroke volumes | 4D 1.403 ± 0.344, 2D 1.326 ± 0.283, catheter 1.499 ± 0.396 |
| Agreement | Spearman 0.95 (valve, peripheral) and 0.93 (stroke volume) against the main artery level; interobserver ICC for Qp/Qs 0.968–0.983 | 4D versus catheter r = 0.962, bias 0.096 (limits −0.127 to 0.319); 2D versus catheter r = 0.894, bias 0.172 (limits −0.201 to 0.546) |
| Direct shunt volume | Measurable in 20 of 29; underestimated by 0.57 L/min (limits −2.71 to 1.57) | Not applicable |
| Acquisition | Scan time median 10.75 min; VENC 150–500 cm/s | Not stated in the extraction |
Limits stated or evident: modest samples, single vendor in [4] and single centre, scanner and software in [5]; no direct comparison with 2D phase-contrast in [4]; the invasive reference is itself imperfect and the MRI was done about an hour before catheterisation under different conditions in [5]; the abstract and results of [4] disagree on the direction of one bias, and [5] reports two sets of 4D results (r = 0.936 and 0.962) without a full explanation [4, 5]. Three patients straddled the 1.5 threshold in [4], which shows how classification near the threshold depends on the measurement level.
7. Reporting Framework
| Item | Content |
|---|---|
| Valve | Morphology on valve-plane cine; leaflet motion; jet direction and visualised extent (qualitative) |
| Regurgitation | Method (direct or indirect), flow level, regurgitant volume and fraction, consistency check of the right and left stroke volumes [2, 3] |
| Thresholds | State which cut-offs were used and that CMR thresholds are not consensus-based [3] |
| Stenosis | Peak velocity (pixel or averaged), plane angle, aliasing check [2] |
| Shunt | Qp/Qs from pulmonary and aortic stroke volumes, level of measurement, direct shunt flow if measurable [2, 4] |
| Ventricular consequences | Volumes, mass and function by cine (see 10109) |
8. Technical Pitfalls
| Pitfall | Consequence | Source |
|---|---|---|
| Mitral regurgitation in arrhythmia or poor breath-holding | Less accurate quantification | [3] |
| Basal slice selection with prolapse or mitral annular disjunction | Error in the stroke volume used for the indirect method | [3] |
| Eccentric or non-laminar aortic jet, or aortic stenosis | Direct measurement less accurate | [3] |
| Indirect method with pulmonary regurgitation or shunt | Invalid | [3] |
| Aliasing, VENC misestimation, phase offset | Wrong velocities and volumes | [2, 3] |
| Large slice thickness | High velocities averaged out | [2] |
| Pacemaker or defibrillator artefact | Reduces tricuspid quantification | [3] |
| Turbulent pulmonary flow in repaired tetralogy or dilated arteries | Regurgitant volume may be underestimated | [3] |
| Jet imaging with bSSFP | Can underestimate the degree of regurgitation | [3] |
9. MRI Technologist Pearls and Quality Checklist
(Reasoning) Plan the flow plane from two orthogonal cine views of the valve and check on the magnitude image that the vessel lies in the middle of the plane before acquiring; write the VENC and the heart rate on the exam sheet, and repeat a series whose phase image is aliased rather than leaving the correction to post-processing.
| Check | Pass criterion |
|---|---|
| Plane | Perpendicular to flow, vessel centred, about 5 mm above the semilunar valve [1] |
| Resolution | 8–16 pixels across the vessel; slice 5–8 mm; at least 11–16 frames [1, 2] |
| Aliasing | Phase-difference images checked after each series [1] |
| Gating | Retrospective gating over the whole cycle for regurgitation planes [1] |
| Repeat | Two or three acquisitions for the aortic forward flow of the indirect mitral method [3] |
| Consistency | Stroke volumes cross-checked with cine [2] |
10. Advanced Technical Parameters
| Parameter | Value stated | Source |
|---|---|---|
| Flow plane distance above semilunar valve | About 5 mm (pulmonary 5 mm to 1 cm in the review) | [1, 3] |
| Slice thickness | 5–8 mm | [1] |
| Temporal resolution | 50 ms or better; 11–16 frames minimum | [1, 2] |
| Pixels across the vessel | 8–16 | [2] |
| VENC example of aliasing shift | 175 cm/s peak with VENC 150 cm/s shifted to −100 to +200 cm/s by software | [2] |
| Bronchial flow | About 5% of aortic output (overestimation in indirect aortic regurgitation methods) | [2] |
| 4D flow in ASD | VENC 150–500 cm/s; median scan time 10.75 min | [4] |
| 4D flow aliasing limit in the review | Aliasing above 5 m/s; long scan times | [3] |
11. Evidence Gaps and Ongoing Debate
Not read: ESC/EACTS 2021 and AHA/ACC 2020 valvular guidelines, the SCMR congenital protocol (Fratz 2013), the SCMR clinical indications document, primary studies behind thresholds, prosthetic valves, endocarditis and perivalvular leak, mitral annular disjunction, tricuspid regurgitation outcome data, stenosis grading and planimetry, 4D flow consensus parameters, and the pulmonary regurgitation thresholds truncated in the review (all to be sourced). Debate: CMR thresholds differ from echo thresholds and among studies; the review concludes values of about 40% (mitral) and 30% (aortic) are very likely significant, pending validation, and notes no reference method exists [3]. Source inconsistencies: the tricuspid row of the protocol document names the pulmonic valve [1]; the 4D flow papers contain internal inconsistencies noted in Section 6.2 [4, 5].
12. Evidence-Based References
References are ordered by level of evidence and then by recency. The reading level of each source is stated in its line; a source read only at abstract level supports only what its abstract states.
A. Guidelines / Consensus / Society Recommendations
B. Systematic Reviews / Meta-analyses
C. Important Prospective / Original Studies
D. Technical MRI Papers
None used on this page.
E. Landmark Historical References
None used on this page.