T1 and T2 Mapping and Extracellular Volume in Cardiac MRI — Technical Deep Dive: Confounders, ROI and Quality Control, ECV Practice, Disease Patterns and Prognosis

up to this point verified by human experts

MRIninja Knowledge Base | Deep Dive — Technical Page
Parent page: Heart master (10101); related child pages: 10102, 10105
Version 1.0 — November 2026

Prerequisite: this page assumes Section 10.3 (sequence schemes and their trade-off, slices and breath-holds, ECV formula and contrast timing, temperature, basic reference-range rules) and Section 10.4 (T2* parameters and risk tiers) of the Heart master, and does not repeat them. It documents what the master does not: the confounders and how to limit them, how to place regions of interest and check the maps, the practical points of ECV (haematocrit, synthetic ECV), T2 mapping specifics, the expected direction of change by disease with its limits, reporting and quality control, the T2* details beyond the master, and the prognostic evidence.

1. Executive Summary

1.1 Added Value over the Master Protocol

Mapping gives numbers, and numbers are only as reliable as the conditions under which they were made. The 2017 consensus states that native T1, T2 and T2* values are not directly comparable between 1.5 T and 3 T and depend on sequence and processing [1]; the 2020 post-processing document asks for a local reference range to be primarily used and for results to be reported as absolute value, Z-score and reference range [2]. This page gathers the technical conditions that make a map interpretable (Section 2), the checks to apply before measuring (Section 3), the practical points of ECV (Section 4), the direction of change expected in each disease and where it is non-specific (Section 6), and the prognostic evidence, which in the one meta-analysis read is limited to heart failure with preserved ejection fraction (Section 8) [3].

1.2 Limits of the Page

Both consensus documents were read as rendered by a summarising tool; the last 261 characters of the 2017 text (reference list) were not read, and the formula of the 2020 document was garbled in the extraction and is not quoted from it [1, 2]. The consensus gives no numeric normal values and no universal cut-offs for T1, T2 or ECV [2]. The evidence base for disease patterns is described by the consensus itself as mostly single-centre studies and several applications (bolus versus infusion ECV, ECV thresholds in dilated and hypertrophic cardiomyopathy, arrhythmogenic cardiomyopathy) are listed as not established [1]. The prognostic meta-analysis concerns one population and has internal inconsistencies [3]. Points without a source are marked (Reasoning).

1.3 Evidence Base and Its Reading Level

SourceTypeReading levelUsed for
Messroghli et al., JCMR 2017 [1]Society consensus statementFull text as rendered by a summarising tool; final 261 characters unreadConfounders, ROI and QC, ECV practice, T2, disease patterns, limits
Schulz-Menger et al., JCMR 2020 [2]Society consensus documentMapping and T2* sections as rendered by a summarising toolROI placement, ECV and T2 reporting, T2* cut-offs
Golukhova et al., J Clin Med 2022 [3]Systematic review and meta-analysisFull text as rendered by a summarising toolPrognosis in HFpEF

2. Confounders and How to Limit Them

ConfounderWhat the consensus statesPractical consequence
Field strength and sequenceNative T1, T2 and T2* are not directly comparable between 1.5 T and 3 T; values differ by sequence and processing [1]Compare only under matching conditions
Off-resonanceAffects MOLLI accuracy; volume-selective B0 shimming focused on the heart is highly recommended at 1.5 T and essential at 3 T [1]Shim before mapping, especially at 3 T
Heart rateInversion-recovery and T2-prepared methods with insufficient waiting time can be heart-rate dependent; saturation methods largely avoid this [1]Check rhythm and rate; record them
Magnetisation transferMOLLI T1 is sensitive to it; saturation schemes largely avoid it [1]Do not mix schemes across studies
Inversion-recovery biasLook-Locker methods show a significant negative offset relative to saturation recovery; correction is required [1]Use the manufacturer's corrected product sequence
Partial volume and fatBlood or fat suppression may help but can alter the measurement; intramyocardial fat gives a wide range of native T1 values [1]Keep ROIs away from borders
TemperatureT1 rises about 1% per 1 °C; T2 shortens as temperature rises [1]Negligible in patients (master 10.3)
Coronary vasodilationMay increase native T1 through intravascular volume expansion [1]Relevant when mapping during or after vasodilator stress (Reasoning)
DevicesMapping is not reliable with pacemakers and implantable defibrillators without special shimming or sequences [1]Flag the limitation in the report
RhythmDiastolic acquisition in regular rhythm; in atrial fibrillation repeat and average [1]Record rhythm

3. Map Analysis and Quality Control

3.1 Visual Check Before Measurement

Review the source images of different T1 weighting for diagnostic quality, then the final map for artefacts; automatically generated quality-control maps (for example T1*) may be used to exclude misregistration, and areas of poor curve fit on them should be flagged [1, 2]. In-plane motion correction is not a replacement for breath-holding and cannot correct through-plane motion, so the motion-corrected raw images must be viewed [1]. Colour maps need site-specific normal ranges; greyscale with image processing can highlight abnormal areas [2].

3.2 Region-of-Interest Placement

SituationRuleSource
Global or diffuse diseaseOne ROI drawn conservatively in the septum on a mid-cavity short-axis map, to limit susceptibility artefact[1, 2]
Mid-cavity artefactual or inconclusiveBasal ROI for validation[1, 2]
Focal diseaseAdd ROIs in visibly abnormal areas[2]
ROI sizeAvoid very small ROIs (under 20 pixels)[1, 2]
Bias controlDraw on greyscale rather than colour maps; check automatic ROIs; draw independently of LGE; for diffuse disease exclude focal LGE from the ROI[1, 2]
SoftwareNo preferred package; the reader must be trained on the local one[2]

No polar-map or 17-segment analysis for mapping, and no border-offset instruction for the ROI, are given in the post-processing document [2].

4. ECV: Practical Points

The formula and contrast dose are in master 10.3. Additional points: the haematocrit should be obtained immediately before the scan if possible, otherwise within 24 h; the 2020 document asks for the same day [1, 2]. If it is not available, haematocrit can be estimated from the native blood T1 (synthetic ECV); the 2017 text calls the agreement with haematocrit-derived ECV preliminary [1, 2]. Post-contrast mapping is performed 10–30 min after the contrast, more than 10 min being needed to approach a steady state; the 2013 consensus used 15 min and the 2017 text accepts 10 min; in split-dose protocols the timing runs from the last dose [1]. The blood-pool ROI is drawn in the centre of the cavity on both maps, excluding papillary muscles and trabeculae, and both maps are processed identically [2]. Inline ECV maps are acceptable if the raw images are checked [2]. Bolus versus infusion protocols are not addressed [1].

5. T2 Mapping

T2-prepared bSSFP or gradient-echo sequences with at least 3 source images are recommended, two-parameter fitting is appropriate, and a saturation pulse may remove T1-recovery bias; GraSE or TSE may be appropriate only if favourable accuracy and precision data are published [1]. T2 is sensitive to T1 and off-resonance, and higher T2 values were reported in women and older subjects in the studies cited [1]. T2 maps are checked, ROI-placed and reported like T1 maps; the report gives the absolute value, the Z-score and the reference range [2].

6. Disease Patterns and Their Limits

ConditionDirection and details statedSource
AmyloidosisNative T1 and ECV are elevated early, before LGE, but initially non-specific; ECV above 55% appears in transmural LGE; diffuse fibrosis rarely raises ECV above 40%[1]
Anderson-Fabry diseaseNative T1 low in around 85% of subjects with left-ventricular hypertrophy and in around half of those without[1]
Iron overloadNative T1, T2 and T2* all decrease; T2* is the main tool (Section 7)[1]
Acute myocarditisT1, ECV and T2 mapping have clinical utility; subacute inflammation mildly, acute inflammation moderately to severely increased[1]
Acute infarctionNative T1 and T2 prolong, post-contrast T1 shortens relative to remote myocardium; ECV expands[1]
Hypertrophic cardiomyopathyNative T1 increases in areas with and without LGE; ECV is raised in the typical phenotype and in asymptomatic relatives[1]
Dilated cardiomyopathySame pattern as above; high native T1 is associated with cardiovascular events and heart failure[1]
Diffuse fibrosisNative T1 and ECV increase, post-contrast T1 decreases[1]

Elevated T1 may reflect fibrosis or oedema (T2), and reduced T1 may reflect iron or off-resonance, so a single map is rarely specific [1]. Severity grades rely on tertiles when outcome data are lacking, which is the case for most applications [1]. Thresholds for ECV in dilated and hypertrophic cardiomyopathy, the role of ECV in thalassaemia, data in arrhythmogenic cardiomyopathy and native T1 in renal impairment are listed as not covered or unresolved [1].

7. T2* Beyond the Master

The master gives parameters and risk tiers. Additions: quantitative analysis is always required, with software that has regulatory approval; the ROI is a full-thickness septal ROI on the mid-ventricular short-axis image, avoiding blood pool and proximal vessels; mean signal against echo time is fitted with a validated algorithm [2]. In heavy iron overload the signal at long echo times falls into background noise, causing a plateau and underestimated T2*, which is handled by truncating late echo times and is not significant with the black-blood sequence [2]. The post-processing document gives normal cardiac T2* as 40 ms, below 20 ms as cardiac iron overload and below 10 ms as increased risk of heart failure at 1.5 T, and states that T2* at 3 T for iron overload cannot be recommended at this time [2]. The consensus adds that a commonly used offset model is fundamentally incorrect and that the truncation model is challenging in severe overload [1].

8. Prognostic Evidence

One systematic review and meta-analysis pooled 7 single-centre observational cohorts (1,930 patients, 1,129 with heart failure with preserved ejection fraction; mean follow-up 25.6 months) for death and/or heart-failure hospitalisation [3]. ECV unadjusted: hazard ratio 1.12 (95% CI 1.06–1.18; I² 53%; 877 patients); ECV adjusted: 1.08 (1.04–1.13; I² 0%); native T1 unadjusted: 1.01 (1.00–1.02; I² 43%) [3]. Post-contrast T1 was reported in one study and not pooled [3]. Limits: few studies, heterogeneous populations and covariates, exclusion of hypertrophic and infiltrative disease, no certainty-of-evidence rating, no sensitivity or subgroup analysis by field strength or technique, and internal inconsistencies (study counts; a native T1 hazard ratio of 2.84 in the text and 2.36 in a table; differing p-values) [3]. The units of the hazard ratios are not explicit; per-unit meaning must be checked in the original (to be sourced).

9. Reporting Framework

ItemContent
TechniqueParameter mapped, sequence and scheme, field strength, shimming, rhythm and rate, contrast timing for post-contrast maps
QualitySource images and quality-control maps checked; artefacts and device limits stated
ROILocation (septal mid-cavity, basal validation, focal ROIs) and size
ResultAbsolute value, Z-score and local reference range; ECV in per cent with haematocrit source (laboratory, same day, or synthetic)
InterpretationDirection of change and its competing explanations; no clinical use of a value without a local range (master 10.3)

10. Technical Pitfalls

PitfallConsequenceSource
Comparing 1.5 T and 3 T values or different sequencesInvalid comparison[1]
Small or border ROIPartial volume, bias[1, 2]
Misregistration, through-plane motionWrong pixel values despite motion-correction[1]
Missing B0 shimming at 3 TOff-resonance error[1]
Haematocrit not recentECV error (24 h limit; same day preferred)[1, 2]
Different processing of native and post-contrast mapsECV error[2]
Heavy iron overload without truncationUnderestimated T2*[2]
Phantom or system not checkedDrift undetected; phantom checks are advised every 3 months and after any system change[1]

11. MRI Technologist Pearls and Quality Checklist

(Reasoning) Write the haematocrit and the time of its sampling into the study record before the examination; run the heart shim before the first map at 3 T; note rhythm and rate on the images; look at the raw source images of each map before leaving the table, so that a repeat is still possible.

CheckPass criterion
ShimCardiac volume-selective shim done (essential at 3 T)
Breath-hold and motionNo breathing artefact; motion-corrected raw images viewed
Map qualityQuality-control map without poor-fit areas in the septum
ECV prerequisitesNative and post-contrast maps with identical processing; haematocrit recent
ReferenceLocal reference range available for this field strength and sequence

12. Advanced Technical Parameters

ParameterValue statedSource
Minimum ROI size20 pixels[1, 2]
Haematocrit timingImmediately before the scan, otherwise within 24 h (same day preferred)[1, 2]
Post-contrast timing10–30 min; 10 min accepted (2013 text: 15 min)[1]
Temperature effectT1 about +1% per 1 °C[1]
Reference range definitionMean ± 2 SD of local normal data; sex and age effect about 0.5 SD[1]
Phantom checksEvery 3 months and after every system change[1]
T2 mappingAt least 3 source images, two-parameter fit; example echo times 0, 25, 50 ms[1]
Amyloid ECV landmarksAbove 55% with transmural LGE; diffuse fibrosis rarely above 40%[1]
T2* cardiacNormal 40 ms; below 20 ms iron overload; below 10 ms heart-failure risk (1.5 T)[2]
T2* at 3 TNot recommended for iron overload[2]

13. Evidence Gaps and Ongoing Debate

No numeric normal values or universal cut-offs are given by the consensus documents [2]. Not established: bolus versus infusion ECV, ECV thresholds in dilated and hypertrophic cardiomyopathy, arrhythmogenic cardiomyopathy, native T1 in renal impairment, reference values for Fabry T1, blood flow as a native T1 confounder [1]. Not read: consensus after 2020 (including a 2025 update if any), vendor-specific reference ranges, original validation studies, outcome studies outside heart failure with preserved ejection fraction (all to be sourced). Prognosis: effect sizes come from one meta-analysis with inconsistencies [3].

14. Evidence-Based References

References are ordered by level of evidence and then by recency. The reading level of each source is stated in its line; a source read only at abstract level supports only what its abstract states.

A. Guidelines / Consensus / Society Recommendations

High
[1] Messroghli DR, Moon JC, Ferreira VM, Grosse-Wortmann L, He T, Kellman P, Mascherbauer J, Nezafat R, Salerno M, Schelbert EB, Taylor AJ, Thompson R, Ugander M, van Heeswijk RB, Friedrich MG. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the Society for Cardiovascular Magnetic Resonance (SCMR) endorsed by the European Association for Cardiovascular Imaging (EACVI). J Cardiovasc Magn Reson. 2017;19:75. DOI: 10.1186/s12968-017-0389-8.
(High — Society consensus statement; full text as rendered by a summarising tool; the last 261 characters (reference list) were not read; numerical values to be checked against the original) Cited in Sections 1, 2, 3, 4, 5, 6, 7, 10, 12, 13.
High
[2] Schulz-Menger J, et al. Standardized image interpretation and post-processing in cardiovascular magnetic resonance – 2020 update. J Cardiovasc Magn Reson. 2020;22:19. DOI: 10.1186/s12968-020-00610-6.
(High — Society consensus document; mapping and T2* sections as rendered by a summarising tool, the ECV formula garbled in the extraction and not quoted from it; numerical values to be checked against the original) Cited in Sections 1, 3, 4, 5, 7, 9, 10, 12, 13.

B. Systematic Reviews / Meta-analyses

Moderate
[3] Golukhova E, Bulaeva N, Alexandrova S, Gromova O, Berdibekov B. Prognostic value of characterizing myocardial tissue by cardiac MRI with T1 mapping in HFpEF patients: a systematic review and meta-analysis. J Clin Med. 2022;11(9):2531. DOI: 10.3390/jcm11092531.
(Moderate — Systematic review and meta-analysis registered in PROSPERO; full text as rendered by a summarising tool; 7 single-centre cohorts; individual quality scores not shown; internal inconsistencies in counts, p-values and one hazard ratio; limited to HFpEF) Cited in Sections 1, 8, 13.

C. Important Prospective / Original Studies

None read for this page.

D. Technical MRI Papers

None read for this page; sequence papers behind the schemes are cited only through [1].

E. Landmark Historical References

None used on this page.