Cine and Real-Time Cardiac MRI — Technical Deep Dive

up to this point verified by human experts

MRIninja Knowledge Base | Deep Dive — Technical Page
Parent page: Heart master (10101)
Version 1.0 — November 2026

Prerequisite: this page assumes Section 4.3 of the Heart master (bSSFP as method of choice, short-axis stack, 6–8 mm slices, optional real-time cine with a preferred temporal resolution of 60 ms or better) and the sequence physics on the bSSFP page, and does not repeat them. It documents what the master does not: how segmented and real-time acquisitions differ in design, what three accelerated cine approaches achieved against breath-hold cine, how irregular rhythm changes the choice, and the contouring conventions that decide whether volumes are comparable. The interdependence protocol for constriction is on the pericardial page and is not repeated.

1. Executive Summary

1.1 Added Value over the Master Protocol

The master states that real-time cine is optional and that absolute volumes are less accurate [2]. This page gives the evidence behind that sentence and its exceptions. In a prospective study of 71 patients with an irregular rhythm at 1.5 T, a single-breath-hold compressed-sensing real-time cine removed arrhythmia-related artefacts that affected a mean of 85.9% of slices with segmented cine, and acquisition fell from 556.7 s to 23.9 s, but right-ventricular parameters were significantly underestimated [5]. In free-breathing accelerated cine, left-ventricular volumes agreed closely with breath-hold cine in healthy volunteers (mean LV end-diastolic volume difference −0.3 mL) [6] and in children (−1.7 ± 4.6 mL, although statistically significant) [4]. The SCMR post-processing document adds the rule that matters most for comparing studies: papillary muscles may be in the blood pool or in the mass, but reference ranges must use the same approach [1].

1.2 Limits of the Page

No source read states a consensus protocol for real-time cine: the SCMR post-processing task force declined to make a statement on analysing real-time images [1]. Each accelerated technique was tested in one centre with a prototype or research sequence [4, 5, 6]. The only study in arrhythmia has no formal Bland–Altman limits and its RV values were not accessible to this review [5]. The 2013 comparison is an abstract [7]. Points without a source are marked (Reasoning) or (to be sourced).

1.3 Evidence Base and Its Reading Level

SourceTypeReading levelUsed for
Schulz-Menger et al., SCMR post-processing 2020 [1]Society consensus documentFull text of the ventricular function section as rendered by a summarising tool (repository copy)Contouring conventions
Kramer et al., SCMR protocols 2020 [2]Society consensus documentCine statements as in the master; summarising toolTargets already in the master
Nayak et al., JMRI 2022 [3]Narrative review (level of evidence 5 as labelled by the journal)Full text as rendered by a summarising toolDefinition, trajectories, acceleration, limits
Steeden et al., JCMR 2018 [4]Prospective validation, 60 childrenArticle page as rendered by a summarising toolSpiral real-time against breath-hold
Longère et al., J Clin Med 2021 [5]Prospective single-centre, 71 patientsFull text as rendered by a summarising toolArrhythmia
Takakado et al., Jpn J Radiol 2022 [6]Prospective, 35 healthy volunteersArticle page as rendered by a summarising toolFree-breathing compressed-sensing cine
Aneja et al., JCMR 2013 [7]Conference abstract, 13 patientsAbstract onlyEarly real-time comparison

2. Segmented Versus Real-Time Acquisition

2.1 Definition and Trade-Offs

Real-time imaging is defined in a review as acquisition that does not use any gating, synchronisation or repetition of the movement; the same review notes that high frame-rate real-time imaging relies on gradient echo and bSSFP, with bSSFP giving higher SNR efficiency and blood–myocardium contrast but sensitivity to off-resonance banding [3]. Spatial resolution, temporal resolution, SNR, artefacts and reconstruction latency trade against each other [3]. Typical acceleration is 2–3 with Cartesian and 3–4 with non-Cartesian trajectories using parallel imaging; compressed sensing allows more but is computationally expensive and its latency often limits clinical adoption [3]. A review cites a recommendation of at least 50 ms (20 frames per second) and 2 mm for functional left-ventricular real-time imaging [3]; this is a secondary citation and differs from the 60 ms preferred in the SCMR protocol document [2].

2.2 Design Differences in Published Comparisons

StudyReference cineReal-time or accelerated cineSource
Aneja 2013 (13 patients, abstract)SSFP 256 × 256, 8 mm, 40 ms, acceleration 2, 11 heartbeats breath-holdSSFP 108 × 160, 10 mm, 60 ms, acceleration 4, free breathing[7]
Steeden 2018 (60 children, 1.5 T)Cartesian bSSFP, 1.5 × 1.5 mm, 29.5 msSpiral bSSFP, tiny golden angle, compressed sensing, about 1.7 × 1.7 mm, about 30 ms[4]
Longère 2021 (71 patients, 1.5 T)Segmented multi-breath-hold bSSFP, 1.3 × 1.3 mm, 41.2 ms, acceleration 2, 20 phasesSingle-shot compressed sensing, 1.6 × 1.6 mm, 49 ms, acceleration 11, 17.0 ± 3.2 phases interpolated to 20, adaptive triggering[5]
Takakado 2022 (35 volunteers, 3 T)Breath-hold bSSFP, 1.9 × 1.9 mm, 40 ms, GRAPPA 3Free-breathing compressed sensing, acceleration 12.5, 43 ms, 12 heartbeats per slice with retrospective non-rigid motion correction[6]

The accelerated techniques are not interchangeable: one is ungated spiral, one is single-breath-hold triggered, one is free-breathing with motion correction [4, 5, 6]. (Reasoning) Results from one cannot be applied to another.

3. Agreement With Breath-Hold Cine

Measure (accelerated minus reference)Steeden 2018 [4]Takakado 2022 [6]Longère 2021 [5]
LV end-diastolic volume−1.7 ± 4.6 mL (significant)−0.3 mL (limits −8.2 to 7.5)−3.6 ± 7.2 mL (p = 0.02)
LV end-systolic volume+1.0 ± 3.5 mL (significant)0.2 mL (limits −5.6 to 5.9)−0.9 ± 6.8 mL (p = 0.51)
LV ejection fraction−1.5 ± 3.0% (significant)−0.3% (limits −3.5 to 3.0)−0.4 ± 1.9% (p = 0.30)
LV mass+3.1 g (significant overestimation)−0.1 g (limits −8.5 to 8.3)Not reported
RV volumesEDV and ESV not significantly differentRVEDV 1.4 mL (limits −15.5 to 18.3)Significantly underestimated; values not accessible
Acquisition of the stack19.8 ± 5.8 s versus 348 ± 79 s135 s versus 198 s (p < 0.001)23.9 ± 7.9 s versus 556.7 ± 145.4 s

Image quality differed by technique. Spiral real-time images in children had more residual aliasing and lower edge definition than breath-hold images (scores 3.8 versus 4.8 and 4.4 versus 4.8) [4]; free-breathing compressed-sensing images had higher SNR and similar edge sharpness [6]; in arrhythmia, edge sharpness was higher with real-time, but no real-time stack was rated excellent, which the authors attribute to interpolation smoothing [5]. The 2013 abstract reports a concordance coefficient of 0.89 for ejection fraction and does not report volumes, bias or limits [7].

4. Irregular Rhythm

The SCMR document warns that in severe arrhythmia end-systolic volumes tend to be overestimated and ejection fraction underestimated, and advises checking for artefacts in irregular heart rates; it mentions real-time cine as possibly beneficial in arrhythmia or limited breath-holding but makes no dedicated statement [1]. A review states that gating fails in about 10% of patients referred for cardiac imaging and that about 10% have difficulty with breath-holding [3]. In the 71-patient study (59.2% atrial fibrillation, 23.9% ventricular hyperexcitability, 16.9% conduction disorders; mean RR coefficient of variation 25.0 ± 9.4%) real-time cine reduced arrhythmia-related artefacts from 85.9 ± 22.7% of slices to none (p < 0.0001) and the artefact score from a median of 3 to 1; non-diagnostic stacks fell from 23 to 10 [5]. Its limits matter for practice: the fixed temporal resolution gives fewer than 20 true frames at higher heart rates (concern above 75 bpm), ten real-time stacks remained non-diagnostic because of failed reconstructions or ECG errors (missed R peaks producing double cycles), the sequence was a prototype, and volume comparisons are difficult because preload and contraction vary beat to beat [5].

SituationOption stated in the sourcesSource
Irregular rhythm, ECG reliable but beat-to-beat variationCheck segmented cine for artefacts; real-time cine may help; interpret volumes with caution[1, 5]
Unreliable ECGPeripheral pulse gating may work for retrospectively gated cine (master)[2]
Limited breath-holdingReal-time or free-breathing accelerated cine[1, 3, 4, 6]
Need for absolute right-ventricular volumesReal-time values may be underestimated; consider segmented acquisition if rhythm allows (Reasoning)[2, 5]

5. Contouring Conventions That Determine Comparability

ElementSCMR 2020 statement
Papillary muscles and trabeculaeMyocardial tissue, ideally in LV mass; commonly kept in the blood pool, which is acceptable; reference ranges must use the same approach. No numerical effect is given [1]
End-diastole and end-systoleLargest and smallest LV blood volume; the same phases for the whole stack, with extra care in dyssynchrony [1]
Basal sliceDescent of the valve plane shifts the basal slice; one alternative defines it as at least 50% of the blood volume surrounded by myocardium; no expert consensus on the method [1]
LV outflow tractIncluded in the LV blood volume [1]
MassEpicardial minus endocardial volume multiplied by 1.05 g/mL [1]
Right ventricleAxial stack identifies the tricuspid plane best; contours up to but not above the pulmonary valve; trabeculae and papillary muscles included in volumes; RV mass not routinely measured [1]
IndexingTo body surface area, all values except ejection fraction [1]
Automated contoursMust be checked by the reader [1]
Wall thicknessBasal short-axis or 3-chamber view; apex from long-axis views [1]
Strain and feature trackingTask force made no dedicated statement [1]

The document gives no numerical reproducibility values [1]; the volunteer study reports intra-observer ICC 0.88–0.99 and inter-observer ICC 0.81–0.99 for its measures [6].

6. Reporting Framework

ItemContent
AcquisitionSegmented breath-hold, real-time or free-breathing accelerated; slices, temporal resolution, rhythm during acquisition
Contouring conventionPapillary muscles in blood pool or in mass; basal slice method; reference range used
Rhythm limitationStatement if arrhythmia may bias end-systolic volume and ejection fraction [1]
Method changeIf a follow-up study used a different acquisition, say so before comparing volumes (Reasoning)

7. Technical Pitfalls

PitfallConsequenceSource
Severe arrhythmia with segmented cineEnd-systolic volume overestimated, ejection fraction underestimated[1]
Different papillary muscle convention against reference rangesNon-comparable volumes and mass[1]
Real-time right-ventricular volumesSignificant underestimation in one study[5]
Spiral real-time cineResidual aliasing and reduced edge definition; LV mass overestimated by 3.1 g[4]
Heart rate above about 75 bpm with fixed temporal resolutionFewer than 20 true frames[5]
Missed R peaks on ECGDouble cycles and non-diagnostic stacks[5]
bSSFP off-resonanceBanding, worse at higher field[3]

8. MRI Technologist Pearls and Quality Checklist

(Reasoning) Look at the ECG trace before the first cine: an irregular trace is the cue to decide the acquisition mode before the stack, not after repeated failed breath-holds. Record the heart rate and whether the sequence is segmented or real-time in the exam notes, because the reader needs it to interpret volumes.

CheckPass criterion
Acquisition modeChosen and recorded; real-time when rhythm or breath-holding prevents segmented cine
CoverageWhole LV in short axis, plus long-axis views (master 4.3)
ArtefactsArrhythmia, banding and aliasing checked on cine loops before leaving the table [1, 3]
Temporal resolutionNoted; compared with 60 ms preferred for real-time (master)
ConventionContouring convention documented for the reader [1]

9. Advanced Technical Parameters

ParameterValue statedSource
Master targetsSlices 6–8 mm; temporal resolution 45 ms or better (segmented); real-time 60 ms or better preferred[2]
Recommendation cited by a reviewAt least 50 ms (20 frames per second) and 2 mm for functional LV real-time imaging[3]
Typical acceleration2–3 Cartesian; 3–4 non-Cartesian (parallel imaging)[3]
Interventional real-time rate2D Cartesian bSSFP at 5–10 frames per second[3]
Free-breathing compressed-sensing cineAcceleration 12.5; 12 heartbeats per slice, five least-motion beats averaged; reconstruction about five minutes[6]
Spiral real-time cineAbout 20 s for a stack of about 13 slices; reconstruction under 30 s on GPU[4]
Single-breath-hold real-time cineAcceleration 11; 49 ms; 17.0 ± 3.2 true phases[5]
Myocardial density for mass1.05 g/mL[1]

10. Evidence Gaps and Ongoing Debate

No consensus was found on the acquisition and analysis of real-time cine [1]. Not read: studies with outcome-relevant endpoints, multicentre validation, vendor-product (not prototype) sequences, reference ranges for real-time volumes, deep-learning reconstruction and automated contouring performance, three-dimensional whole-heart cine, and a numerical effect of the papillary muscle convention (all to be sourced). Population limits: healthy male volunteers only [6]; children from 7 years [4]; arrhythmia in one centre [5]. Debate: whether the 50 ms and 60 ms thresholds apply across heart rates; whether the basal slice should be defined by atrial appearance or by the 50% rule [1]. Source note: the arrhythmia article lists a truncated author list on the page read; the first listed author is Longère [5].

11. Evidence-Based References

References are ordered by level of evidence and then by recency. The reading level of each source is stated in its line; a source read only at abstract level supports only what its abstract states.

A. Guidelines / Consensus / Society Recommendations

High
[1] Schulz-Menger J, Bluemke DA, Bremerich J, et al. Standardized image interpretation and post-processing in cardiovascular magnetic resonance – 2020 update: Society for Cardiovascular Magnetic Resonance (SCMR) Board of Trustees Task Force on Standardized Post-Processing. J Cardiovasc Magn Reson. 2020;22:19. DOI: 10.1186/s12968-020-00610-6.
(High — Society consensus document; ventricular function section read in full as rendered by a summarising tool from a repository copy; no numerical effect of the papillary muscle convention or reproducibility values given; no statement on real-time analysis) Cited in Sections 1, 4, 5, 6, 7, 8, 9, 10.
High
[2] Kramer CM, Barkhausen J, Bucciarelli-Ducci C, Flamm SD, Kim RJ, Nagel E. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson. 2020;22:17. DOI: 10.1186/s12968-020-00607-1.
(High — Society consensus document; cine statements as rendered by a summarising tool; the real-time and gating statements are those already used in the master) Cited in Sections 1, 2, 4, 7, 9.

B. Systematic Reviews / Meta-analyses

Moderate
[3] Nayak KS, Lim Y, Campbell-Washburn AE, Steeden J. Real-time magnetic resonance imaging. J Magn Reson Imaging. 2022;55:81–99. DOI: 10.1002/jmri.27411.
(Moderate — Narrative review, not a systematic review; level of evidence 5 as labelled by the journal; full text as rendered by a summarising tool; general real-time MRI, not cardiac-specific, with mostly congenital and interventional cardiac examples; does not discuss ventricular interdependence) Cited in Sections 1, 2, 4, 7, 8, 9.

C. Important Prospective / Original Studies

Moderate
[4] Steeden JA, Kowalik GT, Tann O, Hughes M, Mortensen KH, Muthurangu V. Real-time assessment of right and left ventricular volumes and function in children using high spatiotemporal resolution spiral bSSFP with compressed sensing. J Cardiovasc Magn Reson. 2018;20:79. DOI: 10.1186/s12968-018-0500-9.
(Moderate — Prospective validation, 60 children with congenital heart disease, 1.5 T; article page as rendered by a summarising tool; author list truncated on the page read; minimum age 7 years; RV mass not assessed) Cited in Sections 1, 2, 3, 5, 7, 9.
Moderate
[5] Longère B, Allard PE, Gkizas CV, et al. Compressed sensing real-time cine reduces CMR arrhythmia-related artifacts. J Clin Med. 2021;10(15):3274. DOI: 10.3390/jcm10153274.
(Moderate — Prospective single-centre paired study, 71 patients with irregular rhythm, 1.5 T; full text as rendered by a summarising tool; non-blinded reading; prototype sequence; no formal Bland–Altman limits; RV values not visible in the text read; three authors employed by the scanner vendor) Cited in Sections 1, 2, 3, 4, 7, 9, 10.
Moderate
[6] Takakado M, Kido T, Ogawa R, et al. Free-breathing cardiovascular cine magnetic resonance imaging using compressed-sensing and retrospective motion correction: accurate assessment of biventricular volume at 3T. Jpn J Radiol. 2023;41(2):142–152 (published online 13 October 2022). DOI: 10.1007/s11604-022-01344-4.
(Moderate — Prospective within-subject study, 35 healthy male volunteers, 3 T; article page as rendered by a summarising tool; short axis only; no patients or wall motion; prototype sequence; two authors employed by the vendor) Cited in Sections 1, 2, 3, 5, 9, 10.
Limited
[7] Aneja A, Ono C, Scandling D, Jin N, Hinton AM, Pennell M, Raman SV, Simonetti OP. Left ventricular function, aortic velocity, and late gadolinium enhancement assessed by real-time and single shot CMR is comparable to breath-held segmented imaging: a prospective study. J Cardiovasc Magn Reson. 2013;15(Suppl 1):O51. DOI: 10.1186/1532-429X-15-S1-O51.
(Limited — Conference abstract, 13 patients in regular rhythm; abstract only; no volumes, bias or limits of agreement; no limitations stated; journal supplement inferred from the DOI) Cited in Sections 1, 2, 3.

D. Technical MRI Papers

None used on this page.

E. Landmark Historical References

None used on this page.

Recent PubMed search for this protocol

Last updated: November 2026