Cine and Real-Time Cardiac MRI — Technical Deep Dive
Linked parent page: MRI Heart for Pericardial Disease
MRIninja Knowledge Base | Deep Dive — Technical Page
Parent page: Heart master (10101)
Version 1.0 — November 2026
Prerequisite: this page assumes Section 4.3 of the Heart master (bSSFP as method of choice, short-axis stack, 6–8 mm slices, optional real-time cine with a preferred temporal resolution of 60 ms or better) and the sequence physics on the bSSFP page, and does not repeat them. It documents what the master does not: how segmented and real-time acquisitions differ in design, what three accelerated cine approaches achieved against breath-hold cine, how irregular rhythm changes the choice, and the contouring conventions that decide whether volumes are comparable. The interdependence protocol for constriction is on the pericardial page and is not repeated.
1. Executive Summary
1.1 Added Value over the Master Protocol
The master states that real-time cine is optional and that absolute volumes are less accurate [2]. This page gives the evidence behind that sentence and its exceptions. In a prospective study of 71 patients with an irregular rhythm at 1.5 T, a single-breath-hold compressed-sensing real-time cine removed arrhythmia-related artefacts that affected a mean of 85.9% of slices with segmented cine, and acquisition fell from 556.7 s to 23.9 s, but right-ventricular parameters were significantly underestimated [5]. In free-breathing accelerated cine, left-ventricular volumes agreed closely with breath-hold cine in healthy volunteers (mean LV end-diastolic volume difference −0.3 mL) [6] and in children (−1.7 ± 4.6 mL, although statistically significant) [4]. The SCMR post-processing document adds the rule that matters most for comparing studies: papillary muscles may be in the blood pool or in the mass, but reference ranges must use the same approach [1].
1.2 Limits of the Page
No source read states a consensus protocol for real-time cine: the SCMR post-processing task force declined to make a statement on analysing real-time images [1]. Each accelerated technique was tested in one centre with a prototype or research sequence [4, 5, 6]. The only study in arrhythmia has no formal Bland–Altman limits and its RV values were not accessible to this review [5]. The 2013 comparison is an abstract [7]. Points without a source are marked (Reasoning) or (to be sourced).
1.3 Evidence Base and Its Reading Level
| Source | Type | Reading level | Used for |
|---|---|---|---|
| Schulz-Menger et al., SCMR post-processing 2020 [1] | Society consensus document | Full text of the ventricular function section as rendered by a summarising tool (repository copy) | Contouring conventions |
| Kramer et al., SCMR protocols 2020 [2] | Society consensus document | Cine statements as in the master; summarising tool | Targets already in the master |
| Nayak et al., JMRI 2022 [3] | Narrative review (level of evidence 5 as labelled by the journal) | Full text as rendered by a summarising tool | Definition, trajectories, acceleration, limits |
| Steeden et al., JCMR 2018 [4] | Prospective validation, 60 children | Article page as rendered by a summarising tool | Spiral real-time against breath-hold |
| Longère et al., J Clin Med 2021 [5] | Prospective single-centre, 71 patients | Full text as rendered by a summarising tool | Arrhythmia |
| Takakado et al., Jpn J Radiol 2022 [6] | Prospective, 35 healthy volunteers | Article page as rendered by a summarising tool | Free-breathing compressed-sensing cine |
| Aneja et al., JCMR 2013 [7] | Conference abstract, 13 patients | Abstract only | Early real-time comparison |
2. Segmented Versus Real-Time Acquisition
2.1 Definition and Trade-Offs
Real-time imaging is defined in a review as acquisition that does not use any gating, synchronisation or repetition of the movement; the same review notes that high frame-rate real-time imaging relies on gradient echo and bSSFP, with bSSFP giving higher SNR efficiency and blood–myocardium contrast but sensitivity to off-resonance banding [3]. Spatial resolution, temporal resolution, SNR, artefacts and reconstruction latency trade against each other [3]. Typical acceleration is 2–3 with Cartesian and 3–4 with non-Cartesian trajectories using parallel imaging; compressed sensing allows more but is computationally expensive and its latency often limits clinical adoption [3]. A review cites a recommendation of at least 50 ms (20 frames per second) and 2 mm for functional left-ventricular real-time imaging [3]; this is a secondary citation and differs from the 60 ms preferred in the SCMR protocol document [2].
2.2 Design Differences in Published Comparisons
| Study | Reference cine | Real-time or accelerated cine | Source |
|---|---|---|---|
| Aneja 2013 (13 patients, abstract) | SSFP 256 × 256, 8 mm, 40 ms, acceleration 2, 11 heartbeats breath-hold | SSFP 108 × 160, 10 mm, 60 ms, acceleration 4, free breathing | [7] |
| Steeden 2018 (60 children, 1.5 T) | Cartesian bSSFP, 1.5 × 1.5 mm, 29.5 ms | Spiral bSSFP, tiny golden angle, compressed sensing, about 1.7 × 1.7 mm, about 30 ms | [4] |
| Longère 2021 (71 patients, 1.5 T) | Segmented multi-breath-hold bSSFP, 1.3 × 1.3 mm, 41.2 ms, acceleration 2, 20 phases | Single-shot compressed sensing, 1.6 × 1.6 mm, 49 ms, acceleration 11, 17.0 ± 3.2 phases interpolated to 20, adaptive triggering | [5] |
| Takakado 2022 (35 volunteers, 3 T) | Breath-hold bSSFP, 1.9 × 1.9 mm, 40 ms, GRAPPA 3 | Free-breathing compressed sensing, acceleration 12.5, 43 ms, 12 heartbeats per slice with retrospective non-rigid motion correction | [6] |
The accelerated techniques are not interchangeable: one is ungated spiral, one is single-breath-hold triggered, one is free-breathing with motion correction [4, 5, 6]. (Reasoning) Results from one cannot be applied to another.
3. Agreement With Breath-Hold Cine
| Measure (accelerated minus reference) | Steeden 2018 [4] | Takakado 2022 [6] | Longère 2021 [5] |
|---|---|---|---|
| LV end-diastolic volume | −1.7 ± 4.6 mL (significant) | −0.3 mL (limits −8.2 to 7.5) | −3.6 ± 7.2 mL (p = 0.02) |
| LV end-systolic volume | +1.0 ± 3.5 mL (significant) | 0.2 mL (limits −5.6 to 5.9) | −0.9 ± 6.8 mL (p = 0.51) |
| LV ejection fraction | −1.5 ± 3.0% (significant) | −0.3% (limits −3.5 to 3.0) | −0.4 ± 1.9% (p = 0.30) |
| LV mass | +3.1 g (significant overestimation) | −0.1 g (limits −8.5 to 8.3) | Not reported |
| RV volumes | EDV and ESV not significantly different | RVEDV 1.4 mL (limits −15.5 to 18.3) | Significantly underestimated; values not accessible |
| Acquisition of the stack | 19.8 ± 5.8 s versus 348 ± 79 s | 135 s versus 198 s (p < 0.001) | 23.9 ± 7.9 s versus 556.7 ± 145.4 s |
Image quality differed by technique. Spiral real-time images in children had more residual aliasing and lower edge definition than breath-hold images (scores 3.8 versus 4.8 and 4.4 versus 4.8) [4]; free-breathing compressed-sensing images had higher SNR and similar edge sharpness [6]; in arrhythmia, edge sharpness was higher with real-time, but no real-time stack was rated excellent, which the authors attribute to interpolation smoothing [5]. The 2013 abstract reports a concordance coefficient of 0.89 for ejection fraction and does not report volumes, bias or limits [7].
4. Irregular Rhythm
The SCMR document warns that in severe arrhythmia end-systolic volumes tend to be overestimated and ejection fraction underestimated, and advises checking for artefacts in irregular heart rates; it mentions real-time cine as possibly beneficial in arrhythmia or limited breath-holding but makes no dedicated statement [1]. A review states that gating fails in about 10% of patients referred for cardiac imaging and that about 10% have difficulty with breath-holding [3]. In the 71-patient study (59.2% atrial fibrillation, 23.9% ventricular hyperexcitability, 16.9% conduction disorders; mean RR coefficient of variation 25.0 ± 9.4%) real-time cine reduced arrhythmia-related artefacts from 85.9 ± 22.7% of slices to none (p < 0.0001) and the artefact score from a median of 3 to 1; non-diagnostic stacks fell from 23 to 10 [5]. Its limits matter for practice: the fixed temporal resolution gives fewer than 20 true frames at higher heart rates (concern above 75 bpm), ten real-time stacks remained non-diagnostic because of failed reconstructions or ECG errors (missed R peaks producing double cycles), the sequence was a prototype, and volume comparisons are difficult because preload and contraction vary beat to beat [5].
| Situation | Option stated in the sources | Source |
|---|---|---|
| Irregular rhythm, ECG reliable but beat-to-beat variation | Check segmented cine for artefacts; real-time cine may help; interpret volumes with caution | [1, 5] |
| Unreliable ECG | Peripheral pulse gating may work for retrospectively gated cine (master) | [2] |
| Limited breath-holding | Real-time or free-breathing accelerated cine | [1, 3, 4, 6] |
| Need for absolute right-ventricular volumes | Real-time values may be underestimated; consider segmented acquisition if rhythm allows (Reasoning) | [2, 5] |
5. Contouring Conventions That Determine Comparability
| Element | SCMR 2020 statement |
|---|---|
| Papillary muscles and trabeculae | Myocardial tissue, ideally in LV mass; commonly kept in the blood pool, which is acceptable; reference ranges must use the same approach. No numerical effect is given [1] |
| End-diastole and end-systole | Largest and smallest LV blood volume; the same phases for the whole stack, with extra care in dyssynchrony [1] |
| Basal slice | Descent of the valve plane shifts the basal slice; one alternative defines it as at least 50% of the blood volume surrounded by myocardium; no expert consensus on the method [1] |
| LV outflow tract | Included in the LV blood volume [1] |
| Mass | Epicardial minus endocardial volume multiplied by 1.05 g/mL [1] |
| Right ventricle | Axial stack identifies the tricuspid plane best; contours up to but not above the pulmonary valve; trabeculae and papillary muscles included in volumes; RV mass not routinely measured [1] |
| Indexing | To body surface area, all values except ejection fraction [1] |
| Automated contours | Must be checked by the reader [1] |
| Wall thickness | Basal short-axis or 3-chamber view; apex from long-axis views [1] |
| Strain and feature tracking | Task force made no dedicated statement [1] |
The document gives no numerical reproducibility values [1]; the volunteer study reports intra-observer ICC 0.88–0.99 and inter-observer ICC 0.81–0.99 for its measures [6].
6. Reporting Framework
| Item | Content |
|---|---|
| Acquisition | Segmented breath-hold, real-time or free-breathing accelerated; slices, temporal resolution, rhythm during acquisition |
| Contouring convention | Papillary muscles in blood pool or in mass; basal slice method; reference range used |
| Rhythm limitation | Statement if arrhythmia may bias end-systolic volume and ejection fraction [1] |
| Method change | If a follow-up study used a different acquisition, say so before comparing volumes (Reasoning) |
7. Technical Pitfalls
| Pitfall | Consequence | Source |
|---|---|---|
| Severe arrhythmia with segmented cine | End-systolic volume overestimated, ejection fraction underestimated | [1] |
| Different papillary muscle convention against reference ranges | Non-comparable volumes and mass | [1] |
| Real-time right-ventricular volumes | Significant underestimation in one study | [5] |
| Spiral real-time cine | Residual aliasing and reduced edge definition; LV mass overestimated by 3.1 g | [4] |
| Heart rate above about 75 bpm with fixed temporal resolution | Fewer than 20 true frames | [5] |
| Missed R peaks on ECG | Double cycles and non-diagnostic stacks | [5] |
| bSSFP off-resonance | Banding, worse at higher field | [3] |
8. MRI Technologist Pearls and Quality Checklist
(Reasoning) Look at the ECG trace before the first cine: an irregular trace is the cue to decide the acquisition mode before the stack, not after repeated failed breath-holds. Record the heart rate and whether the sequence is segmented or real-time in the exam notes, because the reader needs it to interpret volumes.
| Check | Pass criterion |
|---|---|
| Acquisition mode | Chosen and recorded; real-time when rhythm or breath-holding prevents segmented cine |
| Coverage | Whole LV in short axis, plus long-axis views (master 4.3) |
| Artefacts | Arrhythmia, banding and aliasing checked on cine loops before leaving the table [1, 3] |
| Temporal resolution | Noted; compared with 60 ms preferred for real-time (master) |
| Convention | Contouring convention documented for the reader [1] |
9. Advanced Technical Parameters
| Parameter | Value stated | Source |
|---|---|---|
| Master targets | Slices 6–8 mm; temporal resolution 45 ms or better (segmented); real-time 60 ms or better preferred | [2] |
| Recommendation cited by a review | At least 50 ms (20 frames per second) and 2 mm for functional LV real-time imaging | [3] |
| Typical acceleration | 2–3 Cartesian; 3–4 non-Cartesian (parallel imaging) | [3] |
| Interventional real-time rate | 2D Cartesian bSSFP at 5–10 frames per second | [3] |
| Free-breathing compressed-sensing cine | Acceleration 12.5; 12 heartbeats per slice, five least-motion beats averaged; reconstruction about five minutes | [6] |
| Spiral real-time cine | About 20 s for a stack of about 13 slices; reconstruction under 30 s on GPU | [4] |
| Single-breath-hold real-time cine | Acceleration 11; 49 ms; 17.0 ± 3.2 true phases | [5] |
| Myocardial density for mass | 1.05 g/mL | [1] |
10. Evidence Gaps and Ongoing Debate
No consensus was found on the acquisition and analysis of real-time cine [1]. Not read: studies with outcome-relevant endpoints, multicentre validation, vendor-product (not prototype) sequences, reference ranges for real-time volumes, deep-learning reconstruction and automated contouring performance, three-dimensional whole-heart cine, and a numerical effect of the papillary muscle convention (all to be sourced). Population limits: healthy male volunteers only [6]; children from 7 years [4]; arrhythmia in one centre [5]. Debate: whether the 50 ms and 60 ms thresholds apply across heart rates; whether the basal slice should be defined by atrial appearance or by the 50% rule [1]. Source note: the arrhythmia article lists a truncated author list on the page read; the first listed author is Longère [5].
11. Evidence-Based References
References are ordered by level of evidence and then by recency. The reading level of each source is stated in its line; a source read only at abstract level supports only what its abstract states.
A. Guidelines / Consensus / Society Recommendations
B. Systematic Reviews / Meta-analyses
C. Important Prospective / Original Studies
D. Technical MRI Papers
None used on this page.
E. Landmark Historical References
None used on this page.