MRI Thoracic Vessels – Generic Standard Protocol
MRIninja Knowledge Base | Master / General Protocol Page
Related pages: Phase-contrast MRI · Triggered non-contrast MRA · Inflow IR bSSFP MRA · Contrast media · GBCA safety · Cardiac implantable devices · Heart master (technique and safety pages, not parents)
Scope: adult MR angiography (MRA) and morphological MRI of the thoracic aorta and, in a separate section, the pulmonary veins. Congenital, paediatric, coronary, pulmonary artery, supra-aortic and peripheral territories are outside this page. Indication-specific pages are not yet published.
Version 1.0 — November 2026
1. Executive Summary
1.1 What This Page Provides
The Society for Cardiovascular Magnetic Resonance (SCMR) 2020 protocol document gives the only structured generic protocol for thoracic aortic MRA that was accessible for this page [4]. It is a short sequence of modules: three-plane localisers, a single-shot transaxial survey, balanced steady-state free precession (bSSFP) cine along the aorta, transaxial T1-weighted imaging of the wall, a three-dimensional contrast-enhanced MRA (CE-MRA) and, optionally, post-contrast fat-suppressed T1-weighted imaging for aortitis [4]. This page sets those modules out, adds the parameters the document states for the CE-MRA and for the pulmonary veins, and then adds what two further sources contribute: the 2023 SCMR consensus on four-dimensional (4D) flow [2] and two original studies of non-contrast MRA of the thoracic aorta [5, 6]. Guideline statements on thoracic aortic disease [1, 3] are reported only as far as they were accessible, which was through summary pages and not the full guideline texts (Section 9).
Flow physics, velocity encoding and non-contrast MRA mechanisms are not repeated here: see Phase-contrast MRI, Triggered non-contrast MRA and Inflow IR bSSFP MRA. Gadolinium class, dosing in general and safety are on Contrast media and GBCA safety; only the aortic dose and timing stated by the SCMR document are given here.
1.2 Limits of the Evidence
The generic protocol is a consensus document and not a trial-based recommendation [4]. The full texts of the 2022 ACC/AHA aortic disease guideline [3] and the 2024 ESC peripheral arterial and aortic diseases guideline [1] were not accessible, so no statement about MRI use, class of recommendation or surveillance interval is made here beyond the fragments printed on summary pages. The SCMR document does not state slice thickness for the aortic 3D acquisition, the breath-hold duration, a dedicated pre-contrast black-blood module, or any protocol specific to dissection, aneurysm, wall disease or the pulmonary arteries [4]. The two original studies of non-contrast MRA are small (n = 25 and n = 30) and single-centre or retrospective [5, 6].
1.3 Reading Level of Sources
| Source | Role on this page | Reading level |
|---|---|---|
| [4] Kramer, SCMR 2020 | Generic aortic and pulmonary vein protocol | Angiography section as rendered by a summarising tool |
| [2] Bissell, SCMR 2023 | 4D flow acquisition and processing | First 100,000 of 124,113 characters, via a summarising tool |
| [3] Isselbacher, ACC/AHA 2022 | Measurement, reporting and growth thresholds | ACC “Ten Points to Remember” page only; full guideline not read |
| [1] Mazzolai, ESC 2024 | Imaging and surveillance statements | ACC key-points summary only; full guideline not read |
| [5] Zeilinger 2024 | Non-contrast 3D MRA versus CT angiography | Article as rendered by a summarising tool |
| [6] Snel 2020 | ECG-triggered bSSFP versus CE-MRA dimensions | Abstract only |
2. Clinical Question and Modality Choice
The summaries of the ESC 2024 guideline report, as Class I, Level C, that thoracic aortic imaging is performed initially with CT or MRI, and that imaging is also indicated when CT/MRI and echocardiography results disagree and when the aortic diameter exceeds 45 mm; the exact scope of the printed phrase “CT or MRI of the thoracic aorta initially” could not be checked against the guideline text [1]. Neither accessible ACC page states a preference between CT and MRI, a position on MRI for surveillance, or a statement on gadolinium [3]. Both guidelines are therefore not used here to rank the modalities; the comparison is limited to the one original study identified.
In 25 patients, a free-breathing, ECG-triggered, navigator-gated 3D non-contrast MRA at 3 T with Dixon water-fat separation gave aortic diameters and areas at 11 levels that did not differ significantly from high-pitch ECG-triggered CT angiography (p = 0.08–0.94), with most differences below 0.5 mm and three outliers of 1.5–3.0 mm at the sinotubular junction and the diaphragm level [5]. The authors state that MRA is unsuitable for emergencies because of its long scan time and limited patient monitoring [5]. Radiation and iodinated contrast were avoided with MRA; the CT dose-length product was 299.3 mGy·cm for the arterial phase alone and 519.1 mGy·cm for arterial plus venous phases [5]. Pulmonary veins were rated better on CT (p < 0.0001) and the venae cavae better on MRA (p < 0.05) [5].
(Reasoning) Where the question is urgent, the patient is unstable or cannot be monitored adequately, a multi-minute gated MRA is a poor fit; this follows from the statement in [5] and is not a guideline recommendation.
3. Generic Protocol Modules
Patient preparation, ECG set-up, coil selection, breath-hold coaching, and implant screening are not specific to the aorta: see Heart master for gating and breath-hold management and Cardiac implantable devices for cardiac implantable devices.
| Step | Module as stated by the SCMR document | Purpose |
|---|---|---|
| 1 | Three-orientation localisers | Planning |
| 2 | Single-shot black-blood or bSSFP transaxial survey of the thorax in one breath-hold | Anatomy overview, extra-aortic findings |
| 3 | bSSFP cine parallel to the aortic midline | Aortic morphology in motion (Reasoning: also the plane for valve-related questions) |
| 4 | Transaxial T1-weighted fast spin-echo or spoiled gradient echo through the aorta | Intramural haematoma, dissection |
| 5 | Aortic valve imaging per the valvular protocol | When the valve is involved: see Valve and shunt page |
| 6 | 3D CE-MRA (Section 4) | Lumen and branch anatomy |
| 7 | Optional post-contrast fat-suppressed T1-weighted imaging | Aortitis |
All rows from [4]. The SCMR text as rendered does not specify the slice thickness, the cine slice number or a dedicated pre-contrast black-blood sequence for the aorta (to be sourced).
4. Contrast-Enhanced MRA
4.1 Timing
Three options are listed: a test bolus of 2 mL of gadolinium-based contrast agent (GBCA) followed by 20 mL of saline; bolus triggering; or rapid multiphase 3D acquisition without timing [4]. Selection between them is not guided by evidence in the source (to be sourced).
4.2 Injection and Acquisition
| Item | Value stated | Source |
|---|---|---|
| GBCA dose | 0.1–0.2 mmol/kg | [4] |
| Injection rate | 2–3 mL/s | [4] |
| Saline flush | 20 mL at 2–3 mL/s | [4] |
| Spatial resolution | At least 1–1.5 mm | [4] |
| Acceleration and gating | Parallel acquisition and ECG gating if available | [4] |
| Acquisitions | At least two after injection | [4] |
Which GBCA, renal function thresholds and pregnancy are outside this page: see Contrast media and GBCA safety. A mid-axis maximum intensity projection can hide wall thickening and thrombus; see MIP for the failure modes of MIP display.
5. Non-Contrast and ECG-Gated Alternatives
The SCMR angiography section, as rendered, describes non-contrast options only in the peripheral section, and gives no non-contrast thoracic aortic protocol [4]. The mechanisms of triggered and inflow-based non-contrast MRA are on Triggered non-contrast MRA and Inflow IR bSSFP MRA; only the thoracic aortic data are given here.
| Study | Technique | Design and N | Findings relevant to the aorta |
|---|---|---|---|
| Zeilinger 2024 [5] | 3 T, free-breathing, 3D isotropic 1.2 mm, ECG-triggered end-diastolic, liver-dome navigator (±4 mm, end-expiration), compressed sensing (R = 11), two-point Dixon | Prospective comparison with high-pitch CT angiography, n = 25 of 83 contacted; interval 7 ± 5 months | No significant diameter difference at 11 levels; diagnostic image quality in all 25; navigator acceptance 42.7 ± 12.4%, net acquisition 8 min 4 s ± 2 min 52 s; one Stanford B dissection showed both lumina hyperintense (flow-insensitive sequence); aortic implants excluded |
| Snel 2020 [6] | ECG-triggered bSSFP MRA at 1.5 T | Retrospective comparison with non-ECG-triggered CE-MRA, n = 30, nine landmarks, two observers | r = 0.99; mean difference 0.1 mm; limits of agreement −1.9 to +1.9 mm (abstract level) |
The aneurysm subgroup (seven patients) in [5] was not analysed separately, and aortitis was not addressed; no study of non-contrast MRA in aortitis, intramural haematoma or the post-repair aorta was read (to be sourced). Both studies are small, and [6] is a retrospective single-centre study read at abstract level only. Agreement of diameters does not establish equivalence for detecting dissection or wall disease.
6. Flow Imaging in the Aorta
Technique, velocity encoding and phase-offset errors are on Phase-contrast MRI. The 2023 SCMR 4D flow consensus has no dedicated thoracic aorta section and does not cover pulse wave velocity; its general recommendations apply to the aorta [2].
| Parameter | Recommendation in [2] |
|---|---|
| Role in aortic aneurysm | Adjunctive; complete the diagnostic sequences first |
| Scan time | 5–10 minutes |
| Spatial resolution | Isotropic, at least 6 voxels across the vessel; adult aim 2 mm³, not above 2.5 mm³ |
| Temporal resolution | Aim 30 ms, not above 50 ms |
| VENC | 150 cm/s for aorta and pulmonary artery; false lumen in dissection 50–150 cm/s; 250 cm/s as a starting value when stenosis is suspected without prior 2D data; no more than 25% above the maximum expected velocity |
| Coverage | At least the valves and the aortic sinuses |
| Gating | Retrospective, full cardiac cycle, preferred |
| Respiratory navigation | Optional; self-gating recommended once readily available |
| Contrast | Useful but not essential; consider in dissection and other challenging cases |
| Flip angle | About 7° without contrast; 12–25° after contrast; not above 25° |
Processing steps are background phase-offset correction, anti-aliasing, segmentation with planes orthogonal to the vessel, optional visualisation, and quantification with an internal consistency check [2]. Flow volumes and peak velocities are the core clinical parameters; wall shear stress, kinetic energy, vorticity and flow eccentricity are described as mostly research parameters needing separate validation [2]. Quality assurance suggested for initial validation is comparison with 2D flow in the ascending and descending aorta, mass conservation (aortic against pulmonary flow, adding about 5% for coronary flow when measured above the sinuses; superior vena cava plus descending aortic flow equal to ascending aortic flow), and 2–4 ascending aortic planes whose flows should differ by less than 5% [2]. Flow measurements can be unreliable in ascending aortic aneurysm or turbulence, and compressed sensing can underestimate peak velocity and flow [2].
7. Pulmonary Veins
The SCMR document gives a breath-held coronal 3D CE-MRA with 1–2 mm slices, in-plane resolution of 1–1.5 mm, 60–80 slices, and two to three acquisitions (pre-contrast for subtraction, first pass, and optionally a later phase), with 0.1–0.2 mmol/kg at 2–3 mL/s [4]. The indications, the thresholds for stenosis and the post-ablation surveillance scheme are not stated in the rendered text (to be sourced). In [5], CT was rated better than non-contrast MRA for pulmonary veins [5].
8. Measurement and Reporting
The ACC/AHA 2022 summary states that CT, MRI and echocardiography should use standardised measurement techniques at reproducible anatomical landmarks, and that CT and MRI reports should include maximal aortic diameters, wall changes, luminal stenosis, findings suggestive of acute aortic syndrome, branch vessel involvement, prior repair, comparison with prior examinations and disease classification [3]. For patients smaller or taller than average, intervention thresholds may use size indexed to body surface area, to height squared, or cross-sectional area indexed to height [3]. Whether diameters are measured inner-to-inner or outer-to-outer, and in which plane, is not stated in the pages read (to be sourced).
Growth thresholds quoted for rapid growth
| Population | Rapid growth defined as |
|---|---|
| Non-syndromic heritable or sporadic thoracic aortic aneurysm | ≥0.5 cm/year, or ≥0.3 cm/year for 2 consecutive years |
| Marfan syndrome | ≥0.3 cm/year |
| Bicuspid aortic valve | ≥0.3 cm/year |
From [3] (summary page). Measurement-method sensitivity matters: the 2024 study of [5] reports that the guideline growth threshold is 5 mm, and that most MRA-CT differences were under 0.5 mm with outliers up to 3.0 mm.
Proposed report items (Reasoning): modality and sequence used, whether gated, maximum diameters at stated landmarks (annulus, sinuses, sinotubular junction, mid-ascending, arch, descending, diaphragm) with the plane used, comparison with prior studies measured with the same method, wall findings, branch involvement, extra-aortic findings from the survey.
9. Guideline Statements and Surveillance
These statements come from summary pages and are secondary sources.
| Statement | Class / level as printed | Source |
|---|---|---|
| CT or MRI of the thoracic aorta initially; imaging when CT/MRI and echocardiography disagree, and when diameter exceeds 45 mm | I / C | [1] |
| Bicuspid aortic valve with aorta >40 mm: serial echocardiography after 1 year, then every 2–3 years | I / C | [1] |
| Surgery for bicuspid valve or ascending aneurysm to be considered at ≥45 mm; ascending aneurysm surgery at ≥55 mm | IIa / C; I / B | [1] |
| Ionising radiation kept as low as feasible while maintaining diagnostic quality | Not stated | [3] |
| Genetic testing for syndromic features, family history or onset before age 60; family imaging for relatives | Not stated | [3] |
The sources read give no MRI-specific surveillance interval and no statement on MRI in young or syndromic patients (to be sourced). The ACC/AHA summaries note that surgical thresholds for sporadic root and ascending aneurysms were lowered from 5.5 cm to 5.0 cm in selected patients at centres with multidisciplinary aortic teams [3]. Controversy: the two guidelines differ in structure and recommendations on imaging; a published comparison exists but could not be read (to be sourced).
10. Technical Pitfalls
- Contrast timing and breath-hold: CE-MRA quality depends on bolus timing and the patient’s ability to hold the breath [5].
- Motion on gated non-contrast MRA: motion was the main reason for a one-point image quality downgrade in 4 of 25 examinations [5]; navigator efficiency (42.7%) makes acquisition time unpredictable [5].
- bSSFP sensitivity: off-resonance effects and steady-state disruption are noted for bSSFP; see bSSFP and MRI artefacts [5].
- Flow-insensitive lumen signal: on flow-independent MRA both lumina of a dissection appear bright [5], so false lumen flow direction is not shown.
- Single VENC and acceleration in 4D flow: a single VENC limits dynamic range; compressed sensing can underestimate peak velocity [2].
- Ascending aneurysm and turbulence: unreliable flow; alternative net forward flow from the superior vena cava plus descending aorta [2].
- Implants: aortic implants were excluded from [5]; artefact behaviour of stents and grafts is not covered (to be sourced).
- Display: MIP can mask wall findings; see MIP.
11. MRI Technologist Pearls and Quality Checklist
(Reasoning) The following are practical points derived from the modules above and are not trial-based.
- Acquire the transaxial survey first; it shows extra-aortic findings and the extent of disease before the contrast is planned.
- Choose the timing method before the injector is connected, and prepare the saline flush (20 mL) with the contrast line.
- Rehearse the breath-hold at the end of expiration or inspiration used for the acquisition and keep it the same in all acquisitions.
- Check the ECG trace for gating errors before the gated acquisitions; see Heart master.
- Record the table time, the contrast dose in mL and mmol/kg, the delay or trigger used, and whether the acquisition was gated.
| Checklist item | Done |
|---|---|
| Three-plane localisers and transaxial survey acquired | ☐ |
| Cine along the aortic midline acquired | ☐ |
| Transaxial T1-weighted wall imaging acquired when haematoma or dissection is possible | ☐ |
| Timing method, dose and saline flush recorded | ☐ |
| At least two acquisitions after injection | ☐ |
| Post-contrast fat-suppressed T1 acquired if aortitis is in question | ☐ |
| Source images and reformats available to the reader, not MIP alone | ☐ |
12. Advanced Technical Parameters
| Parameter | Value stated | Source |
|---|---|---|
| Test bolus | 2 mL GBCA + 20 mL saline | [4] |
| Aortic CE-MRA dose and rate | 0.1–0.2 mmol/kg at 2–3 mL/s; 20 mL saline at 2–3 mL/s | [4] |
| Aortic CE-MRA resolution | At least 1–1.5 mm | [4] |
| Pulmonary vein MRA | Coronal 3D; 1–2 mm slices; 60–80 slices; in-plane 1–1.5 mm; 2–3 acquisitions | [4] |
| Non-contrast 3D MRA (3 T) | 1.2 × 1.2 × 1.2 mm³; end-diastolic trigger; navigator ±4 mm; compressed sensing R = 11; Dixon | [5] |
| 4D flow | VENC 150 cm/s; 2 mm³; 30 ms; 5–10 min; flip angle ≈7° (12–25° with contrast) | [2] |
| Agreement of bSSFP and CE-MRA dimensions | Mean difference 0.1 mm; limits ±1.9 mm | [6] |
13. Evidence Gaps and Ongoing Debate
Not read: full texts of the ACC/AHA 2022 and ESC 2024 guidelines [1, 3], the SCMR post-processing document for aortic measurement, 4D flow consensus text beyond the first 100,000 characters [2], the comparison of European and American guidelines on imaging, and any systematic review of MRA accuracy in dissection, aneurysm or aortitis (all to be sourced). Not specified in the sources read: slice thickness and breath-hold duration of aortic CE-MRA, navigator parameters for aortic 4D flow, pulmonary artery MRA, MRI surveillance intervals, measurement convention (inner versus outer wall), the effect of aortic grafts and stents on MRA, and gadolinium-free strategies in renal impairment. Debate: whether non-contrast MRA can replace CE-MRA for follow-up is supported by two small studies only [5, 6]; 4D flow parameters beyond flow volume and velocity remain research tools [2].
14. Evidence-Based References
References are ordered by level of evidence and then by recency. The reading level of each source is stated in its line; a source read only at abstract level or through a summary supports only what that text states.
A. Guidelines / Consensus / Society Recommendations
B. Systematic Reviews / Meta-analyses
None used on this page.
C. Important Prospective / Original Studies
D. Technical MRI Papers
None used on this page.
E. Landmark Historical References
None used on this page.
Child Protocols
Clinical pages derived from this master protocol. These pages document what changes for specific indications.
No child protocols have been published yet.
Related Protocols
Recent PubMed search for this protocol