ASL-Based Non-Contrast MR Angiography — Labeling Strategies (PASL, pCASL, Time-Encoded, Superselective), Time-Resolved Readouts, Philips 4D-TRANCE and pCASL-Based 4D-PACK / 4D-S-PACK
Linked parent page: Triggered Non-Contrast-Enhanced MR Angiography — Cardiac-Gated Subtraction, Inflow-Inversion and Quiescent-Interval Techniques (TRANCE, FBI, NATIVE, Inhance Deltaflow, QISS and Equivalents)
ASL-Based Non-Contrast MR Angiography — Labeling Strategies (PASL, pCASL, Time-Encoded, Superselective), Time-Resolved Readouts, Philips 4D-TRANCE and pCASL-Based 4D-PACK / 4D-S-PACK
MRIninja Knowledge Base | Sequence/Technique Deep Dive Builds on: ASL Labeling Scheme Techniques and Parameters · Triggered Non-Contrast-Enhanced MRA (Section 9) · ASL Generic Standard Protocol Version 1.0 — October 2026
1. Introduction and Scope
Two pages already cover parts of this subject. The ASL labeling parameters page (Section 6) documents a Philips 4D-TRANCE console protocol parameter by parameter, and the triggered NCE-MRA page (Section 9) places ASL-based MRA in the taxonomy of non-contrast angiography. This page does not repeat either. It addresses how ASL is turned into an angiographic technique: how the angiographic use differs from perfusion ASL, how time resolution is created, how the labeling families (pulsed, pseudo-continuous, time-encoded, superselective) behave in angiography, and — as a focused section — what distinguishes the Philips pCASL-based angiographic techniques (4D-PACK, 4D-S-PACK, 4D-APACK) from the pulsed 4D-TRANCE.
Information status. The clinical literature on ASL angiography is dominated by single-centre studies, conference abstracts and one narrative review. No guideline or consensus statement specific to ASL angiography was found. Parameters are quoted only where a source reports them, with platform and field strength, as examples of reported practice and not as recommended settings.
2. Principle — Perfusion ASL and Angiographic ASL Use the Same Labeling Physics for Opposite Goals
ASL-based MRA is a subtraction technique: inflowing arterial blood is labelled with RF pulses, a control image is acquired without labeling, and the label–control difference leaves signal only from labelled blood, in the manner of digital subtraction angiography [1,2]. The labeling physics is common to perfusion ASL; the use of the signal is opposite.
| Aspect | Perfusion ASL | ASL angiography |
|---|---|---|
| Signal of interest | Labelled blood that has exchanged into tissue | Labelled blood still inside the arteries |
| Timing | Post-labeling delay set "just longer than the longest value of ATT present in the subject" so that labelled blood has reached tissue [4] | Short delays and multiple time points, to catch blood while it is in the vessels [1,3,5] |
| Intravascular signal | Treated as a transit artefact; the consensus protocol uses a segmented 3D readout without vascular crushing gradients [4] | Is the wanted signal |
| Output | Parametric CBF map | Vessel images, often time-resolved ("4D") |
Reasoning, not a quoted statement: a perfusion protocol and an angiographic protocol of the same labeling scheme are therefore tuned in opposite directions — long label and delay versus short label or delay and several sampled phases — which is why the console parameters of the two differ so markedly on the ASL labeling parameters page.
3. Labeling Families Used in Angiography
| Family | Named implementations (as reported) | How the angiographic time axis is created | Source |
|---|---|---|---|
| Pulsed ASL (PASL) | Philips 4D-TRANCE; Canon mASTAR | A single labeling event, then a series of readouts at increasing delays after labeling (post-labeling delay varied) | [1,3] |
| Pseudo-continuous ASL (pCASL) | Philips 4D-PACK, 4D-S-PACK, 4D-APACK | Labeling for a variable duration (label duration varied from 100 ms to about 2200 ms) with keyhole and view-sharing | [1,5,6,7] |
| Time-encoded pCASL | Philips research implementation | The labeling train is split into sub-boli whose label/control state follows a Hadamard matrix, decoded into images with different effective label durations and delays | [8] |
| Superselective pCASL | Philips 4D-S-PACK | pCASL with perpendicular gradients that restrict labeling to a small spot over one vessel | [7,9] |
The Gomyo review lists the vendor attribution as Philips for 4D-TRANCE (PASL) and for 4D-PACK and 4D-S-PACK (pCASL), and Canon for mASTAR (PASL) and its ultrashort-TE and high-temporal-resolution variants [1]. The Philips 4D-TRANCE product page describes it as "a time-resolved technique for non-contrast angiography" with temporal resolution "down to 160 msec" and does not name the labeling scheme [10]; the PASL attribution therefore comes from the review [1] and from the site's own console review of a "STAR Angio" protocol on the labeling parameters page. The ISMRM comparison of 4D-TRANCE with 4D-SNAP likewise states that the labeling pulse of 4D-TRANCE is not specified in its abstract [3]. Time-encoded pCASL is described in the review only as under development for simultaneous perfusion and angiography and not clinically established [1]. Silent MRA (GE) and VASC-ASL (Hitachi) are listed as ASL-based on the vendor-name table of the triggered NCE-MRA page [2] and are not developed here, because no source describing their labeling scheme was verified for this page.
4. How Time Resolution Is Created
4.1 Varying the Delay After a Single Label (PASL, 4D-TRANCE)
In the Philips 4D-TRANCE / ASL comparison abstract, 4D-TRANCE "varies the post-labeling delay", with a first PLD of 200 ms and an interval of 150 ms over 8 dynamic phases, using a 3D turbo field echo readout (TFE factor 23) or a 3D TFE-EPI readout (TFE factor 11, EPI factor 5) at 3 T on a Philips Ingenia with a 32-channel head coil [3]. The authors reported pulsation artefacts with the TFE-EPI readout and vessels fading in long-PLD images, and noted that ECG gating could reduce pulsation but would lengthen acquisition [3]. The Gomyo review explains the late-phase fading as signal decay with PASL due to RF saturation [1].
4.2 Varying the Label Duration with Keyhole and View-Sharing (pCASL, 4D-PACK)
4D-PACK combines pseudo-continuous labeling with a CENTRA-keyhole acquisition [5,6]. In the Moyamoya abstract, the label durations were 100, 200, 400, 600, 800, 1200, 1600 and 2200 ms, each phase including label and control images; the 100 ms data served as the reference for the first five phases and the 2200 ms data for the last three, with 75% keyhole and view-sharing [5]. The first abstract describing 4D-PACK states that the keyhole technique "can accelerate acquisition speed, while keeping flow dynamics information" in three volunteers [6]. Reasoning, not a quoted statement: in this scheme each time point corresponds to a bolus of different length, not a different delay, and the keyhole and view-sharing allow the central k-space to be refreshed more often than the periphery, which is what reduces the scan time.
4.3 Time-Encoded pCASL (Hadamard)
In time-encoded pCASL the labeling module is split into N segments whose label or control state changes across N+1 acquisitions following a Hadamard matrix; decoding yields N images with different labeling durations and delays [8]. In the Philips implementation by Suzuki and colleagues, an order-8 matrix was used, with a first sub-bolus of 1800 ms optimised for perfusion and six sub-boli of 130 ms for encoding the passage of the label in the angiogram, with a 3D TFE-EPI angiographic readout followed by a single-shot multislice EPI perfusion readout, at 3 T in six healthy volunteers [8]. The angiographic flip angle was a trade-off: 5° and 7.5° gave no significant CBF overestimation after correction, whereas 10° and above did, and higher flip angles produced ghosting of residual arterial signal; 7.5° was judged the best compromise [8]. Perfusion temporal SNR was significantly lower at all tested flip angles than with conventional time-encoded pCASL [8].
5. Focus — Philips pCASL and Its Distinguishing Characteristics
5.1 What the Verified Sources Establish for Philips
- Perfusion pCASL on the Philips console. The reviewed Philips protocol (documented on the labeling parameters page) exposes label distance (90 mm in the reviewed protocol) and post-label delay (1800 ms), and no separate editable labeling duration field in that protocol. This is a console observation from the site's own review of one protocol and is not a vendor specification.
- Angiographic pCASL implementations. 4D-PACK (pCASL with CENTRA-keyhole and view-sharing), 4D-S-PACK (superselective version) and 4D-APACK (4D-PACK combined with acceleration-selective ASL) are reported on Philips Ingenia 3 T systems with Philips-affiliated authors [5,6,7,9,11].
- Philips 4D-TRANCE is, per the review [1], PASL-based, so on this platform the pulsed and pseudo-continuous angiographic approaches coexist.
5.2 Characteristics of pCASL Relative to Pulsed Labeling
| Characteristic | pCASL | Pulsed (PASL) | Source |
|---|---|---|---|
| Mechanism | Flow-driven adiabatic inversion of blood crossing a labeling plane, using a train of RF pulses with gradients | One short inversion pulse labels a slab | [4] |
| Bolus duration | Longer and set by the operator ("the temporal duration of the labeled bolus is longer in PCASL") | "An unknown and relatively short temporal width" | [4] |
| SNR | Higher SNR and repeatability for single-delay perfusion; recommended as first choice for perfusion | Lower, with one stated advantage of lower RF power deposition | [4] |
| Labeling efficiency used in the quantification model | 0.85 | 0.98 | [4] |
| Off-resonance at the labeling plane | Labeling is sensitive to resonance offsets; "inefficient labeling" can result when pulses are not on resonance; pulse spacing governs this sensitivity (1 ms spacing as a goal) | Not discussed in the retrieved passages | [4] |
| Labeling plane geometry | Plane should be where feeding arteries are "relatively straight and perpendicular" to it | Slab labeling, gap to imaging volume | [4] |
| SAR | Not found to be limiting up to 3 T in perfusion use; long labels may be limited by power deposition | Lower power deposition | [4] |
| Late-phase signal in angiography | "Higher flow signal obtained with pCASL and less saturation effect after labeling" (authors' attribution); signal maintained at delayed phases | Signal decays at late phases due to RF saturation | [1,5] |
The consensus [4] concerns perfusion imaging and is cited here only for the physics of the labeling schemes; it is not an angiography guideline.
5.3 Reported Philips pCASL Angiographic Parameters
| Technique | Reported parameters (3 T Philips Ingenia) | Source |
|---|---|---|
| 4D-PACK (Moyamoya abstract) | 3D T1-TFE; TR/TE 4.9/1.73 ms; flip angle 11°; ETL 60; 80 mm slab; voxel 1.0 × 1.4 × 1.6 mm; SENSE 3.0; label durations 100–2200 ms (8 values); 75% keyhole; acquisition 6 min 5 s | [5] |
| 4D-S-PACK (DAVF abstract) | 3D T1-TFE; TR/TE 5.0/1.81 ms; flip angle 11°; ETL 60; 80 mm slab; voxel 1.0 × 1.4 × 1.6 mm; SENSE 3.0; keyhole 70%; label durations 100–2000 ms (7 values); acquisition about 5 min; labeling spot about 2 cm, gradient moment 0.75 mT/m/ms | [9] |
| 4D-APACK | 4D-PACK with the last time point replaced by acceleration-selective ASL (AccASL) with the same label duration; numeric parameters not available in the retrieved text | [11] |
Reasoning, not a quoted statement: the 80 mm slab in both abstracts implies limited craniocaudal coverage, consistent with intracranial rather than extracranial or body use.
5.4 Superselective Labeling (4D-S-PACK)
In 4D-S-PACK, standard pCASL gradients along the flow direction are supplemented with perpendicular Gx and Gy gradients so that labeling is restricted to a circular or elliptical spot placed over the proximal ICA or ECA [9]. In a retrospective single-centre study of 21 patients with intracranial dural arteriovenous fistulas (DSA reference), feeding arteries were detected in 80.0% and 76.8% (two observers) of the 95 seen on DSA with 4D-S-PACK, against 41.1% and 48.4% with non-selective 4D-PACK; shunt location was identified in all 21 patients with both methods; cortical venous reflux sensitivity and specificity were 90.9% for both; Borden classification was correct in 20 of 21 with both [7]. The authors list as limitations the small sample, motion sensitivity of a small labeling spot, harder ECA labeling, faint signal from unlabelled vessels, operator training for labeling-area planning, and an acquisition of about 20 minutes in total; three authors were Philips employees [7]. Lower CNR with 4D-S-PACK was reported in the earlier seven-patient abstract (feeding arteries 47.7 ± 12.8 versus 58.3 ± 15.3; draining veins 48.7 ± 18.5 versus 71.9 ± 24.7) [9]. The two feeding-artery sensitivities (71.0% in the abstract, about 80% in the paper) come from different cohorts and are not necessarily comparable.
5.5 Slow Flow and Acceleration-Selective ASL (4D-APACK)
The 4D-APACK abstract proposes replacing the last time point of 4D-PACK with AccASL to visualise very slow flow independently of arterial transit time; in six healthy volunteers and one AVM patient, transit-time correlation with 4D-pCASL was R = 0.971, MCA M1–M4 CNR did not differ significantly, and the AVM nidus and feeding artery were visible at the AccASL time point [11]. This is a methodological abstract and not a clinical validation.
6. Evidence Summary
| Study | Design | Technique | Result as reported |
|---|---|---|---|
| Gomyo 2025 [1] | Invited narrative review with institutional examples | PASL (4D-TRANCE, mASTAR) and pCASL (4D-PACK, 4D-S-PACK) | Temporal resolution reported as roughly 250 ms (PASL) and 87 ms (high-temporal-resolution mASTAR), comparable with DSA (about 250–300 ms); pCASL shows distal arteries and collaterals in Moyamoya better than PASL; no pooled accuracy data |
| Togao 2022 [7] | Retrospective, single centre, n = 21 DAVF, DSA reference | 4D-S-PACK vs 4D-PACK | See Section 5.4 |
| Togao 2017 [5] | Conference abstract, 13 Moyamoya patients (21 hemispheres), DSA within 3 months | 4D-PACK vs CINEMA (pulsed ASL, Look-Locker) | Higher CNR with 4D-PACK in distal MCA branches at later phases |
| Guo 2020 [3] | Conference abstract, 2 volunteers | 4D-SNAP vs 4D-TRANCE | 4D-SNAP higher SNR and fewer artefacts than 4D-TRANCE; qualitative; no numerical SNR in the abstract |
| Suzuki 2018 [8] | Technical study, 6 volunteers | Time-encoded pCASL | Simultaneous 4D-MRA and perfusion achieved; perfusion tSNR reduced vs conventional te-pCASL |
| Togao 2020 [9]; Akamine 2018 [11]; Obara 2016 [6] | Conference abstracts, 7 patients / 7 subjects / 3 volunteers | 4D-S-PACK; 4D-APACK; 4D-PACK | See Sections 4 and 5 |
For extracranial arteries, Edelman and Koktzoglou describe ASL-based MRA as well suited to fast-flowing vessels such as the carotids, taking about 5 minutes, and vulnerable to misregistration between acquisitions [2]. No study retrieved for this page reports a head-to-head diagnostic accuracy trial of Philips 4D-TRANCE against DSA or CE-MRA.
7. Pitfalls and Failure Modes
| Problem | Mechanism or evidence | Applies to |
|---|---|---|
| Motion between label and control | Subtraction depends on registration; motion causes artefacts [1,2] | All ASL-MRA |
| Late-phase signal loss | PASL signal decays with RF saturation; limited even with pCASL, so veins and slow-filling lesions are poorly seen; the review notes slow-filling residual AVM nidus and tumour stain seen on contrast-enhanced MRDSA were missed [1] | PASL more than pCASL |
| Labeling failure | Off-resonance at the labeling plane can cause inefficient labeling; plane should be perpendicular to straight arteries [4] | pCASL |
| Pulsation artefacts | Seen with TFE-EPI readout in 4D-TRANCE [3] | EPI-based readouts |
| Small labeling spot | Easily affected by patient movement; ECA labeling harder; operator training [7] | 4D-S-PACK |
| Scan time | pCASL about 8 minutes at the review authors' institution [1]; 4D-S-PACK about 20 minutes in total in the DAVF study [7] | pCASL, multi-vessel superselective |
| Lower spatial resolution than TOF-MRA | Stated limitation [1] | ASL-MRDSA |
| Source-image interpretation | Subtraction makes anatomical assessment from source images difficult [1] | All ASL-MRA |
8. Choosing Between Approaches (Reasoned Synthesis)
Reasoning, not a published algorithm: the following follows from the mechanisms and limitations above and from the single-centre evidence, and should not be read as a guideline.
- Short examination time and hemodynamic phases of fast-filling arteries: a pulsed approach such as 4D-TRANCE, accepting late-phase signal loss.
- Distal vessels, collaterals and slow flow (the Moyamoya setting): pCASL-based 4D-PACK, whose later phases retain signal [5].
- Identification of the feeding artery of a shunt: superselective labeling (4D-S-PACK), accepting longer time and operator-dependent planning [7].
- Very slow flow independent of transit time: AccASL addition (4D-APACK), still methodological [11].
- Fine anatomical detail of the arterial lumen: not the strength of ASL-MRA; the review states lower spatial resolution than TOF-MRA [1].
9. Evidence Gaps and Ongoing Debate
- No guideline or consensus statement for ASL angiography was found. The pCASL recommendation [4] applies to perfusion imaging.
- The labeling scheme of Philips 4D-TRANCE is not stated by the manufacturer page or the ISMRM comparison abstract [3,10]; the PASL attribution rests on a review [1] and on the console review on the labeling parameters page.
- Prospective diagnostic-accuracy trials are lacking. The accuracy figures come from retrospective single-centre work with small samples [7], and several techniques are described only in conference abstracts.
- Vendor-affiliated authors. The 4D-PACK family papers and abstracts involve Philips employees [6,7,8,9,11].
- Parameter values for 4D-APACK, and the 4D-TRANCE parameter tables of the comparison abstract beyond those quoted, were not available in the retrieved text.
- The review (Gomyo 2025) was read through its abstract and a processed summary of the accessible page, not as a full text. Quantitative statements taken from it (for example the temporal-resolution figures) are attributed to it and were not independently checked against the primary studies it cites.
- Extracranial and body ASL-MRA is only summarised from a single review [2]; no primary study was verified for this page.
10. Evidence-Based References
No official guideline, consensus statement specific to ASL angiography, systematic review or meta-analysis was verified. References are listed by evidence tier and recency.
A. Consensus Statement (perfusion ASL; physics of labeling only)
C. Original Studies
Conference Abstracts (vendor-affiliated or abstract-only)
D. Reviews and Technical Sources
End of document — ASL-Based Non-Contrast MR Angiography — MRIninja v1.0 — October 2026