ASL-Based Non-Contrast MR Angiography — Labeling Strategies (PASL, pCASL, Time-Encoded, Superselective), Time-Resolved Readouts, Philips 4D-TRANCE and pCASL-Based 4D-PACK / 4D-S-PACK

ASL-Based Non-Contrast MR Angiography — Labeling Strategies (PASL, pCASL, Time-Encoded, Superselective), Time-Resolved Readouts, Philips 4D-TRANCE and pCASL-Based 4D-PACK / 4D-S-PACK

up to this point verified by human experts

MRIninja Knowledge Base | Sequence/Technique Deep Dive Builds on: ASL Labeling Scheme Techniques and Parameters · Triggered Non-Contrast-Enhanced MRA (Section 9) · ASL Generic Standard Protocol Version 1.0 — October 2026


1. Introduction and Scope

Two pages already cover parts of this subject. The ASL labeling parameters page (Section 6) documents a Philips 4D-TRANCE console protocol parameter by parameter, and the triggered NCE-MRA page (Section 9) places ASL-based MRA in the taxonomy of non-contrast angiography. This page does not repeat either. It addresses how ASL is turned into an angiographic technique: how the angiographic use differs from perfusion ASL, how time resolution is created, how the labeling families (pulsed, pseudo-continuous, time-encoded, superselective) behave in angiography, and — as a focused section — what distinguishes the Philips pCASL-based angiographic techniques (4D-PACK, 4D-S-PACK, 4D-APACK) from the pulsed 4D-TRANCE.

Information status. The clinical literature on ASL angiography is dominated by single-centre studies, conference abstracts and one narrative review. No guideline or consensus statement specific to ASL angiography was found. Parameters are quoted only where a source reports them, with platform and field strength, as examples of reported practice and not as recommended settings.


2. Principle — Perfusion ASL and Angiographic ASL Use the Same Labeling Physics for Opposite Goals

ASL-based MRA is a subtraction technique: inflowing arterial blood is labelled with RF pulses, a control image is acquired without labeling, and the label–control difference leaves signal only from labelled blood, in the manner of digital subtraction angiography [1,2]. The labeling physics is common to perfusion ASL; the use of the signal is opposite.

AspectPerfusion ASLASL angiography
Signal of interestLabelled blood that has exchanged into tissueLabelled blood still inside the arteries
TimingPost-labeling delay set "just longer than the longest value of ATT present in the subject" so that labelled blood has reached tissue [4]Short delays and multiple time points, to catch blood while it is in the vessels [1,3,5]
Intravascular signalTreated as a transit artefact; the consensus protocol uses a segmented 3D readout without vascular crushing gradients [4]Is the wanted signal
OutputParametric CBF mapVessel images, often time-resolved ("4D")

Reasoning, not a quoted statement: a perfusion protocol and an angiographic protocol of the same labeling scheme are therefore tuned in opposite directions — long label and delay versus short label or delay and several sampled phases — which is why the console parameters of the two differ so markedly on the ASL labeling parameters page.


3. Labeling Families Used in Angiography

FamilyNamed implementations (as reported)How the angiographic time axis is createdSource
Pulsed ASL (PASL)Philips 4D-TRANCE; Canon mASTARA single labeling event, then a series of readouts at increasing delays after labeling (post-labeling delay varied)[1,3]
Pseudo-continuous ASL (pCASL)Philips 4D-PACK, 4D-S-PACK, 4D-APACKLabeling for a variable duration (label duration varied from 100 ms to about 2200 ms) with keyhole and view-sharing[1,5,6,7]
Time-encoded pCASLPhilips research implementationThe labeling train is split into sub-boli whose label/control state follows a Hadamard matrix, decoded into images with different effective label durations and delays[8]
Superselective pCASLPhilips 4D-S-PACKpCASL with perpendicular gradients that restrict labeling to a small spot over one vessel[7,9]

The Gomyo review lists the vendor attribution as Philips for 4D-TRANCE (PASL) and for 4D-PACK and 4D-S-PACK (pCASL), and Canon for mASTAR (PASL) and its ultrashort-TE and high-temporal-resolution variants [1]. The Philips 4D-TRANCE product page describes it as "a time-resolved technique for non-contrast angiography" with temporal resolution "down to 160 msec" and does not name the labeling scheme [10]; the PASL attribution therefore comes from the review [1] and from the site's own console review of a "STAR Angio" protocol on the labeling parameters page. The ISMRM comparison of 4D-TRANCE with 4D-SNAP likewise states that the labeling pulse of 4D-TRANCE is not specified in its abstract [3]. Time-encoded pCASL is described in the review only as under development for simultaneous perfusion and angiography and not clinically established [1]. Silent MRA (GE) and VASC-ASL (Hitachi) are listed as ASL-based on the vendor-name table of the triggered NCE-MRA page [2] and are not developed here, because no source describing their labeling scheme was verified for this page.


4. How Time Resolution Is Created

4.1 Varying the Delay After a Single Label (PASL, 4D-TRANCE)

In the Philips 4D-TRANCE / ASL comparison abstract, 4D-TRANCE "varies the post-labeling delay", with a first PLD of 200 ms and an interval of 150 ms over 8 dynamic phases, using a 3D turbo field echo readout (TFE factor 23) or a 3D TFE-EPI readout (TFE factor 11, EPI factor 5) at 3 T on a Philips Ingenia with a 32-channel head coil [3]. The authors reported pulsation artefacts with the TFE-EPI readout and vessels fading in long-PLD images, and noted that ECG gating could reduce pulsation but would lengthen acquisition [3]. The Gomyo review explains the late-phase fading as signal decay with PASL due to RF saturation [1].

4.2 Varying the Label Duration with Keyhole and View-Sharing (pCASL, 4D-PACK)

4D-PACK combines pseudo-continuous labeling with a CENTRA-keyhole acquisition [5,6]. In the Moyamoya abstract, the label durations were 100, 200, 400, 600, 800, 1200, 1600 and 2200 ms, each phase including label and control images; the 100 ms data served as the reference for the first five phases and the 2200 ms data for the last three, with 75% keyhole and view-sharing [5]. The first abstract describing 4D-PACK states that the keyhole technique "can accelerate acquisition speed, while keeping flow dynamics information" in three volunteers [6]. Reasoning, not a quoted statement: in this scheme each time point corresponds to a bolus of different length, not a different delay, and the keyhole and view-sharing allow the central k-space to be refreshed more often than the periphery, which is what reduces the scan time.

4.3 Time-Encoded pCASL (Hadamard)

In time-encoded pCASL the labeling module is split into N segments whose label or control state changes across N+1 acquisitions following a Hadamard matrix; decoding yields N images with different labeling durations and delays [8]. In the Philips implementation by Suzuki and colleagues, an order-8 matrix was used, with a first sub-bolus of 1800 ms optimised for perfusion and six sub-boli of 130 ms for encoding the passage of the label in the angiogram, with a 3D TFE-EPI angiographic readout followed by a single-shot multislice EPI perfusion readout, at 3 T in six healthy volunteers [8]. The angiographic flip angle was a trade-off: 5° and 7.5° gave no significant CBF overestimation after correction, whereas 10° and above did, and higher flip angles produced ghosting of residual arterial signal; 7.5° was judged the best compromise [8]. Perfusion temporal SNR was significantly lower at all tested flip angles than with conventional time-encoded pCASL [8].


5. Focus — Philips pCASL and Its Distinguishing Characteristics

5.1 What the Verified Sources Establish for Philips

  • Perfusion pCASL on the Philips console. The reviewed Philips protocol (documented on the labeling parameters page) exposes label distance (90 mm in the reviewed protocol) and post-label delay (1800 ms), and no separate editable labeling duration field in that protocol. This is a console observation from the site's own review of one protocol and is not a vendor specification.
  • Angiographic pCASL implementations. 4D-PACK (pCASL with CENTRA-keyhole and view-sharing), 4D-S-PACK (superselective version) and 4D-APACK (4D-PACK combined with acceleration-selective ASL) are reported on Philips Ingenia 3 T systems with Philips-affiliated authors [5,6,7,9,11].
  • Philips 4D-TRANCE is, per the review [1], PASL-based, so on this platform the pulsed and pseudo-continuous angiographic approaches coexist.

5.2 Characteristics of pCASL Relative to Pulsed Labeling

CharacteristicpCASLPulsed (PASL)Source
MechanismFlow-driven adiabatic inversion of blood crossing a labeling plane, using a train of RF pulses with gradientsOne short inversion pulse labels a slab[4]
Bolus durationLonger and set by the operator ("the temporal duration of the labeled bolus is longer in PCASL")"An unknown and relatively short temporal width"[4]
SNRHigher SNR and repeatability for single-delay perfusion; recommended as first choice for perfusionLower, with one stated advantage of lower RF power deposition[4]
Labeling efficiency used in the quantification model0.850.98[4]
Off-resonance at the labeling planeLabeling is sensitive to resonance offsets; "inefficient labeling" can result when pulses are not on resonance; pulse spacing governs this sensitivity (1 ms spacing as a goal)Not discussed in the retrieved passages[4]
Labeling plane geometryPlane should be where feeding arteries are "relatively straight and perpendicular" to itSlab labeling, gap to imaging volume[4]
SARNot found to be limiting up to 3 T in perfusion use; long labels may be limited by power depositionLower power deposition[4]
Late-phase signal in angiography"Higher flow signal obtained with pCASL and less saturation effect after labeling" (authors' attribution); signal maintained at delayed phasesSignal decays at late phases due to RF saturation[1,5]

The consensus [4] concerns perfusion imaging and is cited here only for the physics of the labeling schemes; it is not an angiography guideline.

5.3 Reported Philips pCASL Angiographic Parameters

TechniqueReported parameters (3 T Philips Ingenia)Source
4D-PACK (Moyamoya abstract)3D T1-TFE; TR/TE 4.9/1.73 ms; flip angle 11°; ETL 60; 80 mm slab; voxel 1.0 × 1.4 × 1.6 mm; SENSE 3.0; label durations 100–2200 ms (8 values); 75% keyhole; acquisition 6 min 5 s[5]
4D-S-PACK (DAVF abstract)3D T1-TFE; TR/TE 5.0/1.81 ms; flip angle 11°; ETL 60; 80 mm slab; voxel 1.0 × 1.4 × 1.6 mm; SENSE 3.0; keyhole 70%; label durations 100–2000 ms (7 values); acquisition about 5 min; labeling spot about 2 cm, gradient moment 0.75 mT/m/ms[9]
4D-APACK4D-PACK with the last time point replaced by acceleration-selective ASL (AccASL) with the same label duration; numeric parameters not available in the retrieved text[11]

Reasoning, not a quoted statement: the 80 mm slab in both abstracts implies limited craniocaudal coverage, consistent with intracranial rather than extracranial or body use.

5.4 Superselective Labeling (4D-S-PACK)

In 4D-S-PACK, standard pCASL gradients along the flow direction are supplemented with perpendicular Gx and Gy gradients so that labeling is restricted to a circular or elliptical spot placed over the proximal ICA or ECA [9]. In a retrospective single-centre study of 21 patients with intracranial dural arteriovenous fistulas (DSA reference), feeding arteries were detected in 80.0% and 76.8% (two observers) of the 95 seen on DSA with 4D-S-PACK, against 41.1% and 48.4% with non-selective 4D-PACK; shunt location was identified in all 21 patients with both methods; cortical venous reflux sensitivity and specificity were 90.9% for both; Borden classification was correct in 20 of 21 with both [7]. The authors list as limitations the small sample, motion sensitivity of a small labeling spot, harder ECA labeling, faint signal from unlabelled vessels, operator training for labeling-area planning, and an acquisition of about 20 minutes in total; three authors were Philips employees [7]. Lower CNR with 4D-S-PACK was reported in the earlier seven-patient abstract (feeding arteries 47.7 ± 12.8 versus 58.3 ± 15.3; draining veins 48.7 ± 18.5 versus 71.9 ± 24.7) [9]. The two feeding-artery sensitivities (71.0% in the abstract, about 80% in the paper) come from different cohorts and are not necessarily comparable.

5.5 Slow Flow and Acceleration-Selective ASL (4D-APACK)

The 4D-APACK abstract proposes replacing the last time point of 4D-PACK with AccASL to visualise very slow flow independently of arterial transit time; in six healthy volunteers and one AVM patient, transit-time correlation with 4D-pCASL was R = 0.971, MCA M1–M4 CNR did not differ significantly, and the AVM nidus and feeding artery were visible at the AccASL time point [11]. This is a methodological abstract and not a clinical validation.


6. Evidence Summary

StudyDesignTechniqueResult as reported
Gomyo 2025 [1]Invited narrative review with institutional examplesPASL (4D-TRANCE, mASTAR) and pCASL (4D-PACK, 4D-S-PACK)Temporal resolution reported as roughly 250 ms (PASL) and 87 ms (high-temporal-resolution mASTAR), comparable with DSA (about 250–300 ms); pCASL shows distal arteries and collaterals in Moyamoya better than PASL; no pooled accuracy data
Togao 2022 [7]Retrospective, single centre, n = 21 DAVF, DSA reference4D-S-PACK vs 4D-PACKSee Section 5.4
Togao 2017 [5]Conference abstract, 13 Moyamoya patients (21 hemispheres), DSA within 3 months4D-PACK vs CINEMA (pulsed ASL, Look-Locker)Higher CNR with 4D-PACK in distal MCA branches at later phases
Guo 2020 [3]Conference abstract, 2 volunteers4D-SNAP vs 4D-TRANCE4D-SNAP higher SNR and fewer artefacts than 4D-TRANCE; qualitative; no numerical SNR in the abstract
Suzuki 2018 [8]Technical study, 6 volunteersTime-encoded pCASLSimultaneous 4D-MRA and perfusion achieved; perfusion tSNR reduced vs conventional te-pCASL
Togao 2020 [9]; Akamine 2018 [11]; Obara 2016 [6]Conference abstracts, 7 patients / 7 subjects / 3 volunteers4D-S-PACK; 4D-APACK; 4D-PACKSee Sections 4 and 5

For extracranial arteries, Edelman and Koktzoglou describe ASL-based MRA as well suited to fast-flowing vessels such as the carotids, taking about 5 minutes, and vulnerable to misregistration between acquisitions [2]. No study retrieved for this page reports a head-to-head diagnostic accuracy trial of Philips 4D-TRANCE against DSA or CE-MRA.


7. Pitfalls and Failure Modes

ProblemMechanism or evidenceApplies to
Motion between label and controlSubtraction depends on registration; motion causes artefacts [1,2]All ASL-MRA
Late-phase signal lossPASL signal decays with RF saturation; limited even with pCASL, so veins and slow-filling lesions are poorly seen; the review notes slow-filling residual AVM nidus and tumour stain seen on contrast-enhanced MRDSA were missed [1]PASL more than pCASL
Labeling failureOff-resonance at the labeling plane can cause inefficient labeling; plane should be perpendicular to straight arteries [4]pCASL
Pulsation artefactsSeen with TFE-EPI readout in 4D-TRANCE [3]EPI-based readouts
Small labeling spotEasily affected by patient movement; ECA labeling harder; operator training [7]4D-S-PACK
Scan timepCASL about 8 minutes at the review authors' institution [1]; 4D-S-PACK about 20 minutes in total in the DAVF study [7]pCASL, multi-vessel superselective
Lower spatial resolution than TOF-MRAStated limitation [1]ASL-MRDSA
Source-image interpretationSubtraction makes anatomical assessment from source images difficult [1]All ASL-MRA

8. Choosing Between Approaches (Reasoned Synthesis)

Reasoning, not a published algorithm: the following follows from the mechanisms and limitations above and from the single-centre evidence, and should not be read as a guideline.

  • Short examination time and hemodynamic phases of fast-filling arteries: a pulsed approach such as 4D-TRANCE, accepting late-phase signal loss.
  • Distal vessels, collaterals and slow flow (the Moyamoya setting): pCASL-based 4D-PACK, whose later phases retain signal [5].
  • Identification of the feeding artery of a shunt: superselective labeling (4D-S-PACK), accepting longer time and operator-dependent planning [7].
  • Very slow flow independent of transit time: AccASL addition (4D-APACK), still methodological [11].
  • Fine anatomical detail of the arterial lumen: not the strength of ASL-MRA; the review states lower spatial resolution than TOF-MRA [1].

9. Evidence Gaps and Ongoing Debate

  • No guideline or consensus statement for ASL angiography was found. The pCASL recommendation [4] applies to perfusion imaging.
  • The labeling scheme of Philips 4D-TRANCE is not stated by the manufacturer page or the ISMRM comparison abstract [3,10]; the PASL attribution rests on a review [1] and on the console review on the labeling parameters page.
  • Prospective diagnostic-accuracy trials are lacking. The accuracy figures come from retrospective single-centre work with small samples [7], and several techniques are described only in conference abstracts.
  • Vendor-affiliated authors. The 4D-PACK family papers and abstracts involve Philips employees [6,7,8,9,11].
  • Parameter values for 4D-APACK, and the 4D-TRANCE parameter tables of the comparison abstract beyond those quoted, were not available in the retrieved text.
  • The review (Gomyo 2025) was read through its abstract and a processed summary of the accessible page, not as a full text. Quantitative statements taken from it (for example the temporal-resolution figures) are attributed to it and were not independently checked against the primary studies it cites.
  • Extracranial and body ASL-MRA is only summarised from a single review [2]; no primary study was verified for this page.

10. Evidence-Based References

No official guideline, consensus statement specific to ASL angiography, systematic review or meta-analysis was verified. References are listed by evidence tier and recency.

A. Consensus Statement (perfusion ASL; physics of labeling only)

High
[4] Alsop DC, Detre JA, Golay X, Günther M, Hendrikse J, Hernandez-Garcia L, Lu H, MacIntosh BJ, Parkes LM, Smits M, van Osch MJP, Wang DJJ, Wong EC, Zaharchuk G. Recommended implementation of arterial spin-labeled perfusion MRI for clinical applications: a consensus of the ISMRM Perfusion Study Group and the European consortium for ASL in dementia. Magn Reson Med. 2015;73(1):102–116. DOI: 10.1002/mrm.25197. PMID: 24715426.
Relevance: pCASL versus PASL characteristics (Sections 2, 5.2). Concerns perfusion, not angiography. Read via the open-access full text.

C. Original Studies

Limited
[7] Togao O, Obara M, Kikuchi K, Helle M, Arimura K, Nishimura A, Wada T, Murazaki H, Van Cauteren M, Hiwatashi A, Ishigami K. Vessel-selective 4D-MRA using superselective pseudocontinuous arterial spin-labeling with keyhole and view-sharing for visualizing intracranial dural AVFs. AJNR Am J Neuroradiol. 2022;43(3):368–375. DOI: 10.3174/ajnr.A7426.
Relevance: Retrospective single-centre study, n = 21, DSA reference; Sections 5.4, 6, 7. Read from the journal page; three authors were Philips employees.
Technical
[8] Suzuki Y, Helle M, Koken P, Van Cauteren M, van Osch MJP. Simultaneous acquisition of perfusion image and dynamic MR angiography using time-encoded pseudo-continuous ASL. Magn Reson Med. 2018;79(5):2676–2684.
Relevance: Time-encoded pCASL, Philips 3 T, 6 volunteers; Section 4.3. Abstract-level content from a repository copy; DOI not retrieved and therefore not given.

Conference Abstracts (vendor-affiliated or abstract-only)

Limited
[5] Togao O, Hiwatashi A, Obara M, Yamashita K, Kamei R, Honda H. 4D MR angiography with pseudo-continuous arterial spin labelling combined with CENTRA-keyhole (4D-PACK): visualization of distal cerebral arteries in Moyamoya disease. Proc Intl Soc Magn Reson Med. 2017;25:0145.
Relevance: 13 Moyamoya patients, label durations, parameters, CNR comparison with a pulsed ASL technique; Sections 4.2, 5.3, 6.
Limited
[9] Togao O, Hiwatashi A, Obara M, Helle M, Kikuchi K, Momosaka D, Kikuchi Y, Wada T, Murazaki H, Van Cauteren M. Vessel-selective 4D-MRA using superselective pCASL combined with CENTRA-keyhole (4D-S-PACK) for intracranial dural arteriovenous fistulas. Proc Intl Soc Magn Reson Med. 2020;28:1080.
Relevance: 7 DAVF patients; superselective labeling parameters; Sections 3, 5.3, 5.4.
Limited
[3] Guo H, Xiong Y, Li S, Liu S, Yuan C. Comparison of non-contrast enhanced dynamic MRA techniques: 4D-SNAP vs 4D-TRANCE. Proc Intl Soc Magn Reson Med. 2020;28:2160.
Relevance: 2 volunteers, Philips 3 T Ingenia; 4D-TRANCE varies PLD, labeling pulse not specified; Sections 3, 4.1, 6.
Limited
[11] Akamine Y, Obara M, Togao O, Shibukawa S, Yoneyama M, Okuaki T, Van Cauteren M. New 4D-MRA approach named 4D-APACK by integrating 4D-PACK and AccASL … Proc Intl Soc Magn Reson Med. 2018;26:1225.
Relevance: AccASL addition for slow flow; 6 volunteers and 1 AVM patient; Section 5.5. The full title was truncated in the retrieved text and is shown with an ellipsis.
Limited
[6] Obara M, Togao O, Okuaki T, Shibukawa S, Yoneyama M, Van Cauteren M. Non-contrast enhanced 4D intracranial MR angiography based on pseudo-continuous arterial spin labelling (PCASL) with the keyhole technique. Proc Intl Soc Magn Reson Med. 2016;24:4376.
Relevance: First 4D-PACK description; 3 volunteers; Section 4.2.

D. Reviews and Technical Sources

Technical
[1] Gomyo M, Tsuchiya K, Yokoyama K. Noninvasive dynamic vascular imaging: arterial spin labeling-based noncontrast magnetic resonance digital subtraction angiography for cerebral disease diagnoses. Jpn J Radiol. 2025;43(7):1049–1065. DOI: 10.1007/s11604-025-01758-w.
Relevance: Invited narrative review; vendor and labeling attribution, temporal resolution, limitations. Open access; read through the retrieved page summary and abstract, not independently in full.
Technical
[2] Edelman RR, Koktzoglou I. Non-contrast MR angiography: an update. J Magn Reson Imaging. 2019;49(2):355–373. DOI: 10.1002/jmri.26288. PMID: 30566270.
Relevance: Family taxonomy, vendor-name table and the general description of ASL-based MRA (Sections 2, 3, 6).
Technical
[10] Philips. 4D-TRANCE MR clinical application — product page (Philips Lebanon, HCNMRB970; accessed October 2026).
Relevance: Manufacturer description; no labeling scheme stated. Manufacturer material, not peer-reviewed.

End of document — ASL-Based Non-Contrast MR Angiography — MRIninja v1.0 — October 2026