DWI Acquisition Parameters: b-Values, Averages, Directions, Gradient Scheme, Noise Floor and Computed b-Values
MRIninja Knowledge Base | MRI Parameter Deep Dive Parent page: MRI Parameters — Overview and Classification (9501), Sections 8.1 and 8.2 Builds on: Diffusion-Weighted Imaging (DWI) Sequence (9012), Echo Planar DWI (9014) Version 1.0 — November 2026
MRI Parameter Deep Dive
DWI Acquisition Parameters: b-Values, Averages, Directions, Gradient Scheme, Noise Floor and Computed b-Values
What the operator chooses on a diffusion-weighted acquisition and what each choice changes: the set of b-values and the averages given to each, the way the three diffusion directions are combined, monopolar, bipolar and twice-refocused gradient schemes, the Rician noise floor and the thresholds that hide it, computed high b-value images, and fat suppression inside a diffusion sequence. Console names are mapped across manufacturers where a source allows, and the undocumented ones are marked.
1. Scope, Boundary and Information Status
The overview page 9501 (Sections 8.1 and 8.2) defines the b-value, gives the signal equation and the ADC formula, lists typical low, diagnostic and high b-values, tabulates approximate ADC values and states the number of directions of diffusion tensor imaging. Those definitions are not repeated here. This page adds what the operator decides: how many b-values, which, with how many averages, in how many directions and combined how, with which gradient waveform, with which noise handling, and with what fat suppression. The sequence pages 9012, 9013 and 9014 are still placeholders. The general behaviour of signal averages is on 9589; the readout parameters of echo-planar imaging (EPI factor, echo spacing, bandwidth, distortion) are planned for a separate page and are only touched here where they meet the diffusion parameters; the partial Fourier and parallel imaging options that shorten the echo time are on 9563 and 9564; the inversion-recovery principle behind STIR-type fat suppression is on 9020.
This page is the home of the acquisition parameters of a diffusion-weighted series as controls. Protocol values for an organ and the diagnostic use of the images belong to the protocol page of the region.
Information status. No guideline states how the parameters of Section 2 must be set. The strongest sources are three consensus-level documents: the consensus statement of an ISMRM-sponsored workshop on diffusion imaging outside the brain [3], the QIBA Profile for the apparent diffusion coefficient [1] and the 2009 consensus on diffusion imaging as a cancer biomarker [4]. Their recommendations are expert consensus, not outcome evidence. Below them sit technical studies with small samples (the comparison of trace and diagonal acquisition has two volunteers [8]), simulations, narrative reviews and manufacturer training material. Manufacturer parameter semantics are only partly documented: a Siemens training job aid states that it does not explain several options [20], and the operator manuals were not available. Every source is listed in Section 17 with the extent to which it was read; several were read through a summarising tool or as search extracts only. Statements marked (Reasoning) are this page's own deductions from cited formulas or from the owner's console values; statements marked (to be sourced) are claims for which a source was sought and not obtained (Section 15). Console values come from one head-and-neck examination at 3 T [25].
2. Parameter Map: What Each Control Changes
A diffusion series trades four quantities against each other: signal-to-noise ratio (SNR) at the highest b-value, accuracy of the derived ADC, acquisition time and geometric distortion with echo time. The table lists the controls treated on this page, the main effects and the section that develops each. The Siemens column gives the console labels seen on the owner's series or documented in the job aid; it is a mapping, not a definition [20], [25].
| Control | What it sets | Main effects | Siemens label (seen or documented) | Section |
|---|---|---|---|---|
| Set of b-values | Number, minimum and maximum diffusion weighting | Contrast, perfusion contamination of the ADC, noise-floor bias, time | Diff. Weightings; b-value fields | 3 |
| Averages per b-value | Number of repeated acquisitions of each weighting | SNR at high b, time (linear), partial averaging of motion | Averages in each weighting row | 4 |
| Directions and trace combination | How the orthogonal directions are acquired and combined | Time, minimum TE, anisotropy bias, eddy-current differences between directions | Diffusion Mode (4-Scan Trace, 3D Diagonal) | 5 |
| Gradient scheme | Waveform of the diffusion gradients | Minimum TE against eddy-current distortion | Diffusion Scheme (Monopolar, Bipolar) | 6 |
| Noise handling and thresholds | What is excluded from the ADC map | Visible noise in the map, hidden low-signal lesions | ADC Noise Threshold | 7 |
| Computed b-value | A high-b image extrapolated from acquired b-values | Contrast at high b without the time of acquiring it; no new information | Calculated b-value | 8 |
| Fat suppression in the sequence | Method that removes fat signal from the diffusion images | Residual fat, SNR, ADC bias, time | Magnetization preparation: slice-selective IR (TI 210 ms on the owner's series) | 9 |
| Field and eddy-current correction | Correction of distortion from gradients and field | Geometric accuracy | Dynamic Field Correction (listed in the project backlog; state not recorded on the series) | 10 |
3. Choosing the Set of b-Values
3.1 What the consensus documents state
The 2009 consensus on diffusion imaging as a cancer biomarker states that mono-exponential ADC assessment should use two b-values, one above 100 and one between 500 and 1000 s/mm² depending on the application, and that free breathing with multiple acquisitions is superior to complex gating techniques [4]. Only the abstract statements of that document were available to this page (the full-text page returned a browser check that was not circumvented), so its justification is not reproduced. The 2016 workshop consensus on diffusion imaging outside the brain gives a technical table for the abdomen and pelvis: TR of at least 3000 ms; TE as short as possible (90 ms or less); matrix at least 128 × 128, higher for the prostate; slice thickness 3–7 mm (3 mm for the prostate, matching T2-weighted imaging); parallel imaging reduction factor 2; at least two b-values with one or two intermediate values optional; a minimum b-value below 100 s/mm²; a maximum b-value of 800–1000 s/mm², with ultra-high values of 1400–2000 s/mm² for the prostate acquired or computed; three diffusion directions; and enough averages and time for adequate SNR at the maximum b-value [3]. The same document gives organ-specific sets:
| Organ | b-values stated | Purpose stated | Source |
|---|---|---|---|
| Liver | Low b below 100; intermediate 400–500; higher 500–1000 s/mm² | Low b gives black-blood images; b = 0 is avoided for ADC unless IVIM is intended | [3] |
| Liver, IVIM (minimal set cited) | 0, 15, 150, 800 s/mm² | Separation of perfusion and diffusion; recent studies used 5–16 b-values | [3] |
| Pancreas | 0, 150, 1000 s/mm² for lesion detection; 9–11 values from 0 to 1000 for IVIM in selected patients | Lesion detection; IVIM | [3] |
| Breast | Maximum at least 800 s/mm²; two values (for example 0 and 800) may be acceptable clinically | Lesion detection; a non-zero minimum b reduces perfusion contamination | [3] |
| Prostate | Historically a maximum of 700–1000 s/mm²; 1400–2000 s/mm² ultra-high, acquired or computed | Contrast between tumour and benign tissue at high b | [3] |
3.2 What the choice changes
Minimum b-value and perfusion. A b-value of zero keeps the contribution of blood moving in capillaries, which is why the consensus recommends avoiding b = 0 for the ADC unless IVIM is the aim and notes that a non-zero minimum b-value reduces perfusion contamination in the breast [3]. The same document reports that ADC measured between b = 0 and 100 s/mm² was less consistent in repeatability studies [3]. The IVIM fit and its parameters are outside this page.
Maximum b-value and the noise floor. Raising the maximum b-value raises contrast between tissues with different ADC but lowers the signal toward the noise floor (Section 7). A Rician noise simulation for the selection of b-value sets in non-small-cell lung cancer found that the best sets contained lower b-values as the noise level increased [9]. QIBA states that the bias of the measured ADC depends on the b-value range and on the suppression of fat and bone-marrow signal, without giving a quantitative relation [1].
Same protocol for follow-up. The QIBA claims on the change of lesion ADC that can be considered real (95% confidence) are 11% for brain, 26% for liver, 47% for prostate and 13% for breast in the current profile, with proposed updates of 8%, 27%, 27% and 15%; both sets apply only when the same scanner and the same protocol are used for the baseline and the follow-up examination [1]. A change of the b-values or the averages between visits therefore removes the basis of the claim [1].
Console example. The owner's series is named for b = 500 and b = 1000 s/mm², with two weightings listed: b = 500 with 2 averages and b = 1000 with 10 averages [25]. Whether a b = 0 image is acquired in addition, and with how many averages, was not recorded in the transcription of the screenshots (to be sourced from the console). The Siemens job aid for a different diffusion sequence shows two weightings with a b-value field of 1000 s/mm² and a stated range of b ≥ 0 [20].
4. Averages per b-Value
The general relation between averages, SNR and time is on 9589. In diffusion imaging three specific points apply.
More averages at higher b. The workshop consensus states that the number of signal averages is usually increased for higher b-values so that a multi-b-value experiment reaches acceptable SNR in a clinically reasonable time [3]. The reason is the decay of signal with b: with a mono-exponential model the signal at b is S0·exp(−b·ADC) while the noise is unchanged, so SNR per acquisition falls exponentially with b (Reasoning, from the equation on 9501). The Siemens job aid for BLADE diffusion states that normally one average is used at a low b-value and one or two at a high b-value [20]; this is a vendor training statement for that sequence, and the owner's series uses ten at b = 1000 s/mm² [25].
Averages compensate for fewer directions or a lower signal. In the study that compared 3-scan trace with diagonal acquisition, the number of averages of the diagonal acquisition was tripled (except at b = 0) to offset the √3 loss of SNR that follows from acquiring one direction instead of three [8]. Their b-values and averages are an example of the practice: abdomen 50, 400 and 800 s/mm² with 1, 2 and 4 averages for the 3-scan trace acquisition, and 50, 1000 and 1600 s/mm² with 1, 5 and 10 averages for the pelvis [8].
Averaging does not remove the noise floor. Averaging magnitude images reduces the spread of the noise but not the positive bias that magnitude reconstruction gives to low signal, so a signal that sits on the noise floor stays on it (Reasoning, from the Rician description in [17]; a source that tests this for clinical diffusion data was not found). The main practical consequence is that SNR at the highest b-value must be judged on the signal itself, and QIBA sets a requirement on the b = 0 image instead: the SNR in the tissue region must be at least 50 [1].
Time. Each average of each weighting and each direction acquired separately is a further repetition, so the acquisition time grows in proportion to the total number of repetitions (Reasoning; see the check in Section 12). Free-breathing body diffusion with several averages is preferred over complex gating by the 2009 consensus [4].
5. Diffusion Directions and Trace Combination
The number of directions of diffusion tensor imaging and the tensor-derived maps are on 9501 (Section 8.2). Clinical body and brain diffusion imaging uses a small set of directions combined into one orientation-independent image, and the way this is done is a console choice with consequences for TE and for anisotropic tissue. The workshop consensus lists three diffusion directions for abdominal and pelvic diffusion imaging [3].
| Approach | How the directions are applied | Stated advantage | Stated limit | Source |
|---|---|---|---|---|
| 3-scan trace | Three orthogonal directions acquired one after the other; the three signals are combined by the geometric mean into a trace image | Orientation-invariant image | Eddy-current differences between the three acquisitions may blur the trace image | [8] |
| Tetrahedral | Four directions at full gradient strength | Shorter TE than 3-scan trace | Four directions instead of three; switching eddy currents may still affect the images | [8] |
| Diagonal | Maximal gradient strength on all three axes at once, giving one net direction per b-value | Shorter TE; one direction instead of three | With the monopolar scheme, intended mainly for isotropic organs; may bias the ADC in anisotropic tissue (kidney, possibly pancreas and liver) | [8] |
What was measured. A phantom and two-volunteer study (men aged 22 and 23; 1.5 T and 3.0 T; 32-channel body coil) compared 3-scan trace with monopolar diagonal acquisition [8]. The TE of the diagonal acquisition was shorter by only about 2–4 ms (abdomen 73 against 70 ms at 1.5 T and 70 against 68 ms at 3.0 T; pelvis 83 against 79 ms and 71 against 69 ms) [8]. Phantom ADC values agreed (mean coefficient of variation 1.4 ± 1.2% at 1.5 T and 0.7 ± 0.7% at 3.0 T), phantom SNR was similar and in one comparison (b = 900 s/mm² at 3.0 T) significantly lower with diagonal acquisition, and no SNR gain was shown [8]. In volunteers the ADC differences were not significant (P = 0.287–0.429) and a non-significant trend (P = 0.063–0.066) favoured the diagonal acquisition for image quality in some conditions, mainly in the pelvis [8]. The authors describe the study as preliminary, ask for a larger test–retest study and attribute the absent SNR gain to the small TE difference, possibly because the 3-scan trace of that manufacturer already uses oblique directions with high gradient strength [8].
Names. The monopolar diagonal acquisition is called “3D diagonal” by Siemens, “3-in-1” by GE and “gradient overplus” by Philips in that paper [8]. A bipolar isotropic diagonal scheme that accounts for anisotropy is described there as a different technique from the monopolar version, which is the one the study used [8].
Console example. The owner's series shows Diffusion Mode “3D Diagonal” together with a bipolar scheme [25]. Whether this combination corresponds to the bipolar isotropic scheme of the literature or to a bipolar variant of the Siemens diagonal mode is not established by the sources read (to be sourced). The Siemens job aid shows the labels “4-Scan Trace” and “3D Diagonal” in its dropdown and recommends 4-Scan Trace for short TE in BLADE diffusion [20]; that “4-Scan Trace” is the tetrahedral scheme of the table is plausible from the number of scans but is not stated by the sources (Reasoning, to be sourced).
Consequence for protocols. Because the trace is a combination of separately acquired signals, any parameter that differs between the directions of one series (distortion, motion, eddy currents) affects the trace image; a mode that acquires one direction removes the combination but gives the single-direction anisotropy sensitivity of the net direction (Reasoning from [8]). For anisotropic tissue the tensor approach is on 9501 (Section 8.2).
6. Gradient Scheme: Monopolar, Bipolar and Twice-Refocused
The diffusion gradients can be applied as a single lobe on each side of the refocusing pulse (monopolar, the Stejskal–Tanner arrangement) or as paired lobes of opposite polarity that are arranged so that the eddy currents of the individual ramps cancel. The twice-refocused spin echo was introduced to reduce the image distortion that eddy currents from the diffusion gradients create in echo-planar diffusion images [16]. In the description of a later waveform-design paper the twice-refocused scheme balances the eddy currents produced by each gradient ramp by using bipolar pairs [13]. A magnetic-field-monitoring study found increased second-order eddy currents in a unipolar diffusion sequence compared with a bipolar one, with zeroth and linear orders similar in both [14].
The trade. A bipolar or twice-refocused waveform spends more time on gradient lobes for the same b-value, so the minimum TE is longer, and a longer TE lowers SNR (Reasoning; the sources read give the qualitative motivation of shorter-TE waveform designs [13] but no clinical TE difference between the schemes). The Siemens job aid recommends the monopolar scheme for short TE [20]. The choice is therefore between a shorter TE with more eddy-current distortion and a longer TE with less; the weight of each depends on the organ, the field strength and the gradient system (Reasoning).
Link to fat suppression. Slice-selection gradient reversal can be used only in sequences that contain two slice-selective 180° refocusing pulses, which is the structure of the twice-refocused scheme [15] (Section 9).
Console example. The owner's series uses Diffusion Scheme “bipolar” [25]; the job-aid example uses “Monopolar” [20]. What other manufacturers call these options is only partly documented: slides of a 2019 NIH course lecture (TORTOISE/AFNI) state that on Philips systems, on the latest release at the time, a “gradient duration” option corresponds to a twice-refocused scheme and that on GE systems the choice is between dual spin echo and spin echo [23]. The same slides advises setting TE to the minimum as the last step of setting up an acquisition and notes that partial Fourier and parallel imaging reduce TE [23]; these options are treated on 9563 and 9564.
7. Noise Floor, ADC Calculation and Thresholds
7.1 The noise floor
Noise in the complex MR signal is normally distributed, whereas in the magnitude image it follows a Rician distribution [17]. At high SNR the difference is negligible; at low SNR the magnitude signal does not fall to zero but to a floor set by the noise. At high b-values the signal of tissues with a high ADC (fluid, normal tissue that should be suppressed) falls toward that floor, and the decay curve flattens. A conference abstract on computed diffusion images notes that the Rician bias becomes dominant when the SNR is below about 2 [12]. A simulation with data near the noise floor gave a lower fitted ADC than the same tissue fitted from data well above the floor [9]. A figure legend of a 2019 European Radiology article shows a further consequence: because the noise floor changes the decay curves, the order of the measured ADC of two tissues can invert, depending on the signal at b = 0 rather than on diffusion, so that a low SNR may lead to misclassification of the diffusion level of a lesion [7]. Correction schemes for the Rician bias have been proposed (an early example is [18]); their availability on clinical consoles was not established (to be sourced).
7.2 What QIBA specifies
The QIBA profile requires the b = 0 image to have an SNR of at least 50 in the tissue region, a linear relation (slope close to 1) between measured and true ADC over a range of 0.3 to 3.0 × 10−3 mm²/s, and exclusion from the mean of voxels with ADC below 0 or above 3.0 × 10−3 mm²/s (at 37 °C) [1]. The article states no other noise-floor method [1]. The phantom checklist of the profile lists a maximum bias below 4% within 4 cm of the isocentre for the b-value pairs 0–500, 0–1000, 0–1500 and 0–2000 s/mm² [2] (search extract of the checklist only).
7.3 Thresholds on the console
An ADC calculation applied to every pixel, including the air and the low-signal background, produces random values; a threshold that excludes pixels below a signal level keeps the background out of the map. A Philips post-processing guide describes such a “threshold mask”: pixels with values below the mask value are excluded from the functional-map calculation and displayed in blue [22]. That is a workstation function, not a scanner parameter. The owner's Siemens series has “ADC Noise Threshold” set to 10 [25], and the job-aid example shows 40 [20]. The unit of the number (a signal level, a percentage of a reference, a multiple of the noise), the image it is applied to and whether it affects only the map or also the stored images are not stated in the sources read (to be sourced; the job aid itself lists the field without explanation [20]). Two statements follow independent of the definition (Reasoning): a threshold set too low leaves noise in the map; a threshold set too high removes true low-signal tissue (a lesion with short T2 and restricted diffusion at high b) from the map, which would then show no value where the diffusion image shows signal. The values 10 and 40 therefore cannot be compared between systems or software versions.
8. Computed (Calculated) b-Values
Method. An ADC map is calculated voxel by voxel from the acquired b-values and is used with the image at the lowest b-value to extrapolate an image at a higher b-value by the Stejskal–Tanner relation, under the assumption of the same exponential decay [11]. The method needs at least two acquired b-values, because an ADC needs two points (Reasoning).
Why it is used. For acquired images the noise is constant across b-values, whereas for computed images the noise decreases as the computed b-value increases [11]. In a phantom, computed images gave a higher SNR than acquired ones, especially above b = 840 s/mm² [11]; the exact crossover depends on the acquired pair and the tissue (Reasoning). In 10 oncology patients with whole-body diffusion at b = 0 and 900 s/mm², computed images at 1500 and 2000 s/mm² were generated, images at 2000 gave good image quality and background suppression (mean scores 2.8 and 4.0) and the authors concluded that clinical computed diffusion imaging is feasible and may improve disease detection [11]. A prostate study reported that computed images at b ≥ 1000 s/mm² improved image quality and tumour detection compared with acquired standard b-value images, gave greater suppression of benign tissue, less distortion and fewer artefacts than acquired images at the same b-value, and that sensitivity and positive predictive value of computed b = 1500 were comparable to those of direct b = 1500 and better than direct b = 1000 [10].
Limits. The workshop consensus states that synthesised high b-value images are extrapolations and should be used qualitatively [3]. A computed image carries no information that is not in the acquired data: it re-weights the same measurement with the ADC map, so the ADC map itself is unchanged (Reasoning from the method [11]). The extrapolation assumes the mono-exponential decay; where the decay is not mono-exponential (perfusion at low b, non-Gaussian diffusion at high b) the computed contrast departs from the acquired one (Reasoning). Studies are mostly of the prostate and of a few patients; none of the sources read is a multicentre trial.
Names. The Siemens cross-vendor acronym table lists the entries “Computed b-values”, “MAGiC-DWI”, “Computed DWI” and “cDWI (computed DWI)” for this function; the extract read does not allow a safe assignment of each to its manufacturer (to be verified in the original table) [21]. The Siemens console label seen is “Calculated b-value” (set to 1500 s/mm² on the owner's series) [25]; the job aid lists the same label and states that it does not explain how it works [20]. Which acquired b-values and which b = 0 image the console uses to calculate 1500 s/mm² is not stated (to be sourced).
9. Fat Suppression Inside a Diffusion Sequence
Fat has a very low apparent diffusion coefficient, so unsuppressed fat signal lowers the measured ADC of tissue in the same voxel; fat suppression is therefore essential in breast diffusion imaging [5]. The methods differ in what they cost:
| Method | Principle | Stated cost or limit | Source |
|---|---|---|---|
| STIR | Inversion with a TI that nulls fat; not specific to fat | Signal of other tissues is attenuated, SNR falls and ADC may be over- or underestimated; in breast diffusion imaging a fat-specific method (SPAIR) is preferred; in a musculoskeletal comparison STIR has long imaging time, low SNR and high SAR but low sensitivity to B0 and B1 | [5], [19] |
| Spectral inversion methods (SPIR, SPAIR) | Spectrally selective pulse with inversion | Sensitive to B0 inhomogeneity; SPAIR insensitive to B1; both long imaging time and high SAR in the musculoskeletal comparison | [19] |
| Spectral saturation followed by spoiling | Saturation of the fat resonance | Lengthens acquisition and increases SAR | [15] |
| Slice-selection gradient reversal | Opposite polarity of the slice gradients on the two refocusing pulses | Requires two slice-selective 180° pulses; suppressed fat completely without affecting water signal or slice profile in the original description | [15] |
The 2016 body-diffusion study that compared acquisition modes used SPAIR [8]. The owner's series uses, as magnetisation preparation, a slice-selective inversion recovery with TI 210 ms [25]. A TI of 210 ms nulls a tissue with T1 of about 210/ln 2 ≈ 300 ms for a fully recovered longitudinal magnetisation (Reasoning; the arithmetic assumes complete recovery, which the TR of 19.9 s allows). Because an inversion method nulls every tissue with that T1 at the chosen TI, the STIR caveats of the table apply to it in principle; the manufacturer's definition of this option and its behaviour in diffusion were not found (to be sourced).
10. Field and Eddy-Current Correction
Diffusion gradients induce eddy currents that distort echo-planar images in a way that depends on the gradient amplitude and direction, so images of different b-values and directions are distorted differently. Two kinds of remedy exist: waveform design (Section 6) and correction of the images. A study on a 3 T Siemens system in brain diffusion imaging compared a field-map-based correction with uncorrected echo-planar diffusion imaging and reported a significant reduction of geometric distortion in all evaluated brain regions and better conspicuity of ischaemic lesions [6] (abstract and methods extract only). The background of a patent text describes “dynamic field correction” as one image-based method used to reduce eddy-current artefacts [24]; the extract does not show the assignee or tie the term to a console label. The 2019 NIH course slides list, for GE systems, a real-time field adjustment option on software releases after a stated version [23]. The Siemens console label “Dynamic Field Correction” is on the project backlog list of labels for this page, but its state was not recorded in the transcription of the owner's diffusion series [25], and no source read defines what the option corrects, from which measurement, or whether it is the field-map correction of [6] (to be sourced). The echo-planar readout contribution to distortion (echo spacing, bandwidth, parallel imaging) is planned for a separate page.
11. Console Mapping Across Manufacturers
The table gives what the sources read allow. Empty cells mean that no source was found; they do not mean that the function is absent. Siemens entries are labels as seen on one console or in one job aid and are not definitions.
| Function | Siemens | GE | Philips | Status |
|---|---|---|---|---|
| Set of b-values and averages | Diff. Weightings; b-value and averages per weighting [20], [25] | Not found | Not found | Siemens transcribed and in job aid |
| Sequential orthogonal directions combined into a trace | Diffusion Mode “4-Scan Trace” [20]; “3-scan trace” in the literature [8] | Not found | Not found | Mapping of “4-Scan Trace” to tetrahedral not stated |
| Monopolar diagonal (net direction) | “3D Diagonal” [8], [20], [25] | “3-in-1” [8] | “Gradient overplus” [8] | Names from one paper |
| Eddy-current-balanced scheme | Diffusion Scheme “Bipolar” (seen) and “Monopolar” [20], [25] | Dual spin echo against spin echo [23] | “Gradient duration” option for twice-refocused [23] | Lecture slides from 2019; releases not specified in the extract |
| Computed high-b image | “Calculated b-value” [20], [25] | See Section 8 | See Section 8 | Column assignment of the cross-vendor table unresolved [21] |
| Exclusion of background from the ADC map | “ADC Noise Threshold” [20], [25] | Not found | “Threshold mask” in the post-processing application [22] | Definition of the Siemens field not found |
| Field and eddy-current correction | “Dynamic Field Correction” (label from the backlog; not read on the series) | “RT Field Adjust” option [23] | Not found | Equivalence between labels not established |
| Fat suppression in the diffusion sequence | Magnetization preparation “slice-selective IR” with TI [25]; SPAIR used on a Siemens system in [8] | Not found | Not found | Definition of the Siemens option not found |
12. Worked Example on the Owner's Console Series
The series is a 3 T head-and-neck diffusion acquisition with TR 19900 ms, TE 62 ms (greyed, minimum), field of view 350 mm × 81.3%, base resolution 128, slice thickness 4.0 mm, calculated b-value 1500 s/mm², two weightings (b = 500 with 2 averages, b = 1000 with 10 averages), ADC noise threshold 10, bipolar scheme, 3D Diagonal mode, slice-selective inversion recovery with TI 210 ms, acquisition time 5:06 [25]. The checks below are arithmetic on those values (Reasoning).
| Check | Calculation | Result | Reading |
|---|---|---|---|
| Relative SNR of the two weightings (assuming a tissue ADC of 0.8 × 10−3 mm²/s; per-acquisition SNR at b = 0 taken as 1) | b = 500: exp(−0.4) × √2; b = 1000: exp(−0.8) × √10 | 0.95 and 1.42 | Ten averages at b = 1000 give a higher SNR than two at b = 500, consistent with giving more averages to the higher b-value (Section 4) |
| Fat-nulling T1 implied by TI 210 ms | 210 / ln 2 | about 303 ms | Assumes complete recovery between inversions, which a TR of 19.9 s allows |
| Phase matrix | 128 × 0.813 | about 104 lines | Gives square 2.73 mm pixels (350/128 and 284.5/104); halved by interpolation to about 1.4 mm, which matches the displayed voxel; whether the EPI factor shown includes the parallel-imaging reduction (the series name suggests a factor of 2) is not resolved and belongs to the EPI page |
| Time against repetitions | 306 s / 19.9 s; 12 × 19.9 s | 15.4 repetitions; 239 s for the 12 listed averages | The remaining 67 s (about 3.4 repetitions) would fit a b = 0 acquisition with about three averages plus a few seconds of overhead, but the b = 0 averages were not recorded; the explanation is an inference, not a reading |
| Contrast of a computed image (tissue ADC 0.7 against 2.0 × 10−3 mm²/s) | Signal ratio exp(b × 1.3 × 10−3) | 3.7 at b = 1000; 7.0 at 1500; 13.5 at 2000 | Under the mono-exponential assumption the contrast ratio grows exponentially with b, whereas the signal of the low-ADC tissue falls from 0.50 to 0.35 to 0.25 of its b = 0 value; this is the trade that computed images address (Section 8) |
13. Parameter Decisions: Gain, Cost and Evidence
| Question | Change | Gain | Cost | Evidence |
|---|---|---|---|---|
| SNR at the highest b-value is too low | More averages at that b-value | SNR grows with the square root of the averages (Reasoning) | Time grows in proportion; the noise floor is not removed | Consensus on more averages at high b [3] |
| High-b contrast is wanted without the time | Computed b-value | Contrast of a higher b at the SNR of lower acquired b-values | No new information; qualitative use; mono-exponential assumption | Technical studies, mostly prostate [10], [11]; consensus caveat [3] |
| Distortion differs between directions or b-values | Bipolar or twice-refocused scheme | Less eddy-current distortion | Longer minimum TE (Reasoning) | Technical studies [14], [16] |
| A shorter TE is needed | Monopolar scheme, diagonal mode, partial Fourier, parallel imaging | Higher SNR | More eddy-current distortion; diagonal gain only 2–4 ms in the one study | Two-volunteer study [8]; vendor job aid [20] |
| The ADC map shows noise or unexpected values | Check SNR at b = 0 (at least 50), the threshold, the maximum b and exclusion of values outside 0–3.0 × 10−3 mm²/s | ADC closer to the true value | Possible loss of low-signal lesions from the map | QIBA [1]; simulations [9] |
| Serial examinations are compared | Keep scanner, b-values, averages and scheme identical | Basis for the profile claim | No room to optimise at follow-up | QIBA [1] |
14. Limits of the Evidence and Controversies
No guideline sets the parameters of this page, and the consensus documents allow ranges (a maximum b-value of 800–1000 s/mm² for the abdomen, a lower b-value below or above 100 s/mm² depending on the document) rather than single values [3], [4]. The comparison of trace and diagonal acquisition has two volunteers and no test–retest data, and found no SNR gain [8]; claims for diagonal acquisition beyond a small TE reduction are not supported by the sources read. Computed b-value images have been tested mainly in prostate and in small oncologic series, and the consensus restricts them to qualitative use [3], [10], [11]. The choice between STIR-type and spectral fat suppression is argued from breast and musculoskeletal sources, not from head-and-neck studies [5], [19]. The QIBA claims are for brain, liver, prostate and breast and require a fixed protocol [1]. The manufacturer material read is training material, not operator documentation [20]. The Taouli consensus was read through a summarising tool up to about 100,000 of 171,478 characters of the page, which included the abstract, the full text up to Table 1 and part of the references [3].
15. Open Gaps
The following were sought and not obtained; each is marked in the text as (to be sourced).
| Gap | What was tried | Result | Where it matters |
|---|---|---|---|
| Definitions of Siemens “ADC Noise Threshold”, “Calculated b-value”, “Dynamic Field Correction” and a noise-masking option | Web searches for the console labels; the Siemens Academy job aid read in full by a summarising tool | The job aid lists the first two fields and states it does not explain them; the operator manual is not public | Sections 7, 8, 10, 11 |
| GE and Philips console names for b-values, averages, trace modes and thresholds | Searches on vendor terminology; the Siemens cross-vendor table (extract); a Philips post-processing guide (extract); NIH lecture slides (extract) | Partial: monopolar diagonal names, twice-refocused options, one Philips post-processing mask | Section 11 |
| Quantified TE cost of bipolar against monopolar and twice-refocused schemes in clinical systems | Searches on Reese 2003 and later waveform designs | Only the qualitative motivation; the abstract page of the 2003 study returned no text | Section 6 |
| Full text of the 2009 consensus and of the 2011 computed-diffusion study | Direct fetch of the PubMed Central and publisher pages | A browser check (not circumvented) and a fetch failure; search extracts used | Sections 3 and 8 |
| QIBA acquisition specification (Appendix S1 of the 2024 article, and the profile's protocol section) | Full text of the article read through a summarising tool; search extracts of the profile documents | Appendix not part of the text retrieved | Sections 3 and 7 |
| Availability of Rician-bias correction on clinical consoles | Search on noise-floor correction and masking | Methods and patents found, no console documentation | Section 7 |
| Whether a b = 0 image and how many averages are acquired on the owner's series; the state of Dynamic Field Correction and of a noise-masking option | Inventory of the 47 screenshots in the project record | Not transcribed; the screenshots contain a patient name and were not re-opened for this page | Sections 3, 10, 12 |