Flip Angle, Refocusing Flip Angle, Hyperecho and Driven Equilibrium

MRIninja Knowledge Base | MRI Parameter Deep Dive Parent page: MRI Parameters — Overview and Classification (9501), Section 7.1 Builds on: RF Pulses and Pulse Sequence Physics (9204); Echo Spacing, Turbo Factor and Echo Trains in TSE (9561) Version 1.0 — November 2026

MRI Parameter Deep Dive

Flip Angle, Refocusing Flip Angle, Hyperecho and Driven Equilibrium

What a flip angle below 180° does to a spin-echo train: the echo pathways it creates, the RF power it saves, the TE that has to change with it, the side effects on magnetisation transfer, flow and fluid signal, how variable-angle schedules and hyperechoes are built, what a flip-back pulse at the end of the train restores, and which Siemens console labels are documented and which are not.


1. Scope, Boundary and Information Status

The overview page 9501 (Section 7.1) states that the flip angle is the angle by which a pulse tips the magnetisation, gives the Ernst-angle formula for gradient-echo sequences and notes that SAR scales with the square of the angle. This page adds what that section does not contain: the refocusing flip angle of a spin-echo train when it is not 180°, the echo pathways that result, the effective echo time, the trade between SAR and contrast, variable flip-angle schedules, hyperechoes, the flip-back pulse of driven equilibrium, and the console labels. Elsewhere on the site: the Ernst angle and the choice of excitation angle for steady-state signal are on 9204 (Section 3.2) and 9008; the choice of a high excitation angle to suppress stationary tissue in time-of-flight imaging is on 9598 (Section 4) and the ramped excitation profile on 9025; the overview of three-dimensional TSE with variable flip angles (SPACE, CUBE, VISTA) is on 9005 (Section 8.4) and its SAR comments at 3 T in Section 8.6; echo spacing, turbo factor and the T2 blurring of an echo train are on 9561; TR, TE and TI as controls on 9565; SAR physics and limits on 9210; the transmit chain that produces the angle on 9708.

This page is the home of the refocusing flip angle and the flip-back pulse as controls of a spin-echo train. Protocol values for a given examination belong to the protocol page of the region.

Information status. No guideline, consensus statement or systematic review defines how the refocusing flip angle should be set in clinical protocols. The evidence is a body of technical development studies: theory (Hennig and Scheffler, Alsop), volunteer series with a signal model (Weigel and Hennig, 16 volunteers and 5 patients), one prospective clinical series with a non-inferiority design (Keerthivasan and colleagues, 51 patients), one clinical comparison of hyperecho and conventional turbo spin echo in neuroradiology (Tetzlaff and colleagues), and conference abstracts. Most of the foundational work comes from a small number of groups, several of them working on one manufacturer's systems; this page keeps the physics vendor-neutral and gives console labels only in Section 11. Every source is listed in Section 17 with the extent to which it was read; several were read through a summarising tool, and equations that were garbled in the extract were re-derived or are given in words. Statements marked (Reasoning) are the page's own deductions from the cited formulas; statements marked (to be sourced) are claims for which a source was sought and not obtained (Section 15).

2. Definitions: Nominal and Actual Angle, and the Roles of RF Pulses in a Train

A flip angle is the angle through which a pulse rotates the magnetisation. The value typed on the console is the nominal angle: the amplitude and duration the system will transmit if the transmit calibration is correct and the radiofrequency field is uniform. The angle actually produced in tissue is scaled by the local transmit field (B1+). The clearest published demonstration in the sources read is a phantom measurement at 3 T in which B1+ was about 8% above nominal, so that a nominal 150° and a nominal 180° pulse acted as about 162° and 194° and behaved almost identically [2]. The transmit chain, the reference amplitude and the body coil are described on 9708.

In the pulse trains discussed here a pulse can play one of several roles, and the same word “flip angle” on a console may refer to different ones:

RoleClassical nominal valueWhat it doesWhere treated
Excitation90° in spin echo; a free parameter in gradient echoCreates transverse magnetisation; in gradient echo the angle also sets the partial saturation of the longitudinal magnetisation9204 §3; Section 10 here
Refocusing180° in classical spin echo and in conventional turbo spin echo; reduced in many turbo spin echo variantsForms the echo; below 180° it also creates stimulated-echo pathways (Section 3)This page
Inversion180° (adiabatic pulses are used in some sequences)Inverts longitudinal magnetisation before the readout9020; 9565 §8
Flip-back (driven equilibrium)−90° at the end of the echo trainRotates remaining transverse magnetisation back to the longitudinal axis [11]Section 9
T2 preparation+90°, 180°, −90° before the readoutImposes T2 weighting on the magnetisation before imaging; a different module from driven equilibrium [11]Section 9.4

In an ordinary 2D turbo spin echo the excitation is 90° and the refocusing pulses are nominally 180°, the case described on 9005. Reduced refocusing angles are the subject of Sections 3 to 8. The Siemens job aid on SAR reduction names a protocol “flip angle” and says that SAR scales with its second power, but the extract read does not state which pulse of a turbo spin echo that field controls [6]; the field name is therefore treated here as the nominal refocusing angle only where the sequence and the console confirm it (Section 11).

3. What Happens When the Refocusing Pulse Is Below 180°

3.1 Spin echo and stimulated echo from one pair of pulses

A 180° refocusing pulse flips the transverse magnetisation completely: every spin that dephased by field inhomogeneity is rephased and the whole transverse magnetisation contributes to the echo. A refocusing pulse smaller than 180° does two things to the magnetisation present just before it. The part that is refocused forms a spin echo; another part is rotated into the longitudinal direction and stored there, to be tipped back into the transverse plane by a later pulse, where it forms a stimulated echo. Mugler’s review of optimised three-dimensional fast spin echo gives the amplitudes of the pathways from a 90° excitation and refocusing pulses of angle θ: the fraction refocused as a spin echo is sin²(θ/2) and the fraction stored and recalled is ½ sin²θ [7]. At θ = 180° the first is 1 and the second is 0, which is why a train of ideal 180° pulses contains only spin echoes.

Refocusing angle θSpin-echo fraction sin²(θ/2)Stored-and-recalled fraction ½ sin²θConstant-angle pseudo-steady echo amplitude, approximately sin(θ/2)
180°1.0001.00
150°0.930.130.97
120°0.750.380.87
90°0.500.500.71
60°0.250.380.50

The first two columns evaluate the formulas of Mugler [7] for a 90° excitation followed by two pulses of angle θ; the last column evaluates the empirical relation quoted by Alsop, that the pseudo-steady-state echo amplitude is roughly the sine of half the refocusing angle [10]. The numbers in the table are computed by this page (Reasoning); the formulas are the sources’. The pathway fractions describe one pair of pulses and cannot be added to give the signal of a long train.

3.2 Why the stored part matters for contrast

Magnetisation that is stored along the longitudinal axis does not decay with T2; it relaxes with T1 while it is stored. A signal that is partly made of stimulated echoes therefore has a contrast that depends on T1 as well as on T2, and its amplitude at a given time differs from that of a pure spin echo. This is the central fact of the page: lowering the refocusing angle saves power (Section 5), but the signal formed is no longer a pure T2 spin echo, so the echo time shown on the console no longer corresponds to the same T2 weighting (Section 4) [4].

3.3 The echo train: the pseudo-steady state

In a train of many refocusing pulses of constant angle the echo amplitude does not simply decay: it falls from the first echo to a plateau, the pseudo-steady state, whose level depends on the angle, before relaxation takes over. Alsop analysed this for the Carr–Purcell–Meiboom–Gill train with T1 and T2 decay left out, noting that the stability of the train reflects the existence of a steady-state solution of the Bloch equations in that case, and compared the theory with phantoms and with the brain of one volunteer imaged with optimised single-shot low-angle sequences [10]. Two observations from the part of that paper that could be read are useful. First, a constant low angle is the least efficient way to use the available magnetisation, whereas a train whose first angles are high (near 180°) and then decrease reaches a chosen plateau more efficiently. Second, the theoretical plateau exceeded the empirical values at all angles; the author attributed the gap to an assumption, not justified in practice, that all the equilibrium longitudinal magnetisation passes into the pseudo-steady state [10]. The first observation is the origin of the variable-angle schedules of Section 7. Only pages 176–178 of the 9-page paper were available, so the experimental and discussion sections were not read (Section 15).

4. Constant Reduced Refocusing Angles: SAR, Effective TE and the T1 Contribution

4.1 The measured relations

Weigel and Hennig measured the contrast of turbo spin echo trains with constant refocusing angles of 180°, 150°, 120°, 90° and 60° and of several variable-angle schemes, in 16 healthy volunteers at 1.5 T and 3 T, at TR 5000 ms, echo spacing 8.9 ms and echo-train lengths of 15 and 25 [4]. They described the contrast of any train by one parameter, ft: the fraction of the time that the magnetisation spends as transverse magnetisation. In the model the signal follows exp(−ft·TE/T2) times a T1 term for the remaining time, and for T1 much longer than TE the train behaves like a fully refocused experiment with an effective echo time TEeff = ft × TE. The parameter depended only weakly on tissue (it differed by less than 1% between tissues, except 1.6% for the 90° train and 3.3% for the 60° train) and agreed between 1.5 T and 3 T; the predicted T2 contribution was within 1% of the measured one for all tissues [4]. The relative SAR of a train was defined as (1/ETL)·Σ(αi/180°)², normalised to a conventional train of ideal 180° pulses [4].

Constant refocusing angleRelative SAR reported (ETL 15, TE 80 ms)Simple (α/180°)²Measured ft at TE 80 msTEeff at TE 80 msTE needed for TEeff 80 ms
180°100%100%1 (by definition)80 ms80 ms
150°70%69%96.3%77 msabout 83 ms
120°46%44%87.7%70 msabout 91 ms
90°27%25%73.9%59 msabout 108 ms
60°13%11%57.3%45 msabout 140 ms

Columns 2, 4 and 5 are as printed in Tables 1 and 3 of Weigel and Hennig [4] (the ft values are those measured at 3 T; the 1.5 T values differ by 0.1 percentage point or less). Column 3 and column 6 are computed by this page: column 3 evaluates the simple square, and column 6 divides the target of 80 ms by ft (Reasoning). The reported relative SAR is slightly above the simple square at the lower angles; the extract does not state why.

4.2 What the table says

Three conclusions can be read from the table without further assumption. First, a reduction of the angle from 180° to 150° already removes about 30% of the RF power (the same figure is given by Mugler, who states that 150° against 180° reduces SAR by more than 30% [7]), while the effective TE falls by only about 4% at the same nominal TE. Second, below about 120° the effective TE falls steeply: at 60° a nominal TE of 80 ms produces the T2 weighting of a 45 ms echo. Third, to obtain the weighting of a given effective echo time the nominal TE has to be lengthened by roughly 1/ft, and that lengthening is not free: it changes the number of echo spacings in the train and reduces signal from tissues with short T1 and short T2 [4], and it lengthens the minimum TR for a given number of slices (Reasoning). The authors state the T1 contribution directly: with a constant 120° train about 13% of the signal comes from T1-dependent stimulated-echo effects [4].

4.3 Limits of the model

The calculation used hard (rectangular) pulses and neglected slice profiles and B1 variation; only two of the volunteers were women; the T1 ≈ T2 approximation used for CSF is not strictly valid although it worked; TE can be shifted only in multiples of the echo spacing, which leaves a small mismatch of effective TE; results come from a single slice, and the clinical check used five patients [4]. The signal model is also vendor-independent in principle, but all the measurements were made on one manufacturer’s systems.

5. SAR and the Refocusing Angle

5.1 The square law and what it includes

The RF energy delivered by a pulse scales approximately with the square of its flip angle, so a train of refocusing pulses is the main contributor to the SAR of a turbo spin echo. This relation is stated in the Siemens job aid (“SAR scales with the second power of the flip angle, so a small reduction can cut SAR significantly”) [6], in Mugler’s review [7] and in the definition of relative SAR used by Weigel and Hennig [4] and by Keerthivasan and colleagues (the sum of squared refocusing angles divided by the echo-train length) [1]. The same definition is the basis of the (Reasoning) evaluation in Table 4.1: for a constant angle, the relative SAR is (α/180°)².

The square law applies to the refocusing train only. The SAR of a complete acquisition also contains the excitation pulses, any inversion or fat-saturation pulses (the job aid states that inversion for STIR and adiabatic spectral inversion for SPAIR add a high amount of SAR, and that Dixon-based fat suppression increases SAR because two echoes are acquired with additional pulses), saturation bands, and the number of pulses per unit time set by TR, the number of slices, the concatenations and the turbo factor [6]. A reduction of the refocusing angle therefore reduces the total by less than the figure for the train alone when such pulses are present (Reasoning).

5.2 Other levers on SAR, and how they differ

The job aid lists the manufacturer’s levers in the order low-SAR RF pulses, longer TR, reduced flip angles, fewer slices, a lower turbo factor, a different fat-suppression method, fewer saturation bands, concatenations, patient positioning and the SAR assistant [6]. They are not equivalent. “RF Pulse Type: Low SAR” keeps the flip angle and lengthens the pulses to lower their power, and the job aid notes that this may need a slightly longer TE and TR; a longer TR spreads the pulses over time and states no contrast trade-off; a lower turbo factor at constant TR lowers SAR but increases scan time; a reduced flip angle lowers pulse amplitude and, as this page shows, changes the signal formed [6]. The job aid itself does not discuss effects of a reduced flip angle on contrast or SNR [6]; the effects are those of Sections 4 and 6.

5.3 What the SAR saving has been used for

In the volunteer series of Weigel and Hennig, the SAR of the variable-angle schemes ranged from about 29% to 55% of the 180° train in the table of the paper (that is, 45–71% lower) [4]; the abstract and introduction of the same paper speak of reductions of “at least 70%” or “on the order of 70%”, a discrepancy that is listed in Section 14. In the prospective spine series of Keerthivasan and colleagues, a variable-angle train with control angles of 130°, 45°, 110° and 45° against a constant 140° train lowered the scanner-reported SAR value (volunteers 1.69 ± 0.14 against 1.31 ± 0.21; clinical 1.84 ± 0.61 against 1.37 ± 0.34, P < .001; units not stated in the extract read) and allowed an echo train of 56 instead of 21, shortening the lumbar sequence from 3 min 25 s to 1 min 28 s [1]. At 4.7 T, Lebel and Wilman kept the short-term average RF power below 3 W/kg for all their scans with reduced and variable refocusing angles, including a single-shot sequence with an echo-train length of 109 [15]. The SAR limits themselves, and their dependence on field strength, are on 9210.

6. Consequences for Contrast, Magnetisation Transfer, Flow and Fluid Signal

6.1 T2 weighting and tissue contrast

Once the nominal TE is chosen so that the effective TE matches, the contrast between tissues in the volunteer series was reproduced: the paper shows matched-effective-TE images of a volunteer with dilated perivascular spaces and of a child with a pilocytic astrocytoma and states that they show identical contrast to the conventional images, but it reports no contrast-to-noise ratios [4]. In a clinical comparison of hyperecho turbo spin echo with conventional 180° turbo spin echo at 3 T in neuroradiology, the abstract reports equivalence for the contrast-to-noise ratio of pathologies and for the contrast between grey and white matter, and equal signal-to-noise ratio of grey matter, white matter and CSF [3]. The abstract and the beginning of the introduction were the parts read; the number of patients and the methods were not captured in the extract (Section 15).

6.2 Magnetisation transfer in multislice acquisitions

In a 2D multislice turbo spin echo the RF pulses for the other slices act as off-resonance saturation for the slice being read, producing an unintended magnetisation-transfer effect that lowers the signal of tissue with a large macromolecular content, so that contrast changes. Weigel and Hennig quantified this at 3 T as the ratio of the signal in a 5-slice acquisition to that of a single-slice acquisition, for constant angles of 180° to 60°, hyperecho and variable-angle trains, at echo-train lengths of 17 and 27 [18]. The ratio rose from about 0.75 at 180° to about 1.00 at 60°; the effect was inversely linear in the flip angle (correlation 0.998 for grey matter and 0.991 for white matter at an echo-train length of 27), it was larger in white matter than in grey matter, and it was negligible in CSF. For variable-angle trains the arithmetic mean of the angles could be used in the same linear relation (correlations 0.994 and 0.988) [18]. The authors noted that the linear behaviour was unexpected, since the literature links magnetisation transfer to RF power, which scales with the square of the angle, and announced a model and further experiments [18]; a later abstract from the same group describes a model fit of the effect for conventional and hyperecho trains at 1.5 T and 3 T, of which only a search extract was read [19]. At 4.7 T, Lebel and Wilman found that, compared with a single slice, the multislice signal at 180° fell by up to 25% in white matter and 15% in grey matter, and by less than 5% below 60° [15]. A reduced refocusing angle therefore removes part of the multislice magnetisation-transfer contrast of a 180° train, which changes the grey–white contrast independently of the T2 effect of Section 4 (Reasoning); the same sources observe that the effect depends on the angle used, not on a fixed property of the tissue.

6.3 Flow sensitivity

Storey and colleagues studied the flow sensitivity of fast spin echo for non-contrast peripheral angiography in a 3 T flow phantom and in three healthy volunteers at 1.5 T and 3 T (calf). They concluded that the flow sensitivity is governed largely by the refocusing flip angle: lower angles increase it, so that moving blood dephases more readily and arterial signal is lost; the spoiler gradient area had a smaller effect, and the echo time mattered mainly at low angles. They proposed that the cause is the mixing of spin-echo and stimulated-echo pathways, which produces more motion-induced dephasing at low angles [2]. In the volunteers large vessels were best seen at high angles and small branches at low angles, with many invisible at 170°; the optimal range in these subjects was 90°–120°, and ghosting appeared at N/8 pixels in the phase-encoding direction at 170° [2]. The authors caution that three subjects and no pathology limit generalisation [2]. The consequence for ordinary imaging (Reasoning) is that a reduced angle changes the appearance of vessels and of pulsatile CSF, independently of the T2 effect.

6.4 Fluid signal, B1 sensitivity and motion

In the prospective spine series the authors list as limitations a slight reduction of CSF signal from incomplete magnetisation recovery with the variable-angle train, lower signal-to-noise ratio at equivalent resolution, sensitivity of the minimum angle to motion (they chose a minimum of 45° to limit motion-related dephasing), and a possible loss of signal in very large patients from B1 inhomogeneity [1]. The extract of the paper also records that the motion advantage appeared only after nine subjects were excluded, so that the claimed motion robustness is not firmly established [1]. The low-angle sensitivity to B1 follows from the relation between the nominal and actual angle (Section 2) and from Table 3.1: the amplitude of a low-angle train changes faster with a given B1 error than that of a 180° train (Reasoning).

7. Variable Flip-Angle Schedules

7.1 The design problem

A constant low angle in a long train has a drawback that the first echoes are much stronger than the later ones and the plateau is low; a point-spread function with large oscillations follows. The schedules in use start with a high angle to bring the magnetisation to a chosen level, lower it to a minimum, raise it again around the acquisition of the centre of k-space, and let it fall towards the end of the train. Busse and colleagues parameterised the schedule by four control angles, αstart, αmin, αcent and αend (as described in the spine series that used them) [1]; an earlier paper from the same group designed such schedules for very long echo trains and introduced a contrast-equivalent TE to replace the conventional effective TE, which no longer corresponds to the T2 contrast when stimulated and spin echoes mix, together with the use of magnetisation preparation such as T2 preparation to shape the contrast [17]; a later paper describes direct control over the minimum, the centre-of-k-space and the maximum angle to obtain sharp point-spread functions from very long echo trains in three-dimensional fast spin echo [16]. The two Busse papers were available only as abstract extracts.

7.2 Published values

Source and settingAngles usedTrainStated purpose and result
Keerthivasan 2018 [1], 3 T, lumbar spine, TSE-VFAαstart 130°, αmin 45°, αcent 110°, αend 45°; comparison constant 140°ETL 56 (against 21 for the constant-angle comparison), effective TE 105 msScan time 3 min 25 s → 1 min 28 s; relative SAR lower (P < .001); non-inferior for all image-quality criteria except SNR
Keerthivasan 2018 [1], whole-spine single shot, HASTE-VFA130°, 50°, 90°, 45°Single shot1 min 57 s against about 9 min for STIR TSE; single-shot resolution lower (1.25 × 1.0 mm against 0.81 × 0.73 mm)
Lebel and Wilman 2008 [15], 4.7 T, brain, HASTESmooth ramp down to 55° over the first four pulsesETL 109, echo spacing 5.5 ms, TE 49.5 ms, 600 ms per sliceSlower signal decay at low angles allows very long trains; power below 3 W/kg
Lebel and Wilman 2008 [15], 4.7 T, brain, high-resolution stack100°, first pulse 140°ETL 8, echo spacing 20.5 ms, TE 41 ms, TR 6.1 s, 30 slices, 0.4 × 0.4 × 1 mm³6 min 30 s; the first pulse at 140° was used to boost signal and reduce oscillations
Mugler, review of 3D TSE [7] as cited by Weigel and Hennig [4]Angles as low as 20° in long trainsLong echo trains (3D)Not restated: see 9005 §8.4 for the three-dimensional overview

These are examples of what was published, with the field strength, region and sequence of each. They are not recommended protocol values, and the schedules are not unique: the spine authors state that their angles are one of many possible optimal solutions [1].

7.3 What the schedule changes

A schedule trades four things (Reasoning, summarising the sources above): the SAR (lower with lower average squared angle), the length of the echo train that can be used (longer when the plateau is controlled), the sharpness of the point-spread function (controlled by the angles around the centre of k-space and by the ramp at the start; non-optimised schemes can blur, as Weigel and Hennig note for the Busse-type scheme of about 150° until the centre followed by decreasing angles, which preserves pure T2 contrast but gives less SAR reduction and more blurring [4]), and the contrast (through the stimulated-echo fraction, Section 4). In the spine series the benefits were substantial in scan time and the cost was in SNR: the lumbar SNR was 55.45 ± 18.75 with the conventional train and 45.18 ± 14.96 with the variable-angle train (P = .006), and the SNR efficiency per unit time 29.93 against 37.27 [1]. The authors had only five volunteers for the validation and call for a larger clinical study [1]. The T2 blurring that is caused by the length of the train is on 9561.

8. Hyperechoes and TRAPS

8.1 The principle

A hyperecho is a refocusing mechanism in which a sequence of RF pulses arranged symmetrically around a central 180° pulse restores the full coherence of the magnetisation, so that the pulse train as a whole acts like a single 180° pulse: the pulses after the central one undo the rotations produced by the pulses before it [9]. Elster’s educational page states the same construction: a central 180° pulse flanked on each side by mirror-image pulses of reversed polarity and phase, with the formation of the echo depending on the symmetry and not on the exact angles or phases, so that low angles can be used through much of the train and the hyperecho can be timed for the centre of k-space [12]. For the symmetric arrangement of three pulses, the original paper reports that the intensities of the echoes formed add up to the original intensity when relaxation is neglected [9]. The same paper states that, for tissues with T1 longer than T2, a hyperecho and a conventional spin echo do not have the same T2 contrast at the same time, so longer echo times have to be used for identical T2 contrast: if the magnetisation spends a fraction d of the time along the longitudinal axis, the echo time must be lengthened by the factor 1/(1 − d), for example by 50% when d = 1/3 [9]. (The factor is given here as the summarising tool reported it from the paper; the equation was not seen printed, Section 15.)

8.2 TRAPS and the hyperecho turbo spin echo

The original hyperecho scheme is based on an antisymmetric arrangement of flip angles around a central 180° pulse, so that the magnetisation is completely refocused during acquisition of the central phase-encoding lines. TRAPS (transitions between pseudo-steady states) is a more general scheme in which the angles can be varied freely after a transition into a static pseudo-steady state, which allows the point-spread function to be shaped and, according to the authors, can improve sharpness relative to standard turbo spin echo [4]. Sequences built on these schemes are called hyperecho turbo spin echo or hyperTSE in the literature read [3], [4]; the Siemens cross-vendor terminology table lists “Hyperecho” as the Siemens name for hyperechoes with no entry for the other makers [23].

8.3 Evidence of equivalence and of SAR saving

Weigel and Hennig reported that a single effective echo time, calculated with the extended phase graph method, describes the T2 contrast of hyperecho turbo spin echo and is practically independent of field strength, so that matching TEeff restores equal contrast [4]. The SAR saving in their Table 1 was 45–71% for the TRAPS variants (relative SAR 29–55%), and the effective TE was reduced by as much as about 30% at the same nominal TE (for example, 92 ms for a TRAPS variant at a nominal 134 ms); their design procedure starts with the desired TEeff as TE and a TRAPS ramp to 60°, then increases TE iteratively until TEeff matches the target [4]. The clinical comparison in neuroradiology at 3 T reports equivalence with 180° turbo spin echo for the contrast-to-noise ratio of pathologies, for grey–white contrast and for the signal-to-noise ratio of grey matter, white matter and CSF [3]. The evidence therefore supports equivalence of contrast when TE is adjusted; it does not support using the unchanged nominal TE of a 180° protocol [4].

8.4 Limits of the evidence

The two studies are from the same group; the volunteer series is small and single-slice, the clinical study was read at abstract level only, and no independent multicentre comparison was found (to be sourced). The extended phase graph paper that gives the calculation of the angles for hyperecho and TRAPS trains (Hennig, Weigel and Scheffler, Magn Reson Med 2004;51:68–80) is cited by others but was not read.

9. Flip-Back Pulse and Driven Equilibrium

9.1 Mechanism

Driven equilibrium, also called fast recovery, applies a −90° pulse at the end of the echo train to rotate the transverse magnetisation that remains back to the longitudinal axis, restoring longitudinal magnetisation before the next excitation [11]. It is most useful for fluids with a long T1 and a long T2 (CSF, urine, bile): at a short TR such fluids do not recover and their normally bright T2-weighted signal falls, whereas at the end of a long echo train they still hold appreciable transverse magnetisation that the flip-back pulse converts into usable longitudinal magnetisation [11]. Spoiler gradients after the flip-back pulse dephase what is left of the transverse magnetisation [11]. The page of Elster stresses that driven equilibrium does not shorten the true T1: it recycles magnetisation that would otherwise be lost, and it works best when substantial coherent transverse magnetisation remains at the end of the train [11].

9.2 Contrast and use

Tissues with a short T2 have already lost their transverse magnetisation by the end of the train and gain much less [11]; the effect on contrast is therefore to lift fluid relative to solid tissue at a short TR. The page lists use in spinal imaging, MR urography and MR cholangiopancreatography [11]. The Siemens glossary describes the same mechanism under the name of the restore sequence: in a T2-weighted turbo spin echo a 90° pulse at the end of the echo train boosts the signal of long-relaxing substances such as CSF and allows a shorter TR and a shorter measurement time [13]. For the cervical spine, a technical development paper in Radiology reports the use of a driven-equilibrium pulse at the last echo of a three-dimensional fast spin echo to improve recovery of longitudinal magnetisation (the extract read contains the title, the schematic and the caption of an image pair of a volunteer acquired with and without the pulse) [20], and a review in Applied Radiology describes it as a resonant 90° pulse applied when the last 180° pulse has refocused the transverse magnetisation, restoring CSF signal and the myelographic effect at a short TR [22]. A retrospective cross-sectional study of spondylotic myelopathy found no significant differences between three-dimensional driven-equilibrium fast spin echo and two-dimensional fast spin echo in the number, area or transverse dimensions of cord lesions (extract of the abstract only; authors and design details not read) [21].

9.3 A caveat: inversion-recovery sequences

The sign of the flip-back pulse matters when the fluid magnetisation is not on the positive longitudinal axis. In short-TR STIR the fluid magnetisation lies on the negative axis when the next excitation arrives, and the conventional −90° pulse moves the remaining transverse magnetisation onto the negative axis as well, which, according to the authors, delays longitudinal recovery and attenuates the fluid signal; a modified +90° flip-back (a 180° phase change) was proposed to avoid that [14]. In one healthy volunteer at 3 T (lumbar sagittal), the CSF-to-muscle ratio in 2D was 2.7 ± 0.3 without driven equilibrium, 2.0 ± 0.2 with the conventional pulse and 3.9 ± 0.4 with the modified one; in 3D the modified pulse gave 7.8 ± 0.7 against 2.8 ± 0.2 for the reference approach [14]. The abstract is a single-subject report with manual regions of interest, no statistics and a 3D comparison whose reference is labelled inconsistently in the text and the figure caption, and TE, echo-train length and TI were in a table not available as text; it is evidence that the effect can occur, not that it occurs in every implementation [14].

9.4 Driven equilibrium is not T2 preparation

A T2-preparation module is a train of three non-selective pulses (+90°, 180°, −90°) applied before the imaging sequence to impose T2 weighting; driven equilibrium acts after the readout to recover signal [11]. The two should not be confused on a protocol: the first changes contrast, the second recovers magnetisation.

9.5 Names across makers

The sources agree on the Siemens name (RESTORE), the GE name (fast recovery fast spin echo, FRFSE) and the Philips name (DRIVE). They differ on the remaining makers: the mrimaster table lists “T2 Plus FSE” under Toshiba and “Driven Equilibrium” under Hitachi [23]; the Elster page lists “T2 Plus FSE” under Fujifilm and “Driven equilibrium FSE” under Hitachi [11]. The page does not resolve the attribution (Section 14). The generic acronym DEFT is sometimes used for driven equilibrium Fourier transform, though Elster notes that it historically refers to a broader family of methods [11].

10. A Note on the Flip Angle in Gradient-Echo Sequences

For gradient echo the excitation angle is the main variable and is covered on 9204 and 9008; only the points that complete this page are given. In spoiled gradient echo the signal at a given TR peaks at the Ernst angle, but contrast is poor or absent there; larger angles give T1 weighting and smaller angles give proton-density weighting, and in a spoiled FLASH series the signal typically peaks between 30° and 60° [5]. As a computed example (Reasoning), for TR 5 ms and T1 900 ms the formula arccos(e−TR/T1) gives about 6°, so a clinical angle of 10°–15° at that TR lies above the Ernst angle and trades some signal for T1 contrast. In balanced steady-state free precession the contrast is weighted by T2/T1 and the optimum angle depends on T1 and T2; for similar T1 and T2 (CSF, fat) the optimum is around 70°–90° with a signal approaching half of M0, the transient lasts of the order of 50–500 pulses, an α/2 preparation pulse and linearly ramped angles over the first 5–15 pulses reduce the oscillations, and the angles of 50°–80° needed for maximum signal can exceed SAR limits, particularly above 1.5 T [8]. These are the same mechanisms (angle sets the partition of the magnetisation between axes, SAR scales with its square) in a different sequence family; this page does not give guidance on those sequences.

11. Console Labels (Siemens) and Cross-Vendor Names

The labels below are those that the sources name or that appear on the console screenshots supplied by the site owner (one 3 T head-and-neck examination, received 2026-11-23; the images contain a patient name and are not published) [25]. “Status” says whether a source read defines the label. The first column gives the label as printed; no meaning is assumed for a label that no source defines.

LabelWhat the sources saySourceStatus
Flip Angle (protocol parameter)Controls pulse amplitude; SAR scales with its second power. The aid does not say which pulse of a turbo spin echo it controls. Its worked example changes a flip angle from 90° to 65° and TR from 600 ms to 800 ms in a TSE protocol with turbo factor 15[6]Partly documented
Flip Angle ModeSeen on the owner’s screenshots. Its options and its meaning are not in the glossary, the two job aids or the cross-vendor tables; the SAR job aid states explicitly that it does not mention hyperecho, RESTORE or a flip-angle mode[25], [6]Not documented (to be sourced)
“Hyperec…” (truncated on the screenshot)The cross-vendor table lists “Hyperecho” as the Siemens name of hyperechoes. That the truncated label is this option is an inference from the beginning of the word (Reasoning)[25], [23]Probable, unconfirmed
Wrap-up Magn.Seen on the owner’s screenshots. Not defined in any source read. The name suggests recovery of magnetisation at the end of the echo train, but it must not be taken as a synonym of RESTORE without a source (Reasoning)[25]Not documented (to be sourced)
RESTORESiemens name for the 90° flip-back pulse at the end of a T2-weighted TSE echo train; it boosts signal of long-relaxing substances such as CSF and allows a shorter TR and measurement time. Equivalent names are listed in Section 9.5[13], [23]Documented
RF Pulse Type: Low SARLonger pulses with lower power at the same flip angle; may need a slightly longer TE and TR. A different lever from reducing the angle[6]Documented
SAR Assistant (Off, TR, Flip Angle, Flip Angle TR, Allowed Delay TR Flip Angle)The options are listed in the job aid without explanation of each. An extract of a later SAR-assistant job aid says that the parameter card can set a maximum TR or a minimum flip angle. The assistant can therefore change the angle of a protocol without the operator typing it (Reasoning)[6], [24]Partly documented
SPACEThree-dimensional TSE variant with non-selective short refocusing pulse trains of variable flip angles and turbo factors above 100; see 9005 §8.4[13]Documented
HASTESingle-excitation half-Fourier TSE. The glossary describes the echoes as generated by subsequent 180° pulses; reduced-angle single-shot versions are described in the literature (Section 14)[13], [1], [15]Documented, with a caveat

Cross-vendor names of 3D TSE with variable flip angle are listed as SPACE (Siemens), CUBE (GE), VISTA (Philips) and mVox (Toshiba) in the cross-vendor table [23]; the nature of the schedule in each product is not described in the extract read. Request to the site owner: screenshots of the open option lists of “Flip Angle Mode” and of the hyperecho and “Wrap-up Magn.” fields, so that the labels can be recorded exactly as the console prints them.

12. Practical View for the Technologist

The following points are derived from the sources of Sections 2 to 9; those that go beyond them are marked as reasoning.

  • The typed angle is nominal. A B1+ error maps directly into an angle error; a nominal 150° and 180° acted as about 162° and 194° in one 3 T phantom [2]. Low-angle trains are more sensitive to such errors than 180° trains, and the authors of the spine series name B1 inhomogeneity in very large patients as a possible cause of signal loss [1].
  • A lower refocusing angle is not a free reduction of SAR. It lowers the effective TE for the same nominal TE (Table 4.1), adds a T1 contribution, reduces multislice magnetisation-transfer contrast, increases flow sensitivity and, with a variable schedule, can reduce CSF signal and SNR [1], [2], [4], [15], [18].
  • Do not compare TE values between protocols with different refocusing schemes. The “effective TE” defined as the time at which the centre of k-space is sampled no longer corresponds to the expected spin-echo T2 contrast when spin and stimulated echoes mix; Busse and colleagues therefore introduced a contrast-equivalent TE [17], and Weigel and Hennig a TEeff = ft × TE [4].
  • If SAR is the limit, compare the levers of Section 5.2 before changing the angle: a low-SAR pulse type, a longer TR, fewer slices, a lower turbo factor, fewer saturation bands or a different fat-suppression method, each with its own cost [6].
  • When a protocol contains a flip-back pulse, expect brighter CSF, bile or urine at a short TR and little change in solid tissue [11]; in an inversion-recovery sequence the sign of the pulse matters (Section 9.3) [14].
  • Record the refocusing scheme in the protocol notes (the angle or the mode name, and whether a flip-back option is on), because the console name of the option is not self-explanatory (Section 11) (Reasoning).

13. Worked Example: Matching the T2 Weighting of a 180° Train

Take the base protocol of the volunteer series: TR 5000 ms, echo spacing 8.9 ms, echo-train length 15, nominal TE 80 ms with 180° refocusing pulses (so TEeff = 80 ms) [4]. The numbers below use the measured ft values of Table 4.1 and are computed by this page (Reasoning).

SchemeRelative SARTEeff if nominal TE is kept at 80 msNominal TE needed for TEeff 80 msNearest echo times available (multiples of 8.9 ms) and the TEeff they give
Constant 180°100%80 ms80 ms9 × 8.9 = 80.1 ms → 80.1 ms
Constant 150°about 70%77 msabout 83 ms80.1 ms → 77.1 ms; 89.0 ms → 85.7 ms
Constant 120°about 46%70 msabout 91 ms89.0 ms → 78.1 ms; 97.9 ms → 85.9 ms
Constant 60°about 13%45 msabout 140 msNot practical: nominal TE nearly twice the target, so signal of tissues with short T2 and short T1 is lost [4]

Two conclusions follow. At 150° the T2 weighting is nearly that of the 180° train, and the quantisation of TE in steps of the echo spacing (a limitation the authors list [4]) leaves a mismatch of the same order as the effect being corrected. At 120° and below, TE has to be lengthened considerably, which is why Weigel and Hennig designed their variable-angle schemes (starting from the desired TEeff and a ramp to 60°, then increasing TE until TEeff matched) rather than using constant low angles [4]. The SAR saving is real in each case, but the contrast that is obtained at an unchanged nominal TE is not the contrast of the 180° protocol.

14. Discrepancies and Controversies in the Sources

  • Size of the SAR reduction with variable-angle trains. The abstract of Weigel and Hennig speaks of reductions of “at least 70%” and the introduction of “on the order of 70%”, while Table 1 shows relative SAR of 29–55% for the TRAPS variants (reductions of 45–71%), with 70% or more only for the lowest-SAR variants [4]. Not resolved.
  • Relative SAR in the spine series. The text defines relative SAR with squared angles, while the displayed equation omits the squaring; the abstract rounds the scan-time gain to 2× while the printed times give about 2.3×; Table 7 gives a mean SNR score of 4.34 where the abstract says all scores were at least 4.4 [1].
  • The hyperecho paper as read. The text as captured mentions 17 and 18 echoes and echo times of 70 and 84 ms in a way that could not be reconciled from the extract; the discrepancy may be in the paper or in the extraction and should be checked against the printed paper [9].
  • Theory against empirical amplitudes. Alsop’s theoretical pseudo-steady-state amplitude exceeds the empirical values at all angles, explained by an assumption about the transfer of equilibrium magnetisation [10].
  • Magnetisation transfer. The effect was linear in the flip angle although the literature links it to RF power, which scales with the square; a model was announced and not read [18].
  • Relative SAR above the simple square. The relative SAR of constant 120° and 90° trains in Table 1 is 1–2 percentage points above (α/180°)²; the extract does not say why [4].
  • Names of the flip-back sequence. “T2 Plus FSE” is attributed to Toshiba in one table and to Fujifilm in another (Section 9.5) [11], [23]. Not resolved.
  • HASTE. The glossary describes the echoes of HASTE as produced by 180° pulses, while the literature includes single-shot sequences with refocusing angles as low as 45°–55° [1], [13], [15]. The two are consistent if the glossary describes the generic sequence and the options on a given console vary (Reasoning).
  • Driven equilibrium in STIR. The 3D comparison in the abstract is labelled “state-of-the-art approach” in the text and “no driven equilibrium” in the caption [14].
  • Correction on other pages. In preparing this page the Ernst-angle examples on pages 9008 and 9501 were recomputed from the formula: for TR 5 ms the angles are about 6° (T1 900 ms) and about 11° (T1 250 ms), not 11°–12° and 20° as printed; the earlier numbers correspond to a TR of about 15–20 ms. Both pages were corrected in the same release.

15. Open Questions, Evidence Gaps and Sources Attempted

15.1 Open questions

No source read gives a clinical threshold for the lowest refocusing angle that keeps lesion conspicuity equal to that of a 180° train; the studies report image-quality scores and contrast ratios, not diagnostic accuracy for small lesions [1], [3], [4]. The behaviour of low-angle trains where B1+ is strongly non-uniform (large body parts at 3 T) is mentioned as a concern and not measured in the sources read [1]. Whether the console’s “Flip Angle Mode” corresponds to any of the schedules in Section 7, and what “Wrap-up Magn.” changes, is unknown to this page (to be sourced).

15.2 Sources sought and not fully obtained

ItemWhat was triedResultConsequence for the page
Meaning of “Flip Angle Mode”, “Hyperec…”, “Wrap-up Magn.”Siemens MR glossary (whole text read for page 9565; re-read for these terms in the first 100,000 of 127,298 characters), the two SAR job aids, cross-vendor tables, web searchesNot defined; the SAR job aid states that it does not mention hyperecho, RESTORE or flip-angle modeLabels listed as seen, with no meaning assumed (Section 11)
Busse 2006 and 2008 (variable-angle schedule design)Publisher and mirror pagesAbstract extracts only; one mirror unavailable in the regionDescribed at abstract level, and through the spine series that used the scheme [1], [16], [17]
Alsop 1997An exhibit copyPages 176–178 of 176–184; the rest behind a paywallTheory only; no experimental numbers [10]
Hennig, Weigel, Scheffler 2003 (TRAPS) and 2004 (extended phase graph)Cited references onlyNot readTRAPS described through [4] and [12]
Melhem 2001 (Radiology) and the 2014 PLOS ONE comparisonPublisher page and a PMC pagePublisher page refused; the PMC page returned a browser check, which was not circumventedExtracts only [20], [21]
Tetzlaff 2008Abstract page and PDF extractsAbstract and the beginning of the introductionResults quoted at abstract level [3]
Independent clinical trials of hyperecho or variable-angle TSESearchesNone found beyond the cited series (to be sourced)Evidence level stays technical and observational

16. Where Next

Pages that would complete this topic and are not yet written: magnetisation transfer in multislice imaging (a home for Section 6.2); SAR management at protocol level (a single page that orders the levers of Section 5.2 with their consequences); and a page on 3D TSE with variable flip angles (a home for the details of the schedules that are now summarised on 9005 §8.4). The console labels of Section 11 will be completed when the option lists are supplied.

17. References

Prospective clinical study
[1] Keerthivasan MB, Winegar B, Becker JL, Bilgin A, Altbach MI, Saranathan M. Clinical utility of a novel ultrafast T2-weighted sequence for spine imaging. AJNR Am J Neuroradiol. 2018;39(8):1568–1575. doi:10.3174/ajnr.A5713.
Full text through a summarising tool (figure images not seen). Phantoms, 5 volunteers, 51 prospective patients, 3 T; non-inferiority design with two blinded readers. Internal inconsistencies are listed in Section 14.
Experimental study
[2] Storey P, Atanasova IP, Lim RP, Xu J, Kim D, Chen Q, Lee VS. Tailoring the flow sensitivity of fast spin-echo sequences for noncontrast peripheral MR angiography. Magn Reson Med. 2010;64(4):1098–1108. doi:10.1002/mrm.22510.
Full text through a summarising tool. One 3 T flow phantom and three healthy volunteers (calf) at 1.5 T and 3 T; no patients.
Clinical comparison
[3] Tetzlaff RH, Mader I, Küker W, Weber J, Ziyeh S, Schulze-Bonhage A, Hennig J, Weigel M. Hyperecho-turbo spin-echo sequences at 3T: clinical application in neuroradiology. AJNR Am J Neuroradiol. 2008;29(5):956–961. doi:10.3174/ajnr.A0971.
Abstract and the beginning of the introduction only (search extracts of the full-text PDF); number of patients and methods not captured. Same group as [4], [9] and [18].
Experimental study with signal model
[4] Weigel M, Hennig J. Contrast behavior and relaxation effects of conventional and hyperecho-turbo spin echo sequences at 1.5 and 3 T. Magn Reson Med. 2006;55:826–835.
Full text through a summarising tool, from a copy hosted on mriquestions.com; several equations and table cells garbled and reconstructed from the prose; 16 volunteers, 5 patients, single slice, one manufacturer’s systems.
Online book
[5] Rinck PA. Magnetic Resonance in Medicine – The Basics, Chapter 10 “Main contrast factors in MR imaging”, page 10-06 (gradient echo sequences). © 2024 Rinck and The Round Table Foundation.
One chapter page through a summarising tool; the printed signal equation is inconsistent with its stated condition; three tables are images and were not read.
Manufacturer job aid
[6] Siemens Healthcare GmbH. Tips and Tricks for SAR and Stimulation Reduction, MAGNETOM Systems. Job aid effective 2 March 2021 (US version 18 March 2021).
Page text through a summarising tool (the same document as reference 12 of page 9565); screenshots seen as text only; it does not mention hyperecho, RESTORE or a flip-angle mode.
Review
[7] Mugler JP III. Optimized three-dimensional fast-spin-echo MRI. J Magn Reson Imaging. 2014;39:745–767.
Read in parts through a summarising tool; formulas given in words.
Review
[8] Scheffler K, Lehnhardt S. Principles and applications of balanced SSFP techniques. Eur Radiol. 2003;13:2409–2418. doi:10.1007/s00330-003-1957-x.
Full text through a summarising tool; the equations are missing from the supplied text, so only the statements in the prose are used.
Technical study (theory)
[9] Hennig J, Scheffler K. Hyperechoes. Magn Reson Med. 2001;46:6–12.
Full text through a summarising tool; equations not seen printed; some numerical statements could not be reconciled (Section 14).
Technical study (theory and phantoms)
[10] Alsop DC. The sensitivity of low flip angle RARE imaging. Magn Reson Med. 1997;37:176–184.
Pages 176–178 only, from a copy filed as an exhibit in a legal proceeding; the remaining pages were behind a paywall; equations garbled by optical character recognition.
Educational web page
[11] Elster AD. Driven equilibrium (fast recovery) pulses. MRIquestions.com (Questions and Answers in MRI).
Page text through a summarising tool; the collapsed advanced section was only partly captured; undated; the page lists its own references (Becker 1969, van Uijen 1984, Melhem 2001, Kim 2017).
Educational web page
[12] Elster AD. Hyperechoes. MRIquestions.com (Questions and Answers in MRI).
Complete page through a summarising tool; no numerical examples; undated.
Manufacturer glossary
[13] Siemens. MR Glossary (copy hosted at a university address; no date or edition shown).
Entries read through a summarising tool; whole text read for page 9565 and re-read for the terms of this page in the first 100,000 of 127,298 characters.
Conference abstract
[14] von Deuster C, Nanz D. Driven-equilibrium STIR with short TR (title as rendered by the summarising tool). Proc Intl Soc Magn Reson Med. 2023; abstract 0625.
Abstract through a summarising tool; one healthy volunteer; the table of TE, echo-train length and TI was an image and was not read.
Conference abstract
[15] Lebel RM, Wilman AH. Efficient multi-slice fast spin echo imaging with reduced flip angles at high field. Proc Intl Soc Magn Reson Med. 2008; abstract 3510.
Abstract through a summarising tool; 4.7 T; number of subjects not stated; no SAR figures beyond the stated power limit.
Abstract (extract)
[16] Busse RF, Brau ACS, Vu A, Michelich CR, Bayram E, Kijowski R, Reeder SB, Rowley HA. Effects of refocusing flip angle modulation and view ordering in 3D fast spin echo. Magn Reson Med. 2008;60(3):640–649. doi:10.1002/mrm.21680.
Search extract of the abstract of the author manuscript only.
Abstract (extract)
[17] Busse RF, Hariharan H, Vu A, Brittain JH. Fast spin echo sequences with very long echo trains: design of variable refocusing flip angle schedules and generation of clinical T2 contrast. Magn Reson Med. 2006;55:1030 (end page not shown in the extracts).
Abstract extracts only.
Conference abstract
[18] Weigel M, Hennig J. Evaluation of magnetization transfer effects of lower flip angle TSE sequences at 3T. Proc Intl Soc Magn Reson Med. 2005; abstract 2222.
Abstract through a summarising tool; one 3 T system; ROIs in grey matter, white matter and CSF; some labelling inconsistent in Table 1.
Conference abstract (extract)
[19] Hennig J, Weigel M. Investigation and modeling of magnetization transfer effects of conventional and hyperecho TSE sequences at 1.5T and 3T. Proc Intl Soc Magn Reson Med. 2006; abstract 3501.
Search extracts only; the author order differs between the archive page (Hennig, Weigel) and the PDF (Weigel, Hennig).
Technical development (extract)
[20] Melhem ER, Itoh R, Folkers PJM. Cervical spine: three-dimensional fast spin-echo MR imaging — improved recovery of longitudinal magnetization with driven equilibrium pulse. Radiology. 2001;218(1):283–288.
Search extract (title, schematic and image caption) only; the publisher page was not retrieved.
Retrospective study (extract)
[21] Characteristics of spondylotic myelopathy on 3D driven-equilibrium fast spin echo and 2D fast spin echo magnetic resonance imaging: a retrospective cross-sectional study. PLoS One. 2014 (published 15 July 2014). doi:10.1371/journal.pone.0100964.
Abstract extract only; authors not shown; the PMC full-text page returned a browser check that was not circumvented.
Review article (extract)
[22] Lighvani [initials not shown], Melhem ER. Advances in high-field MR imaging of the spine. Applied Radiology (date not shown).
Search extract only; the two authors are identified from the author note in the extract.
Terminology tables (extract)
[23] Siemens Healthineers. MRI acronyms: cross-vendor comparison (PDF); and mrimaster.com, MRI cross vendor terminology (MRI acronyms).
Search extracts of the tables; the two tables differ on some names (Section 9.5).
Manufacturer job aid (extract)
[24] Siemens Healthineers Academy. SAR Assistant – syngo MR XA51 (job aid; effective dates shown in the extract: 10 Aug 2021 and 3 Sep 2022).
Search extract only.
Owner’s data
[25] Siemens console screenshots of one 3 T head-and-neck examination, received 2026-11-23 (47 images).
Used through the inventory table of the project record, not re-opened for this page; the images contain a patient name and are not published.

Recent PubMed search for this protocol

Last updated: November 2026