Pre-scan Adjustments: Centre Frequency, B0 Shim, Transmit Reference Amplitude, Receiver Gain and Image Scaling

MRIninja Knowledge Base | MRI Parameter Deep Dive Parent page: MRI Parameters — Overview and Classification (9501) Builds on: RF Transmit System and Body Coil (9708); MRI Shim System Hardware (9706) Version 1.0 — November 2026

MRI Parameter Deep Dive

Pre-scan Adjustments: Centre Frequency, B0 Shim, Transmit Reference Amplitude, Receiver Gain and Image Scaling

What the scanner calibrates for every patient and series before the sequence starts, what the operator can change on the adjustment and transmit/receive cards of the protocol, and what goes wrong when a calibration is wrong: the water peak and the fat and silicone offsets, the shim volume and shim modes, the transmit reference amplitude and the two-channel B1 shim, receiver gain against image scaling, and the cases that need a manual check (breast, neck, off-centre anatomy). Console names are given as a mapping table; the meaning of several options is not documented in the sources read and is marked.


1. Scope, Boundary and Information Status

The hardware of the shim system (stages, coil orders, currents, switching) is on 9706; the transmit chain, the body coil, the two-channel drive and the physical principle of the transmit calibration are on 9708 (Section 7); field-strength effects on homogeneity are on 9208; the effect of a flip-angle error on contrast is on 9204 and, for gradient-echo sequences, on 9566; fat-suppression methods are on the fat and water pages listed in Related Content, and their dependence on B0 is summarised here only for the centre frequency. This page adds the operator-facing side: which quantities are set before the series, which are visible and adjustable at the console, and what they change in the image.

Information status. No guideline or consensus statement defines pre-scan settings. The generic principles rest on an educational abstract by a scanner-physics author [1], a question-and-answer text [6], [7] and a review of shimming methods [10], [11], [12]. The console options rest on manufacturer training material (job aids) of one vendor [2], [3], [4], [5], [17]; these are training documents and one of them states that the operator manual is the authoritative reference [3]. The operator manual was not available. The image-scaling section rests on one peer-reviewed study [8]. Two patent texts are used only for what they state as background [18], [19]. Several sources were read only through search extracts; each is marked in Section 12. Console names for GE and Philips were not found and are not given.

Protocol values for an organ belong to the protocol page of the region. This page contains no settings to copy.

2. What the Pre-scan Sets and Why It Is Patient-Specific

Scanners are first calibrated with phantoms for effects that do not depend on the patient (large-scale B0 inhomogeneity, eddy currents, RF output power, gradient gain, gradient non-linearity); the patient-dependent quantities are then set at the pre-scan: the transmit gain, the receiver gain and the centre frequency, together with the shim [1]. A question-and-answer text lists the typical steps as quick shimming, coil tuning and matching, centre frequency, transmitter gain, receiver gain and dummy cycles to approach steady state [6]. The process is normally automatic and can be run manually when needed, for example for spectroscopy, chemical-shift fat suppression or imaging near strong susceptibility variation [6].

QuantityWhat is setConsequence of a wrong valueSource
Centre frequencyOperating frequency of the protons of interest; usually the water peakSlice-direction localisation error; in EPI a shift of the object in the phase-encoding direction; banding in balanced SSFP; blurring in spiral imaging; poor fat suppression with some methods[1]
Transmit gain (reference amplitude)RF power that gives the intended flip angle at the calibration locationLoss of signal and contrast, poor uniformity, poor fat saturation[1]
Receiver gainAttenuation of the analogue signal before digitisationToo high: saturation of the converter with shading and intensity distortion. Too low: quantisation noise, added noise and artefacts[1]
ShimCurrents of the active shim channels that reduce B0 inhomogeneity in the chosen volumeDistortion in EPI, failure of chemical-shift fat suppression, line broadening in spectroscopy[6], [11], [12]

The transmit gain depends on patient size and composition and is therefore calibrated at the scanning location; recalibration is needed only if the loading changes substantially [1]. The numerical tolerances and the criteria used by each platform to decide when to repeat the calibration were not found in the sources (to be sourced). A per-patient reference transmit level is described in a patent text as the value from which the transmitter gains of all later pulses are calculated [19], which is consistent with the statement that other pulses are handled by scaling from a calibration pulse [1].

3. Centre Frequency: Water Peak, Fat Offset, Silicone Offset and the Confirm Step

3.1 Principle

Water and fat protons resonate a few hundred hertz apart at clinical field strengths, and the scanner must lock on the intended peak for correct localisation and fat suppression [6], [7]. The centre frequency is usually found as the dominant peak of a spectrum measured at the pre-scan [6]. It can be difficult to set in regions with poor B0 homogeneity such as the neck, where susceptibility changes quickly over short distances; the author suggests setting it over a small localised volume, ideally after or together with local shimming [1]. In breast imaging the fat peak can be larger than the water peak, so a robust fat-elimination method helps identify the water peak [1].

3.2 Peak separations

Separation from water1.5 T3 TSource
Fatabout 220 Hzabout 440 Hz[3], [4], [7]
Silicone (breast implants)about 300 Hzabout 600 Hz[3], [4]

The spectrum is displayed with silicone, fat and water from left to right; the fat line usually overlaps the silicone line [4]. The same job aid states that only the position of the peak, not its amplitude, determines the centre frequency [4]. Arithmetic check (Reasoning): 3.5 ppm at 123.26 MHz is about 431 Hz, consistent with the quoted 440 Hz; 600 Hz corresponds to about 4.9 ppm.

3.3 Automatic assumption and manual confirmation

A patent text states that systems of the earlier art record the highest peak of the spectrum and assign it to water, which fails when fat or silicone gives a higher peak, as in breast imaging [18] (patent background; assignee not shown in the extract). The Siemens job aids describe the corresponding console options on the adjustment card: a Confirm Frequency option that opens a window showing the spectrum after the shim so that the operator can check or move the chosen peak [2], [3], [4], and an Assume Silicone option that uses a three-peak model to calculate the resonance frequency and sets the frequency on the water peak without user interaction; it is to be used only for patients with known silicone implants [4]. An older-software job aid also lists Assume Dominant Fat for breasts that are mostly fat [3].

Option (Siemens label)Values reportedWhat the source saysSource
Confirm FrequencyNever, Always, When Changed (current software); “Only after freq. change” (older software)The confirmation window appears after the shim; with “only after/when changed” it appears only if an inline adjustment changed the resonance frequency. For most breast protocols the job aid chooses Always[2], [3], [4]
Assume SiliconeCheckboxThree-peak model; water peak set automatically; use only with known silicone implants; with it enabled use Always or When Changed for the confirmation[4]
Assume Dominant FatCheckbox (older software)For breasts that are mostly fat, such as non-dense breasts[3]

The job aids give the manual procedure: after the shim, the spectrum is shown with a tick mark at the chosen peak; for fat suppression the tick must lie over the water peak, otherwise the operator magnifies the display, clicks the intended peak, applies, checks the new system frequency value, and continues [3], [4]. The adjustment must target the water peak (the right-hand peak) except when the aim is to remove the silicone signal, in which case the silicone peak is selected; errors may occur with certain oral marking products or where fat may dominate water, as in the breast [2].

Sensitivity of fat suppression to the frequency error. In an educational abstract on adiabatic spectrally selective inversion (SPAIR) at 3 T in the breast, fat was suppressed with water preserved for offsets of about −110 to +110 Hz; at negative offsets fat saturation was lost, and at positive offsets the pulse overlapped the water resonance and reduced the signal of both [9]. The abstract states no precise failure threshold per failure mode and did not test a correction; the offsets were produced deliberately [9]. Consequence for practice (Reasoning): a frequency error of a fraction of the fat–water separation is enough to change the result, so the peak position is worth confirming where water is not the dominant peak.

4. B0 Shim: Adjustment Strategy, Shim Mode, Adjust Volume and Tolerance

4.1 Principle

Automated shimming is a standard part of the clinical pre-scan, at least for the constant linear shim terms [12]. The linear terms are generated by static offset currents of the three gradient coils; higher-order terms need dedicated shim coils [21] (hardware on 9706). Two classes of automatic method exist in the shimming literature: projection-based shimming, fast and efficient for small volumes such as a single spectroscopy voxel but possibly blind to localised inhomogeneity, and image-based shimming, which acquires a three-dimensional field map over a region of interest and suits arbitrarily shaped volumes such as the brain [11]. Which method a given clinical platform uses at the pre-scan is not stated in the sources read (to be sourced). Field inhomogeneity near air–tissue interfaces (sinuses, ear cavities) is prominent in the brain [10], and for the neck the centre frequency is also hard to set [1].

4.2 What the console offers (Siemens labels, one manufacturer)

The following options were read in a manufacturer training document on field uniformity [2]. The document carries effective dates of 2020 and 2021 and the software generation it refers to is not stated in the text read; several options appear in the screenshots without explanation, and the table gives only what the text states.

OptionWhat the source statesSource
Adjustment Strategy: SliceAdjustSlice-by-slice adjustment instead of adjustment of one volume; available for certain sequences (TSE and spin-echo EPI)[2]
Adjustment Strategy: Standard, FastViewAppear as labels without description[2]
Adjustment Strategy: OffcenterSeparate entry, see Section 4.4[5]
B0 Shim: Tune-upNo adjustment at sequence start; uses values from system installation; prioritised for sequences that do not need correction of patient-induced non-uniformity (spin echo or gradient echo without fat saturation)[2]
B0 Shim: StandardThe adjustment volume is superimposed on the acquisition volume[2]
B0 Shim: Standard NeckFor fat saturation in the neck, sagittal direction only; uses the neck coil only[2]
B0 Shim: AdvancedAdditional iterations (longer adjustment) for diffusion and spectroscopy; available in certain sequences only[2]
B0 Shim: Cardiac, Foot/Ankle, AbdomenAdapted to cardiac measurements, foot and ankle examinations, and transverse abdominal examinations[2]
B0 Shim: Brain, Breast, ProstateOptimise spectroscopy at these locations; certain sequences only[2]
B0 Shim: Whole BodyFor whole-body diffusion; certain sequences only[2]
Adjustment ToleranceShown as “Auto” with a truncated second word; no options or values explained[2], [4]
CoilShimValues Off, Coil and Combined appear in a screenshot; not described[2]

4.3 Adjust volume

The adjust volume (green box) is usually equal to the acquisition volume; it can differ, rarely, for example with saturation bands. Its position matters because a judicious position limits the effect of field non-uniformity, and the box must be checked and adjusted after manual placement [2]. In automatic mode the size is reset according to the shim mode chosen on the adjustment card; in manual mode the volume set when the sequence was recorded is applied. The document warns that a manually saved shim volume gives a message when the sequence is reopened and strongly recommends re-adjusting the volume manually or returning to automatic mode [2].

4.4 Off-centre anatomy

A job aid on off-centre imaging describes an adjustment strategy named Offcenter available from a stated 3 T software version and later, only for turbo spin echo, TSE-Dixon, BLADE and spin-echo sequences; it must be set manually on the adjustment card of each sequence. The adjustment runs in a central transverse slice at the isocentre, and the B1 evaluation and flip-angle determination use the current field of view. After enabling it, the excitation and refocusing flip angles can be adjusted manually; the document gives the example of lowering the flip angle by about 15 % to improve signal uniformity for the elbow [5]. The job aid also recommends keeping the extremity as close to the isocentre as possible and prefers T2 STIR or Dixon-based sequences to spectral fat saturation off-centre, because of B1 variation and distortion at the margins [5]. It gives no guidance for SliceAdjust, shim mode or adjust volume for the elbow or wrist [5].

The magnitude of the signal-uniformity benefit and its dependence on field of view were not quantified in the source (to be sourced).

5. Transmit Reference Amplitude and the Two-Channel B1 Shim

5.1 Reference amplitude

The calibration measures the flip angle produced by a given transmit gain with a standard pulse in a slice at the isocentre, and other pulses follow by scaling; B1 non-uniformity prevents the nominal flip angle from being reached everywhere, and a trade-off with speed is sometimes needed [1] (detail on 9708, Section 7). On the console the result appears on the transmit/receive card as a frequency for 1H and a reference amplitude in volts, together with an image scaling value and a correction factor [17]. The manufacturer job aid on the ACR phantom labels the frequency as the centre frequency and the reference amplitude as the transmitter gain [17] (extract only). The definition of the volt value beyond that (for example the pulse and the coil to which it refers) is not stated in the sources read (to be sourced).

5.2 B1 shim

At 3 T the body coil can produce a B1 distribution with shading; a manufacturer describes anatomy-specific B1 shimming with two body-coil channels of different amplitude and phase, against conventional circular polarisation with equal amplitude and a 90° phase [13], [14]. An independent abstract defines RF shimming as a subject-specific adjustment of the relative amplitudes and phases of the transmit channels to improve B1+ homogeneity or reduce SAR [16] (8-channel research body coil of another manufacturer). A manufacturer document of that other vendor describes the circularly polarised field of an empty coil as quadrature drive and multi-channel transmission as a way to improve B1 uniformity at high field [15] (extract only). The console option B1 Shim: TrueForm appears in the Siemens job aids [2], [4], and in the breast job aid it is part of the recommended adjustment card [4]. Its description in the field-uniformity job aid is limited to the name [2]. What the option changes in each protocol (preset or patient-calibrated shim) is discussed on 9708 (Section 6.2), where two manufacturer descriptions differ.

5.3 Adjust with body coil

The option Adjust with Body Coil appears on the adjustment card in all the job aids read [2], [4], [5] and is ticked in some and not in others. None of the sources read describes what it does. Hypothesis (Reasoning): the name suggests that the adjustment measurements are made with the body coil rather than with the local receive coil; this is not established and is not recommended as a rule (to be sourced from the operator manual).

6. Receiver Gain and Image Scaling

6.1 Receiver gain

The receiver gain is the attenuation applied to the analogue signal before digitisation. If it is too high, the signal saturates the converter and causes shading and intensity distortion; if it is too low, quantisation error dominates and adds noise and artefacts. The signal peaks at the centre of k-space, so a run without phase encoding can find the maximum; the ideal setting lies somewhat below saturation while avoiding quantisation error, and because some receive coils have enough phase variation that the maximum is not at zero phase encoding, the author recommends setting it somewhat below the zero-phase-encoding maximum. The dynamic-range demand is highest for high-SNR scans: high field, 3D, and coils with large sensitive volumes [1]. The receiver gain is normally invisible to the operator; the sources read give no console label for it (to be sourced).

6.2 Image scaling is a different quantity

Image scaling is a numerical multiplication and offset of the pixel values after reconstruction, used to fill the storage bit depth (for example 16-bit integers). The parameters are stored in DICOM tags: public rescale slope and intercept, and private scale slope and intercept for one manufacturer. A multi-centre working-group study found that scale factors derive from each series' pixel distribution and can differ between series acquired with identical settings, and that a quantitative tool that ignores them can give an intensity bias near 100 % compared with unscaled images. Receiver gain was held constant in the experiments, and the bias persisted, so the effect is independent of the receiver gain [8].

Finding in the scaling studyDetailLimitSource
Which vendor scaledSeries-specific scaling was detected for one manufacturer (Philips); the other two (GE, Siemens) did not scale the standard spoiled gradient-echo images testedStudy published in 2014; software versions not shown in the extract read, so not applicable to current versions without checking; one sequence type[8]
Effect on analysis tools13 standard tools (of 16 tested) gave apparent signal changes above 80 % in constant objects with the scaled data; the three customised tools were correctPhantom study[8]
ADC maps11 of 16 tools returned values close to the expected value; four returned about half the value; one about ten times too highOne manufacturer's maps[8]
Recommended actionSoftware to apply the scale parameters before combining series; users to validate each source–software combination with a constant-signal phantom—[8]

The “Image Scaling” and “Correction Factor” fields of the Siemens transmit/receive card [17] carry the same word, but no source read states that they correspond to the DICOM scale parameters of [8]; they are not to be equated with them (to be sourced). Values seen on the owner's series are given in Section 8.

7. Console Mapping (Siemens Labels and Generic Terms)

Generic termSiemens label (card)Status of the meaningSource
Centre frequencyFrequency 1H (Tx/Rx card)Labelled as centre frequency in a job aid; value in MHz[17]
Transmit gain / reference levelRef. Amplitude 1H (Tx/Rx card)Labelled as transmitter gain in a job aid; precise definition not found[17]
Shim mode and volumeB0 Shim, Adjust Volume (Adjustments card)Modes described in Section 4.2[2]
Slice-wise shim and transmit adjustmentAdjustment Strategy: SliceAdjust, OffcenterSliceAdjust and Offcenter described; Standard and FastView not described[2], [5]
Two-channel transmit shimB1 Shim: TrueFormName only in the job aids; manufacturer description in Section 5.2[2], [13], [14]
Manual check of the water peakConfirm Frequency (Never, Always, When Changed)Described[3], [4]
Silicone and fat-dominant spectrum modelsAssume Silicone, Assume Dominant FatAssume Silicone described; Assume Dominant Fat only in an older job aid[3], [4]
Adjustment accuracy criterionAdjustment ToleranceNot described[2], [4]
Local receive-coil shimCoilShim (Off, Coil, Combined)Values only; not described[2]
Adjustment with the body coilAdjust with Body CoilNot described (Section 5.3)[2], [4], [5]
Pixel-value scaling of imagesImage Scaling, Correction Factor (Tx/Rx card)Labels only; relation to DICOM scaling not established[17], [8]
Other manufacturers—Console names for GE and Philips not found (to be sourced)—

8. Worked Check on One Series

The following values were read from console screenshots of one 3 T head-and-neck examination on a Siemens system [20]. The reading level is the screenshot; the images contain a patient name and are not published. The interpretation is the editor's.

Card and labelValue seenReading (Reasoning)
Adjustments: Adjustment Strategy; Adjustment ToleranceStandard; AutoDefaults described only as labels [2]
Adjustments: B0 ShimStandard NeckThe mode is described as sagittal, neck coil only, for fat saturation in the neck [2]; whether the coil set-up of this series fulfilled that condition is not recorded
Adjustments: B1 Shim; Adjust with Body Coil; CoilShimTrueForm; Off; OffNames only (Sections 5.2, 5.3, 4.2)
Adjustments: Confirm Frequency; Assume SiliconeNever; offNo confirmation window: the frequency chosen by the system is used without operator check; appropriate or not depends on whether water is the dominant peak in the region, which for the head and neck is not a breast-type problem but is a poor-B0 region [1]
Tx/Rx: Frequency 1H123.261512 MHz and 123.261494 MHz in two capturesDifference 18 Hz, about 0.15 ppm (arithmetic); small against the 440 Hz fat–water separation [3], [4]
Tx/Rx: Ref. Amplitude 1H436.083 V and 442.632 V in two capturesDifference about 1.5 % (arithmetic). If the flip angle scales linearly with the amplitude, as in the standard relation between B1 and flip angle (Reasoning), the calibrations differ by that fraction
Tx/Rx: Image Scaling; Correction Factor2.000 and 5.000 in two captures; 1.00Meaning not sourced (Section 6.2)

Which capture each value belongs to, the sequence behind it and whether the two captures were taken in the same examination step are not recorded in the project record (to be checked against the images).

9. Technologist View (Reasoning)

The statements below connect the sources to what is seen at the console; they are derived by the editor and are not recommendations.

SituationComponent behind itBasis and limit
Breast, especially fat-dominant breasts or implantsWater may not be the highest peak[1], [3], [4], [18]; the job aid chooses a confirmation of the frequency for most breast protocols [4]
Neck and skull baseRapid susceptibility change; centre frequency and shim both difficult[1], [10]; small adjust volume placement matters [2]
Extremity far from the isocentreB1 variation and distortion at the margins[5]; off-centre adjustment strategy and flip-angle reduction on the stated sequences only
Fat suppression incomplete or unevenFrequency error, B0 offset, B1 error[1], [9]; the cause cannot be told from the image alone
Quantitative comparison of two seriesImage scaling can differ between series[8]; applies to the software and sequences tested; check the vendor
Table moved or coil set-up changed between seriesLoading and local field change[1] states the loading criterion only; platform rules not found (to be sourced)

10. Discrepancies Between Sources

TopicSource ASource BHandling
Confirm Frequency optionsOlder software job aid: “only after frequency change” [3]Current-generation job aids: Never, Always, When Changed [2], [4]Probably the same option under different software versions (Reasoning); not verified
Fat–water separation at 3 TAbout 440 Hz [3], [4]3.5 ppm at 123.26 MHz is about 431 Hz (arithmetic)Rounded values; no conflict
Date of the off-centre job aidFetched page text: job aid effective 22 Aug 2025Search extract: effective 2 Sep 2024Dates not relied on; only the content is used

11. Evidence Gaps and Items to Complete

GapStatus
Operator-manual definitions of Adjustment Tolerance, Adjust with Body Coil, CoilShim, Image Scaling, Correction Factor, Ref. Amplitude(to be sourced): manual not available; job aids give labels only
Console names and options on GE and Philips systems(to be sourced)
Criteria and numerical tolerances for repeating the transmit and frequency calibration(to be sourced); [1] gives only the loading criterion
Shim method (projection or image-based) of clinical platforms at the pre-scan(to be sourced)
Quantified uniformity benefit of B1 shimming and of the off-centre strategy(to be sourced); only a qualitative description and a 15 % example
Image scaling behaviour of current software of each vendor(to be sourced); [8] is a 2014 study
Receiver-gain console label and display(to be sourced)
Which values belong to which capture in the owner's seriesTo be checked against the images

12. References

Educational abstract (scanner-physics author)
[1] King K. Prescan: transmit/receive gain settings, frequency calibration. Proc Intl Soc Mag Reson Med. 2016;24.
Full text of the abstract through a summarising tool; qualitative; no numerical tolerances; image scaling and body-coil versus local-coil adjustment not covered.
Manufacturer training material
[2] Siemens Healthineers Academy. MR Management of Field Uniformity — Numaris X (job aid; effective dates 10 Sep 2020 and 1 Dec 2021 in the text).
Full text through a summarising tool; many options appear only as screenshot labels; training material, not the operator manual.
Manufacturer training material
[3] Siemens Healthineers Academy. Confirm Frequency Adjustment — Eline (job aid; effective 8 March 2020).
Full text through a summarising tool; states that the operator manual is authoritative; older software generation.
Manufacturer training material
[4] Siemens Healthineers Academy. MR Breast Silicone Implant — Numaris X (job aid; effective 11 April 2025 as read; master template 5 March 2020).
Full text through a summarising tool; training material.
Manufacturer training material
[5] Siemens Healthineers Academy. Off-center imaging (job aid; effective dates in 2025 as read).
Full text through a summarising tool; some numerals garbled in the extract; applies to stated sequences and software versions.
Educational text
[6] Elster AD. Prescan (automatic prescan). Questions and Answers in MRI (mriquestions.com); reference text Elster AD, Burdette JH. Questions and Answers in MRI, 2nd ed., Mosby, 2001.
Full page through a summarising tool; no numerical values except “a few hundred hertz”; the page itself names no author.
Educational text (extract)
[7] Elster AD. Center frequency. Questions and Answers in MRI (mriquestions.com).
Search extract only; about 220 Hz fat–water difference at 1.5 T.
Peer-reviewed study
[8] Chenevert TL, Malyarenko DI, Newitt D, et al. Errors in quantitative image analysis due to platform-dependent image scaling. Transl Oncol. 2014;7(1):65–71.
Full text through a summarising tool; phantom study; software versions not shown in the extract.
Educational abstract
[9] Sakhardande JD, Eddin AS, Allen TJ, Henze Bancroft LC, Strigel RM, Hsieh SK, Holmes JH. Artifacts in breast imaging when using SPAIR fat suppression in the presence of large B0 variations. Proc Intl Soc Mag Reson Med. 2023;31:abstract 5382.
Full text of the abstract through a summarising tool; figures not viewed; no correction tested; the scanner system is not stated in the extract.
Review (extract)
[10] Review of high-field B0 shimming of the brain in rodents and humans (PMC2802018; title and authors not shown in the extract).
Search extract only.
Technical study (extract)
[11] Automated high-order shimming for neuroimaging studies (PMC10748357; authors and journal not shown in the extract).
Search extract only; deep-learning brain extraction and image-based high-order shimming; a method study, not a description of clinical pre-scans.
Review (extract)
[12] Off-resonance artifact correction for magnetic resonance imaging: a review (arXiv 2205.01028; authors not shown in the extract).
Preprint; search extract only.
Manufacturer material (extract)
[13] Siemens Healthineers. TimTX TrueShape (product web page).
Search extract only; promotional page; used only for the description of B1 shimming as a concept.
Manufacturer article (extract)
[14] Siemens Healthineers. TimTX TrueShape and syngo ZOOMit aspects. MAGNETOM Flash 1/2012.
Search extract only.
Manufacturer document (extract)
[15] Philips. MultiTransmit parallel RF transmission technology (document hosted at documents.philips.com; date not shown).
Search extract only.
Technical abstract
[16] Study of RF shimming with an 8-channel parallel-transmit body coil on a 3 T Philips Achieva system. MAGMA (Springer, doi 10.1007/s10334-012-0360-5; authors and title not shown in the extract).
Abstract extract only.
Manufacturer training material (extract)
[17] Siemens Healthineers Academy. ACR MR Phantom — locate centre frequency and transmitter gain — Numaris X.
Search extract only; used for the card labels.
Patent text (background)
[18] US patent 9335395: magnetic resonance system and method for frequency calibration of the magnetic resonance system.
Search extract only; assignee not shown; used only for the statement of prior practice.
Patent text (background)
[19] US patent 12510616: magnetic resonance imaging method, system and computer program product utilizing B1 amplitude-based corrections.
Search extract only; assignee not shown; used only for the statement on the per-patient reference level.
Owner's data
[20] Siemens console screenshots of one 3 T head-and-neck examination, received 2026-11-23 (47 images).
Used through the inventory table of the project record, not re-opened for this page; the images contain a patient name and are not published.
Patent text (background)
[21] US patent 10705172: magnetic resonance apparatus and method for dynamic adjustment thereof with multiple adjustment parameters.
Search extract only; assignee not shown; used only for the statement that the linear shim terms are produced by gradient offset currents.

Recent PubMed search for this protocol

Last updated: November 2026

Report an error on this page