MRI Kidneys – Generic Standard Protocol

up to this point verified by human experts

MRIninja Knowledge Base | Master / General Protocol Page
Related pages: Liver master (5101) · Biliary / MRCP master (5201) · Pancreas master (5301) · Adrenal glands master (5402) · Small bowel master (5501) (companions, not parents)
Version 1.0 — October 2026

1. Executive Summary

This page covers magnetic resonance imaging (MRI) of the kidneys. The dominant clinical use in the sources read is the characterisation of an indeterminate renal mass, in particular the cystic mass classified with the Bosniak system, version 2019, which for the first time defines MRI criteria [1, 5]. A second, technical body of work concerns renal diffusion-weighted imaging (DWI) and other functional techniques, for which a consensus project (COST Action PARENCHIMA) exists [3, 4]. The adrenal glands, although examined in the same retroperitoneal field, have a separate master page: Adrenal glands master (5402).

The page describes the generic adult examination. It is built on the Bosniak version 2019 criteria [1], the ACR Appropriateness Criteria rating table for the indeterminate renal mass [2], the PARENCHIMA consensus on renal DWI [3] and a meta-analysis of MRI for small renal masses [6]. Three limits must be stated at the outset. First, no society technical consensus on a complete clinical renal mass MRI protocol was found in the sources read; the sequence table in Section 4 is a single-centre example [8], not a consensus (to be sourced). Second, the ACR table was read without its narrative text. Third, renal functional MRI (T1 / T2 mapping, arterial spin labelling, blood oxygenation level dependent imaging) is the subject of a consensus project [4] whose technique-level recommendations were read only for DWI [3].

1.1 Core Strengths

Defined MRI criteria for the cystic renal mass: Bosniak version 2019 defines MRI features for classes I to IV and defines enhancement as visible on comparison of unenhanced and enhanced images or as a signal intensity increase of 15% or more [1]. On unenhanced MRI, a mass that is homogeneously markedly hyperintense on T2-weighted imaging, or markedly hyperintense on unenhanced T1-weighted imaging, is Bosniak II and can generally be disregarded even without a full renal mass protocol [1, 5].

A pathway when a contrast agent is contraindicated: the ACR table rates unenhanced MRI of the abdomen as "usually appropriate" for the indeterminate renal mass when both iodinated and gadolinium-based contrast are contraindicated, and MRI without and with contrast as "usually appropriate" when only iodinated contrast is contraindicated [2]. MRI carries no radiation in the ACR table.

Reproducible quantitative diffusion technique: a modified Delphi consensus (21 respondents, 8 countries) defines b-values, echo time, resolution, fat suppression and acceleration for renal DWI [3] (Section 10.3).

1.2 Intrinsic Limitations of the Generic Protocol

The small renal mass is only moderately well characterised: a meta-analysis of 10 studies (860 small masses of 4 cm or less in 815 patients; nine retrospective) reports pooled sensitivity 0.85 (95% CI 0.79–0.90) and specificity 0.83 (0.67–0.92), with substantial heterogeneity (I² 68.3% and 86.7%) [6].

The protocol evidence is expert-level: Bosniak version 2019 is an expert classification (its review abstract is labelled Level of Evidence 5 [5]) and the protocol table is an example from one centre [8]. The renal DWI consensus targets clinical translation of functional measurements and its settings are not a routine mass protocol [3].

MR urography is not the mass examination: the ACR table rates MR urography without and with contrast "usually not appropriate" for the indeterminate renal mass in all three variants [2].

Where this page adds an interpretation that no source states it is marked Reasoning; where no source was read it is marked (to be sourced).

1.3 Evidence Base and Its Reading Level

SourceTypeReading levelUsed for
Bosniak classification version 2019 [1]Expert classification (Radiology)MRI criteria table and text read; the consensus method was not described in the part readCystic renal mass criteria on MRI
ACR Appropriateness Criteria, Indeterminate Renal Mass, 2020 [2]Guideline (rating tables)Rating tables read; narrative sections not present in the page readChoice of examination by contraindication
PARENCHIMA renal DWI consensus 2020 [3]Modified Delphi consensus, 21 respondents, threshold 75%Full text readRenal DWI parameters
PARENCHIMA overview [4]Consensus-process reportManuscript read; no sequence-level recommendations in this paperScope of the functional MRI consensus
MRI review of cystic renal masses 2021 [5]Invited review (Level of Evidence 5)Abstract onlyRole of MRI after CT
Meta-analysis, small renal masses [6]Systematic review with meta-analysisFull text readAccuracy, field strength
Excretory MR urography study [7]; protocol table [8]Technical study; technical example[7] abstract only; [8] supplementary table readMR urography; protocol example

2. Main Clinical Indications

2.1 Indications with Graded Statements

IndicationStatement in the sourceSource
Indeterminate renal mass, no contraindication to either contrast agentMRI abdomen without and with IV contrast: "usually appropriate"; MRI without IV contrast: "may be appropriate"; MR urography without and with contrast: "usually not appropriate"Guideline [2]
Indeterminate renal mass, contraindication to both iodinated and gadolinium-based contrastMRI abdomen without IV contrast: "usually appropriate"; MRI without and with contrast: "usually not appropriate"Guideline [2]
Indeterminate renal mass, contraindication to iodinated contrast onlyMRI abdomen without and with IV contrast: "usually appropriate"Guideline [2]
Cystic renal mass left uncertain by CTMRI recommended for masses with thick or nodular calcifications, for homogeneous hyperattenuating nonenhancing masses of at least 3 cm, and for heterogeneous nonenhancing masses (abstract of the review)Expert review [5]
Renal parenchymal and functional diseaseTechniques prioritised in the consensus project: T1 and T2 mapping, arterial spin labelling, DWI, BOLD. Clinical indications are not given in the part readConsensus process [4]
Renal artery, renal vein and renal transplant imaging; MR urography beyond the mass pathwayNot covered by the sources read (to be sourced)—

2.2 Red Flags and Escalation

For Bosniak version 2019, enhancing nodules (a convex protrusion of 4 mm or more with obtuse margins, or of any size with acute margins) define class IV [1]. Reasoning: a finding in this category should be communicated directly to the referring clinician, but no source read states a communication standard (to be sourced).

3. Preparation Reference

General patient screening, safety and positioning are in Universal MRI Preparation and MRI Safety Master (9800). Only what the sources read state for the kidney is given here.

3.1 Preparation Table

ItemStatementSource
Fasting before kidney MRINo statement in the sources read (to be sourced)—
Hydration (renal DWI)Scan in normal hydration status when clinically appropriate: consensus (76.2% agreement; 94.1% excluding abstentions)Consensus [3]
Salt intake (renal DWI)Consensus (91.7% excluding abstentions) that a controlled, standardised salt intake is not requiredConsensus [3]
Diet before renal DWINo consensus (46.7% agreed that diet must be controlled)Consensus [3]
Breathing for renal DWIConsensus against an expiration breath-hold (90% excluding abstentions); overall breathing mode: no consensus; respiratory gating (70%) or free breathing (66.7%) are preferencesConsensus [3]
Renal function and gadoliniumEligibility, group of agent and screening are in Gadolinium Contrast Agent Safety (9806) and Contrast Media in MRI (9101); they are not repeated hereInternal pages
SpasmolyticNot described for the kidney in the sources read (to be sourced)—

The hydration and diet statements come from a consensus on renal DWI for clinical translation of functional measurements; whether they apply to a routine renal mass examination is not stated (to be sourced).

4. Standard Protocol Design

4.1 Core Sequences: Renal Mass Example Protocol

The only protocol-level table found is the supplementary Table E1 of a 2021 Radiology publication, "Example of Renal Mass MRI protocol at 3T" [8]. The parent article title and authors were not retrieved from the page read, the table gives one example and not a recommendation, and its only stated design rationale is the reordering of sequences to reduce examination time [8].

OrderPlane and sequenceNote stated in the table
1Coronal 2D T2-weighted single-shot FSE (SSFSE)5 mm slice / 1 mm gap
2Axial 2D T1-weighted GRE, in-phase / opposed-phase5 mm / 1 mm; the purpose is not stated in the table
3Axial 2D T2 SSFSE5 mm / 1 mm
4Coronal 3D T1-weighted spoiled GRE with Dixon, dynamicPre-contrast, then corticomedullary, early nephrographic and late nephrographic phases; water, in-phase, opposed-phase and fat images reconstructed
5Axial 2D T2-weighted fat-suppressed SSFSE, post-contrastSPAIR fat suppression; acquired after the dynamic series
6Axial 2D DWI, fat-suppressed, post-contrastb = 50 and 800 s/mm²; ADC map; SPIR fat suppression
7Axial 3D T1-weighted spoiled GRE with Dixon, post-contrastExcretory phase

Contrast in this protocol is a single weight-based dose of an extracellular gadolinium-based agent at 2 mL/s followed by a 20 mL saline flush at 2 mL/s, with a phased-array body coil [8]. The T2 fat-suppressed and DWI series are acquired after the dynamic imaging and before the axial 3D T1 series, during the excretory phase [8]. Numerical parameters are in Section 10.

4.2 Mapping the Protocol to Bosniak Version 2019

Bosniak version 2019 states that enhancement is either clearly visible when unenhanced and contrast-enhanced images are compared or measurable as an increase of 15% or more in signal intensity; a feature too small to measure and not clearly visible is considered nonenhancing [1]. Visual assessment on subtraction images is accepted, and a renal mass protocol MRI with subtraction images can help to confirm that there is no enhancing component, especially in masses with thick calcification [1]. The criteria are tabulated in Section 7.1.

4.3 Sequence Rationale and Cross-References

Sequence principles are in TSE/FSE (9005) (fast spin echo), GRE / FLASH (9008) (gradient echo), DWI (9012) and EPI (9010) (diffusion and echo planar imaging) and Parallel Imaging (9588) (parallel imaging). The physics of the in-phase / opposed-phase pair is in Chemical Shift and Fat–Water Physics (9206), Chemical Shift: Theoretical Foundations (9596) and Chemical Shift Artefact, Second Kind (9608). Reasoning: the example protocol places the non-contrast T1 and T2 sequences first so that unenhanced signal characteristics (for example high T1 signal in a cyst) are recorded before gadolinium is given; the source states only the time-saving reordering [8].

4.4 Conditional Sequences: MR Urography

The ACR table rates MR urography without and with contrast "usually not appropriate" for the indeterminate renal mass in all three contrast variants [2]. For excretory MR urography of nondilated upper urinary tracts, one study of 22 patients (all given 250 mL of intravenous saline; 11 also given 10–20 mg of intravenous furosemide; images 5 minutes after gadolinium 0.1 mmol/kg with 3D and 2D breath-hold spoiled gradient echo) found that furosemide significantly improved calyceal and renal pelvis distension (P < 0.005) and reduced artefacts in all upper tract segments (P < 0.001) compared with saline alone; the authors concluded that saline alone is insufficient for diagnostic studies of nondilated upper tracts [7]. Only the abstract was read, the study is small, and a diuretic is not a consensus recommendation (to be sourced).

5. Optimisation Strategy

5.1 Artefacts and Countermeasures

ProblemCountermeasure stated in a sourceSource
Respiratory motion in renal DWIRespiratory gating or free breathing with post hoc motion correction (preferences); no cardiac gating; expiration breath-hold not recommendedConsensus [3]
Fat signal and misregistration in renal DWISPAIR fat suppression (consensus); STIR is not recommended (consensus against)Consensus [3]
Susceptibility and distortion of echo planar DWISingle-shot EPI (consensus); parallel imaging acceleration factor 2 (consensus), factors above 2 not recommended; echo time as short as the hardware allowsConsensus [3]
Enhancement assessment of masses with thick calcificationSubtraction imagesBosniak 2019 [1]
Other motion, flow, metal, wrap-around, dielectric artefactsNot specific to the kidney in the sources read; see Physics of MRI Artefacts (9209) (to be sourced)—

5.2 Efficiency

The example protocol shortens the examination by acquiring the fat-suppressed T2 and DWI series during the excretory phase instead of before contrast [8]. Whether post-contrast DWI changes the ADC values in the kidney is not addressed in the sources read (to be sourced).

5.3 Field Strength

The renal DWI consensus accepts both 1.5 T and 3.0 T (81% agreement) [3]. In the meta-analysis of small renal masses, studies at 1.5 T had higher pooled sensitivity than studies at 3.0 T (0.91 versus 0.83), a difference that was not significant (P = 0.06), and meta-regression found no significant effect of field strength [6]. The example protocol was performed at 3 T [8].

6. Contrast Use: Principles Specific to Renal MRI

Agent classes, nephrogenic systemic fibrosis, retention, hypersensitivity, pregnancy and lactation are in Gadolinium Contrast Agent Safety (9806) and Contrast Media in MRI (9101). This section covers only organ-specific use.

6.1 When Gadolinium Is Needed

For the indeterminate renal mass the ACR table rates the examination without and with contrast "usually appropriate" when neither contrast agent is contraindicated and the unenhanced examination "may be appropriate" [2]. Bosniak version 2019 uses enhancement as the main discriminator between classes (Section 7.1) [1]. Unenhanced MRI alone suffices to assign class II to homogeneously very T2-hyperintense or very T1-hyperintense masses [1, 5].

6.2 Renal Phases in the Example Protocol

PhaseIn the example protocolDelay
Pre-contrast3D T1 Dixon, coronal—
Corticomedullary3D T1 Dixon, coronalNot stated in the table read (to be sourced)
Early nephrographic3D T1 Dixon, coronalNot stated (to be sourced)
Late nephrographic3D T1 Dixon, coronalNot stated (to be sourced)
Excretory3D T1 Dixon, axial (after post-contrast T2 and DWI)Not stated (to be sourced)

7. Reporting Essentials

7.1 Bosniak Version 2019: MRI Criteria

ClassMRI criteria as tabulated in the source
IWell-defined, smooth wall of about 2 mm or less, which may enhance; homogeneous fluid with signal similar to cerebrospinal fluid; no septa or calcifications
II(a) 1–3 thin (≤2 mm) smooth enhancing septa (non-enhancing septa and any calcification allowed); (b) homogeneous mass markedly hyperintense to cerebrospinal fluid on unenhanced T2-weighted imaging; (c) homogeneous mass markedly hyperintense on unenhanced T1-weighted imaging, about 2.5 times normal parenchymal signal
IIF(a) Smooth, minimally thickened (≤3 mm) enhancing wall or septum, or four or more smooth thin enhancing septa; (b) mass heterogeneously hyperintense on unenhanced fat-suppressed T1-weighted imaging that does not meet class III or IV criteria
IIIOne or more enhancing thick (4 mm) walls or septa, or enhancing irregular walls or septa with obtuse convex protrusions of 3 mm or less
IVOne or more enhancing nodules: convex protrusion of 4 mm or more with obtuse margins, or of any size with acute margins

Thickness definitions in the source: thin is 2 mm or less, minimally thickened 3 mm, thick 4 mm or more [1]. The risk of malignancy per class is not given in the part read (to be sourced). A web calculator and mobile apps called "Bosniak Calculator" are mentioned in the review abstract [5]. Interobserver agreement of version 2019 was not assessed in the sources read (to be sourced).

7.2 Solid Renal Mass and Incidental Findings

The meta-analysis of MRI for small malignant renal masses describes "moderate diagnostic performance" [6]. An algorithmic MRI approach to the solid renal mass ("virtual biopsy") was found in a search but its full text could not be read (to be sourced). An adrenal mass found incidentally on a renal examination is handled on Adrenal glands master (5402). Other incidental findings are not covered (to be sourced).

8. MRI Technologist Pearls

The sources read are consensus, guideline and diagnostic-accuracy papers; none is a technologist-practice study. Items below are therefore either direct source statements or marked Reasoning.

8.1 Before the Examination

Check from the request whether a renal mass (cystic or solid), parenchymal or functional disease, or a urinary tract question is asked, because the protocol differs (Sections 4.1 and 4.4). Check whether iodinated or gadolinium-based contrast is contraindicated, because the ACR table recommends different examinations in each case [2]; screening and agent selection follow Gadolinium Contrast Agent Safety (9806).

8.2 During the Examination

Acquire the unenhanced series before contrast: Bosniak enhancement depends on comparison with unenhanced images [1]. Acquire the dynamic series as one coherent set with the same sequence before and after contrast, so that subtraction is possible [1, 8]. For DWI, free breathing or respiratory gating is preferred to an expiration breath-hold [3].

8.3 Common Errors (Reasoning)

ErrorConsequence
Omitting the unenhanced T1 series, or changing parameters between pre- and post-contrastEnhancement cannot be assessed against the 15% signal criterion and subtraction becomes unreliable [1]
Applying research DWI settings of the consensus [3] to every clinical examinationThe consensus targets functional parenchymal measurement; the clinical example protocol uses b = 50 and 800 s/mm² [8]
Incomplete coverage of both kidneysA mass or its extent may be missed; the DWI consensus lists full kidney coverage as a consensus item [3]

9. Quality Control Checklist

CheckBasis
Clinical question recorded (renal mass, functional, urinary tract)Section 2
Contraindications to iodinated and gadolinium-based contrast checked[2]; Gadolinium Contrast Agent Safety (9806)
Both kidneys fully covered on T2 and on the dynamic series[3] (DWI); Section 4 (Reasoning)
Unenhanced T1 and T2 series acquired before contrast[1, 8]
Pre- and post-contrast T1 series acquired with identical parameters; subtraction available[1, 8]
Contrast dose weight-based, injection rate and saline flush recorded[8]
Phases of the dynamic series identified (corticomedullary, early and late nephrographic, excretory)[8]
DWI: free breathing or respiratory gating; ADC map generated; fat suppression applied[3, 8]
Motion, wrap-around and susceptibility artefacts checked on each seriesSection 5; Physics of MRI Artefacts (9209)

10. Advanced Technical Parameters

This section is intended for MRI technologists and protocol optimisation specialists. All values come from the single-centre example protocol at 3 T [8] or from the renal DWI consensus [3]; they are examples and consensus items, not validated optima, and vendor-specific equivalents are not given (to be sourced).

10.1 T1-Weighted Dual-Echo In-Phase / Opposed-Phase

The example protocol includes this series; its purpose in the renal mass protocol is not stated in the table [8]. Physics: Chemical Shift and Fat–Water Physics (9206), Chemical Shift: Theoretical Foundations (9596), Chemical Shift Artefact, Second Kind (9608).

ParameterValue stated in the example protocol (3 T)
Plane and dimensionAxial, 2D GRE
Slice thickness / gap5 mm / 1 mm
Pixel size1 × 1.4 mm
TR180 ms
TE1.15 / 2.3 ms
Flip angle55°

10.2 T2-Weighted Single-Shot FSE

ParameterCoronal (unenhanced)Axial (unenhanced)Axial fat-suppressed (post-contrast)
Slice / gap5 / 1 mm5 / 1 mm5 / 1 mm
Pixel size1.2 × 1.5 mm1.3 × 1.6 mm1.3 × 1.5 mm
TR1250 ms1250 ms1250 ms
TE80 ms80 ms85 ms
Flip angle90°90°90°
Fat suppressionNoneNoneSPAIR

Source: [8]. Sequence principles: TSE/FSE (9005).

10.3 Diffusion-Weighted Imaging

ParameterClinical example protocol [8]Renal DWI consensus [3]
b-values (ADC)50 and 800 s/mm²Include b below 200 (consensus, 85.7%); highest b = 800 preferred (61.9%); highest b = 600 disagreed (81%). Recommended settings in the paper: b = 0, 100, 200, 800 with 3 directions
Fat suppressionSPIRSPAIR (consensus, 81.3% excluding abstentions); STIR against
TR / TE1425 ms / shortest (about 60 ms)TE below 100 ms and minimum allowed (consensus); TR 2000–4000 ms preferred (66.7%)
Slice / gap5 / 1 mmSlice thickness 2 mm or less disagreed (consensus); above 4 mm preferred (61.9%); 2–4 mm no consensus
In-plane resolution3 × 3 mm2–3 mm (consensus, 90.5%)
AcquisitionAxial, after contrastSingle-shot EPI; oblique coronal orientation; full kidney coverage; matrix above 128; acceleration factor 2; no cardiac gating; monopolar gradients
BreathingNot statedRespiratory gating or free breathing; expiration breath-hold disagreed
Units—10⁻³ mm²/s (consensus, 81%)

The consensus paper's own Table 4 mixes consensus items and preferences, and the 3-direction choice for ADC and IVIM is drawn from the literature and not from a consensus question; the Results text and Table 3 of the paper disagree on the IVIM b-value count [3]. Sequence principles: DWI (9012), EPI (9010), EPI-DWI (9014).

10.4 T1-Weighted 3D Dixon, Pre- and Post-Contrast

ParameterCoronal dynamic seriesAxial excretory series
Sequence3D spoiled GRE with Dixon; water, in-phase, opposed-phase and fat imagesSame
Slice / gap3 / 0 mm (1.5 mm interpolation)3 / 0 mm (1.5 mm interpolation)
Pixel size1.5 × 1.7 mm1.5 × 1.5 mm
TRShortest (about 4.4 ms)Shortest (about 3.7 ms)
TEShortest, 2 echoesShortest, 2 echoes
Flip angle10°10°
ContrastPre-contrast and three post-contrast phasesAfter T2 fat-suppressed and DWI

Source: [8]. Sequence principles: GRE / FLASH (9008); parameter concepts: Slice Thickness (9583), Field of View (9581).

11. Evidence Gaps and Ongoing Debate

Bosniak version 2019: the sources read describe the criteria [1, 5] but not interobserver agreement, the malignancy rate by class, or validation against pathology (to be sourced). The abstract of the MRI review is labelled Level of Evidence 5 [5].

Solid renal mass: the pooled accuracy for small masses is moderate and heterogeneous, nine of ten studies were retrospective, and patient selection was at high risk of bias in three because lesions under 1 cm were excluded [6]. The MRI algorithm for histological subtypes, microscopic fat and angiomyolipoma were not read (to be sourced).

Renal functional MRI: the PARENCHIMA project produced 166 statements across four techniques, with a consensus threshold of 75%, but none of the vendor protocols submitted fully complied with the recommendations [4]; only the DWI recommendations were read [3], and in that paper diet control and the choice between gating and free breathing did not reach consensus.

Protocol standardisation: the only protocol table is a single-centre example [8]. No society consensus on technique for renal mass MRI was found (to be sourced).

MR urography: a small study supports a diuretic [7], and the ACR table rates MR urography "usually not appropriate" for the indeterminate mass [2]; MR urography for other indications was not read (to be sourced).

Not covered by any source read: renal artery and vein MR angiography, renal transplant, renal cell carcinoma staging and thrombus, paediatric renal MRI, incidental findings, and fasting or spasmolytic practice (to be sourced).

12. Evidence-Based References

References are ordered by level of evidence and then by recency. The reading level of each source is stated in its line; a source read only at abstract level supports only what its abstract states.

A. Guidelines / Consensus / Society Recommendations

High
[1] Silverman SG, Pedrosa I, Ellis JH, Hindman NM, Schieda N, Smith AD, Remer EM, Shinagare AB, Curci NE, Raman SS, Wells SA, Kaffenberger SD, Wang ZJ, Chandarana H, Davenport MS. Bosniak Classification of Cystic Renal Masses, Version 2019 (title as given in the search listing; subtitle, if any, not verified). Radiology. 2019;292(2):475–488. DOI: 10.1148/radiol.2019182646.
(High — Expert classification; MRI criteria table and text read; the consensus method is not described in the part read) Cited in Sections 1, 2, 4, 6, 7, 8, 11.
High
[2] American College of Radiology. ACR Appropriateness Criteria: Indeterminate Renal Mass. New 2020. acsearch.acr.org (topic 31).
(High — Guideline; rating tables of the three variants read; authors, journal and DOI not shown on the page; narrative not available) Cited in Sections 1, 2, 4.4, 6, 8, 11.
High
[3] Ljimani A, Caroli A, Laustsen C, et al. Consensus-based technical recommendations for clinical translation of renal diffusion-weighted MRI. Magn Reson Mater Phys Biol Med. 2020;33:177–195. DOI: 10.1007/s10334-019-00790-y.
(High — Modified Delphi consensus, 21 respondents from 21 institutions in 8 countries, 87 agree/disagree questions of which 57 reached consensus at 75% or more; full text read) Cited in Sections 1, 3, 5, 8, 9, 10, 11. Some internal inconsistencies of the paper are noted in Section 10.3.
Moderate
[4] Mendichovszky I, et al. Technical recommendations for clinical translation of renal MRI: a consensus project of the Cooperation in Science and Technology Action PARENCHIMA. (Journal, volume, pages and DOI not verified in the file read.)
(Moderate — Consensus-process report: 166 statements (36 T1/T2 mapping, 14 BOLD, 57 DWI, 59 ASL), threshold 75%; manuscript read; contains no sequence-level recommendations) Cited in Sections 1, 2, 11. The citation details must be completed before the page is verified.
Expert
[5] Krishna S, Schieda N, Pedrosa I, Hindman N, Baroni RH, Silverman SG, Davenport MS. Update on MRI of Cystic Renal Masses Including Bosniak Version 2019. J Magn Reson Imaging. 2021;54(2):341–356. DOI: 10.1002/jmri.27364.
(Expert — Invited review, Level of Evidence 5; abstract only read) Cited in Sections 1, 2, 6, 7, 11.

B. Systematic Reviews / Meta-analyses

Moderate
[6] Shang W, Hong G, Li W. MRI for the detection of small malignant renal masses: a systematic review and meta-analysis. Front Oncol. 2023;13:1194128. DOI: 10.3389/fonc.2023.1194128.
(Moderate — Systematic review and meta-analysis, 10 studies, 860 masses, 815 patients; full text read) Cited in Sections 1, 5.3, 7.2, 11.

C. Important Prospective / Original Studies

Technical
[7] Ergen FB, Hussain HK, Carlos RC, Johnson TD, Adusumilli S, Weadock WJ, Korobkin M, Francis IR. 3D excretory MR urography: improved image quality with intravenous saline and diuretic administration. J Magn Reson Imaging. 2007;25(4):783–789. DOI: 10.1002/jmri.20875.
(Technical — Comparative technical study, 22 patients, blinded readers; abstract read) Cited in Sections 4.4, 11.

D. Technical MRI Papers

Technical
[8] Supplementary Table E1, "Example of Renal Mass MRI protocol at 3T". Radiology. 2021. DOI: 10.1148/radiol.210034 (parent article title and authors not shown on the page read).
(Technical — Single-centre example protocol; supplementary table read in full; not a recommendation) Cited in Sections 1, 4, 5, 6, 8, 9, 10, 11.

E. Landmark Historical References

No landmark historical reference was read for this page (to be sourced).

Child Protocols

Clinical pages derived from this master protocol. These pages document what changes for specific indications.

No child protocols have been published yet.

Recent PubMed search for this protocol

Last updated: October 2026

Report an error on this page

The report stores the page address, the text above and, if you give it, your e-mail. It does not store your IP address, browser type or any cookie. See the Disclaimer, section 7.