MRI Anal Canal for Perianal Fistula

Required Protocol at a Glance

Mandatory core sequences for this examination. Detailed rationale, conditional additions and optimisation notes are provided later in the protocol.

View full protocol design ↓
Sequence Status in the ESGAR consensus Purpose
T2-weighted, axial Mandatory for most members; with or without fat suppression (level V) Track, abscess, sphincter anatomy
T2-weighted, coronal As above Levator ani, ischioanal fossa, supralevator extension
T2-weighted, sagittal As above Anterior and posterior extensions
Fat-suppressed T1 with gadolinium Optional Separation of fluid from granulation tissue; activity
STIR Optional Fluid-sensitive alternative
DWI Optional; no parameters given Abscess, activity (emerging)
up to this point verified by human experts

MRIninja Knowledge Base | Child Page — Pathology-Specific Protocol
Parent page: Anal canal master (6401)
Version 1.0 — October 2026

Prerequisite: this page assumes the anal canal anatomy of Anal canal master (Section 3.2: sphincter complex, dentate line estimated in thirds) and does not repeat it. The parent page is written for anal cancer, so most of its preparation, staging and response content does not apply here; this page documents what is different when the clinical question is a perianal fistula or other anal sepsis: preparation, coil, planes, coverage, sequences, contrast, timing after surgery, the report minimum dataset and the activity scores used in Crohn's disease. Universal safety and contrast screening are on the general pages and are not repeated.

1. Executive Summary

1.1 Added Value over the Parent Protocol

MRI of a perianal fistula answers a different set of questions from MRI of an anal tumour. It must show the course of the track in relation to the sphincter complex (Parks classification), the position of the internal opening, any abscess or secondary extension, and, in Crohn's disease, the activity of the disease [1, 6]. The ESGAR consensus of 2020 (full text read) judges MRI the most clinically useful modality (evidence level II) and advises against computed tomography as the primary investigation and against contrast fistulography where MRI or anal ultrasound is available (level V) [1]. The European Society of Coloproctology (ESCP) guideline of 2024 recommends preoperative MRI for a fistula suspected to be complex on clinical assessment, with a very low threshold for imaging (MRI or endoanal ultrasound) to separate simple from complex fistulas [2].

The technical profile is also different from the parent. The ESGAR statements on preparation are unusually restrictive: no fasting, no rectal emptying or distension, no anal marker, no intravenous smooth-muscle relaxant, a surface coil and no endoanal coil [1]. At least one acquisition must cover the supralevator and ischioanal spaces, and the axial and coronal planes must be aligned to the anal canal axis [1]. Gadolinium and diffusion-weighted imaging (DWI) are optional in the consensus [1].

1.2 Limits of the Dedicated Protocol

The consensus is expert opinion for most technical points: the sequence, preparation and contrast statements of the ESGAR consensus carry evidence level V (expert opinion), and slice thickness and field-of-view dimensions are not specified [1]. Accuracy data are heterogeneous: the pooled specificity of MRI (69%) and of endoanal ultrasound (43%) quoted in the secondary sources comes from a meta-analysis that the sources themselves caution is limited by heterogeneity [3, 6, 7]. Activity indices are research tools: the MAGNIFI-CD index has a retrospective external validation in 65 patients and its use in routine practice is not established [8, 10]. The definition of radiological healing is the subject of a Delphi process whose protocol states that no consensus definition existed [5]; the definition described in Section 6.3 comes from a news report of a conference presentation.

1.3 Evidence Base and Its Reading Level

SourceTypeReading levelUsed for
ESGAR consensus on imaging of fistula-in-ano and other anal sepsis, 2020 [1]Consensus (modified Delphi, RAND/UCLA; 13 members; 53 statements; threshold 11 of 13)Full text readPreparation, acquisition, sequences, timing, reporting, pitfalls
ESCP guideline on cryptoglandular anal fistula, 2024 [2]Guideline (AGREE-S, GRADE)Imaging section and methods read; document truncated at section 8.11Role of MRI and endoanal ultrasound
Dutch surgical guideline, Proctologie, 2015 [3]National guideline moduleWeb page read; validity review was due by 2019Choice of MRI versus endoanal ultrasound
Pediatric Crohn's trial consensus, 2024 [4]; Delphi protocol on radiological healing, 2024 [5]Consensus abstract; study protocol without resultsAbstract; protocolIndex endorsement; absence of a definition of healing
Arkenbosch et al., 2025 [6]; Habeeb et al., 2025 [7]; Anand et al., 2025 [8]Narrative reviewsFull text as rendered by a tool (the third truncated in the references)Classification, protocol, indices, follow-up
Hindryckx et al., 2019 [9]; Beek et al., 2025 [10]; VALIDATE-PERIANAL, 2026 [11]Index development (abstract); external validation (full text); conference study (news report)Abstract; full text; secondary reportMAGNIFI-CD; radiological healing

2. Clinical Context

2.1 Classification Used in the Report

SystemContentSource
Parks (anatomical)Reported for every fistula in the ESGAR minimum dataset (level V)[1]
St James's University Hospital (MRI-based)May be added if used locally, but not at the expense of Parks (level V)[1]
Simple versus complex (review)Simple: superficial, low intersphincteric or low transsphincteric, involving under 30% of the distal sphincter complex, usually one internal opening. Complex: high transsphincteric (over 30% of the external sphincter), supra- or extrasphincteric, horseshoe, multiple side branches, or related to radiation or malignancy[6]
Complexity (ESCP)Extrasphincteric, suprasphincteric or high transsphincteric (over 30% of the external sphincter); intersphincteric (over 50% of the internal sphincter); rectovaginal or anterior in women; recurrent; horseshoe or multiple extensions[2]
Horseshoe (ESGAR definition)A semilunar region of sepsis spreading horizontally on either side of an internal opening, involving two or more adjacent quadrants; ischioanal, intersphincteric or supralevator[1]

The two complexity definitions differ in detail (for example the 50% intersphincteric criterion and the inclusion of recurrent and anterior fistulas in women appear in the ESCP text but not in the review). The report should state which system is used (Reasoning).

2.2 Clinical Settings

Cryptoglandular fistula: the setting of the ESCP guideline [2]. Perianal fistulising Crohn's disease: about 1 in 5 patients with Crohn's disease have perianal disease over their lifetime, about two-thirds undergo perianal surgery, and radiological healing can lag clinical remission by up to 12 months [8]. Another review quotes perianal fistulising disease in 30–50% of Crohn's patients, which differs from the first figure; the sources do not reconcile the two [6, 8]. Other anal sepsis: acute ischioanal abscess, complex pilonidal sinus and hidradenitis suppurativa where cryptoglandular infection is in the differential, anal pain in immunocompromised patients and anal necrotising fasciitis (ESGAR indications, level II–III) [1]. Fistula and neoplasia: a fistula related to radiation or malignancy is classed as complex [6]; soft tissue inside a long-standing fistula should raise concern for a fistula-associated tumour, see Anal canal master Section 2.2. Occult disease: dedicated MRI found occult fistulas in up to 39% of patients with chronic anal or perianal pain referred to a tertiary centre [6].

3. Indications, Modality Choice and Escalation

3.1 Indications with Graded Statements

QuestionStatement in the sourceSource
Potential anal sepsis, diagnosis and classification of fistula-in-ano, acute ischioanal abscessMRI may be used (level II)ESGAR [1]
Therapy response in Crohn's disease; abscess before biological therapyMRI may monitor response and identify abscesses (level II)ESGAR [1]
Most useful modalityMRI (level II); CT not as primary investigation where MRI or anal ultrasound is available, acceptable with intravenous contrast when neither is available or in emergencies (level V); contrast fistulography not where MRI or anal ultrasound is available (level V)ESGAR [1]
Complex fistula, preoperative"Preoperative MRI should be used" for a fistula suspected on clinical assessment to be complex, or when endoanal ultrasound is insufficient or unavailable (strong wording; moderate certainty)ESCP [2]
Threshold for imagingMRI or endoanal ultrasound "with a very low threshold" to separate simple from complex (certainty upgraded by expert opinion)ESCP [2]
Examination under anaesthesia alone"Should not be used" as the only diagnostic tool in complex fistula (moderate certainty)ESCP [2]
Preoperative endoanal ultrasound"Can be considered" as a primary tool for most patients (expert opinion)ESCP [2]
First-choice modalityMRI, with endoanal ultrasound as an alternative in experienced handsDutch guideline 2015 [3]
Paediatric Crohn's trialsmVAI and MAGNIFI-CD endorsed as appropriate; MRI must be part of outcome assessment; intravascular contrast in trials rated appropriate (15-member panel)Consensus abstract [4]

3.2 MRI versus Endoanal Ultrasound: Numbers as Quoted

All the figures below are quoted from the guideline or review texts, not from the primary studies, which were not read.

FindingValueSource (quoted in)
Single outcome-based study (104 patients): accuracy of MRI, endoanal ultrasound, digital examination90%, 81%, 61%; kappa for primary tract 0.84, 0.68, 0.38[2]
Systematic review: sensitivity for complex fistulaMRI 0.97; examination under anaesthesia 0.75; endoanal ultrasound 0.92[2]
Meta-analysis: sensitivity and specificitySensitivity 87% for both; specificity MRI 69%, endoanal ultrasound 43%; heterogeneity noted[3, 6, 7]
Meta-analysis of 12 studies of 3D endoanal ultrasoundAccuracy 91%; internal opening sensitivity 97%, specificity 61%; the authors conclude that ultrasound alone is not sufficient for complex fistula[2]
Added value of MRI over clinical examination85% of patients, particularly complex fistulas; negative predictive value up to 100%[3]
InconsistencyMRI specificity is given as 76–100% in one section of a review and 69% in another[6, 8]

Endoanal ultrasound is preferred for the internal opening, the sphincter complex and when MRI is contraindicated; MRI is better for suprasphincteric and extrasphincteric fistulas and for extensions beyond the sphincter complex [1, 2, 8]. Endoanal ultrasound has a limited field of view, depends on operator skill, requires an endoluminal probe that is painful for many patients and has difficulty separating active fistula from fibrosis [3, 8].

3.3 Red Flags and Escalation

Pre-treatment MRI can identify abscesses that preclude biological therapy in Crohn's disease (level II) [1]. Reasoning (no source read grades these as urgent findings): soft tissue in a long-standing fistula (possible neoplasm), supralevator or horseshoe extension and an unsuspected abscess merit direct communication with the surgical or gastroenterology team; the route and timing of communication are not defined in the sources (to be sourced).

4. Dedicated Protocol Design

4.1 Mandatory Core Sequences

SequenceStatus in the ESGAR consensusPurpose
T2-weighted, axialMandatory for most members; with or without fat suppression (level V)Track, abscess, sphincter anatomy
T2-weighted, coronalAs aboveLevator ani, ischioanal fossa, supralevator extension
T2-weighted, sagittalAs aboveAnterior and posterior extensions
Fat-suppressed T1 with gadoliniumOptionalSeparation of fluid from granulation tissue; activity
STIROptionalFluid-sensitive alternative
DWIOptional; no parameters givenAbscess, activity (emerging)

Source of the table: [1]. The consensus states that sequences are flexible and chosen to show sepsis and its relationship to the sphincter complex (level V). Sequence principles: TSE/FSE (9005), DWI (9012).

4.2 Protocol Delta vs the Anal Canal Master (6401)

ElementAnal canal master (tumour)Perianal fistula (this page)
Fasting and bowel preparationNone required; micro-enema and gel not examined in anal cancerNo fasting; no rectal emptying or distension; no anal lumen marker such as a catheter (level V) [1]
SpasmolyticUsed by some, not found necessary in the essayNo intravenous smooth-muscle relaxant such as hyoscine butylbromide (level V) [1]. One validation study nevertheless gave intravenous scopolamine butylbromide [10]: practice varies
PositionNot statedSupine [1]
CoilExternal phased-array; endorectal coil avoidedSurface coil improves quality (level III); endoanal receiver coils not used, unnecessary (level III) [1]. A review adds that endoluminal coils can improve resolution but limit the field of view and may cause discomfort [6]
Field strength1.5 T or 3 T in the studies1.5 T or 3 T, local availability and preference (level V) [1, 6]
PlanesHigh-resolution T2 in three planes aligned to the canalMinimum: axial, coronal, sagittal; axial and coronal aligned to the anal canal axis (level III) [1]
CoveragePromontory or aortic bifurcation to the anal margin and inguinal areas, for nodesAt least one acquisition must cover remote extensions (supralevator and ischioanal); whole-pelvis coverage is one way (level II) [1]
ContrastDebatedOptional; gadolinium 0.1 mmol/kg is sufficient (level V) [1]; see 4.3
DWIImportant at responseOptional [1]
ChildrenNot coveredNo special modifications needed (level V) [1]
TimingAfter chemoradiotherapy: 3 to 6 months in the studiesAt least 4 weeks after therapeutic examination under anaesthesia (level V) [1]

4.3 Contrast Strategy

The sources agree that contrast is not essential for the diagnosis of an uncomplicated fistula and disagree on how often it should be used. ESGAR lists fat-suppressed T1 with gadolinium as optional and states that 0.1 mmol/kg is sufficient [1]. A review states that gadolinium is generally unnecessary in uncomplicated cases but suggests it for chronic persistent Crohn's fistulas, to delineate tracts and to screen for malignancy [6]. Another review gives contrast as optional at many centres with non-standardised protocols and notes that contrast-enhanced T1 distinguishes fluid, pus and granulation tissue [8]. The MAGNIFI-CD index depends on post-contrast findings, which makes contrast part of the scoring in trials [6, 7, 8], and a paediatric consensus rated intravascular contrast appropriate in trials [4]. The validation study used axial fat-suppressed T1 after 0.1 mmol/kg [10]. In practice this means that contrast is a protocol decision linked to the purpose: optional for diagnosis, needed if an index is scored. Agent choice, renal function and consent follow Contrast Media in MRI (9101) and Gadolinium Contrast Agent Safety (9806). The ESGAR consensus describes the activity assessment (T2 signal and/or enhancement) as optional and dependent on local usefulness (level III) [1].

4.4 Diffusion-Weighted Imaging

DWI is optional in the consensus and no parameters are given [1]; the reviews describe DWI and dynamic contrast-enhanced imaging as promising or emerging, not as protocol items [6, 8]. Primary studies of the added value of DWI in fistula were not read (to be sourced). Sequence principles: DWI (9012), EPI (9010).

4.5 Timing

After surgery: MRI at least 4 weeks after therapeutic examination under anaesthesia, because post-surgical cavities are hard to distinguish from untreated extensions (expert opinion, level V) [1]. In Crohn's disease: a review recommends repeating MRI at 6 to 12 months after surgery or advanced therapy, with 6 months for early response and 12 months to confirm healing [7]; the ESGAR consensus gives no follow-up interval beyond the 4 weeks after examination under anaesthesia [1].

5. MRI Assessment of the Fistula

5.1 Elements to Assess

ElementWhat to determineSource
Primary trackCourse relative to the internal and external sphincters, Parks type; tortuous tracks may have internal and external openings in different quadrants, so an approximate primary track location may be needed[1]
Internal openingExact radial position (clock face); height relative to the dentate line, puborectalis or anal verge[1]
External openingExact clock-face position[1]
Extensions and abscessLocation (intersphincteric, ischioanal, supralevator, other); size as maximal cavity diameter excluding the fibrous wall; horseshoe component[1]
SphinctersIntegrity of the internal and external sphincter, especially after previous surgery[1]
SetonPresent or not (may be hard to identify)[1]
RectumProctitis; other less common findings such as sacral osteomyelitis[1]
ActivityT2 signal and/or gadolinium enhancement of the track (optional)[1]

5.2 Pitfalls and Mimickers

Dentate line: it is a histological boundary and is not visible on conventional imaging; it usually lies about one-third along the anal canal, 1–2 cm above the verge, and inexperienced readers often place it too high [1]. Oblique planes: planes not aligned to the anal canal axis hinder diagnosis [1]. Early post-operative MRI: treated cavities are confused with untreated disease [1]. Seton: difficult to identify [1]. Abscess versus track: no accepted definition separates them [1]. Post-treatment confusion: endoanal ultrasound has difficulty distinguishing active fistula from fibrosis, which is one reason MRI is preferred for suspected recurrence [3]. Reader experience: about 50 mentored MRI examinations are suggested for independent reporting (level II) [1].

6. Reporting Framework

6.1 ESGAR Minimum Dataset

ItemRequirement
Clinical detailsIncluded
Number of fistulasTotal number stated, then each one described
ClassificationParks for every fistula; St James's optional, not at the expense of Parks
Internal openingClock-face position (a quadrant may be added but not substituted); height relative to the dentate line, puborectalis or anal verge; measurement in mm optional
External openingClock-face position, quadrant optional; multiple or shared openings stated with number and location
ExtensionsLocation and maximal cavity diameter; horseshoe described
Length of the fistulaOptional
SetonStated if present
Disease-free regionsSummarised, not described individually
Sphincter integrity and other findingsStated; comparison with prior MRI where available

Source: [1] (all level V). No structured report template is provided by the review that discusses classification [6].

6.2 Activity Indices in Crohn's Disease

IndexContent as describedReliability and validationSource
Van Assche index (2003)Six parameters (tract location, abscess, proctitis, T2 signal); range 0–22; no contrast neededInter-rater ICC 0.68; intra-rater 0.81[7, 6, 9]
Modified Van Assche (2017)Adds T1 post-contrast hyperintensity and dominant tract features; range 0–19.5Inter-rater ICC 0.67; intra-rater 0.81[7, 6, 9]
MAGNIFI-CD (2019)Six items; range 0–25; contrast needed. The reviews describe the items differently (number of tracts, post-contrast hyperintensity, T1 features, dominant feature; or fistula length, wall thickness and abscess); the development abstract does not name themDevelopment: 160 participants; inter-rater ICC 0.74 (0.67 and 0.68 for the Van Assche indices); intra-rater 0.85; standardised effect size 1.02 vs 0.84 and 0.68[6, 7, 8, 9]

External validation (retrospective, 65 patients with complex fistulising Crohn's disease at two centres; baseline and follow-up MRI 3–24 months apart): interobserver ICC 0.87; baseline median 18; responders fell from 18.0 to 9.0; proposed response cut-off a decrease of at least 2 points; proposed remission cut-off a score of 6 or less at follow-up (AUROC 0.81 against one reference and 0.90 against another); at 12 months none of the patients meeting the remission cut-off had a re-intervention against 76% of the others. Correlation with change in the patient's visual analogue scale was moderate (rho 0.59) and with C-reactive protein and faecal calprotectin negligible. The authors list the retrospective design, unvalidated clinical response standards, tertiary-centre selection and a subjective reference standard as limitations [10]. A narrative review notes that MAGNIFI-CD is limited by its sensitivity to short-term change, by contrast dependence and by time-consuming scoring, and that 50% of radiologists surveyed saw research-only utility [8]; indices are not routinely used in practice [8].

6.3 Radiological Healing

The protocol of the Delphi study states that no consensus definition of a radiologically healed fistula on pelvic MRI existed; candidate features were fibrosis of the tract, absence of collections under 2 cm and absence of T2 hyperintensity [5]. A news report of a retrospective multicentre study presented in 2026 (VALIDATE-PERIANAL) describes the TOpClass consensus definition endorsed with 95% agreement as requiring all of: no T2 hyperintensity, a completely fibrotic tract and, when contrast is used, no enhancement; of 40 patients reported as healed locally only 14 met the strict definition on re-review, and those 14 had more sustained clinical remission at one year (93% against 78%) and less recurrence (7% against 22%), though a higher return to the operating room (30% against 7%) [11]. This source is a secondary report of a conference presentation and the original publication was not read. Radiological healing was reported in one meta-analysis at 25% against 47% clinical remission at 52 weeks, and a reduction of at least 16% in active fistula volume predicted clinical remission (sensitivity 84.6%, specificity 81%); both numbers are quoted in a review [7].

7. Technical Pitfalls

Unaligned planes: axial and coronal planes not aligned to the canal axis (ESGAR minimum standard) [1]. Insufficient coverage: no acquisition with a field of view that includes supralevator and ischioanal spaces [1]. Unnecessary preparation: a rectal filling, a marker or a spasmolytic added by habit from rectal or prostate protocols, contrary to the ESGAR statements [1]; the rectal-gel considerations of Rectum master are not transferable here. Endoluminal coil: reduces the field of view and may cause discomfort [1, 6]. Imaging too soon after surgery: less than 4 weeks [1]. Contrast mismatch: a trial or follow-up scored with a contrast-dependent index but acquired without contrast (Reasoning). Inconsistent technique: planes, sequences and contrast changed between baseline and follow-up make comparison of activity unreliable (Reasoning).

8. MRI Technologist Pearls

8.1 Before the Examination

Confirm the indication (fistula, abscess, Crohn's disease, recurrence after surgery) and the date of any recent examination under anaesthesia [1]. Ask whether the study is for diagnosis or for activity scoring, since this decides contrast (Reasoning). No fasting, no enema, no catheter or marker, no routine spasmolytic [1]. Explain that the examination is performed supine with a surface coil, without an internal probe [1, 3].

8.2 During the Examination

Plan the axial and coronal series on a sagittal localiser aligned to the canal axis [1]. Include the ischioanal fossae and the supralevator space in at least one series [1]. Add sagittal T2 [1]. If contrast is used, 0.1 mmol/kg [1, 10]. Use the same plane set and, in Crohn's disease, the same sequences at each follow-up (Reasoning).

8.3 Common Errors (Reasoning)

Stopping at an axial T2; narrow field of view that misses a supralevator extension; adding rectal gel; omitting the sagittal plane; not noting the date of surgery on the request. None is documented as an error rate in a source read.

9. Quality Control Checklist

CheckBasis
Indication, previous surgery and date of examination under anaesthesia recorded; at least 4 weeks since therapeutic surgery[1]
No rectal filling, marker, endoanal coil or routine spasmolytic[1]
Surface coil; supine[1]
T2 in axial, coronal and sagittal planes; axial and coronal aligned to the canal axis[1]
At least one acquisition includes supralevator and ischioanal spaces[1]
Contrast given or omitted according to purpose; dose 0.1 mmol/kg if given[1]; Section 4.3
Same technique as the prior examination when activity is comparedReasoning
Report contains the ESGAR minimum dataset (Section 6.1)[1]

10. Advanced Technical Parameters

This section is intended for MRI technologists and protocol optimisation specialists. The ESGAR consensus does not specify slice thickness, field-of-view dimensions, repetition time, echo time, matrix or DWI settings [1]; no source read gives them (to be sourced). The only complete technique description read is that of one validation study, shown as an example and not as a recommendation.

10.1 T2-Weighted Sequences

Tissue Contrast Logic

Fluid-containing tracks and abscesses are bright on T2 with or without fat suppression, against the lower signal of sphincter muscle and fibrosis; fat suppression or STIR increases the conspicuity of fluid against ischioanal fat [1]. Sequence principles: TSE/FSE (9005). Reasoning: the physical wording is mine.

Key Parameters

ParameterValue statedSource
PlanesAxial, coronal, sagittal; axial and coronal aligned to the anal canal axis[1, 6]
Fat suppressionT2 with and without fat suppression in the validation study; with or without in ESGAR[10]; [1]
Field of viewAt least one acquisition covering supralevator and ischioanal spaces (whole pelvis is one way)[1]
Slice thickness, matrix, TR, TENot specified—
CoilSurface coil; endoanal coil not used[1]

10.2 Contrast-Enhanced T1

ParameterValue statedSource
DoseGadolinium 0.1 mmol/kg[1, 6, 10]
SequenceFat-suppressed T1, post-contrast, axial in the validation study; at least one plane in ESGAR[10]; [1]
PurposeFluid versus granulation tissue; activity; MAGNIFI-CD scoring[8]; [6]

10.3 Example from a Validation Study

ItemTechnique statedSource
Scanners1.5 T (Siemens Avanto Fit) or 3.0 T (Philips Alpha and Omega) at two tertiary centres[10]
SequencesT2-weighted sagittal, coronal and axial with and without fat suppression; axial T1 fat-suppressed after gadolinium 0.1 mmol/kg[10]
MedicationIntravenous scopolamine butylbromide given (contrasting with the ESGAR statement against smooth-muscle relaxants)[10]
Population65 patients with complex perianal fistulising Crohn's disease; baseline and follow-up 3–24 months apart (2007–2021)[10]

11. Evidence Gaps and Ongoing Debate

Technical consensus is expert opinion: the preparation, coil, sequence and contrast statements are level V or III, and the slice thickness and field of view are not specified [1].

Spasmolytic: ESGAR advises against intravenous smooth-muscle relaxants, while a validation study used scopolamine butylbromide [1, 10]; no comparative study of image quality with and without spasmolytic in fistula was found (to be sourced).

Contrast: optional for diagnosis, needed for contrast-based scoring; its added diagnostic value in uncomplicated cryptoglandular fistula is not quantified in the sources read [1, 6, 8].

DWI: optional, no parameters, and primary studies on added value were not read.

Accuracy of MRI against endoanal ultrasound: the quoted meta-analysis is heterogeneous; the ESCP evidence rests on a single outcome-based study in its GRADE table [2, 3].

Activity scoring and healing: MAGNIFI-CD has one retrospective external validation; the descriptions of its items differ between reviews; no consensus definition of radiological healing had been published when the Delphi protocol was registered, and the definition described here comes from a news report [5, 6, 8, 10, 11]. Follow-up intervals are suggested by a review [7] and not by the ESGAR consensus [1].

Sources not read: the full ESCP guideline beyond section 8.11 [2]; the original studies behind the quoted accuracy figures; the original publication of the healing definition; the Parks and St James's original classifications; the ECCO-ESGAR guidance (known only from the Dutch guideline [3]) (to be sourced).

Not covered by any source read: rectovaginal and rectourethral fistula protocols, hidradenitis and pilonidal disease in detail, post-radiation fistula, fistula in children beyond the ESGAR statement of no modification, 3 T versus 1.5 T comparison, artificial intelligence (to be sourced).

12. Evidence-Based References

References are ordered by level of evidence and then by recency. The reading level of each source is stated in its line; a source read only at abstract level supports only what its abstract states.

A. Guidelines / Consensus / Society Recommendations

High
[1] Halligan S, Tolan D, Amitai MM, Hoeffel C, Kim SH, Maccioni F, Morrin MM, Mortele KJ, Rafaelsen SR, Rimola J, Schmidt S, Stoker J, Yang J. ESGAR consensus statement on the imaging of fistula-in-ano and other causes of anal sepsis. Eur Radiol. 2020. DOI: 10.1007/s00330-020-06826-5.
(High — Consensus, modified Delphi on the RAND/UCLA method, 13 members, 53 statements, threshold 11 of 13; full text read (author's accepted manuscript); volume and pages not stated in the text read; the manuscript gives the threshold as 85% while a table caption says more than 80%) Cited throughout.
High
[2] Reza L, Gottgens K, Kleijnen J, Breukink S, Ambe PC, Aigner F, et al. European Society of Coloproctology: Guidelines for diagnosis and treatment of cryptoglandular anal fistula. Colorectal Dis. 2024;26(1):145–196. DOI: 10.1111/codi.16741.
(High — Guideline, AGREE-S and GRADE; methods and imaging section (pp. 157–159) read; the text read is truncated at section 8.11) Cited in Sections 1, 2, 3, 11.
Moderate
[3] Nederlandse Vereniging voor Heelkunde. Richtlijn Proctologie, module "Voorkeur EUS of MRI aanvullende beeldvorming". 3 August 2015. richtlijnendatabase.nl.
(Moderate — National guideline module; web page read; the page states the validity was to be reviewed by 2019, so the current status is unknown; it quotes ECCO-ESGAR 2013, which was not read) Cited in Sections 1, 3, 5, 11.
Limited
[4] Crowley E, Ma C, Guizzetti L, Zou G, Lewindon PJ, Gee MS, et al. Recommendations for standardizing MRI-based evaluation of perianal fistulizing disease activity in pediatric Crohn's disease clinical trials. Inflamm Bowel Dis. 2024;30(3):357–369. DOI: 10.1093/ibd/izad134. PMID: 37524088.
(Limited — Consensus, modified RAND/UCLA, 79 statements, 15 panellists; abstract only read, partly garbled; paediatric trial context) Cited in Sections 3, 4.3.
Limited
[5] Anand E, Devi J, Joshi S, Antoniou A, Doering M, Stoker J, Lung P, Hart AL, Ballard DH, Deepak P, Tozer P. Consensus definition of a radiologically healed fistula on magnetic resonance imaging in perianal Crohn's disease: an international Delphi study (protocol). BMJ Open. 2024;14(11):e087919. DOI: 10.1136/bmjopen-2024-087919.
(Limited — Study protocol without results; full protocol read) Cited in Sections 1.2, 6.3, 11.

B. Systematic Reviews / Meta-analyses

No systematic review of MRI technique in perianal fistula was read for this page (to be sourced). The meta-analyses quoted in Section 3.2 are cited through [2, 3, 6, 7].

Expert
[6] Arkenbosch JHC, van Ruler O, de Vries AC, van der Woude CJ, et al.; Dwarkasing RS. The role of MRI in perianal fistulizing disease: diagnostic imaging and classification systems to monitor disease activity. Abdom Radiol (NY). 2025;50(2):589–597. DOI: 10.1007/s00261-024-04455-w.
(Expert — Narrative review, open access; full text read as rendered by a tool; author list truncated) Cited throughout.
Expert
[7] Habeeb H, Chen L, De Kock I, Bhatnagar G, Kutaiba N, Vasudevan A, Srinivasan AR. Imaging in perianal fistulising Crohn's disease: a practical guide for the gastroenterologist. World J Gastroenterol. 2025;31(34):110611. DOI: 10.3748/wjg.v31.i34.110611. PMID: 40937454.
(Expert — Narrative review; full text read as rendered by a tool; coil, slice thickness and DWI not covered) Cited in Sections 3, 4.5, 6.2, 6.3, 11.
Expert
[8] Anand E, et al.; Lung P. Current practice and innovations in diagnosing perianal fistulizing Crohn's disease (pfCD): a narrative review. Tech Coloproctol. 2025;29:102. DOI: 10.1007/s10151-025-03122-6.
(Expert — Narrative review, open access; full text read as rendered by a tool, truncated in the references; the tool reported the author order ambiguously, so the author list should be verified; its MRI specificity figures are internally inconsistent (69% and 76–100%)) Cited in Sections 2, 3, 4.3, 6.2, 11.

C. Important Prospective / Original Studies

Moderate
[9] Hindryckx P, Jairath V, Zou G, Feagan BG, Sandborn WJ, Stoker J, Khanna R, et al. Development and validation of a magnetic resonance index for assessing fistulas in patients with Crohn's disease. Gastroenterology. 2019;157(5):1233–1244.e5. DOI: 10.1053/j.gastro.2019.07.027.
(Moderate — Index development and internal validation on trial data, 160 participants; abstract only read; items and score range not named in the abstract) Cited in Sections 6.2, 11.
Limited
[10] Beek KJ, Mulders LGM, van Rijn KL, et al. External validation of the MAGNIFI-CD index in patients with complex perianal fistulising Crohn's disease. Eur Radiol. 2025;35:1428–1439. DOI: 10.1007/s00330-024-11029-3.
(Limited — Retrospective multicentre cohort, 65 patients; full text read as rendered by a tool; the median age differs between abstract and results) Cited in Sections 1.2, 4, 6.2, 10, 11.
Limited
[11] VALIDATE-PERIANAL (Samaan S, presenting). Retrospective multicentre study of the TOpClass radiological healing definition, presented at the 2026 Crohn's & Colitis Congress. Reported in: GI & Hepatology News, 2 February 2026.
(Limited — Conference study; news report only read; not peer reviewed; no journal publication found; 40 patients, 14 radiologically healed) Cited in Sections 6.3, 11.

D. Technical MRI Papers

No dedicated technical sequence paper on perianal fistula was read for this page (to be sourced).

E. Landmark Historical References

No landmark historical reference was read for this page (to be sourced).

Recent PubMed search for this protocol

Last updated: October 2026

Report an error on this page

The report stores the page address, the text above and, if you give it, your e-mail. It does not store your IP address, browser type or any cookie. See the Disclaimer, section 7.